K192406 · Medwaves, Inc. · NEY · Nov 25, 2020 · General, Plastic Surgery
Device Facts
Record ID
K192406
Device Name
AveCure BT Microwave Ablation System
Applicant
Medwaves, Inc.
Product Code
NEY · General, Plastic Surgery
Decision Date
Nov 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AveCure BT Microwave Ablation System is intended for: - Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body. - Coagulation and ablation of tissue in bone during surgical procedures, including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy.
Device Story
System uses 915MHz microwave energy to induce coagulation necrosis in bone tissue. Components include a generator-controller and sterile, single-patient-use antenna probes. Operator selects antenna size, temperature set-point (60-90°C), and duration (0-450s) via interface. During ablation, system automatically modulates power, frequency, and duration based on real-time temperature feedback from an antenna-integrated sensor. Frequency is adjusted ~1Hz to minimize reflected power, ensuring efficient energy transfer. No active cooling required due to low transmission loss. Used by clinicians in surgical settings to treat metastatic lesions. Output is thermal energy; clinical benefit is pain palliation through targeted tissue ablation.
Clinical Evidence
Bench testing only. Comparison of ablation performance (volume, length, width) between subject MWA and predicate RFA systems in bone. Results demonstrate equivalent ablation dimensions (e.g., max volume 7260mm³ vs 7018mm³) with significantly shorter cycle times (up to 50% reduction). No clinical data provided.
Technological Characteristics
Microwave generator (915MHz ±13MHz ISM band); max 40W power. Probes feature distal antennas with integrated temperature sensors. Connectivity: standalone generator with wired probe/cable interface. Materials: biocompatible (ISO 10993-1). Sterilization: probes/accessories sterile (ISO 11135-1). Software: firmware-controlled ablation parameters. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-2-6, ISO 14971.
Indications for Use
Indicated for palliative treatment of metastatic malignant lesions in vertebral bodies and coagulation/ablation of bone tissue for pain palliation in patients with metastatic bone lesions who failed or are not candidates for standard therapy.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Baylis Medical OsteoCool® V-3 RF Ablation System (K161949)
Reference Devices
MedWaves AveCure Microwave Ablation System (K143203)
MedWaves Microwave Coagluation/Ablation System (K070356)
Submission Summary (Full Text)
{0}------------------------------------------------
November 25, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MedWaves, Inc. Theodore Ormsby President/CEO 16760 West Bernardo Drive San Diego, California 92127
Re: K192406/S002
Trade/Device Name: AveCure BT Microwave Ablation System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NEY Dated: October 20, 2020 Received: October 27, 2020
Dear Mr. Ormsby:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192406
Device Name AveCure BT Microwave Ablation System
#### Indications for Use (Describe)
The AveCure BT Microwave Ablation System is intended for:
- Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.
- · Coagulation and ablation of tissue in bone during surgical procedures, including palliation of pain associated with
metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:120%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:120%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for MedWaves Incorporated. The text "MedWaves" is in large, red, stylized font. Below that, the word "INCORPORATED" is in smaller, blue font. To the right of the text are red, curved lines that resemble radio waves.
## 510(k) Summary
16760 W Bernardo Dr San Diego, CA 92127
#### Date: November 25, 2020
Type of Submission: Traditional 510(k) Notification
#### 510(K) submitter information:
MedWaves Incorporated 16760 West Bernardo Drive San Diego, CA 92127 Phone: 858-946-0015 Fax: 858-946-0016
Contact: Theodore Ormsby President/CEO 760-807-1000 Phone: Email: tedormsby@avecure.com
| Regulation: | 21 CFR 878.4400. Electrosurgical Cutting and Coagulation Device and Accessories |
|----------------------|-------------------------------------------------------------------------------------------|
| Panel: | General & Plastic Surgery |
| The Device: | |
| Trade name: | AveCure BT Microwave Ablation System |
| Common name: | Microwave ablation system and accessories |
| Classification name: | Class II, 21 CFR 878.4400, Electrosurgical Cutting and Coagulation Device and Accessories |
| Product Code: | NEY |
The MedWaves' AveCure BT Microwave Ablation System is comprised of a microwave (915MHz ±13MHz) generator-controller, two antenna sizes mounted on the distal end of probes or catheters of various lengths, and flexibility, and a set of extension cables, one for microwave (MW) transmission and another for information communication between the antenna and the generator. The accessories (probes, catheters and extension-cable set) are supplied sterile and the generator-controller is the durable component of the system and is supplied non-sterile.
