← Product Code [FBK](/productcode/FBK) · K192498

# Single Use Injector (K192498)

_Olympus Medical Systems Corp. · FBK · Apr 28, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K192498

## Device Facts

- **Applicant:** Olympus Medical Systems Corp.
- **Product Code:** [FBK](/productcode/FBK.md)
- **Decision Date:** Apr 28, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

This instrument has been designed to be used with an Olympus endoscope to deliver injectable materials into the urinary bladder wall during the transurethral endoscopic procedures.

## Device Story

Single-use endoscopic injection needle; used with Olympus endoscope for transurethral delivery of injectable materials into urinary bladder wall. Device consists of needle section, coil sheath, and tube sheath; includes stylet. Operation: device inserted into endoscope in retracted state; needle protrudes from distal end of sheath upon manual activation of operation portion; external syringe (not included) attached to proximal end for injection. Used in clinical settings by physicians. Benefits: enables targeted delivery of therapeutic agents to bladder wall during endoscopic procedures.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including endoscope compatibility, needle slidability, puncture performance, liquid leakage, durability, flow rate (BS EN 1618:1997), and corrosion testing (ISO 9626:2016). Biocompatibility testing performed per ISO 10993-1.

## Technological Characteristics

Single-use endoscopic injector; 27G needle; dual-structure sheath (outer coil, inner tube); 971mm working length; includes 0.3mm diameter stylet. Materials evaluated for biocompatibility per ISO 10993-1. Corrosion resistance tested per ISO 9626:2016. Sterile, single-use device.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- BONEE NEEDLE FOR BLADDER INJECTIONS NB1070 ([K090217](/device/K090217.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 28, 2020

Olympus Medical Systems Corp. Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610

Re: K192498

> Trade/Device Name: Single Use Injector NM-221C-0427 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FBK Dated: March 19, 2020 Received: March 20, 2020

Dear Sheri Musgnung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Martha W. Betz, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K192498

Device Name Single Use Injector NM-221C-0427

Indications for Use (Describe)

This instrument has been designed to be used with an Olympus endoscope to deliver injectable materials into the urinary bladder wall during the transurethral endoscopic procedures.

Type of Use (Select one or both, as applicable)

| <label><input checked="checked" type="checkbox"/> Research Use (Part 81 CFR 801.3 Subject to Review)</label> |
|--------------------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Genetic Testing (81.420)</label>                                             |

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are all capitalized and evenly spaced. A registered trademark symbol is located to the right of the letter "S".

March 19, 2020

Single Use Injector NM-221C-0427

# Revised 510(k) Summary

### 1 GENERAL INFORMATION

| ■ 510(k) Submitter: | OLYMPUS MEDICAL SYSTEMS CORP.<br>2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507,<br>Japan |
|---------------------|--------------------------------------------------------------------------------------------|
|---------------------|--------------------------------------------------------------------------------------------|

- Contact Person: Sheri L. Musgnung Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 Fax: 484-896-7128 Email: sheri.musgnung@olympus.com
- Manufacturing site: Aomori Olympus Co., Ltd. ■ 2-248-1 Okkonoki, Kuroishi-shi, Aomori 036-0357, Japan

#### 2 DEVICE IDENTIFICATION

| ■ Device Name          | Single Use Injector NM-221C-0427                      |
|------------------------|-------------------------------------------------------|
| ■ Common Name          | endoscopic injection needle, gastroenterology-urology |
| ■ Regulation Number    | 21 CFR 876.1500                                       |
| ■ Regulation Name      | Endoscope and accessories                             |
| ■ Regulatory Class     | II                                                    |
| ■ Product Code         | FBK                                                   |
| ■ Classification Panel | Gastroenterology/Urology                              |

#### 3 PREDICATE DEVICE

#### ■ Predicate device

| Device name                                   | 510(k) Submitter | 510(k) No. |
|-----------------------------------------------|------------------|------------|
| BONEE NEEDLE FOR BLADDER<br>INJECTIONS NB1070 | COLOPLAST A/S    | K090217    |

{4}------------------------------------------------

#### 4 DEVICE DESCRIPTION

#### - General Description of the subject device

This instrument has been designed to be used with an Olympus endoscope to deliver injectable materials into the urinary bladder wall during the transurethral endoscopic procedures. The subject device is sterilized and packaged in a sterilization package.

#### Principle of Operation

When the operation portion of the needle section is pushed into the Sheath section, the needle protrudes from the distal end of the sheath section. Pulling the operation portion of the needle section toward the proximal end with respect to the sheath section pulls the needle into the coil sheath. The subject device is inserted into the endoscope in the retracted state. After the subject device reaches the urinary bladder wall, puncture is performed with the needle protruding from the coil sheath. A commercially available marketed syringe (not included in the subject device) is attached to the proximal end of the needle section to deliver injectable materials to the target site of the urinary bladder wall.

#### 5 INDICATIONS FOR USE

This instrument has been designed to be used with an Olympus endoscope to deliver injectable materials into the urinary bladder wall during the transurethral endoscopic procedures.

