The Wattson™ temporary pacing guidewire is intended to introduce and position catheters and other interventional devices within the chambers of the heart, including those used within transcatheter aortic valve replacement (TAVR) procedures and balloon aortic valvuloplasty (BAV), while transmitting an electrical signal from an external pulse generator to the heart. The temporary pacing guidewire is not intended to remain in place following the clinical procedure.
Device Story
Dual-purpose 0.035" guidewire; facilitates catheter delivery and temporary rapid cardiac pacing. Features distal pigtail shape; bipolar electrode configuration (distal tip positive, three proximal negative electrodes). Includes proprietary adapter converting proximal end to terminal pins for standard external pulse generator connection. Used in cardiac catheterization labs during TAVR/BAV procedures; operated by interventionalists. Provides guidewire support for device delivery; transmits pacing signals to induce rapid pacing (reducing blood pressure for valve placement). Fluoroscopically visible. Benefits include procedural efficiency by combining guidewire support and pacing functions in one device.
Clinical Evidence
Prospective, single-center study of 20 subjects undergoing TAVR. Primary endpoints: guidewire support and pacing reliability. Results: device provided predictable support and reliable bipolar pacing at low thresholds, facilitating successful heart valve delivery. No clinical complications reported.
Technological Characteristics
Stainless steel wire; fluoropolymer (FEP) jacket; stainless steel electrodes. 0.035" OD; 280 cm length. Bipolar electrode configuration. Ethylene oxide sterilization. Single-use. Radiopaque. Connects to external pulse generators via proprietary adapter.
Indications for Use
Indicated for patients undergoing TAVR or BAV procedures requiring catheter positioning and temporary cardiac pacing. Contraindicated for use in coronary arteries and cerebrovasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
January 15, 2020
Vascular Solutions Nancy Frame Sr. Regulatory Product Specialist 6464 Sycamore Court North Minneapolis, Minnesota 55369
Re: K192454
Trade/Device Name: Wattson Temporary Pacing Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: December 18, 2019 Received: December 20, 2019
Dear Nancy Frame:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hetal Patel Assistant Director Implantable Electrophysiology Devices Team Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192454
#### Device Name Wattson™ temporary pacing guidewire
#### Indications for Use (Describe)
The Wattson™ temporary pacing guidewire is intended to introduce and position catheters and other interventional devices within the chambers of the heart, including those used within transcatheter aortic valve replacement (TAVR) procedures and balloon aortic valvuloplasty (BAV), while transmitting an electrical signal from an external pulse generator to the heart. The temporary pacing guidewire is not intended to remain in place following the clinical procedure.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> |
|----------------------------------------------|-------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <input type="checkbox"/> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Vascular Solutions. The logo features a square with a gradient of black squares above it, resembling a stylized image of smoke. Below the square is the text "Vascular" in a bold, serif font, with the word "SOLUTIONS" in a smaller, sans-serif font underneath.
# 510(k) Summary
[As required by 21 CFR 807.92]
510(k) NUMBER: K192454
### SUBMITTER
Vascular Solutions 6464 Sycamore Court North Minneapolis, MN 55369 USA Establishment Registration # 2134812 Phone: 763.656.4230 (direct) Fax: 763.251.0363 Contact Person: Nancy Frame, Sr. Regulatory Product Specialist Date Prepared: September 5, 2019
### DEVICE
Name of Device: Wattson™ temporary pacing guidewire Common or Usual Name: Catheter Guidewire Classification Name: Catheter Guidewire (21 CFR 870.1330) Regulatory Class: II Primary Product Code: DQX Secondary Product Code: LDF
### PREDICATE DEVICES
### Primary Predicate:
Manufacturer: Medtronic Device Name: Confida™ /Breaker Guidewire 510(k) No: K181001 (cleared May 3, 2018), Class II, Product Code DQX; classified per 21 CFR 870.1330
{4}------------------------------------------------
### Secondary Predicate:
Manufacturer: Bard
Device Name: Bard® Temporary Pacing Electrode Catheter
510(k) No: K800298 (cleared April 16, 1980). Class II, Product Code LDF; classified per 21 CFR 870.3680
# DEVICE DESCRIPTION
The Wattson temporary pacing guidewire is a dual-purpose 0.035" guidewire designed for the delivery of devices and for temporary rapid pacing of the heart. The distal end of the device has an atraumatic pigtail shape. The shaft of the device has a silicone lubricant applied to the outer surface.
