K192493 · Becton Dickinson Infusion Therapy Systems, Inc. · FOZ · Jan 24, 2020 · General Hospital
Device Facts
Record ID
K192493
Device Name
BD Cathena (TM) Safety IV Catheter
Applicant
Becton Dickinson Infusion Therapy Systems, Inc.
Product Code
FOZ · General Hospital
Decision Date
Jan 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BD Cathena™ Safety IV Catheters are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. These devices are suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa).
Device Story
Over-the-needle peripheral IV catheter; includes radiopaque polyurethane catheter, needle, grip, passive safety needle shield, and flash chamber with vent plug. Features BD Instaflash™ technology for immediate blood visualization in catheter; optional multi-access BD Multiguard™ technology stops blood flow from hub until Luer connection. Used in clinical settings by healthcare providers for vascular access. Needle tip passively protected upon removal to prevent needlestick injuries. Hubs color-coded by gauge. Benefits include reduced needlestick risk, improved visualization of vessel entry, and controlled blood flow.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included flashback visualization time, adhesion/drag forces, burst pressure, catheter separation force, leak time, flow rate, and power injection testing per ISO 10555-1. Biocompatibility evaluated per ISO 10993-1:2018. Particulate testing performed per USP <788>.
Technological Characteristics
Peripheral IV catheter. Materials: Polyurethane (catheter), Stainless Steel (needle), Polypropylene (grip/hub/adapter), Acrylonitrile Butadiene Styrene (safety shield). Lubricant: 1-part silicone. Sterilization: Ethylene Oxide (SAL 10^-6). Dimensions: 16G-24G, 0.75-2.00 inch lengths. Passive safety mechanism. Complies with ISO 10555-1, ISO 11135:2014, and USP <788>.
Indications for Use
Indicated for insertion into peripheral vascular system for short-term blood sampling, blood pressure monitoring, or fluid administration. Suitable for any patient population based on vascular anatomy, procedure, fluid type, and therapy duration. Compatible with power injectors up to 325 psi.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
BD Insyte™ Autoguard™ BC Safety IV Catheter (K110443)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
January 24, 2020
Becton Dickinson Infusion Therapy Systems, Inc. Henry Boland Staff Regulatory Affairs Specialist 9450 South State Street Sandy, Utah 84070
Re: K192493
Trade/Device Name: BD Cathena™ Safety IV Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: December 19, 2019 Received: December 20, 2019
Dear Henry Boland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Geeta Pamidimukkala Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192493
Device Name BD Cathena™ Safety IV Catheter
#### Indications for Use (Describe)
BD Cathena™ Safety IV Catheters are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. These devices are suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa).
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="font-size:20px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="font-size:20px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
bd.com
Image /page/3/Picture/2 description: The image shows the BD logo. The logo consists of an orange circle with white rays emanating from the center, resembling a sun. To the right of the circle are the letters "BD" in a bold, blue font. The letters are large and easily readable.
# 510(k) Summary (21 CFR §807.92)
# BD Cathena™ Safety IV Catheter
| Submitter | Submitter Name: | Becton Dickinson Infusion Therapy Systems Inc. |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Information | Submitter Address: | 9450 South State Street<br>Sandy, UT 84070 |
| | Contact Person: | Henry Boland<br>Staff Regulatory Affairs Specialist |
| | Email Address: | henry.boland@bd.com |
| | Phone Number: | (801) 565-2550 |
| | Date of Preparation: | January 23, 2020 |
| Subject Device | Trade Name: | BD Cathena™ Safety IV Catheter |
| | 510(k) Reference: | K192493 |
| | Common Name: | Peripheral Intravascular or IV Catheter |
| | Regulation Number: | 21 CFR §880.5200 |
| | Regulation Name: | Catheter, intravascular, therapeutic, short-term less than<br>30 days |
