K192460 · Boston Scientific Corporation · PUM · Feb 19, 2020 · Cardiovascular
Device Facts
Record ID
K192460
Device Name
Sentinel Cerebral Protection System
Applicant
Boston Scientific Corporation
Product Code
PUM · Cardiovascular
Decision Date
Feb 19, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1251
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sentinel Cerebral Protection System is indicated for use as an embolic protection device to capture and remove thrombus/debris while performing transcatheter aortic valve replacement procedures. The diameters of the arteries at the site of filter placement should be between 9 - 15 mm for the brachiocephalic and 6.5 - 10 mm in the left common carotid.
Device Story
Sentinel CPS is a 6 French, 95 cm, single-use, percutaneously delivered embolic protection device; inserted via radial or brachial artery. System deploys two filters: proximal filter in brachiocephalic artery; distal filter in left common carotid artery. Captures and removes debris dislodged during TAVR procedures. At procedure completion, filters and captured debris are recaptured into catheter and removed from patient. Used in clinical setting by physicians. Benefits patient by reducing risk of embolic events during TAVR.
Clinical Evidence
Bench testing only. Verification activities included simulated use deployment/retrieval force testing, dimensional verification of filter pore size, tensile characterization of filter film, and shelf-life evaluation. Biocompatibility testing conducted per ISO 10993-1, including cytotoxicity, sensitization, reactivity, systemic injection, pyrogenicity, hemolysis, complement activation, and PTT/platelet/leucocyte testing.
Technological Characteristics
6 French, 95 cm catheter; dual-filter system (proximal 15 mm, distal 10 mm). Materials include polyurethane filter film. Delivered percutaneously via radial or brachial artery. Sterile, single-use. No electronic components or software.
Indications for Use
Indicated for patients undergoing transcatheter aortic valve replacement (TAVR) procedures to capture and remove embolic thrombus/debris. Requires target vessel diameters of 9-15 mm for the brachiocephalic artery and 6.5-10 mm for the left common carotid artery.
Regulatory Classification
Identification
This device is a single use percutaneous catheter system that has (a) blood filter(s) at the distal end. This device is indicated for use while performing transcatheter intracardiac procedures. The device is used to filter blood in a manner that may prevent embolic material (thrombus/debris) from the transcatheter intracardiac procedure from traveling towards the cerebral circulation.
Special Controls
In combination with the general controls of the FD&C Act, the Sentinel® Cerebral Protection System is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Simulated-use testing in a clinically relevant bench anatomic model to assess the following:
(A) Delivery, deployment, and retrieval, including quantifying deployment and retrieval forces, and procedural time; and
(B) Device compatibility and lack of interference with the transcatheter intracardiac procedure and device.
(ii) Tensile strengths of joints and components, tip flexibility, torque strength, torque response, and kink resistance.
(iii) Flow characteristics.
(A) The ability of the filter to not impede blood flow.
(B) The amount of time the filter can be deployed in position and/or retrieved from its location without disrupting blood flow.
(iv) Characterization and verification of all dimensions.
(2) Animal testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be assessed:
(i) Delivery, deployment, and retrieval, including quantifying procedural time.
(ii) Device compatibility and lack of interference with the transcatheter intracardiac procedure and device.
(iii) Flow characteristics.
(A) The ability of the filter to not impede blood flow.
(B) The amount of time the filter can be deployed in position and/or retrieved from its location without disrupting blood flow.
(iv) Gross pathology and histopathology assessing vascular injury and downstream embolization.
(3) All patient contacting components of the device must be demonstrated to be biocompatible.
(4) Performance data must demonstrate the sterility of the device components intended to be provided sterile.
(5) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(6) Labeling for the device must include:
(i) Instructions for use;
(ii) Compatible transcatheter intracardiac procedure devices;
(iii) A detailed summary of the clinical testing conducted; and
(iv) A shelf life and storage conditions.
