K192392 · DiaDent Group International · KLE · May 26, 2020 · Dental
Device Facts
Record ID
K192392
Device Name
DiaPlus
Applicant
DiaDent Group International
Product Code
KLE · Dental
Decision Date
May 26, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
As a dentine/enamel total etching bonding system for direct adhesion, it is used in the adhesion of all direct restoration substances. - Bonding of direct composite - Bonding to composite and set amalgam - Bonding of indirect restoration-Porcelain, Composite (Inlays, Onlays, Veneers, Crowns)
Device Story
DiaPlus is a 5th generation, light-cured, single-component dentine and enamel adhesive system. It is used by dental professionals in clinical settings to bond restorative materials to tooth structure. The device is applied to the tooth surface following a total-etching procedure; it acts as a primer and adhesive in one bottle. Upon exposure to a dental curing light, the resin polymerizes to create a bond between the tooth and the restorative material (composite, amalgam, or indirect restorations). The device facilitates the adhesion of direct and indirect dental restorations, potentially improving the retention and longevity of the dental repair.
Clinical Evidence
No clinical data was collected or provided. Substantial equivalence is supported by bench testing, including film thickness, depth of cure, sensitivity to ambient light, bond strength (enamel/dentin/metal/veneer), and dentinal tubule occlusion, alongside biocompatibility testing (cytotoxicity, sensitization, oral mucosa irritation, and acute systemic toxicity) per ISO standards.
Technological Characteristics
5th generation, light-cured, single-component resin adhesive. Composition: Ethanol, UDMA, BisGMA, TEGDMA, HEMA succinate, and camphorquinone. Conforms to ISO 4049 (polymer-based restorative materials) and ISO 10993-1 (biocompatibility). Supplied as a 5ml bottle. RX only.
Indications for Use
Indicated for use as a dentine/enamel total etching bonding system for direct adhesion of composite, set amalgam, and indirect restorations (porcelain, composite inlays, onlays, veneers, crowns).
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
May 26, 2020
Diadent Group International Kab Sun Lee Quality Assurance Manager 16, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu Cheongju-si, 28161 KOREA
Re: K192392
Trade/Device Name: DiaPlus Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: February 21, 2020 Received: February 27, 2020
Dear Kab Sun Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K192392
Device Name DIAPLUS
Indications for Use (Describe)
As a dentine/enamel total etching bonding system for direct adhesion, it is used in the adhesion of all direct restoration substances.
- Bonding of direct composite
- Bonding to composite and set amalgam
- Bonding of indirect restoration-Porcelain, Composite (Inlays, Onlays, Veneers, Crowns)
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K192392
# DIADENT GROUP INTERNATIONAL
# 510(k) Summary
# 1 Application Information
Date Prepared: May 25 2020
## Manufacturer:
DiaDent Group International
16, Osongsaengmyeong 4-ro, Osong-eup,
Heungdeok-gu, Cheongju-si,
Chungcheongbuk-do, 28161,
Republic of Korea
Contact Person: Kab Sun, Lee
Quality Assurance Manager
Phone: +82-43-266-2315
Fax: +82-43-235-2315
Email: diadent32@diadent.co.kr
# 2 Device Information
| Device Type: | Agent, Tooth Bonding, Resin |
|-------------------------|-----------------------------|
| Regulation Description: | Resin tooth bonding agent. |
| Review Panel: | Dental |
| Regulation Number: | 21 CFR 872.3200 |
| Product Code: | KLE |
| 510(k) Number | K192392 |
| Device Class: | II |
| Device Name: | DIAPLUS |
### 3 Predicate Devices
| 510(k) Number: | K043562 (Primary Predicate |
|--------------------|----------------------------|
| Applicant: | VERICOM Co., Ltd. |
| Device Name: | BC Plus |
| Regulation Number: | 872.3200 |
| Product Code: | KLE |
| Device Class: | Class II |
#### 4 Products configuration
DiaPlus : 1 bottle (5ml)
{4}------------------------------------------------
# K192392
# DIADENT GROUP INTERNATIONAL
# 5 Device Description
DiaPlus is a photopolymerization type dentine and enamel adhesive as a 5th generation dentine adhesion system that can be applied to all types of composite resin restorations.
