K192533 · Health Line International Corporation · FOZ · Dec 10, 2019 · General Hospital
Device Facts
Record ID
K192533
Device Name
Health Line CT CVC
Applicant
Health Line International Corporation
Product Code
FOZ · General Hospital
Decision Date
Dec 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HEALTH LINE CT CVC is indicated for short term (less than 30 days) access to the central venous system for intravenous administration of fluids, medications, blood products, and/or nutritional therapy solutions when prescribed. Blood sampling and power injection of contrast media may also be conducted with the HEALTH LINE CT CVC.
Device Story
Central venous catheter (CVC) for short-term (less than 30 days) vascular access; used for infusion, blood sampling, and power injection of contrast media. Catheter inserted percutaneously into central vein (typically internal jugular or subclavian) via Seldinger technique. Device consists of radiopaque polyurethane tubing, injection-molded polyurethane hub, extension legs with Luer lock fittings, and optional suture wing. Operates by providing a conduit for fluid delivery or withdrawal; withstands power injection up to 5 ml/sec at 300 psi. Used by medical professionals in clinical settings. Benefits include reliable central venous access for therapy and diagnostic imaging.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and leveraging of biocompatibility and performance data from the identical predicate device (K113622).
Technological Characteristics
Radiopaque polyurethane catheter; polyurethane hub; rigid PVC female Luer connectors; acetal pinch clamps; ABS clamp inserts. Dimensions: 16G/14G single lumen, 5Fr/7Fr dual lumen, 7Fr triple lumen; 13-30 cm length. Sterilization: Ethylene Oxide (EtO) per ISO 11135-1:2007 and AAMI TIR 28. Biocompatibility: ISO 10993. Non-powered, mechanical device.
Indications for Use
Indicated for adult patients requiring short-term (less than 30 days) central venous access for fluid/medication/blood/nutritional therapy, blood sampling, and power injection of contrast media.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 10, 2019
Health Line International Corporation Joel Faulkner CEO 5675 W 300 S Salt Lake City, Utah 84117
Re: K192533
Trade/Device Name: Health Line CT CVC Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: September 12, 2019 Received: September 16, 2019
Dear Joel Faulkner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Geeta Pamidimukkala Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192533
Device Name Health Line CT CVC
#### Indications for Use (Describe)
The HEALTH LINE CT CVC is indicated for short term (less than 30 days) access to the central venous system for intravenous administration of fluids, medications, blood products, and/or nutritional therapy solutions when prescribed. Blood sampling and power injection of contrast media may also be conducted with the HEALTH LINE CT CVC.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <div> <span> X Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
(As required by 21 CFR 807.92)
#### SUBMITTER:
Health Line International Corporation 5675 West 300 South Salt Lake City, Utah 84104
#### ESTABLISHMENT REGISTRATION NUMBER:
3006097687
### CONTACT:
Joel K. Faulkner CEO Telephone: 801-773-7798 Fax: 801-820-8007 Email: jkfaulkner@hlic.net
# DATE PREPARED:
November 20, 2019
### NAME OF MEDICAL DEVICE:
| Proprietary Name: | <i>Health Line CT CVC</i> |
|--------------------|------------------------------------------------------------------------------------------------|
| Regulation Name: | Intravascular Catheter |
| Common/Usual Name: | Central Venous Catheter (CVC), single, double and triple lumen<br>short-term less than 30 days |
### DEVICE CLASSIFICATION:
| Classification Panel: | General Hospital |
|-----------------------|------------------|
| Regulatory Class: | Class II |
| Product Code: | FOZ |
| Regulation Number: | 21 CFR 880.5200 |
### PREDICATE DEVICE:
Proprietary Name: Regulation Name: Common/Usual Name:
Classification Panel: Regulatory Class: Product Code: Regulation Number: Orion™ II CT CVC (K113622)
Intravascular Catheter Central Venous Catheter (CVC), single, double and triple lumen, short-term less than 30 days General Hospital Class II FOZ 21 CFR 880.5200
{4}------------------------------------------------
# DEVICE DESCRIPTION:
The HEALTH LINE CT CVC is a family of central venous catheters designed to perform infusion, intravenous therapy, blood sampling and also power injection of contrast media studies. The catheters, made of radiopaque polyurethane tubing, are inserted in a central vein. Each HEALTH LINE CT CVC has a kink resistant, gradual tapered catheter design. The HEALTH LINE CT CVC kit includes a catheter and introduction components. The catheter is supplied sterile and nonpyrogenic in a variety of kit configurations.
The HEALTH LINE CT CVC is indicated for dwell times shorter than 30 days. The HEALTH LINE CT CVC catheter assemblies have been tested to withstand power injection of worst-case viscosity injection media at 5 ml/sec with a maximum power injector pressure of 300 psi.