## 6 COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEVICE

The NM-221C-0427 has the same technological characteristics and design as the predicate device except for the following new features:

- Needle design
- Sheath design
- Working length

Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety or effectiveness of the subject device.

{5}------------------------------------------------

K192498

A side by side comparison of the subject device and the predicate device is provided below.

|                                       | Subject Device                                                                                                                                                                                         | Predicate Device                                                                                                                                                                |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item                                  | Single Use Injector<br>NM-221C-0427                                                                                                                                                                    | BONEE NEEDLE FOR<br>BLADDER INJECTIONS<br>NBI070                                                                                                                                |
| Indications for Use                   | This instrument has been<br>designed to be used with an<br>Olympus endoscope to<br>deliver injectable materials<br>into the urinary bladder<br>wall during the transurethral<br>endoscopic procedures. | The Bonee Needle for<br>Bladder Injections is used<br>to deliver injectable<br>materials into the urinary<br>bladder wall during the<br>transurethral endoscopic<br>procedures. |
| Regulation Number                     | 21 CFR 876.1500                                                                                                                                                                                        | 21 CFR 876.1500                                                                                                                                                                 |
| Regulation Name                       | Endoscope and accessories                                                                                                                                                                              | Endoscope and accessories                                                                                                                                                       |
| Regulatory Class                      | II                                                                                                                                                                                                     | II                                                                                                                                                                              |
| Product Code                          | FBK                                                                                                                                                                                                    | FBK                                                                                                                                                                             |
| Classification Panel                  | Gastroenterology/Urology                                                                                                                                                                               | Gastroenterology/Urology                                                                                                                                                        |
| Compatible Olympus<br>cystoscope      | Working length: less than<br>380mm<br>Channel inner diameter:<br>2.0mm or more                                                                                                                         | Working length: unknown<br>Channel inner diameter: 5<br>French(= 1.7mm) or more                                                                                                 |
| Needle width                          | 27G                                                                                                                                                                                                    | 22G                                                                                                                                                                             |
| Typical needle length                 | 4mm                                                                                                                                                                                                    | 4mm                                                                                                                                                                             |
| Maximum insertion portion<br>diameter | φ1.84mm                                                                                                                                                                                                | unknown                                                                                                                                                                         |
| Working length                        | 971 mm                                                                                                                                                                                                 | 700mm                                                                                                                                                                           |
| Sheath type                           | Sheath structure: dual<br>structure (outer: coil sheath,<br>inner: tube sheath)                                                                                                                        | Sheath structure: single<br>structure (tube sheath only)                                                                                                                        |
| Stylet                                | Equipped (Diameter:<br>φ0.3mm, Length: 700mm)                                                                                                                                                          | Not equipped                                                                                                                                                                    |
| Tip shape of needle                   | Tip shape: Lancet Point                                                                                                                                                                                | Tip shape: Unknown                                                                                                                                                              |

{6}------------------------------------------------

### 7 PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

#### 1) Sterilization/Shelf-life testing

Sterilization/shelf-life testing for the NM-221C-0427 were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".

Accelerated aging test for the NM-221C-0427 was conducted in accordance with ASTM F1980-16, the standard guide for accelerated aging of sterile barrier systems for medical devices. The real-time aging test for three-years will be performed to demonstrate longer stability and support the results of the accelerated aging test.

#### 2) Biocompatibility testing

Biocompatibility testing for the NM-221C-0427 were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff. Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The Single Use Injector NM-221C-0427 is considered an externally communicating medical device in contact with Tissue/bone/dentin. The contact duration is limited exposure (i.e. contact is up to 24 hours). The biocompatibility testing included the following tests:

- Cytotoxicity Study Using the Colony Assay ।
- ISO Intracutaneous Study in Rabbits ।
- ISO Guinea Pig Maximization Sensitization Test
- ISO Acute Systemic Toxicity Study in Mice
- USP Rabbit Pyrogen Study, Material Mediated ।

#### 3) Performance testing - Bench

Bench testing for the NM-221C-0427 as listed below was conducted to ensure that the subject device performs as intended and meet design specifications.

- Endoscope compatibility
- Needle slidability
- Puncture performance
- Liquid leakage
- Durability
- Flow rate (BS EN 1618: 1997 Catheters other than intravascular catheters Test methods for common properties)
- Corrosion Testing (ISO 9626: 2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are all capitalized and have a slightly three-dimensional appearance. A registered trademark symbol is located to the right of the letter "S".

K192498 Page 5 of 5

#### 4) Performance testing - Animal

No animal study was performed to demonstrate substantial equivalence.

#### 5) Performance testing - Clinical

No clinical study was performed to demonstrate substantial equivalence.

#### 6) Risk analysis

Risk analysis for the NM-221C-0427 was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.

#### 8 CONCLUSIONS

Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate device, the NM-221C-0427 raise no new issue of safety and effectiveness and are substantially equivalent to the predicate device in terms of safety, effectiveness and performance.

---

**Source:** [https://fda-staging.innolitics.com/device/K192498](https://fda-staging.innolitics.com/device/K192498)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).