Kitazato ET Catheters are intended for ultrasound guided introduction of embryos into the uterine cavity following in vitro fertilization.
Device Story
Kitazato ET Catheters are sterile, single-use devices for embryo transfer procedures. The system consists of a transfer catheter (3Fr or 4Fr), a guide catheter, and optional obturators or stylets. The catheter holds and delivers embryos to the uterine cavity. Some versions include an echo-reflective chip in the tip to aid visualization under ultrasound guidance; others include a stainless steel or polyimide core for rigidity. The guide catheter assists in navigating the cervical canal. The stylet provides additional rigidity during guide insertion. The device is used by clinicians in a clinical setting. The output is the physical delivery of embryos into the uterus. The device benefits patients by facilitating precise, ultrasound-guided embryo placement during IVF, potentially improving procedure success.
Clinical Evidence
Bench testing only. Studies included biocompatibility (ISO 10993), sterilization validation (ISO 11137), package integrity (ASTM F1886, F1929, F88), and mechanical performance (tensile strength, luer taper, echo chip visualization). Mouse Embryo Assay (MEA) confirmed embryotoxicity safety (≥80% blastocysts at 96h).
Indicated for patients undergoing in vitro fertilization requiring ultrasound-guided embryo transfer into the uterine cavity.
Regulatory Classification
Identification
Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
Predicate Devices
Guardia™ Access Embryo Transfer Catheter Sets (K173686)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 1, 2020
Kitazato Corporation % Audrey Swearingen Regulatory Affairs Manager/Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, TX 78746
Re: K192540
> Trade/Device Name: Kitazato ET Catheters (Type 1 [Versions 1-4], Type 2 [Versions 1-5], and Type 3 [Versions 1-4]) Regulation Number: 21 CFR§ 884.6110 Regulation Name: Assisted Reproduction Catheters Regulatory Class: II Product Code: MQF Dated: April 1, 2020 Received: April 2, 2020
Dear Audrey Swearingen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192540
#### Device Name
Kitazato ET Catheters (Type 1 [Versions 1-4], Type 2 [Versions 1-5], and Type 3 [Versions 1-4])
#### Indications for Use (Describe)
Kitazato ET Catheters are intended for ultrasound guided introduction of embryos into the uterine cavity following in vitro fertilization.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## Kitazato ET Catheter
### 1. Submission Sponsor
Kitazato Corporation Shibakoen Building 1-1-8 Shibadaimon Minato-ku Tokyo 105-0012 Japan Contact: Mr. Futoshi Inoue Phone: +81 35 800 6055 Title: President and Representative Director
## 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327-9997 Contact: Audrey Swearingen Title: Regulatory Affairs Manager/Senior Consultant
#### 3. Date Prepared
April 28, 2020
## 4. Device Identification
| Trade/Proprietary Name: Kitazato ET Catheters (Type 1 [Versions 1-4], Type 2 [Versions 1-5], and Type<br>3 [Versions 1-4]) | |
|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| Common/Usual Name: | Embryo Transfer Catheter |
| Regulation Name: | Assisted Reproduction Catheters |
| Regulation Number: | 884.6110 |
| Product Code: | MQF (Catheter, Assisted Reproduction) |
| Class: | Class II |
## 5. Legally Marketed Predicate Device(s)
{4}------------------------------------------------
Guardia™ Access Embryo Transfer Catheter Sets (Cook Incorporated), K173686.
This predicate device has not been subject to any design related recalls.
## 6. Indication for Use Statement
Kitazato ET Catheters are intended for ultrasound guided introduction of embryos into the uterine cavity following in vitro fertilization.
## 7. Device Description
Kitazato ET Catheters are sterile, single-use catheters for use in embryo transfer procedures.
The catheter is used to hold and deliver embryos to the uterine cavity during a procedure. Catheters are available in 3Fr and 4Fr sizes, both of which are provided in lengths of 220 mm and 250 mm, as well as one version that is 3Fr and 400 mm length that has the same uterine insertion depth as the other device versions, but allows the user to be positioned further from the patient during a procedure. Devices also include versions with and without an echo reflective chip in the tip of the catheter. The echo reflective chip aids in visualization of the catheter during a transfer procedure; all transfer procedures are to be done under ultrasound guidance. Both the Type 2 devices also include a stainless steel or polyimide core to provide additional rigidity to the catheter.
The guide is used to navigate the pathway through the cervical canal to allow for easier catheter insertion Guides are provided with a 30° pre-curved distal shaft and are available in lengths of 170 mm and 200 mm. Guides also include a styrene elastomer stopper to aid in positioning the targeted uterine insertion depth. The guides are also available with a straight or bulb tip design, and with and without an echo reflective chip in the tip of the guide. An obturator with a length of 170 mm or 200 mm is provided with the Type 2 versions of the device to prevent ingress of fluid into the guide lumen during guide insertion.
