Low-speed (vane motor) handpiece is used for drilling of teeth in a variety of dental procedures.
Device Story
Midwest Rhino XE and Air Motor M are reusable, air-powered, low-speed vane motors used for drilling teeth in dental procedures. Operated by dental professionals in clinical settings, these devices connect to dental handpieces via ISO-couplers or Midwest 4-hole/Borden 2/3-hole couplings. The motors transform pneumatic energy into mechanical rotation via planetary gearing to drive dental burs. Air Motor M includes an optional external spray port for irrigation. Devices are steam-autoclave sterilizable. Output is mechanical rotation used for tooth preparation, removal of decay, and finishing/polishing. Clinical benefit is the facilitation of standard dental restorative and surgical procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including performance verification per ISO 14457, sterilization validation per ISO 17665-1/ANSI/AAMI ST79, cleaning validation per AAMI TIR 30/ISO 15883-5, and biocompatibility assessment per ISO 10993-1 and ISO 7405.
Technological Characteristics
Air-powered vane motor; planetary gearing; steam autoclave sterilization; ISO-coupler, Midwest 4-hole, or Borden 2/3-hole coupling; external spray port (Air Motor M only). Compliant with ISO 14457, ISO 9168, ISO 3964, and ISO 13504.
Indications for Use
Indicated for drilling of teeth in a variety of dental procedures. No specific patient population (age/gender) or contraindications provided.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 13, 2020
Dentsply Sirona Karl Nittinger Vice President, Corporate Regulatory Affairs 221 W Philadelphia Street. Suite 60W York, Pennsylvania 17401
Re: K192409
Trade/Device Name: Midwest Rhino XE and Air Motor M Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: Class I, reserved Product Code: EFB Dated: October 14, 2019 Received: October 15, 2019
Dear Karl Nittinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Midwest Rhino XE and Air Motor M
Indications for Use (Describe)
Low-speed (vane motor) handpiece is used for drilling of teeth in a variety of dental procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
Image /page/3/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a stylized, abstract shape on the left, resembling a curved, segmented form. To the right of the shape are the words "Dentsply" on the top line and "Sirona" on the bottom line, with the two words stacked vertically. The text and the shape are in a dark gray color.
## SECTION 5. 510(k) SUMMARY for
Midwest Rhino XE and Air Motor M (K192409)
Submitter Information:
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
Contact Person: Karl Nittinger Telephone Number: 717-849-4424 Fax Number: 717-849-4343 Email: karl.nittinger@dentsplysirona.com
872.4200
-
EFB
Date Prepared:
January 13, 2020
Midwest Rhino XE and Air Motor M
Dental Handpiece and Accessories
Handpiece, Air-Powered, Dental
### Device Name:
- Proprietary Name: ●
- Common Name: .
- Classification Name: ●
- CFR Number: .
- Device Class:
- Product Code: .
## Predicate Device:
| Predicate Device Name | 510(k) | Company Name |
|---------------------------------------------------------------------------------------|---------|---------------------|
| LS-100 | K792302 | Dentsply Sirona |
| Reference Device Name | 510(k) | Company Name |
| W&H Air-Powered<br>Handpieces and Handpiece<br>Attachments (Air Motor AM-<br>25-E RM) | K162926 | W&H Dentalwerk GmbH |
#### Description of Device:
The Midwest Rhino XE and Air Motor M are low speed air motors. The motors are reusable and can be sterilized in an autoclave. Air Motor M includes an external spray port whereas the Midwest Rhino XE does not have the external sprayport.
## Indications for Use:
Low-speed (vane motor) handpiece is used for drilling of teeth in a variety of dental procedures.
{4}------------------------------------------------
### Substantial Equivalence:
The subject devices have identical indications for use as the predicate device, LS-100 (K792302). Table 5.1 show a comparison of the indications for use and technological features of the proposed devices and the predicate device.