#### 6 Indications for Use
As a dentine/enamel total etching bonding system for direct adhesion, It is used in the adhesion of all direct restoration substances.
- Bonding of direct composite
- Bonding to composite and set amalgam
- Bonding of indirect restoration-Porcelain, Composite (Inlays, Onlays, Veneers, Crowns)
#### 7 Substantial Equivalence discussion
#### <Comparison table>
| | Subject Device | Primary predicate Device | Discuss |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Product name | DiaPlus | BC Plus | |
| Manufacturer | DiaDent Group International | VERICOM Co., Ltd. | |
| 510K Number | K192392 | K043562 | |
| Product Code | KLE | KLE | |
| Device<br>Description | DiaPlus is a photopolymerization type<br>dentine and enamel adhesive as a 5th<br>generation dentine adhesion system that can<br>be applied to all types of composite resin<br>restorations. | BC Plus is a single component bonding<br>agent designed to bond composite to dentin,<br>enamel, cast metals, treated porcelain and<br>set amalgam. BC Plus is an ethanol based<br>formulation of light-activated, adhesive<br>resin. | |
| Indications for<br>Use | As a dentine/enamel total etching bonding<br>system for direct adhesion, it is used in the<br>adhesion of all direct restoration substances.<br>- Bonding of direct composite<br>- Bonding to composite and set amalgam<br>- Bonding of indirect restoration-Porcelain,<br>Composite (Inlays, Onlays, Veneers,<br>Crowns) | BC Plus is a light curing single component<br>bonding agent use in restorative adhesive<br>dentistry specifically developed for bonding<br>resin-based filling materials (e.g. composite,<br>compomers) to hard dental tissues. Other<br>indications include bonding of amalgam and<br>laboratory-produced restorations. BC Plus<br>permits priming and bonding to be carried<br>out in single step | |
| Chemical<br>Composition | - Ethanol<br>- UDMA (Urethane Dimethacrylate)<br>- BisGMA (bisphenolA-glycidyl methacrylate)<br>-TEGDMA (Triethylene glycol<br>dimethacrylate)<br>- HEMA succinate<br>- Champhor quinone | - Ethanol<br>- BisGMA (bisphenolA-glycidyl<br>methacrylate)<br>- HEMA ((Hydroxyethyl)methacrylate)<br>- Champhor quinone | |
{5}------------------------------------------------
# K192392
# DIADENT GROUP INTERNATIONAL
| Principle of<br>operation | Light cured | Light cured | equivalent |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|------------|
| Physical and<br>Mechanical<br>properties | Conformed in accordance with ISO 4049<br>-Bonding strength<br>-Depth of cure<br>-Sensitivity to ambient light | Conformed in accordance with ISO 4049<br>Bonding strength<br>Depth of cure | equivalent |
| Bonding<br>Agent Type | 5th generation bonding agent<br>(Primer & Adhesive combined in one bottle) | 5th generation bonding agent<br>(Primer & Adhesive combined in one bottle) | equivalent |
| Biocompati<br>bility | Yes<br>ISO 10993-1 :2009 | Yes<br>ISO 10993-1 :2009 | equivalent |
| RX Only/<br>OTC | RX Only | RX Only | equivalent |
#### Indication for use -
| | Subject device (DiaPlus) | Primary Predicate Device (BCPlus) |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description | DiaPlus is a photopolymerization type dentine and<br>enamel adhesive as a 5th generation dentine<br>adhesion system that can be applied to all types of<br>composite resin restorations. | BC Plus is a single component bonding agent<br>designed to bond composite to dentin, enamel, cast<br>metals, treated porcelain and set amalgam. BC Plus<br>is an ethanol based formulation of light-activated,<br>adhesive resin. |
| Indication For Use | As a dentine/enamel total etching bonding system for<br>direct adhesion, it is used in the adhesion of all direct<br>restoration substances.<br>- Bonding of direct composite<br>- Bonding to composite and set amalgam<br>- Bonding of indirect restoration-Porcelain,<br>Composite (Inlays, Onlays, Veneers, Crowns) | BC Plus is a light curing single component bonding<br>agent use in restorative adhesive dentistry<br>specifically developed for bonding resin-based filling<br>materials (e.g. composite, compomers) to hard<br>dental tissues. Other indications include bonding of<br>amalgam and laboratory-produced restorations. BC<br>Plus permits priming and bonding to be carried out in<br>single step. |
Discussion:
DiaPlus and BCPlus are photopolymerization type dentine and enamel adhesive using light curing.
