← Product Code [IYN](/productcode/IYN) · K192388

# SonoVision Ultrasound Imaging System (K192388)

_Tissue Differentiation Intelligence, LLC · IYN · Mar 12, 2020 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K192388

## Device Facts

- **Applicant:** Tissue Differentiation Intelligence, LLC
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Mar 12, 2020
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML

## Indications for Use

The SonoVision™ Ultrasound Imaging System is a general purpose ultrasound system intended for use by a qualified physician for the visualization and evaluation of nerves, vascular and other anatomical structures. The system provides imaging assistance in spinal procedure applications.

## Device Story

SonoVision™ Ultrasound Imaging System with Beluga1™ probe; general-purpose ultrasound for intraoperative spinal procedures. System inputs: ultrasonic signals via 10 MHz transducer. Processing: B-mode imaging, Doppler/color flow mapping, and image processing software to define anatomical features. Output: real-time 2D bone and tissue visualization on console monitor. Operated by qualified physicians in clinical/intraoperative settings. Healthcare providers use visual output to guide spinal procedures and evaluate anatomical structures. Benefits: real-time imaging assistance for nerve/vascular/bone identification during surgery.

## Clinical Evidence

In vivo testing conducted to demonstrate device performance in human applications. Bench testing confirmed compliance with electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), acoustic output (IEC 60601-2-37, NEMA UD 2), biocompatibility (ISO 10993-1), and sterilization (ISO 14937).

## Technological Characteristics

10 MHz transducer probe; B-mode and Doppler/color flow imaging; non-user adjustable acoustic output; touchscreen operation; data capture/external storage. Electrical safety/EMC per IEC 60601-1/1-2; acoustic output per IEC 60601-2-37/62359 and NEMA UD 2. Biocompatibility per ISO 10993-1; sterilization per ISO 14937. Proprietary software for image processing.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Accuro™ 3000 Ultrasound System ([K132736](/device/K132736.md))

## Reference Devices

- Esaote 6200 and 6250 Ultrasound System ([K153277](/device/K153277.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 12, 2020

Tissue Differentiation Intelligence, LLC % Lorry Huffman Principal Consultant, Regulatory Affairs Qserve Group, US, Inc. 303 Twin Dolphin, Suite 600 REDWOOD SHORES CA 94065

### Re: K192388

Trade/Device Name: SonoVision™ Ultrasound Imaging System with Beluga1™ Ultrasound Probe Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: February 5, 2020 Received: February 7, 2020

### Dear Lorry Huffman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K192388

Device Name

SonoVision™ Ultrasound Imaging System with Beluga1™ Ultrasound Probe

#### Indications for Use (Describe)

The SonoVision™ Ultrasound Imaging System is a general purpose ultrasound system intended for use by a qualified physician for the visualization and evaluation of nerves, vascular and other anatomical structures. The system provides imaging assistance in spinal procedure applications.

| Type of Use (Select one or both, as applicable)                                                                               |                                                                                                                              |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span> <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

TDi, LLC March 6, 2020

### Section 5: 510(k) Summary – [21 cFR 807.92]

Image /page/3/Picture/3 description: The image shows the alphanumeric string 'K192388' in a bold, sans-serif font. The string is horizontally oriented and appears to be a code or identifier. The characters are uniformly sized and spaced, with the 'K' being the only letter and the rest being numerical digits.

807.92(a)(1), (2), (3)

| Date Prepared:       | March 6, 2020                                                                                                                                           |                                        |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Submitter's Name:    | Tissue Differentiation Intelligence, LLC<br>Christian Zaal, Vice President Operations<br>234 NE 4th Avenue<br>Delray Beach, FL 33486<br>+1-305-351-2011 |                                        |
| Device Trade Name:   | SonoVision™ Ultrasound System with Beluga1™ Ultrasound Probe                                                                                            |                                        |
| Common Name:         | Ultrasound pulsed doppler imaging system                                                                                                                |                                        |
| Review Panel:        | OIR                                                                                                                                                     |                                        |
| Classification Name: | Ultrasound pulsed doppler imaging system: 892.1550                                                                                                      |                                        |
| Classification Code: | IYN: Ultrasound pulsed doppler imaging system                                                                                                           |                                        |
|                      | IYO: Ultrasonic pulsed echo imaging system                                                                                                              |                                        |
|                      | ITX: Diagnostic ultrasonic transducer                                                                                                                   |                                        |
| Predicate Device(s): |                                                                                                                                                         |                                        |
|                      | Primary Predicate                                                                                                                                       | Reference Predicate                    |
|                      | Accuro™ 3000 Ultrasound System                                                                                                                          | Esaote 6200 and 6250 Ultrasound System |
| 510(k) Number        | K132736                                                                                                                                                 | K153277                                |
| Classification Code  | IYO, ITX                                                                                                                                                | IYN, IYO, ITX                          |
| Regulation           | 892.1550                                                                                                                                                | 892.1550                               |

#### Device Description 807.92(a)(4):

The SonoVision™ Ultrasound Imaging System with Beluga1™ transducer probe is a general purpose, Track 1, ultrasound system intended for use by a qualified physician for the visualization and evaluation of nerves, vascular and other anatomical structures. The Beluga1™ is an ultrasound imaging probe that has been ergonomically designed for use in intraoperative procedures.

