The INFINITY Angioplasty Balloon Catheter™ is indicated for Percutaneous Transluminal Angioplasty (PTA) in the peripheral vasculature, including iliac, femoral, popliteal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary or cerebrovascular arteries.
Device Story
Intravascular angioplasty balloon catheter; 6Fr system; 150cm shaft length. Features retractable sheath allowing variable balloon inflation lengths up to 250mm. Radiopaque marker bands denote 0mm, 100mm, 200mm, and 250mm lengths. Semi-compliant balloon; 5.0-5.5mm nominal diameter; 15atm rated burst pressure; 4-12atm clinical use range. Used in peripheral vasculature by physicians to dilate obstructive lesions. Physician controls balloon length via outer sheath placement. Output is mechanical dilation of vessel/fistula; benefits patient by restoring blood flow.
Clinical Evidence
No human clinical data. Evidence consists of performance bench testing and pre-clinical animal testing. Bench tests included balloon compliance, nominal diameter, length, fatigue, marker visibility, inflation/deflation time, tracking/delivery/retrieval, torque strength, kink resistance, joint strength, rated burst pressure, and simulated use.
Technological Characteristics
6Fr OTW catheter; 150cm shaft. Semi-compliant balloon; 5.0-5.5mm diameter. Retractable sheath for variable length (up to 250mm). 5 radiopaque marker bands. Sterilization: Ethylene Oxide (ISO 11135:2018). Biocompatibility: ISO 10993-1:2018. Mechanical standards: ISO 10555:2018.
Indications for Use
Indicated for Percutaneous Transluminal Angioplasty (PTA) in peripheral vasculature (iliac, femoral, popliteal, infra-popliteal arteries) and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae. Contraindicated for coronary or cerebrovascular arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Boston Scientific Corporation Sterling Over-the-Wire PTA Balloon Dilation Catheter (K141112)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue rectangle with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 20, 2020
Infinity Angioplasty Balloon Company, LLC Ms. Tiffini Wittwer Regulatory Affairs 6865 N. Reynolds Rd Suite 200 Toledo, Ohio 43615
## Re: K192399
Trade/Device Name: Infinity Angioplasty Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: April 13, 2020 Received: April 15, 2020
# Dear Ms. Wittwer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192399
Device Name INFINITY Angioplasty Balloon Catheter
#### Indications for Use (Describe)
The INFINITY Angioplasty Balloon Catheter™ is indicated for Percutaneous Transluminal Angioplasty (PTA) in the peripheral vasculature, including iliac, femoral, popliteal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary or cerebrovascular arteries.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span style="font-size: 20px;">☑</span> </div> |
|----------------------------------------------|------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span style="font-size: 20px;">☐</span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submitter: | INFINITY Angioplasty Balloon Company, LLC | | INFINITY Angioplasty<br>Balloon Catheter<br>(Subject Device) | Boston Scientific<br>Corporation<br>Sterling™ Over-the-<br>Wire PTA Balloon<br>Dilation Catheter<br>(Predicate Device) | Comment |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tiffini Wittwer<br>Regulatory Affairs | 510(k) Number | K192399 | K141112 | |
| | Phone: 707.799.6732<br>E-mail: twittwer@mededge.io | Manufacturer | INFINITY Angioplasty<br>Balloon Co., LLC | Boston Scientific<br>Corporation | |
| Trade Name: | INFINITY Angioplasty Balloon Catheter | Classification | Class II | Class II | Same |
| Common Name: | Percutaneous catheter | Product Code | LIT | LIT | Same |
| Classification: | Class II | Regulation | 21 CFR 870.1250 | 21 CFR 870.1250 | Same |
| Product Code: | LIT | Indications for Use | The INFINITY<br>Angioplasty Balloon<br>Catheter is indicated for<br>Percutaneous<br>Transluminal Angioplasty<br>(PTA) in the peripheral<br>vasculature, including<br>iliac, femoral,<br>popliteal, infra-popliteal<br>arteries, and for the<br>treatment of obstructive<br>lesions of native or<br>synthetic arteriovenous<br>dialysis fistulae. This<br>catheter is not for use in<br>coronary or<br>cerebrovascular arteries. | The Sterling OTW PTA<br>Balloon Dilatation<br>Catheter is indicated for<br>Percutaneous<br>Transluminal Angioplasty<br>(PTA) in the peripheral<br>vasculature, including<br>iliac, femoral,<br>popliteal, infra-popliteal.<br>and renal arteries, and for<br>the treatment of<br>obstructive lesions of<br>native or synthetic<br>arteriovenous dialysis<br>fistulae. This device is<br>also indicated for post-<br>dilatation<br>of balloon expandable and<br>self-expanding stents in<br>the peripheral vasculature. | Subject device has<br>narrower indication for<br>use than the predicate.<br>This difference limits use<br>of the device compared to<br>the predicate and does not<br>raise new questions of<br>safety or effectiveness. |
| Regulation | 21 CFR 870.1250 | Principle of operation | Inflation of semi-<br>compliant balloon for<br>dilation | Inflation of semi-<br>compliant balloon for<br>dilation | Same |