Rapid pacing is achieved through a bipolar electrode configuration integrated in the shaft of the guidewire. The electrodes terminate in a single positive electrode at the distal tip of the device and three discrete negative electrodes located proximal to the tip within the pigtail section. The entire device is visible using standard fluoroscopic methods.
The device is packaged with a proprietary adapter that locks on to the proximal end of the guidewire and converts it to discrete positive and negative terminal pins. The terminal pins of the adapter are compatible with standard external pulse generators.
The Wattson temporary pacing guidewire is sterilized with ethylene oxide.
# INDICATIONS FOR USE
The Wattson temporary pacing guidewire is intended to introduce and position catheters and other interventional devices within the chambers of the heart, including those used within transcatheter aortic valve replacement (TAVR) procedures and balloon aortic valvuloplasty (BAV), while transmitting an electrical signal from an external pulse generator to the heart. The temporary pacing guidewire is not intended to remain in place following the clinical procedure.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
This submission is seeking the clearance of the Wattson temporary pacing guidewire which has similar uses, fundamental technologies, principals of operation and performance as the identified predicate devices. The Wattson temporary pacing guidewire is similar to the Confida Brecker wire in that it is used to introduce and position catheters during diagnostic and interventional procedures within the chambers of the heart. It is similar to the Bard Temporary Pacing catheter in that it is used to transmit an electrical signal from an external pulse generator to the heart. Where there are technological differences, results of bench and animal tests confirm that the Wattson temporary pacing guidewire is substantially equivalent to the predicate devices. A comparison of the technological characteristics for the Wattson temporary pacing guidewire against the predicate devices is as follows:
{5}------------------------------------------------
### Substantial Equivalence Comparison
| Characteristic | Subject Device | Primary Predicate<br>Device | Secondary Predicate<br>Device |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Wattson temporary<br>pacing guidewire | Confida Guidewire | Temporary Pacing<br>Electrode Catheter |
| 510(k) | K19XXXXX | K181001 | K800298 |
| Manufacturer | Vascular Solutions | Medtronic | Bard |
| Product Class | II | II | II |
| Product<br>Classification | 870.1330 | 870.1330 | 870.3680 |
| Product Code | DQX; LDF | DQX | LDF |
| Indications for Use | The WattsonTM<br>temporary pacing<br>guidewire is intended to<br>introduce and position<br>catheters and other<br>interventional devices<br>within the chambers of<br>the heart, including<br>those used within<br>transcatheter aortic<br>valve replacement<br>(TAVR) procedures and<br>balloon aortic<br>valvuloplasty (BAV),<br>while transmitting an<br>electrical signal from an<br>external pulse generator<br>to the heart. The<br>temporary pacing<br>guidewire is not<br>intended to remain in<br>place following the<br>clinical procedure. | The Medtronic<br>ConfidaTM Brecker<br>guidewire is intended<br>for use to introduce<br>and position catheters<br>during diagnostic and<br>interventional<br>procedures within the<br>chambers of the heart,<br>including transcatheter<br>aortic valve<br>replacement (TAVR). | Bard® Temporary Pacing<br>Catheters are designed to<br>transmit an electrical<br>signal from an external<br>pulse generator to the<br>heart or from the heart to<br>a monitoring device.<br>When an internal lumen<br>is present (other than the<br>one used for balloon<br>inflation), it may be used<br>for fluid infusion,<br>pressure monitoring, or<br>blood sampling. |
| Contraindication | The guidewire is<br>contraindicated for use<br>in the coronary arteries<br>and in the<br>cerebrovasculature. | The Medtronic<br>ConfidaTM Brecker<br>Guidewire is<br>contraindicated for<br>patients presenting<br>with an intolerance to<br>anticoagulation<br>therapy and | None |
| Characteristic | Subject Device | Primary Predicate<br>Device | Secondary Predicate<br>Device |
| | Wattson temporary<br>pacing guidewire | Confida Guidewire | Temporary Pacing<br>Electrode Catheter |
| | | unheparinized<br>patients. The<br>guidewire is<br>contraindicated for use<br>in the coronary arteries<br>and in the<br>cerebrovasculature. | |
| Anatomical sites | Left Ventricle | Left Ventricle | Right Ventricle |
| Materials of<br>construction | Stainless Steel Wire<br>with fluoropolymer<br>jacket and stainless-steel<br>electrodes | Stainless Steel wire<br>with fluoropolymer<br>coating. | Woven or extruded<br>polyurethane jacketed<br>shaft with platinum or<br>stainless-steel electrodes |