The system is comprised of the SonoVision™ console & monitor, 10 MHz transducer imaging probe (Beluga1™), and control software. The system includes image processing software that can be used to define anatomical features in an ultrasound B-mode image. The Beluga1™ Imaging Probe has been developed for specific application in spine procedures.

#### Intended Use 807.92(a)(5):

The SonoVision™ Ultrasound Imaging System is a general purpose ultrasound system intended for use by a qualified physician for the visualization and evaluation of nerves, vascular and other anatomical structures. The system provides imaging assistance in spinal procedure applications.

{4}------------------------------------------------

### Technological Characteristics 807.92(a)(6):

The SonoVision™ Ultrasound Imaging System has the same fundamental technological characteristics as the primary and reference predicates. Similar features are as follows:

- Intraoperative clinical use ●
- Spine application
- B-mode imaging
- Doppler (echographic image)/color flow mapping
- Real-time 2-D bone and tissue scanning (differentiation)
- . Radiofrequency (RF) energy for its internal function resulting in very low RF emissions
- Non-user adjustable acoustic output
- Proprietary software performance
- Data capture and external storage
- Electrical safety/EMC/acoustic output meets standards ●
- . Reusable probe

# Summary of Non-Clinical Test - Performance and Safety Testing 807.92(b)(1)

The SonoVision™ Ultrasound Imaging System was evaluated for performance, acoustic output, biocompatibility, cleaning and sterilization as well as thermal, electromagnetic, and mechanical safety, and found to conform with the following FDA recognized standards:

- IEC 60601-1-2: 4th edition (19-8) Medical Electrical Equipment Part 1-2: general requirements . for basic safety and essential performance - collateral standard: electromagnetic disturbances requirements and tests
- IEC 60601-1-1: 2005/(R)2012 And A1:2012 (19-4) Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: general requirements [including: corrigendum 1 (2011)]

# Acoustic output

- . IEC 60601-2-37 ED 2.1 2015 medical electrical equipment – Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- . IEC 62359 Ed 2.1 2017-09 Ultrasonics - field characterization - test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields [Including: technical corrigendum 1 (2011)]
- NEMA UD 2 revision 3 2009 Acoustic output measurement standard for diagnostic ultrasound equipment

# Temperature rise of the device is in accordance with:

- IEC 60601-2-37 ED 2.1 2015 medical electrical equipment Part 2-37: Particular requirements ● for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment

{5}------------------------------------------------

TDi, LLC March 6, 2020

### Biocompatibility

ISO 10993-1 5th edition 2018-08 for limited exposure, external communicating device. ISO 10993-1 Annex A Table A.1 – Evaluation tests for consideration indicate biological effects cytotoxicity, sensitization, irritation or intracutaneous reactivity be evaluated for the medical device category of the transducer.

#### Sterilization

ISO 14937:2009 Sterilization of healthcare products-General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices.

#### Summary of Clinical Tests 807.92(b)(1)

In vivo testing was conducted to demonstrate performance of the device in human applications.

#### Conclusion 807.92(b)(1):

SonoVision™ Ultrasound Imaging System with Beluga1™ transducer probe and the Accuro™ 3000 Ultrasound System both provide general purpose ultrasound tissue imaging, ability to highlight and enhance bone anatomy and state of the art features such as reusable probe, touchscreen operation, and data capture and external storage. Doppler imaging and intraoperative features are supported by the reference predicate Esaote 6000/6250.

The predicate comparison table and performance and safety testing provided in this 510(k) is sufficient to demonstrate that SonoVision™ Ultrasound Imaging System with Beluga1™ transducer probe is substantially equivalent to preamendment devices, Accuro™ 3000 ultrasound system, and Esaote 6000/6250 Ultrasound System and meets requirements for Track 1 as described in FDA Final Guidance Marketing Clearance of Diagnostic Ultrasound Systems and Transducers-June 27, 2019.

---

**Source:** [https://fda-staging.innolitics.com/device/K192388](https://fda-staging.innolitics.com/device/K192388)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