The Stylet is an optional accessory used when additional rigidity is needed during insertion of the guide through the cervix, and is provided in two lengths, 170 mm and 200 mm. The stainless steel core of the Stylet can be shaped to aid in guide delivery through the cervix.
| Type | Size | Version Components | | |
|----------|------|--------------------|------------------------------------|-----------|
| | | Catheter | Guide | Obturator |
| Type1-v1 | 3Fr | 220 mm ET Catheter | 170 mm Guide (curved) bulb tip | |
| Type1-v2 | 3Fr | 250 mm ET Catheter | 200 mm Guide (curved) bulb tip | |
| Type1-v3 | 4Fr | 220 mm ET Catheter | 170 mm Guide (curved) straight tip | |
| Type1-v4 | 4Fr | 250 mm ET Catheter | 200 mm Guide (curved) straight tip | |
A summary of the different device versions included in this submission are shown in the tables below:
{5}------------------------------------------------
| Type2-v1 | 3Fr | 220 mm ET Catheter | 170 mm Guide (curved) bulb tip | 170 mm Obturator |
|----------|-----|--------------------|------------------------------------|------------------|
| Type2-v2 | 3Fr | 250 mm ET Catheter | 200 mm Guide (curved) bulb tip | 200 mm Obturator |
| Type2-v3 | 4Fr | 220 mm ET Catheter | 170 mm Guide (curved) straight tip | 170 mm Obturator |
| Type2-v4 | 4Fr | 250 mm ET Catheter | 200 mm Guide (curved) straight tip | 200 mm Obturator |
| Type2-v5 | 3Fr | 400 mm ET Catheter | 200 mm Guide (curved) bulb tip | 200 mm Obturator |
| Type | Description |
|----------|----------------------------|
| Type3-v1 | 3Fr Stylet for 170mm Guide |
| Type3-v2 | 3Fr Stylet for 200mm Guide |
| Type3-v3 | 4Fr Stylet for 170mm Guide |
| Type3-v4 | 4Fr Stylet for 200mm Guide |
## 8. Substantial Equivalence Comparison
The following table compares the Kitazato ET Catheter to the predicate device.
| Manufacturer | Kitazato Corporation | Cook Medical, Inc. | Device Comparison |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Kitazato ET Catheter<br>K192540 | Guardia Access Embryo<br>Transfer Catheter Sets<br>K173686 | |
| Indications for Use | Kitazato ET Catheters are<br>intended for ultrasound<br>guided introduction of<br>embryos into the uterine<br>cavity following in vitro<br>fertilization. | Used to place in vitro<br>fertilized (IVF) embryos into<br>the uterine cavity. | The indications for use<br>statements are not identical;<br>however, the intended use<br>of the subject and predicate<br>devices are the same<br>(delivery of embryos into the<br>uterine cavity). |
| Variations/Models | Type1: Catheter with<br>supporting core and Guide<br><br>Type2: Catheter with<br>supporting core and Guide<br>with Obturator<br><br>Type3: Stylet | Catheter and Guide –<br>internal supporting cannula<br>available in one version | Different: Both offer transfer<br>and guide catheters. All of<br>the subject devices include<br>an additional internal<br>support, which is<br>comparable to a version of<br>the predicate device. The<br>subject device also includes<br>obturators and stylets. The |
| Manufacturer | Kitazato Corporation | Cook Medical, Inc. | Device Comparison |
| Trade Name | Kitazato ET Catheter<br>K192540 | Guardia Access Embryo<br>Transfer Catheter Sets<br>K173686 | |
| Design Features | Catheter: Open-end with<br>depth markings, Luer lock<br>hub, with and without echo<br>chip marker;<br>Guide: Precurved, depth<br>marks, styrene elastomer<br>stopper, straight or bulb tip,<br>with and without echo tip | Transfer Catheter: Open-<br>ended with depth markings,<br>Luer lock hub, with and<br>without echo tip;<br>Guide: Precurved, bulb tip,<br>stopper, no depth markings<br>or echo tip | Different. Both offer<br>catheter models and guides<br>with the same basic design<br>features. The differences<br>shown do not raise different<br>questions of safety and<br>effectiveness. |
| | | | inclusion of obturators and<br>stylets do not raise different<br>questions of safety and<br>effectiveness. |
| Materials | Transfer Catheter - Silicone,<br>stainless steel/polyimide<br>core, ABS, nylon<br>Guide - Nylon,<br>polycarbonate, styrene<br>elastomer, stainless steel<br>Obturator - Nylon,<br>polycarbonate<br>Stylet - Nylon, stainless<br>steel, polycarbonate | Transfer catheter -<br>Polyethylene, polyurethane,<br>stainless steel<br>Guide - Polyurethane,<br>silicone, polymethylpentene | Different: Different materials<br>are used in the subject and<br>predicate devices; however,<br>they do not raise different<br>questions of safety and<br>effectiveness. |