# Predicate Devices: Baylis Medical OsteoCool® V-3 RF Ablation System: K161949: Class II, 21 CFR 878.4400, GEI
## Reference Devices: MedWaves AveCure Microwave Ablation System MedWaves Microwave Coagluation/Ablation System
K143203: Class II, 21 CFR 878.4400, NEY K070356: Class II, 21 CFR 878.4400, NEY
## Indications for Use:
The AveCure BT Microwave Ablation System is intended for:
- Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.
- . Coagulation and ablation of tissue in bone during surgical procedures, including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for MedWaves Incorporated. The word "MedWaves" is written in red, with the "M" being slightly larger than the other letters. Below "MedWaves" is the word "INCORPORATED" in smaller, blue letters. To the right of the word "MedWaves" are three red concentric circles, resembling radio waves.
16760 W Bernardo Dr San Diego, CA 92127
## Technology Characteristics
MedWaves AveCure® BT Ablation System uses MW energy to treat soft-tissue tumor or tumor-like lesions. The system consists of MW generator-controller operating in the frequency range of 902MHz to 928MHz (915MHz ±13MHz ISM band), array of probes equipped with distally disposed antennas tuned to transfer energy to adjacent tissue in contact with it, and accessory extension cables connected them together. The microwave generator-controller is considered durable and the probes and accessories, which are provided sterile are considered single patient use and are disposed following a treatment. The generator-controller produces and manages appropriate power-train levers and durations to reach and maintain programmed temperature of tissue in contact with an antenna for the treatment durations. The AveCure BT Ablation System is based upon the systems cleared in K070356 and K143203, but only allows users to select from two antenna sizes with limited temperature settings.
The generator-controller is powered by 47-63Hz (50-60Hz ±3Hz), 100-240VAC international current range. It generates and controls up to 40 watts of microwave (MW) power signal from 902 MHz (915MHz ±13MHz ISM band). The generator-controller is composed of digital and analog circuits, and control firmware for user-interface and ablation control factors such as temperature, power, frequency and duration. In standby mode, the operator can select antenna size (mini, small) being used, and the ablation parameters, such as, temperature set-point (60°C - 90°C) and duration (0 - 450seconds). During the ablation processes, MW power-train duration, frequency and level are automatically modulated to achieve and maintain tissue temperature in contact with antenna to a programmed set-point based upon the user input and feedback from a temperature sensor located at the antenna. The power frequency is scanned adjusted approximately once a second to minimize reflected-power (reverse-power) within the band for maximum antenna-tissue energy transfer. Since its MW transmission loss is small, it requires relatively low power level, up to 40 watts, and hence, no need for a cooling mechanism. Therefore, the generator-controller is not blinded from seeing the tissue temperature at the antenna for safe and effective ablation processes.
The single-patient use antenna-probe assemblies are equipment with temperature sensor at the antenna for feedback to the generator-controller for power control. The antenna is used to detect reflected (reverse) power feedback signal during an ablation for generator-controller to use to select the frequency for optimum energy delivery to the nearby tissue. The generator-controller scans the allocated ISM band spectrum to select a frequency which minimizes the reverse-power levels for the most efficient energy transfer from the antenna to targeted tissue as the tissue conductivity and permittivity are modified from the temperature rise due to microwave power absorption, electrical effect and electromagnetic field interacting with the tissue.
## Substantially Equivalent Discussion:
The features of the MedWaves AveCure BT Microwave Ablation (MWA) System are substantially equivalent to the predicate devices by Baylis Medical OsteoCool V-3 RFA System (K161949). Both systems deliver energy to a volume of targeted soft-tissue surrounding an active tip to raise its temperature for coagulate-ablation necrosis effect under temperature feedback control. The indication for use of the subject system is identical to the K161949. The ablation performance of the subject device is substantially equivalent to the predicate K161949. The subject device results in similar performance (tissue necrosis dimensions) in shorter cycles compared to the predicated device. Although the subject device comes in
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for MedWaves Incorporated. The logo is red and features the company name in a stylized font. To the right of the name are three curved lines that resemble radio waves.
16760 W Bernardo Dr San Diego, CA 92127
two antenna sizes versus four electrode sizes offered by the predicate device, the similar ablation performance dimensions can be obtained by overlapping applications. The subject device offers a greater variety of shaft length and flexibility. The subject system is based on the reference devices cleared under K143203 and K070356; it utilizes a subset of the antenna selections offered by the prior systems. The system meets all design specifications, design risk-analysis, and medical device standards for electrical safety-EMC (IEC60601-1:2006, IEC60601-1-2:2014, IEC60601-2-2:2009, and IEC 60601-2-6:2012), biocompatibility (ISO10993-1:2018), sterility (ISO11135-1:2014), Risk management ISO14971:2019, and packaging shelf-life (ISO11607-1: 2006). Applicable software verification were completed per IEC 62304:2006.
