KeraStat Cream
K192386 · Keranetics · KGN · Jul 16, 2020 · SU
Device Facts
| Record ID | K192386 |
| Device Name | KeraStat Cream |
| Applicant | Keranetics |
| Product Code | KGN · SU |
| Decision Date | Jul 16, 2020 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
KeraStat® Cream is intended to maintain a moist wound environment. KeraStat® Cream is indicated for management of a number of partial thickness skin wounds such as: partial thickness (first and second degree) burns, severe sunburns, superficial injuries, cuts, abrasions, and incisions/surgical wounds. Under the direction of a healthcare professional, KeraStat® Cream also may be used in the management of dry, light, and moderately exuding partial thickness wounds including: pressure (stage I-II) ulcers, venous stasis ulcers, ulcers caused by mixed vascular etiologies, diabetic ulcers, radiation dermatitis, donor sites, and grafts. KeraStat® Cream is not indicated for full thickness or third degree burns. This device will be available by prescription.
Device Story
KeraStat® Cream is a non-sterile, multi-use topical wound dressing containing 5% keratin protein in a cream base. Applied topically by healthcare professionals or patients (under direction) to partial thickness wounds; device maintains a moist wound environment to support healing. Provided in 1 oz (29.6 mL) screw-top tubes. Output is the physical barrier/moist environment provided by the cream application. Benefits include support of wound healing process in various dermal wound types.
Clinical Evidence
Clinical evidence includes a Radiation Dermatitis Clinical Study, Repeat Insult Patch Test, and Skin Prick Test. Non-clinical testing included biocompatibility (ISO 10993-5, -6, -10, -11, -17, -18, USP <151>) and performance testing (protein content, viscosity, pH, microbial contamination, preservative effectiveness, SEC, stability).
Technological Characteristics
Non-sterile, multi-use cream emulsion. Composition: purified water, glycerin, mineral oil, 5% keratin, caprylic triglyceride, dimethicone, sodium polyacrylate, hydrogenated polydecene, trideceth-6, sodium stearoyl glutamate, phenoxyethanol, ethylhexylglycerin. Topical application. No energy source. No software.
Indications for Use
Indicated for management of partial thickness skin wounds (first/second degree burns, severe sunburns, superficial injuries, cuts, abrasions, surgical wounds) and, under professional guidance, dry/light/moderately exuding partial thickness wounds (stage I-II pressure ulcers, venous stasis ulcers, mixed vascular etiology ulcers, diabetic ulcers, radiation dermatitis, donor sites, grafts). Not for full thickness or third-degree burns.
Predicate Devices
Reference Devices
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters, and the word "ADMINISTRATION" in smaller, blue letters below.
July 16, 2020
KeraNetics Luke Burnett CEO and CSO 200 East First Street, Suite 101, Box#4 Winston Salem, North Carolina 27101
Re: K192386
Trade/Device Name: KeraStat Cream Regulatory Class: Unclassified Product Code: KGN, Dated: June 12, 2020 Received: June 15, 2020
Dear Luke Burnett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anjana Jain, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192386
Device Name KeraStat® Cream
#### Indications for Use (Describe)
KeraStat® Cream is intended to maintain a moist wound environment. KeraStat® Cream is indicated for the management of a number of partial thickness skin wounds such as: partial thickness (first and second degree) burns, superficial injuries, cuts, and incisions/surgical wounds. Under the direction of a healthcare professional, KeraStat® Cream also may be used in the management of dry, light, and moderately exuding partial thickness wounds including: pressure (stage I-II) ulcers, venous stasis ulcers caused by mixed vascular etiologies, diabetic ulcers, radiation dermatitis, donor sites, and grafts.