| Length | 280 cm | 260 cm | 100, 110, 115, 125 cm |
| Wire OD | 0.035"/0.89 mm | 0.035"/0.89 mm | N/A |
| Method of<br>Sterilization | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| Single Use or<br>Reusable | Single Use | Single Use | Single Use |
| Coating | Fluropolymer outer<br>Jacket (FEP) with<br>Silicone Oil Lubricant. | Fluoropolymer (PTFE)<br>coating | Polyurethane Outer<br>Jacket |
| Radiopacity | Yes | Yes | Yes |
| Tip Configuration | Pigtail | Preformed 360°<br>curved tip (pigtail) | Straight, Preshaped<br>angle tip |
| External<br>Pacemaker<br>Connections | Removable guidewire<br>adapter terminating in<br>two shrouded<br>positive/negative<br>connectors | N/A | Two incorporated<br>shrouded<br>positive/negative<br>connectors. |
| Pulse Generator<br>compatibility | Compatible with<br>standard external pulse<br>generators | N/A | Compatible with<br>standard external pulse<br>generators |
{6}------------------------------------------------
{7}------------------------------------------------
Page 5 of 6
### PERFORMANCE DATA
## Biocompatibility Testing
The biocompatibility evaluation for the Wattson temporary pacing guidewire was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The Wattson temporary pacing guidewire is considered an externally communicating device in contact with circulating blood and tissue for a limited period of time (<24 hours) during use. The battery of tests included the following:
- Cytotoxicity
- Sensitization
- Irritation ●
- Acute Systemic Toxicity
### Performance Testing- Bench
The device design was verified through the following tests:
- Dimensional Verification .
- Visual Inspection ●
- Tensile Strength
- Track Force
- Guidewire Support Profile
- Distal Flex Force
- Shaft Friction Force ●
- Electrode Size
- Electrode Location
- Device Continuity ●
- Particulate Evaluation ●
- Corrosion Resistance ●
- Radiopacity ●
### Performance Data – Animal:
Animal testing was performed to demonstrate the substantial equivalence of the Wattson temporary pacing guidewire to the predicates Confida Breaker Guidewire and Bard Temporary Pacing Catheter in a porcine model. The testing was performed in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP) for Non-Clinical Laboratory Studies. The subject Wattson temporary pacing guidewire shares the same fundamental scientific technology as the predicate devices.
A GLP study demonstrated that the Wattson temporary pacing guidewire and the predicate (control) devices performed similarly with regard to adverse events and other animal health concerns. No animals experienced sustained arrhythmia in the post-rapid pacing monitoring period, all animals returned to normal sinus rhythm after rapid pacing was turned off, and there were no histological findings that indicate any differences between the test and control group.
-
- Pyrogenicity ●
- . Hemolysis
- Complement Activation ●
- Thrombogenicity ●
{8}------------------------------------------------
A second GLP study successfully demonstrated that the Wattson temporary pacing guidewire and the predicate (control) devices are similar in performance and handling during simulated clinical use in the swine model. In all cases: the Wattson wire provided a clinically significant reduction in blood pressure to adequately allow for successful BAV inflation; and the interventionalist was able to properly position the device, successfully deliver the BAV device and successfully connect to the external pacing programmer. Animals were monitored for 30 minutes post-procedure and demonstrated a return to normal sinus rhythm and were free of any concerning cardiac events.
# Performance Testing - Clinical:
Twenty subjects underwent TAVR procedures and were treated with the Wattson temporary pacing guidewire in a single-center, pre-market, prospective study.
The Wattson temporary pacing guidewire offered predictable guidewire support with concomitant reliable bipolar pacing at low thresholds to allow safe transcatheter heart valve delivery in the treated cohort. In this experience the Wattson wire performance indicates the device has the potential to make TAVR a more efficient procedure in patients.
# CONCLUSIONS
The results of bench and animal tests confirm that the Wattson temporary pacing guidewire, subject of this 510(k), is substantially equivalent to the predicate devices. The Wattson temporary pacing guidewire is similar to the predicate devices in its indications for use and substantially equivalent in technology, materials, and performance to the predicates. Performance data demonstrates that the Wattson temporary pacing guidewire performs as well as the Confida Breaker Guidewire and Bard Temporary Pacing Catheter. Therefore, it can be concluded that the Wattson temporary pacing guidewire is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.