The indication for Use of both devices are direct composite and indirect restoration!t seems that the indication of use are similar.
{6}------------------------------------------------
#### . Raw material
| Subject device (DiaPlus) | Primary Predicate Device (BCPlus) |
|------------------------------------------------|---------------------------------------------|
| - Ethyl alcohol | - Ethyl alcohol |
| - UDMA (Urethane Dimethacrylate) | |
| - BisGMA (bisphenolA-glycidyl methacrylate) | - BisGMA (bisphenolA-glycidyl methacrylate) |
| -TEGDMA (Triethylene glycol<br>dimethacrylate) | -HEMA((Hydroxyethyl)methacrylate) |
| - HEMA succinate | - Champhor quinone |
| - Champhor quinone | |
#### Discussion :
The predicate device is composed of Solvent, resin, primer, Photoinitiator for light curing as does the subject device. Results of bench and biocompatibility testing completed in alignment with ISO 4049 and ISO 10993-1, respectively demonstrate that any material differences between the subject device and predicate device do not raise any new questions as to safety and effectiveness. Therefore, it is concluded that DiaPlus is substantially equivalent to the predicate device.
#### 8 Non-Clinical Performance Data
This device has demonstrated conformance with non-clinical performance requirements through evaluation and testing in accordance with the following harmonized standards:
- -ISO 4049 Dentistry -- Polymer-based restorative materials
- Dentistry Evaluation Of Biocompatibility Of Medical Devices Used In Dentistry -ISO 7405 [Including: Amendment 1 (2013)]
- -ISO 29022 Dentistry - Adhesive - Notched-edge sheer bond strength test
- -ISO/TS 11405 Dentistry - Testing of adhesion to tooth structure
- -ISO 10993-1 Biological evaluation of medical devices - Part 1:Evaluation and testing within a risk management process
- -ISO 10993-3 Biological evaluation of medical devices - Part 1:Evaluation and testing within a risk management process
- -ISO 10993-5 Biological evaluation of medical devices - Part 5. Tests for in vitro Cytotoxicity
- -ISO 10993-10 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- -ISO 10993-11 Biological Evaluation Of Medical Devices - Part 11: Tests For Systemic Toxicity
{7}------------------------------------------------
# DIADENT GROUP INTERNATIONAL
The following Performance tests were conducted.
- -Film Thickness
- -Depth of cure
- -Sensitivity to ambient light
- -Bond strength (Enamel/Dentin)
- -Bond strength (Metal/Veneer)
- -Dentinal tubule occluding property
- = Biocompatibility test
#### Biocompatibility test:
| Test | Standard | Test result |
|--------------------------------|----------------------------------------------------------------------------------------------------------------|-------------|
| Cytotoxicity Study | •10993-5 – Biological Evaluation of Medical Devices –<br>Part 5: Tests for In Vitro Cytotoxicity | Pass |
| Sensitization Test | •10993-10 – Biological Evaluation of Medical Devices –<br>Part 10: Tests for Irritation and Skin Sensitization | Pass |
| Oral Mucosa<br>Irritation Test | •10993-10 – Biological Evaluation of Medical Devices –<br>Part 10: Tests for Irritation and Skin Sensitization | Pass |
| Acute Systemic<br>Toxicity | •10993-11 – Biological Evaluation of Medical Devices –<br>Part 11: Tests for Systemic Toxicity | Pass |
# 9 Clinical Performance Data
No clinical data was collected or provided to support substantial equivalence between the subject and predicate devices.
#### 10 Conclusions
Based on the above information and all data provided in this submission, the comparison of intended uses, technological characteristics, and non-clinical performance testing demonstrates that the subject device is substantially equivalent to the legally marketed devices identified in this submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.