KeraStat® Cream is not indicated for full thickness or third degree will be available by prescription.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------------------------------------------|
| | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY--- KERASTAT® CREAM
| Submitter Name: | KeraNetics, Inc. |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 200 East First Street, Box #4, Suite 102<br>Winston Salem, NC 27101 |
| Phone Number: | 1.336.202.1307 |
| Fax Number: | 1.336.725.0619 |
| Contact Person: | Luke Burnett, PhD<br>CEO and CSO, KeraNetics, Inc. |
| Phone Number: | 1.336.202.1307 |
| Fax Number: | 1.336.725.0619 |
| Email: | lburnett@keranetics.com |
| Date Prepared: | July16, 2020 |
| Device Trade Name: | KeraStat® Cream |
| Device Common Name: | Wound Dressing |
| Classification Number: | Unclassified |
| Product Code: | KGN |
| Product Type Name: | Dressing, Wound, Collagen |
| Predicate Device: | KeraStat® Gel<br>KeraNetics Inc.<br>K162759<br>Cleared on June 2, 2017 |
| Reference Device: | Biafine®<br>Valeant Pharmaceuticals<br>K173549<br>Cleared on August 13, 2018 |
| Device Description: | KeraStat® Cream is a non-sterile, non-implantable wound<br>dressing intended to provide a moist environment in the<br>management of a variety of partial thickness dermal<br>wounds. KeraStat® Cream is provided in a screw top tube<br>for multiple uses. Each tube contains 1 oz (29.6 mL) of<br>KeraStat® Cream, which contains 5% keratin protein<br>incorporated into a cream base. |
{4}------------------------------------------------
#### Statement of Intended Use :
KeraStat® Cream is intended to maintain a moist wound environment. KeraStat® Cream is indicated for management of a number of partial thickness skin wounds such as: partial thickness (first and second degree) burns, severe sunburns, superficial injuries, cuts, abrasions, and incisions/surgical wounds. Under the direction of a healthcare professional, KeraStat® Cream also may be used in the management of dry, light, and moderately exuding partial thickness wounds including: pressure (stage I-II) ulcers, venous stasis ulcers, ulcers caused by mixed vascular etiologies, diabetic ulcers, radiation dermatitis, donor sites, and grafts.
KeraStat® Cream is not indicated for full thickness or third degree burns. This device will be available by prescription.
## Comparison to the Predicate and Reference Devices:
KeraStat® Cream and the predicate wound dressing, KeraStat® Gel, are both designed for dry, light, and moderately exuding partial thickness wounds. The wound dressings are applied to the wound and function to provide a moist environment for wound management. The operational principles of the subject and predicate device are identical, and they contain identical concentrations and composition of human derived keratin protein. The only differences between KeraStat® Cream and the KeraStat® Gel predicate device are that KeraStat® Cream is provided in an emulsion base, is nonsterile, and is intended for multiple uses.
KeraStat® Cream and the reference wound dressing, Biafine®, have the same intended use to provide a moist environment for wound management including wounds caused by radiation dermatitis. Biafine® has similar technological characteristics as KeraStat® Cream, as both are non-sterile, multi-use cream emulsions composed of a protein/amine, emollients, emulsifiers, thickeners, and preservatives.
## Performance Data Summary of Non-Clinical Tests:
The following testing was performed to demonstrate substantial equivalence:
- Biocompatibility testing: .
- Cytotoxicity (ISO 10993-5) o
- Sensitization (ISO 10993-10) o
- o Irritation/Intracutaneous reactivity (ISO 10993-10)
- Acute systemic toxicity (ISO 10993-11) O
- Toxicology Risk Assessment (ISO 10993-17) O
- Chemical Characterization (ISO 10993-18) o
- Implantation Test (ISO 10993-6) o
- o Rabbit Pyrogenicity Test (USP Chapter <151>)
{5}------------------------------------------------
- Performance testing:
- o Protein content
- o Viscosity
- o pH
- Microbial contamination o
- Preservative effectiveness o
- o Size-exclusion chromatography (SEC)
- o Stability
## Summary of Clinical Tests:
The following human clinical testing was performed:
- Radiation Dermatitis Clinical Study ●
- Repeat Insult Patch Test ●
- Skin Prick Test
## Testing Summary:
Biocompatibility, performance, and clinical testing support the safety and effectiveness of KeraStat® Cream.
#### Subs tantial Equivalence :
KeraStat® Cream has the same intended use, same principles of operation, and similar technological characteristics as the predicate device KeraStat® Gel (K162759). While the subject device differs from the predicate device with respect to its additional components and multi-use design, both devices share the same mode of action in that they provide a moist environment to support wound healing. Any differences in technological characteristics outlined in Table 1 do not raise new questions about the safety and effectiveness of KeraStat® Cream.
KeraStat® Cream has similar intended use, same principles of operation, and similar technological characteristics as the reference device Biafine® (K173549). Both devices share the same mode of action in that they provide a moist environment to support wound healing, including wounds due to radiation dermatitis. Biafine® has similar technological characteristics as KeraStat® Cream, as both are non-sterile, multiuse cream emulsions composed of a protein/amine, emollients, emulsifiers, thickeners, and preservatives. Any differences in technological characteristics outlined in Table 1 do not raise new questions about the safety and effectiveness of KeraStat® Cream.