(7) Clinical performance testing must demonstrate:
(i) The ability to safely deliver, deploy, and remove the device;
(ii) The ability of the device to filter embolic material while not impeding blood flow;
(iii) Secure positioning and stability of the position throughout the transcatheter intracardiac procedure; and
(iv) Evaluation of all adverse events including death, stroke, and vascular injury.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
February 19, 2020
Boston Scientific Corporation Eric Elliott Regulatory Affairs Manager Three Scimed Place Maple Grove, Minnesota 55311-1566
Re: K192460
Trade/Device Name: Sentinel Cerebral Protection System Regulation Number: 21 CFR 870.1251 Regulation Name: Temporary Catheter for Embolic Protection During Transcatheter Intracardiac Procedures Regulatory Class: Class II Product Code: PUM Dated: January 17, 2020 Received: January 21, 2020
Dear Mr. Elliott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jaime Raben, Ph.D. Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K192460
Device Name Sentinel Cerebral Protection System
Indications for Use (Describe)
The Sentinel Cerebral Protection System is indicated for use as an embolic protection device to capture and remove thrombus/debris while performing transcatheter aortic valve replacement procedures. The diameters of the arteries at the site of filter placement should be between 9 - 15 mm for the brachiocephalic and 6.5 - 10 mm in the left common carotid.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary per 21 CFR §807.92
| Sponsor | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA |
|---------------------------------|-------------------------------------------------------------------------------------------------------|
| Contact Name and<br>Information | Eric Elliott<br>Three Scimed Place<br>Maple Grove, MN 55311-1566<br>Phone: 763-494-1654 |
| Date Prepared | September 06, 2019 |
| Proprietary Name | Sentinel™ Cerebral Protection System |
| Common Name | Temporary catheter, embolic protection, transcatheter<br>intracardiac procedures |
| Product Code | PUM |
| Classification | Class II, 21 CFR Part 870.1251 |
| Predicate Device | Sentinel Cerebral Protection System<br>DEN160043 June 1, 2017 |
### Device Description
The Sentinel™ Cerebral Protection System (Sentinel CPS) is a 6 French, 95 cm working length, single use, temporary, percutaneously delivered embolic protection device, inserted into the radial or brachial artery. The system is designed to capture and remove debris dislodged during transcatheter aortic valve replacement (TAVR) procedures. The Sentinel CPS utilizes an embolic filter delivered to the brachiocephalic artery (Proximal Filter) and a second embolic filter delivered to the left common carotid artery (Distal Filter). At the completion of the procedure, the filters and debris are recaptured into the catheter and removed from the patient. The device is provided sterile and is single-use only.
The Sentinel CPS is available with a Proximal Filter size of 15 mm (target vessel size of 9 - 15 mm) and a Distal Filter size of 10 mm (target vessel size of 6.5 - 10 mm).
### Indications for Use / Intended Use
The Sentinel Cerebral Protection System is indicated for use as an embolic protection device to capture and remove thrombus/debris while performing transcatheter aortic valve replacement procedures. The diameters of the arteries at the site of filter placement should be between 9 – 15 mm for the brachiocephalic and 6.5 – 10 mm in the left common carotid.
{4}------------------------------------------------
## Comparison of Technological Characteristics
Replacement of filter film polyurethane material with a similar polyurethane is required due to material obsolescence by the supplier. Comparisons of the new and currently marketed predicate device cleared under DEN160043 show that the technological characteristics such as filter performance, materials, design, sterilization, packaging, and intended use/ indications for use are substantially equivalent. The subject Sentinel Cerebral Protection System indications for use, design, and principles of operation are identical to the currently marketed predicate cleared under DEN160043.
### Non-clinical Performance Data
Modifications to the predicate device were assessed according to risk-based failure mode effects analysis and with consideration of FDA Guidance. Coronary and Carotid Embolic Protection Devices - Premarket Notification [510(k)] Submission (issued February 15, 2008). The following non-clinical testing was successfully completed on the modified device.
- Simulated Use Proximal and Distal Filter Deployment Force ●
- Simulated Use Proximal and Distal Filter Retrieval Force .
- Dimensional Verification Filter Pore Size ●
- Tensile Characterization Filter Film .
- . Shelf Life Evaluation
Testing demonstrated that the modified Sentinel Cerebral Protection System met all previously established acceptance criteria. No new safety or effectiveness issues were raised during verification and validation activities; thereby supporting a determination of substantial equivalence.
Biocompatibility testing was also assessed with consideration of evaluation recommendations per ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process and FDA Guidance on ISO 10993-1 (issued June 16, 2016) for devices categorized as externally communicating with limited (<24 hours) direct circulating blood contact. The following is a list of biocompatibility tests conducted on the modified Sentinel Cerebral Protection System.
- MEM Elution Cytotoxicity .
- . Guinea Pig Maximization Sensitization
- Intracutaneous Reactivity ●
- Acute Systemic Injection ●
- . Material Mediated Rabbit Pyrogen
- Hemolysis (direct and extract methods) ●
- . Complement Activation (SC5b-9 method)
- USP Physiochemical test <661> ●
- Natural Rubber Latex ELISA Inhibition Assay for Antigenic Protein ●
- Partial Thromboplastin Time (PTT) ●
- Platelet/Leucocyte testing
Results confirm that the modified device remains biocompatible for its intended use.
{5}------------------------------------------------
## Conclusion
Successful design verification and biocompatibility testing support substantial equivalency of the Sentinel Cerebral Protection System to the currently marketed predicate device. There were no new safety or effectiveness issues raised during verification and testing activities.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.