Intellifuse Administration and Intellifuse Blood Administration Sets
K192375 · Smiths Medical Asd, Inc. · FPA · Sep 18, 2020 · General Hospital
Device Facts
Record ID
K192375
Device Name
Intellifuse Administration and Intellifuse Blood Administration Sets
Applicant
Smiths Medical Asd, Inc.
Product Code
FPA · General Hospital
Decision Date
Sep 18, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intellifuse™ Administration Sets are indicated for the delivery of fluids from a container to a patient's vascular system. Intellifuse™ Blood Administration Sets are indicated for the delivery of fluids including but not limited to blood and blood products from a container to a patient's vascular system.
Device Story
Intellifuse™ Administration and Blood Administration Sets are single-use, sterile, disposable gravity-flow devices; used to deliver fluids, medications, blood, and blood products from a container to a patient's vascular system. Components include tubing, spike assemblies, drip chambers, check valves, clamps, needleless access connectors, injection sites, in-line filters, and luer connectors. Operated by healthcare professionals in clinical settings; device facilitates fluid administration via gravity. Output is the controlled delivery of fluids to the patient; clinical benefit is the safe, sterile infusion of necessary therapeutic substances.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ISO 8536-4 (infusion sets), ISO 1135-4 (transfusion sets), ISO 594-1/2 (luer fittings), USP <788> (particulate assessment), ISO 8536-12 (check valves), ISO 8536-5 (burette chambers), ISO 10993-1 (biocompatibility), and ISO 11135 (sterilization).
Technological Characteristics
Materials: ABS, DEHT, PVC, HDPE, Acrylic, Nylon, LDPE, MABS, PC, Silicone, Flurosilicone, Polyisoprene, E-PTFE, Polyurethane. Gravity-fed fluid path. Components: spike, drip chamber, check valve, clamp, needleless connector, filter, luer lock. Sterilization: Ethylene Oxide (SAL 10⁻⁶). Standards: ISO 8536-4, ISO 1135-4, ISO 594-1/2, ISO 8536-12, ISO 8536-5, ISO 10993-1, ISO 11135, ISO 11607-1.
Indications for Use
Indicated for the delivery of fluids, blood, and blood products from a container to the vascular system of the general patient population.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 18, 2020
Smiths Medical ASD. Inc. % Danielle Besal Principal Consultant MRC Global, LLC 9085 East Mineral Circle, Suite 110 Centennial. CO 80112
Re: K192375
Trade/Device Name: Intellifuse™ Administration and Intellifuse™ Blood Administration Sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: September 9, 2020 Received: September 11, 2020
Dear Danielle Besal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K192375
Device Name
Intellifuse™ Administration Sets and Intellifuse™ Blood Administration Sets
#### Indications for Use (Describe)
Intellifuse™ Administration Sets are indicated for the delivery of fluids from a container to a patient's vascular system.
Intellifuse™ Blood Administration Sets are indicated for the delivery of fluids including but not limited to blood and blood products from a container to a patient's vascular system.
| Type of Use (Select one or both, as applicable): |
|----------------------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# smiths medical
bringing technology to life
## 510(k) Summary K192375
| Date: | September 18, 2020 |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Smiths Medical<br>6000 Nathan Lane North, Minneapolis, MN 55442, USA<br>Phone: 763.383.3000 |
| Establishment Registration: | 3012307300 |
| Primary Contact: | Danielle Besal<br>Principal Consultant MRC Global<br>Phone: 901.827.8670 |
| Proprietary Name: | Intellifuse™ Administration Sets and Intellifuse™ Blood Administration<br>Sets |
| Regulation Name: | Intravascular administration set |
| Regulation: | 21 CFR 880.5440 |
| Product Codes: | FPA |
| Device Class: | II |
| Panel: | General Hospital |
| Predicate Device: | Hospira Administration Sets: Hospira Primary Sets, Hospira<br>Extension Sets, Hospira Burette Sets, and Hospira Blood Sets<br>(K160870, 21 CFR 880.5440, FPA) |
#### Device Description:
Intellifuse™ Administration Sets and Intellifuse™ Blood Administration Sets are single use, disposable, sterile fluid path, intravenous administration sets used to deliver fluids, medications, blood and blood products from a container into a patient's vascular system depending on the configuration selected. These sets are available in multiple configurations and may be comprised of various components which are broadly used throughout industry tubing, spike assemblies, drip chambers, check valves, clamps, needleless access connector, injection sites, in-line filters, luer connectors, and a tube frame assembly. The administration sets are intended to administer fluids with gravity flow. The IV Administration sets are configured to ensure the intended use of the device is met. The devices are intended for the general patient population.