## Ablation Range comparison between the Predicate and Subject devices:
#### Comparison Summary:
Table A compares the ablation performance ranges in terms of volume, length, and widths. The ablation volume ranges from 7018mm³ to the 320mm³ for predicate and 7260mm³ for the subject. The ablation length (Z) of 31mm to 10mm for the predicate and 30mm to 10mm for the subject. The ablation width (X & Y) ranges from 22mm for the predicate and 22mm to 8mm for the subject. The accumulated matching ablation cycle time ranges from 15minutes for the predicate and 7.5minutes to 0.5minutes for the subject. The subject K192406 device can achieve substantially equivalent ablation performances in 50% or less cycle times.
| Category | Predicate OsteoCool V-3 RFA | Subject AveCure BT MWA | Difference | I/SE |
|-----------------|-----------------------------|------------------------|------------|------|
| Volume Max | 7018mm³ (15minutes) | 7260mm³ (7.5minutes) | 3.4% | SE |
| Volume Min | 320mm³ | 320mm³ | 0% | I |
| Length (Z) Max | 31mm | 30mm | -3% | SE |
| Length (Z) Min | 10mm | 10mm | 0% | I |
| Width (X&Y) Max | 22mm | 22mm | 0% | I |
| Width (X&Y) Min | 8mm | 8mm | 0% | I |
## Ablation Performance Range Comparison Table A:
Volumes are estimates based on the calculation of volume for an ellipsoid.
## I: Identical. SE: Substantially equivalent.
## Conclusion:
The intended use of the subject device AveCure BT MW Ablation System is substantially equivalent to the predicate Medtronic (Baylis Medical) OsteoCool V-3 RF Ablation System K161949. The subject device identical to the MedWaves MWA systems approved under K070356 and K143203 with the following exceptions. The firmware configuration is modified to restrict its operating parameters for Mini and Small antennas. The power levels, temperature setting, and limits are configured to accommodate safe and effective bone ablation applications. The device configuration files and menu selection information which determines the applicable devices, and their operating control characteristics such as maximum power and maximum temperature are modified to reflect the new intended use. The foundation software
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for MedWaves Incorporated. The text "MedWaves" is in large, red, bold font. Below that, the word "INCORPORATED" is in smaller, dark blue font. To the right of the text is a graphic of concentric circles, also in red, that resemble radio waves.
16760 W Bernardo Dr San Diego, CA 92127
remains the same as the previously cleared device under K143203 and K070356, but the narrower operating characteristics are implemented for the new indications for use by customizing the configuration firmware. The power levels and user temperature settings range are modified to match the single-patient use devices with the system for the intended use.
The regulation number and description, product Class, indications for use, user prescription, anatomic site, access methods, the principle of operation, mechanism of action, feedback mechanism, and mechanism of control are substantially equivalent to the predicate device. Bench testing in bone has demonstrated substantially equivalent ablation performance between the two systems. The predicate RFA and subject MWA systems can perform ablations of comparable volumes, lengths and widths. The subject device can achieve them in ablation cycle 50% shorter for the largest volume and 87.5% shorter for the smallest volume as described in the Substantial Equivalence discussion.
Unintended tissue ablation is important for both predicate and subject device to avoid. Data from experimental protocols show the subject ablation device behaves substantially equivalent to the predicate system in and around the bone and soft tissue structures, and the primary driver for heat distribution is through thermal conduction. All safety precautions derived from these studies have been incorporated into the labeling.
The differences in design and technology characteristics of the subject devices and predicate devices do not raise any new questions of safety and effectiveness and support the substantial equivalence. The subject device employs MW energy in the low microwave frequency range of 915MHz ±13MHz versus predicate, 465KHz ±3%, and due to their physical nature, the subject device employee antennas versus bipolar electrodes in the predicate. The predicate device active elements (electrodes) are in the form of semirigid probes whereas the subject device active elements (antennas) mounted as the distal tip of either semirigid probes or flexible catheters. Both applicators are inserted into the patient to deliver energy in order to raise temperatures for coagulation necrosis. Both RF and Microwave generators produce up to 40 watts of power, with the predicate device typically using 20 watts, and the subject device using 16 or 24 watts. However, due to the way the energy is delivered through the transmission line, the subject device does not require a cooling system along the shaft. The subject device produces electromagnetic field about its antennas and therefore no net current flow. The predicate RFA system and subject MWA system are substantially equivalent in their ablation performance, and both systems' ablation parameters can be adjusted as necessary to match one another.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.