| Sterilization method | Gamma irradiation | Ethylene oxide | Different: Different<br>questions of safety and<br>effectiveness are not raised<br>by the use of different<br>sterilization methods. |
| Single-Use | Yes | Yes | Same |
| Shelf Life | 3 years | 3 years | Same |
| Length of Catheters | Transfer catheter - 22 cm,<br>25 cm and 40cm<br>Guide catheter - 17 cm, 20<br>cm | Transfer catheter - 23 - 25<br>cm<br>Guide catheter - 16.7 cm,<br>17.3 cm | Different: The dimensions<br>are comparable for the<br>majority of the subject<br>devices with the exception<br>of the Type 2 v 5. This<br>version is longer than the<br>predicate devices, but it |
| Manufacturer | Kitazato Corporation | Cook Medical, Inc. | Device Comparison |
| Trade Name | Kitazato ET Catheter<br>K192540 | Guardia Access Embryo<br>Transfer Catheter Sets<br>K173686 | |
| | | | does not raise different<br>questions of safety and<br>effectiveness as the device<br>can only be inserted the<br>same depth as the other<br>devices. The additional<br>length is provided to allow<br>the physician to be<br>positioned further from the<br>subject during a procedure. |
| Outer Diameter | Catheter – 1.0mm/3 Fr,<br>1.35mm/4 Fr<br>Guide – 1.9mm, 2.2mm | Transfer Catheter – 2.8 Fr<br>Guide Catheter - 6.6 Fr | Different: The differences<br>in outer diameter between<br>the subject and predicate<br>do not raise different<br>questions of safety and<br>effectiveness. |
| Depth Markings | Transfer Catheter - located<br>at 1, 2, 3 and 4 cm from<br>connector.<br>Guide catheter marks -<br>located at 4, 5, 6, 7 and 8cm<br>from the tip. | Transfer catheter includes<br>depth marks (spacing not<br>known).<br>No depth markings on Guide<br>catheter. | Different: The subject and<br>predicate catheter<br>components include depth<br>markings, while only the<br>subject device Guide<br>catheter includes depth<br>markings. Differences in<br>depth marking between the<br>subject and predicate device<br>do not raise different<br>questions of safety and<br>effectiveness. |
| Mouse Embryo<br>Assay | 1-cell, ≥ 80% blastocyst at<br>96h | ≥ 80%expanded blastocytes<br>within 96 hours | Similar |
| Endotoxin | < 20 EU/device | < 20 EU/device | Same |
{6}------------------------------------------------
{7}------------------------------------------------
## 9. Non-Clinical Performance Data
The following studies have been performed to support substantial equivalence to the predicate device:
- . Sterilization validation per ISO 11137-1:2006(R)2010 &A1:2013 and ISO 11137-2:2013
- . Biocompatibility studies conducted in accordance with ISO 10993-1: 2018 and the 2016 FDA Guidance Document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1:Evaluation and testing within a risk management process."
{8}------------------------------------------------
K192540 Page 6 of 6
- Cytotoxicity ISO 10993-5:2009 O
- O Sensitization ISO 10993-10:2010
- Intracutaneous reactivity ISO 10993-10:2010 O
- Endotoxin testing per AAMI/ANSI ST72:2011; USP <85> (<20 EU/device) ●
- . Transportation Simulation study per ASTM D4169-16
- Assessment of packaging and contents after conditioning o
- Bubble leak test per ASTM F2096-11 after conditioning O
- . Package Integrity testing:
- Visual Inspection per ASTM F1866/F1886M-09 o
- Dye Penetration test per ASTM F1929-15 O
- O Seal Strength test per ASTM F88/F88M-15
- Mouse Embryo Assay (MEA) before and after aging (1-Cell MEA: ≥80% blastocysts at 96h)
- . Bench Performance studies before and after aging demonstrated that all predetermined acceptance criteria were met in the following tests:
- Dimensional Verification Devices were measured and verified against device input o requirements.
- Appearance Devices were visually inspected to ensure that no burrs, scratches, or foreign o objects observed.
- o Tensile Strength: Testing demonstrates that the tensile strength value is greater than the pre-determined acceptance criterion.
- Luer Taper Inspection: Testing is visual inspection to assess connection of the shaft to a 6% o tapered female connector.
- O Colorfastness – Testing demonstrates that the depth marking is legible undergoing friction testing (100 rubs at 200 g load weight).
- o Echo Test: Testing demonstrates maintenance of the echo chip band on the catheter and guide in accordance with a pre-determined acceptance criterion.
## 10. Statement of Substantial Equivalence
The subject and predicate devices have the same intended use and comparable technological characteristics. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.