| Predicate Device(s): | The subject device is substantially equivalent to the following device:<br>● Boston Scientific Corporation Sterling Over-the-Wire PTA<br>Balloon Catheter (K141112) | Balloon Diameter (mm) | 5.0 | 2.0, 2.5, 3.0, 3.5, 4.0, 5.0,<br>6.0, 7.0, 8.0, 9.0, 10.0 | Same as one predicate<br>device size |
| Device Description: | The INFINITY Angioplasty Balloon Catheter is a intravascular<br>angioplasty balloon catheter with a retractable sheath allowing a<br>variable length inflatable balloon; up to 250mm. The over the wire<br>(OTW) catheter has a retractable sheath allowing the balloon to be<br>inflated at various lengths as determined by the physician. The balloon<br>lengths are graduated with radiopaque marker bands denoting 0mm,<br>100mm, 200mm, and 250mm. The INFINITY Balloon Catheter is a<br>6Fr catheter system with a shaft working length of 150cm.<br>The device uses a semi-compliant balloon with a rated burst pressure<br>of 15atm and an indicated clinical use range of 4atm - 12atm. The<br>balloon expands to a nominal diameter of 5.0 to 5.5mm. | Balloon Length (mm) | Variable<br>Marker bands located at<br>100mm, 200mm, and<br>250mm on balloon length | 20, 30, 40, 60, 80, 100,<br>200, 220 | Dimensional testing,<br>simulated use, rated burst<br>pressure, and in vivo<br>testing demonstrates the<br>difference does not create<br>additional risk to safety<br>and effectiveness of the<br>subject device |
| Indication for Use: | The INFINITY Angioplasty Balloon Catheter is indicated for<br>Percutaneous Transluminal Angioplasty (PTA) in the peripheral<br>vasculature, including iliac, femoral, popliteal, infra-popliteal arteries,<br>and for the treatment of obstructive lesions of native or synthetic<br>arteriovenous dialysis fistulae. This catheter is not for use in coronary<br>or cerebrovascular arteries. | Balloon Length Variability | Yes | No 1 fixed balloon length<br>per device size | Simulated use and in vivo<br>testing demonstrate the |
INFINITY Angioplasty Balloon Company, LLC
Page 1 of 4
Premarket Notification for the INFINITY Angioplasty Balloon Catheter™
{4}------------------------------------------------
INFINITY Angioplasty Balloon Company, LLC
Page 2 of 4
{5}------------------------------------------------
| | Balloon length controlled<br>by outer sheath placement<br>relative distal tip of the<br>balloon. | | difference does not create<br>additional risk to safety<br>and effectiveness of the<br>subject device |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Outer Sheath | Yes | Yes | Same |
| Number of radiopaque<br>markerbands | 5<br>Proximal and distal ends<br>of balloon, 100mm and<br>200mm balloon length,<br>and distal tip of outer<br>sheath | 2<br>Proximal and distal ends<br>of balloon | Visibility, marker band to<br>balloon alignment, and in<br>vivo testing demonstrate<br>the difference does not<br>create additional risk to<br>safety and effectiveness<br>of the subject device |
| Location markers | Yes Markers designate<br>location of the sheath<br>relative to the catheter<br>shaft | Markers designate<br>locations on catheter tip to<br>the guiding catheter | Marker band to balloon<br>alignment, and simulated<br>use demonstrate the<br>difference does not create<br>additional risk to safety<br>and effectiveness of the<br>subject device |
| Catheter Shaft Lengths<br>(cm) | 150 | 40, 90, 135, 150 | Same |
| Guidewire compatibility | 0.018" | 0.014", 0.018" | Same |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
{6}------------------------------------------------
| Functional Testing: | To verify that the device design meets its functional and<br>performance requirements, representative samples of the<br>device underwent performance bench testing and pre-clinical<br>animal testing. As a result of verification and validation<br>activities and risk assessment, testing ensures the device<br>design meets its functional and performance requirements. The<br>following tests were performed:<br>• Balloon compliance<br>• Balloon nominal diameter<br>• Balloon length<br>• Balloon Fatigue<br>• Marker band visibility<br>• Inflation and deflation time<br>• Device tracking, delivery, and retrieval<br>• Torque strength<br>• Kink resistance<br>• Joint strength testing<br>• Rated burst pressure<br>• Simulated use<br>The following standards were used in testing:<br>• ISO 11135:2018, Sterilization of health-care products - Ethylene<br>oxide - Requirements for the development, validation and routine<br>control of a sterilization process for medical devices<br>• ISO 10993-1:2018, Biological evaluation of medical devices<br>• ISO 10555:2018, Intravascular catheters – Sterile and single use<br>intravascular catheters. |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | The INFINITY Angioplasty Balloon Catheter™ intended use, indication<br>for use, materials, and fundamental scientific technology is similar to the<br>predicate device. Performance bench testing and pre-clinical testing<br>demonstrate that the differences between the subject device and the<br>predicate device do not raise new risks of safety and effectiveness. |
INFINITY Angioplasty Balloon Company, LLC
Page 4 of 4
Premarket Notification for the INFINITY Angioplasty Balloon Catheter™
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.