The HEALTH LINE CT CVC product line has catheters in 16 G and 14 G single lumen, 5 Fr and 7 Fr dual lumen and 7 Fr triple lumen. Catheters range from approximately 13-30 cm long. The catheters are attached to an injection-molded polyurethane hub that has extension legs with Luer lock fittings for access attachment. All HEALTH LINE CT CVC products have a maximum recommended infusion rating is 5 ml/sec. The maximum pressure or pounds per square inch (psi) of the power injector utilized should not exceed 300 psi.
# INDICATIONS FOR USE:
The HEALTH LINE CT CVC is indicated for short term (less than 30 days) access to the central venous system for intravenous administration of fluids, medications, blood products, and/or nutritional therapy solutions when prescribed. Blood sampling and power injection of contrast media may also be conducted with the HEALTH LINE CT CVC.
# TECHNOLOGICAL COMPARISON TO PREDICATE DEVICE:
The HEALTH LINE CT CVC is IDENTICAL to the predicate device the Orion™ II CT CVC (K113622) in ALL aspects except:
- New labeling, Instructions for use (IFU) (with additional instructions), addition suture wing components (to provide clinicians with additional options, if desired) in the kit configuration.
- Catheter length - Exactly same as predicate.
- . Insertion site – Exactly the same as predicate Orion™ II CT CVC and is placed in a central vein.
- Dwell time – Exactly the same as predicate Orion™ II CVC is indicated for short term access.
The changes do not impact the performance or safety and effectiveness of the device compared to the predicate.
{5}------------------------------------------------
The HEALTH LINE CT CVC subject device and the predicate device Orion™ II CT CVC (K113622) are manufactured by the same company, Health Line International Corporation, the submitter of this submission.
| Comparison<br>Area | Submission Device<br>HEALTH LINE CT CVC | Predicate Device<br>Orion™ II CT CVC K113622 |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use: | The Health Line CT CVC is indicated for<br>short term (less than 30 days) access to<br>the central venous system for<br>intravenous administration of fluids,<br>medications, blood products, and/or<br>nutritional therapy solutions when<br>prescribed. Blood sampling and power<br>injection of contrast media may also be<br>conducted with the Health Line CT CVC. | The Orion™ II CT CVC is<br>indicated for short term (less<br>than 30 days) access to the<br>central venous system for<br>intravenous administration of<br>fluids, medications, blood<br>products, and/or nutritional<br>therapy solutions when<br>prescribed. Blood sampling and<br>power injection of contrast<br>media may also be conducted<br>with the Orion™ II CT CVC.<br><br>Different- Updated with name<br>of catheter |
| Intended Use: | Intended to be used by medical<br>professionals for short term access to<br>the central venous system for infusion,<br>intravenous therapy, blood sampling<br>and power injection of contrast media | Identical |
| Target Population | Adults | Identical |
| Duration of Use | Less than 30 days | Identical |
| Insertion Method | Seldinger Technique | Identical |
| Tip Placement Location | Located just before the junction of the<br>superior vena cava and the right atrium | Identical |
| Insertion Site | Inserted percutaneously and primarily<br>placed in the internal jugular vein.<br>Alternate insertion sites include the<br>subclavian vein. | Identical |
| Sizes | 16G & 14G Single Lumen, 5 & 7 Fr Dual<br>Lumen and 7 Fr Triple Lumen | Identical |
| Labeling | New Health Line CT CVC logo | Orion™ II CT CVC logo. |
| Instructions for Use | Added optional use of suture wing to<br>instructions, added priming volumes | Without suture wing<br>instructions and priming<br>volumes. |
| Components | Added suture wing securement device<br>to kit components. | Without suture wing. |
| Biocompatibility | ISO-10993 | Identical |
| Sterilization Method | EtO (SAL 10-6) | Identical |
| <b>Materials</b> | | |
| Female Luer connectors | Rigid PVC | Identical |
| Hub | Polyurethane | Identical |
| Catheter Tubing | Polyurethane | Identical |
| Extension Leg Tubing | Polyurethane | Identical |
| Pinch Clamps | Acetal | Identical |
| Informational Clamp Inserts | ABS | Identical |
{6}------------------------------------------------
### PERFORMANCE TESTING
The subject device is identical to the predicate device (K113622) in its final finished form. Therefore, the biocompatibility and performance data were leveraged from the predicate device. The subject device complies with FDA's Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, dated 3/16/95.
The subject device complies with sterilization requirements of ISO 11135-1:2007, Sterilization of Health Care Products – Ethylene Oxide – Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices. It also complies with AAMI TIR 28, Product adoption and process equivalence for ethylene oxide sterilization.
### CONCLUSION
Based on the Indications for Use, technological characteristics and performance testing; the HEALTH LINE CT CVC is substantially equivalent to the following predicate device Orion™ II CT CVC (K113622).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.