| | Regulatory Class: | II |
| | Product Code: | FOZ |
| | Classification Panel: | General Hospital |
| Predicate | Trade Name: | BD Cathena™ Safety IV Catheter |
| Device | 510(k) Reference: | K172506, cleared 17 September 2017 |
| | Common Name: | Peripheral Intravascular or IV Catheter |
| | Regulation Number: | 21 CFR §880.5200 |
| | Regulation Name: | Catheter, intravascular, therapeutic, short-term less than<br>30 days |
| | Regulatory Class: | II |
| | Product Code: | FOZ |
| | Classification Panel: | General Hospital |
| Reference | Trade Name: | BD Insyte™ Autoguard™ BC Safety IV Catheter |
| Device | 510(k) Reference: | K110443, cleared 19 July 2011 |
| | Common Name: | Peripheral Intravascular or IV Catheter |
| | Regulation Number: | 21 CFR §880.5200 |
| | Regulation Name: | Catheter, intravascular, therapeutic, short-term less than<br>30 days |
| | Regulatory Class: | II |
| | Product Code: | FOZ |
| Classification Panel: | General Hospital | |
| Reason for Submission | The reason for this submission is the introduction of a performance specification<br>(due to a change in material supplier and grade of silicone lubricant),<br>modification of catheter tubing dimensions, and modifications to product<br>labeling. | |
| Device Description | BD Cathena™ Safety IV Catheters are over-the-needle, intravascular (IV)<br>catheters. These devices include a radiopaque BD Vialon™ catheter, needle, grip,<br>passive safety needle shield, and flash chamber with removable vent plug. The<br>needle and catheter are protected by a needle cover. These devices have BD<br>Instaflash™ Needle Technology, allowing for immediate visualization of blood<br>along the catheter. The flash chamber provides confirmation that the device has<br>entered the vessel. The needle tip is passively protected when the needle is<br>removed, reducing the risk of accidental needlestick injury.<br><br>These devices are available with or without multi-access BD Multiguard<br>technology, which is designed to stop the flow of blood from the catheter hub<br>until a Luer connection is made. Once a connection is made, fluids or blood can<br>flow through the catheter hub in either direction.<br><br>These devices are available with or without wings. The catheter hub and wings<br>are color coded to indicate the catheter gauge size (24 GA (0.7 mm) = Yellow, 22<br>GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA<br>(1.7 mm) = Grey). These devices are not made with natural rubber latex. | |
| Indications for Use<br>(21 CFR §<br>807.92(a)(5)) | The subject device Indications for Use is identical to the predicate BD Cathena™<br>Safety IV Catheter, with the exception that 'catheters' was changed to 'devices'<br>in some cases.<br><br>BD Cathena™ Safety IV Catheters are intended to be inserted into a patient's<br>peripheral vascular system for short term use to sample blood, monitor blood<br>pressure, or administer fluids. These devices may be used for any patient<br>population with consideration given to adequacy of vascular anatomy, procedure<br>being performed, fluids being infused, and duration of therapy. These devices are<br>suitable for use with power injectors set to a maximum pressure of 325 psi (2240<br>kPa) | |
{4}------------------------------------------------
{5}------------------------------------------------
#### bd.com
Technological characteristics of the subject device are substantially equivalent to the subject Technological Characteristics BD Cathena™ Safety IV Catheter achieves its intended use based on the same technology and principles of operation as the predicate device.
> The changes to the device include the introduction of a performance specification (due to a change in material supplier and grade of the needle lubricant from a 2-part silicone material), modification of catheter tubing dimensions for the 18, 20, 22, and 24G catheters, and modifications to product labeling. There were no changes to the product performance specifications as a result of the catheter tubing dimension changes. Biocompatibility evaluation was performed for the change in silicone, performance testing was performed to support the modifications in the catheter tubing.
A comparison of the subject and predicate device technological characteristics is provided in the table below.