| Element | Proposed Devices | | Predicate Device | Reference Device | Differences | |
|--------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|------------------------------|---------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Midwest<br>Rhino XE | Air Motor<br>M | LS-100 (K792302) | W&H Air-Powered Handpieces and<br>Handpiece Attachments (Air Motor<br>AM-25 E RM) (K162926) | | |
| Indications<br>for Use | Low-speed (vane motor)<br>handpiece is used for<br>drilling of teeth in a variety<br>of dental procedures. | | Low-speed (vane motor)<br>handpiece is used for<br>drilling of teeth in a<br>variety of dental<br>procedures. | The turbine handpiece is intended<br>for the following applications:<br>removal of decayed materials,<br>cavities and crown preparation,<br>removal of fillings, finishing of tooth<br>and restoration surfaces.<br>The dental handpiece/contra-angle is<br>intended for the following applications:<br>removal of decayed materials, cavities<br>and crown cement, removal of fillings,<br>finishing and polishing of tooth and<br>restorations surfaces. | Identical for the predicate device as<br>used in a variety of dental activities.<br>Similar to the reference device<br>indications for use for which are detailed<br>as to the specific application the device<br>is used. | |
| Operational<br>Mode | Air-powered | | Air-powered | Air-powered | Identical | |
| Operating<br>Pressure (psi) | 42 $ ± $ 2.9 | | 40-45 | 42 | There is no significant difference in<br>the operating pressure of the<br>proposed air motors when comparing<br>them to the predicate air motor. The<br>proposed devices fall within the<br>cleared range of the<br>predicate device. | |
| Motor Speed<br>(rpm) | 0 - 24,000 | $ ≥ $ 18,000 | 0-8,000 | 2.2 bar: 5,000-20,000*<br>3 bar: 5,000-25,000* | The difference of the motor speed<br>comes through the planetary gearing<br>of the proposed Midwest Rhino XE<br>and Air Motor M. Conformity testing to<br>the requirements of ISO 14457<br>requirements are included to support<br>substantial equivalence. | |
| Table 5.1- Comparison of proposed devices and the predicate device | | | | | | |
| Element | | Proposed Devices | Predicate Device | Reference Device | Differences | |
| | | Midwest Rhino<br>XE | Air Motor M | LS-100<br>(K792302) | W&H Air-Powered Handpieces and<br>Handpiece Attachments (Air Motor<br>AM-25 E RM) (K162926) | |
| Torque (N/cm) | | ~2.0 | ~2.1 | ~6 | 4 | The difference of the motor torque<br>comes through the planetary gearing<br>of the Midwest Rhino XE and Air<br>Motor M. It changes speed into<br>torque. Conformity testing to the<br>requirements of ISO 14457<br>requirements are included to support<br>substantial equivalence. |
| External<br>Spray Port | | NA | Yes | NA | Yes | The external spray port in the design<br>of the proposed Air Motor M is an<br>optional feature to facilitate<br>connection of dental instruments with<br>an external spray connection. |
| Spray Air<br>Pressure in<br>bar (psi) | | NA | $2.3 \pm 0.2$<br>$33.4 \pm 2.9$ | NA | 2.2-3* | |
| Spray Water<br>Pressure in<br>bar (psi) | | NA | $2.0 \pm 0.2$<br>$29.0 \pm 2.9$ | NA | 1.5-2.5* | |
| Sterilization | | Steam Autoclave | | Steam Autoclave | Steam Autoclave | Identical |
| Lubricant | | Sirona T1 Spray<br>(K150750) | Sirona T1<br>Spray<br>(K150750) | Sirona T1 Spray<br>(K150750) | W&H Service Oil F1, MD-400 | Identical |
| Type of Chuck | | ISO-Coupler | ISO-Coupler | Midwest-Coupler | ISO 3964 connection | The proposed devices are identical to<br>the predicate device. |
| Coupling | | Midwest 4-hole<br>coupling | Midwest 4-hole<br>coupling<br>Borden 2/3-<br>hole coupling | Midwest 4-hole<br>coupling | Fixed connection standard 4-hole* | The Air Motor M is also available in<br>version with a Borden 2/3-hole<br>coupling. This coupling is another type<br>of connection that can be used to<br>connect the<br>handpiece to the air motor hose. |
{5}------------------------------------------------
{6}------------------------------------------------
Non-Clinical Performance Data.
Testing to verify the performance requirements of the proposed devices, Midwest Rhino XE and Air Motor M, was conducted and included in this premarket notification. The results of the performance testing support substantial equivalence.
Tests included in this premarket notification verify the conformity of the proposed devices, Midwest Rhino XE and Air Motor M, with the requirements of:
- EN ISO 17665-1- Sterilization of health care products-Moist heat Part 1: ● Requirements for the development, validation and routine control of a sterilization process for medical devices.
- ANSI/AAMI ST79:2010 and A1:2010 and A2:2011 and A3:2012 and A4:2013-.
Comprehensive guide to steam sterilization and sterility assurance in health care facilities.
- Reprocessing validation per the FDA Guidance Document entitled, ● "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" dated March 17, 2015
- Cleaning Process Validation per AAMI TIR 30 and ISO 15883-5 ●
- ISO 14457:2-17-10- Dentistry-Handpiece and motors ●
- ISO 9168-Dental Handpieces-Hose Connections .
- ISO 3964-Dental Handpiece-Coupling dimensions .
- Biocompatibility equivalency with the reference device (K150750) as far as . material composition is included in support of substantial equivalence. The equivalency statement is made in reference to the requirements of ISO 10993-1:2010-Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and ISO 7405:2013-Dentistry-Evaluation of biocompatibility ofmedical devices used in dentistry.
- ISO 13504:2012-Dentistry-General requirements for instruments and related ● accessories used in dental implant placement and treatment
- Guidance for Industry and FDA Staff-Dental Handpieces-Premarket . Notification [510(k)] Submissions
## Risk Analvsis
Risk analysis was performed on the proposed devices based on ISO 14971. The results of the risk analysis performed concluded that all device design controls and process controls will be able to mitigate known potential failures andeffect.
#### Clinical Performance Data.
No data from human clinical studies has been included to support the substantial equivalence of the proposed Midwest Rhino XE and Air Motor M handpieces.
#### Conclusion Regarding Substantial Equivalence
The proposed Midwest Rhino XE and Air Motor M handpieces are air-powered dental handpieces which are used for drilling of teeth in a variety of dental procedures. The proposed Midwest Rhino XE and Air Motor M handpieces have identical indications for use as those cleared for the predicate device, LS-100 (K792302). The proposed devices incorporate the same fundamental technology and have same intended uses as the predicate device, LS-100 (K792302) and the reference device, W&H Air-Powered Handpieces and Handpiece Attachments (Air Motor AM-25 E RM) (K162926). Test data according to ISO 14457:2017-10 to verify the performance of the Midwest Rhino XE and Air Motor M handpieces has been provided and the results of this testing, combined with the design and intended use comparison with the predicate device, support substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.