{6}------------------------------------------------
| Table 1. Substantial Equivalence Information | | | |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Subject Device:<br>KeraStat® Cream | Predicate Device:<br>KeraStat® Gel | Reference Device:<br>Biafine® |
| 510(k) No. | | K162759 | K173549 |
| Intended Use | Wound dressing for<br>management of partial<br>thickness wounds | Wound dressing for<br>management of partial<br>thickness wounds | Wound dressing for<br>management of partial and<br>full thickness wounds |
| Indications | Dry, light, and moderately<br>exuding partial thickness<br>wounds such as: first and<br>second degree burns, severe<br>sunburns, superficial injuries,<br>cuts, abrasion, and surgical<br>wounds; may also be used<br>under the guidance of a health<br>care professional in the<br>management of pressure<br>(stage I-II) and venous stasis<br>ulcers, ulcers caused by<br>mixed vascular etiologies,<br>diabetic ulcers, radiation<br>dermatitis, donor sites and<br>grafts | Dry, light, and moderately<br>exuding partial thickness<br>wounds such as: first and<br>second degree burns, severe<br>sunburns, superficial injuries,<br>cuts, abrasion, and surgical<br>wounds; may also be used<br>under the guidance of a health<br>care professional in the<br>management of pressure<br>(stage I-II) and venous stasis<br>ulcers, ulcers caused by mixed<br>vascular etiologies, diabetic<br>ulcers, donor sites and grafts | OTC indications:<br>Management of superficial<br>wounds such as minor cuts,<br>minor scrapes, minor<br>irritations, minor<br>abrasions, minor blisters, 1st<br>degree burns including<br>sunburns, minor skin<br>irritations following post non<br>ablative laser therapy<br>procedures,<br>microdermabrasion therapy or<br>superficial chemical peels.<br>May also be used for relief of<br>itch, pain and burning from<br>minor skin irritations,<br>lacerations, abrasions and<br>minor burns.<br>Rx Indications: management<br>of full thickness wounds,<br>pressure sores, dermal ulcers<br>including lower leg ulcers,<br>radiation dermatitis, donor<br>sites and 2nd degree burns.<br>May also be used for relief of<br>itch, pain and burning from<br>minor skin irritations,<br>lacerations, abrasions and<br>minor burns. |
| Mode of Action | Provides a moist environment<br>that is supportive of wound<br>healing | Absorbs exudate and provides<br>a moist environment that is<br>supportive of wound healing | Provides a moist environment<br>that is supportive of wound<br>healing |
| Technological<br>Characteristics | Mixture of human hair<br>derived keratin proteins in a<br>cream base intended to a<br>create a moist environment<br>and packaged as a non- | Mixture of human hair<br>derived keratin proteins<br>intended to absorb exudate<br>and create a moist<br>environment and packaged as | Trolamine/sodiumalginate in<br>a cream base intended to a<br>create a moist environment<br>and packaged as a non-sterile,<br>multi-use dressing |
| Trade Name | Subject Device: | Predicate Device: | Reference Device: |
| | KeraStat® Cream | KeraStat® Gel | Biafine® |
| 510(k) No. | | K162759 | K173549 |
| Components | Purified water, glycerin,<br>mineral oil, keratin, caprylic<br>triglyceride, dimethicone,<br>sodiumpolyacrylate,<br>hydrogenated polydecene,<br>trideceth-6, sodiumstearoyl<br>glutamate, phenoxyethanol,<br>and ethylhexylglycerin | Purified water, keratin,<br>phenoxyethanol, carbomer,<br>sodiumhydroxide, and<br>ethylhexylglycerin | Purified water, liquid paraffin,<br>ethylene glycol monostearate,<br>stearic acid, propylene<br>glycol, paraffin wax.<br>squalane, avocado oil,<br>trolamine/sodiumalginate,<br>triethanolamine,cetyl<br>palmitate, methylparaben<br>(sodiumsalt), sorbic acid (as<br>potassiums alt ), propyl<br>paraben (sodium<br>salt), and fragrance |
| Form of Wound<br>Dressing | Cream | Hydrogel | Cream |
| Application<br>Method | Topical | Topical | Topical |
| Sterility | Non-sterile | Sterile | Non-sterile |
| Numberof<br>Uses | Multi-use | Single use | Multi-use |
| Prescription<br>Use | Yes | Yes | Yes (also OTC) |
Table 1. Substantial Equivalence Information
{7}------------------------------------------------
# Conclusion:
KeraStat® Cream demonstrates substantial equivalence to the predicate device, KeraStat® Gel (K162759) and there are not different questions of safety and effectiveness.