| Attribute | SUBJECT<br>BD Cathena™ Safety IV Catheter | PREDICATE (K172506)<br>BD Cathena™ Safety IV Catheter | Comparison |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Classification | 21 CFR §880.5200<br>Class II<br>FOZ - Intravascular Catheter | 21 CFR §880.5200<br>Class II<br>FOZ - Intravascular Catheter | Identical |
| Indication for<br>Use | BD Cathena™ Safety IV Catheters are<br>intended to be inserted into a patient's<br>peripheral vascular system for short term use<br>to sample blood, monitor blood pressure, or<br>administer fluids. These devices may be used<br>for any patient population with consideration<br>given to adequacy of vascular anatomy,<br>procedure being performed, fluids being<br>infused, and duration of therapy. These<br>devices are suitable for use with power<br>injectors set to a maximum pressure of 325 psi<br>(2240 kPa). | BD Cathena™ Safety IV Catheters are intended<br>to be inserted into a patient's peripheral<br>vascular system for short term use to sample<br>blood, monitor blood pressure, or administer<br>fluids. These catheters may be used for any<br>patient population with consideration given<br>to adequacy of vascular anatomy, procedure<br>being performed, fluids being infused, and<br>duration of therapy. The catheters are<br>suitable for use with power injectors set to a<br>maximum pressure of 325 psi (2240 kPa). | Identical, with the exception that 'catheters'<br>was changed to 'devices' in some cases. |
| Attribute | SUBJECT<br>BD Cathena™ Safety IV Catheter | PREDICATE (K172506)<br>BD Cathena™ Safety IV Catheter | Comparison |
| Fundamental<br>Scientific<br>Technology | Peripheral intravascular catheter designed<br>with a passive needlestick safety mechanism<br>and a multi-use blood control septum.<br>Incorporates BD Instaflash™ technology to<br>assist with flashback visualization. | Peripheral intravascular catheter designed<br>with a passive needlestick safety mechanism<br>and a multi-use blood control septum.<br>Incorporates BD Instaflash™ technology to<br>assist with flashback visualization. | Identical |
| Primary<br>Device<br>Components /<br>Materials | Safety Shield<br>Acrylonitrile Butadiene Styrene | Safety Shield<br>Polystyrene | Change in safety shield and silicone lubricant<br>materials |
| | Grip / Needle Hub<br>Polypropylene | Grip / Needle Hub<br>Polypropylene | |
| | Needle<br>Stainless Steel | Needle<br>Stainless Steel | |
| | Needle Lubricant<br>1-part Silicone | Needle Lubricant<br>2-part Silicone | |
| | Catheter Adapter<br>Polypropylene<br>Catheter Tubing<br>Polyurethane with radiopaque barium sulfate | Catheter Adapter<br>Polypropylene<br>Catheter Tubing<br>Polyurethane with radiopaque barium sulfate | |
| Catheter<br>Dimensions | Catheter Diameters<br>16 G, 18 G, 20 G, 22 G, 24 G<br>Catheter Lengths<br>0.75 IN, 1.00 IN, 1.25 IN, 1.75 IN, 2.00 IN | Catheter Diameters<br>16 G, 18 G, 20 G, 22 G, 24 G<br>Catheter Lengths<br>0.75 IN, 1.00 IN, 1.25 IN, 1.75 IN, 2.00 IN | Change in catheter tubing dimensions for the<br>18 G, 20 G, 22 G, and 24 G catheters |
| Shelf-Life | 3 years | 1 year | Change in shelf life from 1 year to 3 years. |
| Attribute | SUBJECT<br>BD Cathena™ Safety IV Catheter | PREDICATE (K172506)<br>BD Cathena™ Safety IV Catheter | Comparison |
| Sterilization<br>Method | EO (SAL 10^-6) | EO (SAL 10^-6) | Identical |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### bd.com
Performance tests completed on the subject device were limited to those tests Summary of Performance required to support a determination of substantial equivalence to the predicate Tests device. A risk analysis was conducted to assess the impact of the proposed modifications to the predicate devices. When technological characteristics between the subject and predicate devices were found to be identical, results of performance testing conducted on the predicate devices were applied to the subject device. The performance tests listed below were conducted to ensure that the subject device meets pre-determined design requirements:
# BD Internal Specification
- . Time to visualize flashback in flash chamber
- Force to break adhesion between catheter unit and needle (initial adhesion) ●
- Force to remove needle from catheter unit (average system drag) .
- Device burst pressure
- Catheter separation force
- Time to visualize flashback in catheter adapter ●
- Procedural leak time
## Standards Compliance
- . Flow rate (ISO 10555-1 Intravascular catheters - Sterile and single-use catheters -Part 1: General requirements)
- . Power injection (ISO 10555-1 Intravascular catheters — Sterile and single-use catheters -Part 1: General requirements)
The subject device complies with sterilization requirements of ISO 11135:2014. Sterilization of Health Care Products – Ethylene Oxide – Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices.
The subject device complies with particulate testing - USP <788> Particulate Matter in Injections.
A biocompatibility evaluation, in accordance with 1) ISO 10993-1:2018, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing and 2) FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (issued June 16, 2016), was conducted. Biocompatibility data was leveraged from the reference device in addition to performing additional endpoints when required.
{9}------------------------------------------------
| | Per design control requirements specified in 21 CFR 820.30, the subject device |
|-------------|-----------------------------------------------------------------------------------|
| | met all predetermined acceptance criteria for the above-listed performance tests, |
| | demonstrating substantial equivalence to the predicate device. |
| Summary of | Based on the indications for use, technological characteristics, and results of |
| Substantial | performance testing, the subject BD Cathena™ Safety IV Catheters has been |
| Equivalence | demonstrated to be substantially equivalent to the predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.