#### Indications for Use:
Intellifuse™ Administration Sets are indicated for the delivery of fluids from a container to a patient's vascular system.
Intellifuse™ Blood Administration Sets are indicated for the delivery of fluids including but not limited to blood and blood products from a container to a patient's vascular system.
#### Substantial Equivalence:
The technology for Intellifuse™ Administration Sets and Intellifuse™ Blood Administration Sets is similar between the subject and the predicate devices as described in the table below.
{4}------------------------------------------------
| | Subject:<br>Intellifuse™ Administration<br>Sets & Intellifuse Blood<br>Administration Sets<br>(K192375) | Predicate:<br>Hospira Administration<br>Sets<br>(K160870) | Substantial Equivalence<br>Discussion |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Smiths Medical | Hospira | N/A |
| Indications<br>for Use | Intellifuse™ Administration Sets<br>are indicated for the delivery of<br>fluids from a container to a<br>patient's vascular system.<br><br>Intellifuse™ Blood<br>Administration Sets are indicated<br>for the delivery of fluids<br>including but not limited to blood<br>and blood products from a<br>container to a patient's vascular<br>system. | Hospira Primary sets are<br>indicated for the delivery of<br>fluids from a container to a<br>patient's vascular system.<br><br>Hospira Extension sets are<br>indicated for the delivery of<br>fluids from a container to a<br>patient's vascular system.<br><br>Hospira Burette sets are<br>indicated for the delivery of<br>fluids from a container to a<br>patient's vascular system.<br><br>Hospira Blood sets are<br>indicated for the delivery of<br>fluids including but not<br>limited to blood and blood<br>products from a container to<br>a patient's vascular system | The subject Intellifuse™<br>Administration Sets encompass<br>various configurations (including<br>primary, secondary, and burette<br>sets) that differ slightly from the<br>predicate set configurations, but all<br>are equivalently intended for the<br>delivery of fluids from a container<br>to a patient's vascular system.<br><br>The subject blood sets have the<br>same indications as the blood sets<br>for the predicate device.<br><br>Substantially equivalent. |
| Usage | Single Patient Use, Disposable | Single Patient Use,<br>Disposable | Identical; substantially equivalent |
| Available<br>Design<br>Features | PVC tubing, low light absorption<br>tubing, bag spike and cap drip<br>chamber, bag spike handle, drip<br>chamber, burette subassembly,<br>check valve, pinch clamp, roller<br>clamp, injection site assembly,<br>needleless access connector<br>(nSyte™), filter, male luer<br>adapter with cap, blood chamber<br>Gravity Flow | Male and female luer adapter<br>with cap, piercing pin<br>connector, tubing, flow<br>control device, filter, in-line<br>adapter, injection site<br>assembly, luer activated<br>needleless valve connector<br>(Nuitiv™ Connector), check<br>valve, burette chamber,<br>blood chamber<br>Gravity Flow | Equivalent common components are<br>used in all sets and performance<br>testing shows compliance with ISO<br>requirements and acceptability for<br>device's intended use. Substantially<br>equivalent |
| Priming<br>Volume | Primary Sets: 24-29 mL<br>Blood Sets: 29-30 mL<br>Secondary Set: 14 mL | Not disclosed | Priming testing supports<br>acceptability of device for its<br>intended use. Substantially<br>equivalent. |
| Drops / mL | Burette Sets: 60 drops/mL<br>All other sets: 20 drops/mL | Not disclosed | Performance testing supports<br>acceptability of device for its<br>intended use. Substantially<br>equivalent. |
| | Subject:<br>Intellifuse™ Administration<br>Sets & Intellifuse Blood<br>Administration Sets<br>(K192375) | Predicate:<br>Hospira Administration<br>Sets<br>(K160870) | Substantial Equivalence<br>Discussion |
| Materials | ABS, DEHT, PVC<br>HDPE, Acrylic & Nylon<br>LDPE, MABS, PC, Silicone,<br>Flurosilicone, Clear ABS, LDPE,<br>Polyisoprene, Acrylic, E-PTFE,<br>Polyurethane (biocompatible,<br>non-DEHP, not manufactured<br>with natural rubber latex) | Polycarbonate, copolyester,<br>silicone, acrylic,<br>polypropylene, non-phthalate<br>PVC, HDPE, ABS, nylon | Similar, all differences supported by<br>ISO 10993 biocompatibility testing.<br>Substantially Equivalent |
| Sterility | Sterile fluid path, EO<br>SAL 10⁻⁶ | Sterile,<br>method not disclosed,<br>SAL 10⁻⁶ | Predicate sterilization method not<br>disclosed; however, both devices<br>are validated to the same sterility<br>assurance level and the subject<br>sterilization has been validated and<br>is acceptable for its intended use;<br>substantially equivalent. |
{5}------------------------------------------------
The differences between the subject and predicate devices are minimal and include differences in the material construction of the various subcomponents. The predicate device also comprises extension sets, which are not part of this submission. Smiths Medical has determined materials differences do not affect the device's intended use, indication for use, or alter the device's fundamental scientific technology for fluid, blood, or blood product administration. In addition, the performance requirements were met by the tests performed on the device.
#### Non-Clinical Performance Data:
Non-clinical testing of the components comprised in each configuration of the subject devices were assessed and tested appropriately to design controls; i.e., design verification, design validations. The test results conclude that the Intellifuse™ Administration Sets and Intellifuse™ Blood Administration Sets to be substantially equivalent to the predicate device. Testing listed below passed and results were verified against requirements:
- ISO 8536-4, "Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed"
- . ISO 1135-4. " Transfusion equipment for medical use - part 4 Transfusion sets for single use, gravity feed"
- . ISO 594-1, "Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General Requirements"
- ISO 594-2, "Conical fittings with a 6% (Luer) taper for syringes, needles and certain other ● medical equipment - Part 2: Lock fittings"
- Particulate assessment (USP <788>)
- Microbial Ingress (Internal test method)
- Pinch Clamp (Internal test method) ●
- Check Valve (ISO 8536-12)
- Needleless Access Connector (Internal test method) .
- Hydrostatic pressure O
- Translucency O
- O Inspection
- Burette Chamber (ISO 8536-5) ●
{6}------------------------------------------------
- Biocompatibility (ISO 10993-1) .
- In addition, biocompatibility of needleless connector for K163172 was leveraged for o needleless access connector
- Pyrogenicity (material mediated/bacterial endotoxins) o
- Sterilization - SAL 10-6
- ISO 11135, "Sterilization of health care Products Ethylene Oxide Requirements for o development, validation and routine control of a sterilization process for medical devices
- Packaging (transport and distribution) ●
- ISO 11607-1, "Packaging for terminally sterilized medical devices Part 1: o Requirements for materials, sterile barrier systems and packaging systems
#### Clinical Performance Data:
Clinical testing was not necessary for the determination of substantial equivalence.
#### Conclusion:
Smiths Medical concludes the differences between the subject and predicate device do not impact the indications for use, intended use, and do not raise new questions of safety or effectiveness imposed on the patient or device use; therefore, the subject device, for Intellifuse™ Administration Sets and Intellifuse™ Blood Administration Sets, is considered substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.