K192372 · Ascend Hit · LLZ · Feb 14, 2020 · Radiology
Device Facts
Record ID
K192372
Device Name
ASCEND Image Viewer
Applicant
Ascend Hit
Product Code
LLZ · Radiology
Decision Date
Feb 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Ascend device is intended for diagnostic review, quantification and reporting of cardiovascular structures and function of patients with suspected disease to support physicians in diagnosis.
Device Story
ASCEND Image Viewer is a software-only medical device for diagnostic review, quantification, and reporting of cardiovascular echocardiograms. It accepts DICOM-formatted ultrasound images (2D, Color Doppler, M-mode, Spectral Doppler). Used in hospitals and clinics by cardiologists and sonographers, the software runs on standard Windows PCs. Clinicians use mouse/keyboard to interact with the interface, perform manual measurements (linear, volume, velocity, etc.), and review cine loops. Output consists of displayed images and measurement data, which support physicians in diagnosing cardiovascular structures and function. The device facilitates clinical decision-making by providing tools for quantitative analysis of cardiac studies.
Clinical Evidence
Clinical testing conducted in US city-hospital environments with resident cardiologists, cardiology fellows, and sonographers. Testing focused on functional performance and equivalence to the predicate device. Results confirmed effective performance for intended use with no late-arising risks or safety concerns identified.
Technological Characteristics
Software-only medical device; operates on off-the-shelf hardware running Microsoft Windows 7 or higher. Supports DICOM standard for image data. Features include manual measurement tools (linear, volume, velocity, etc.) and cine play review. Web-accessible architecture.
Indications for Use
Indicated for trained clinicians to display, view, and perform measurements on echocardiograms from compatible ultrasound systems, including transthoracic, transesophageal, stress echocardiogram, cath, and vascular study types for patients with suspected cardiovascular disease.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
February 14, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
ASCEND HIT % Robert A. Miller Director of Quality Assurance and Regulatory Affairs 801 Warrenville Road, Suite 200 LISLE IL 60532
Re: K192372
Trade/Device Name: ASCEND Image Viewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: January 13, 2020 Received: January 16, 2020
Dear Mr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192372
Device Name ASCEND Image Viewer
Indications for Use (Describe)
ASCEND Image Viewer is a software-only medical device intended for use by trained clinicians to display medical images. ASCEND Image Viewer provides viewing of and measurements on echocardiograms from compatible ultrasound systems that support transthoracic echocardiogram, transesophageal echocardiogram, stress echocardiogram, cath, and vascular study types.
ASCEND Image Viewer is for prescription use only.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="display:inline-block;"> <span style="padding-right:5px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span style="display:inline-block;"> <span style="padding-right:5px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Section 1
K192372
## 510(k) Summary
(as required by 21 CFR 807.92)
#### Date: 20-Jan-2020
| A. | Submitter's Information: | | | |
|----|------------------------------------------|------------------------------------------------------------------------------|--|--|
| | Name: | ASCEND HIT | | |
| | Address: | 801 Warrenville Road<br>Suite 200<br>Lisle, Illinois 60532 USA | | |
| | Telephone Number: | (844) 413-2610 | | |
| | Contact Name: | Andy Miller | | |
| | Contact Email address: | amiller@ascendhit.com | | |
| B. | Device Information | | | |
| | Proprietary Name: | ASCEND Image Viewer | | |
| | Common or Usual Name: | Image Processing System | | |
| | Classification:<br>Classification Panel: | Class II 21 CFR 892.2050 LLZ<br>Radiology | | |
| | Substantial Equivalence: | ASCEND Image Viewer is substantially equivalent to TomTec<br>Arena (K132544) | | |
Indications for Use:
ASCEND Image Viewer is a software-only medical device intended for use by trained clinicians to display medical images. ASCEND Image Viewer provides viewing of and measurements on echocardiograms from compatible ultrasound systems that support transthoracic echocardiogram, transesophageal echocardiogram, stress echocardiogram, cath, and vascular study types.
ASCEND Image Viewer is for use by prescription only.
Intended Use:
{4}------------------------------------------------
The Ascend device is intended for diagnostic review, quantification and reporting of cardiovascular structures and function of patients with suspected disease to support physicians in diagnosis.
# Device Description:
The ASCEND Image Viewer is a Medical Device for viewing and measuring cardiovascular echocardiograms in DICOM format.
The software has features for displaying and measuring cardiovascular images acquired from echocardiograms. Images supported include:
- . 2D
- Color Doppler
- M-mode
- . Spectral Doppler (Pulsed Wave and Continuous Wave)
The ASCEND Image Viewer provides Cine play review and control.
Standard Review supported study types include:
- TTE
- TEE
- Cath
- . Stress Echo
- Vascular
# Supported Measurements include:
- Linear
- . Volume (method of discs)
- Velocity time integral
- Pressure half-time
- . Area
- Time
- Velocity
- Angle (measurement of two lines with a common vertex drawn on the image) .
The user can make measurements which persist for later use within the ASCEND Image Viewer.
## Operator profile
The device is generally used by medical professionals such as doctors and sonographers in need of displaying echocardiograms for diagnostic purposes.
Intended use environment
{5}------------------------------------------------
The software is intended to be used in Health Care facilities including Office environments within hospitals and clinics.
# Operating principle
- . On desktop PCs the interaction with the software is mainly performed with mouse and/or keyboard
## Primary operating functions
Diagnostic viewing of echocardiograms in ultrasound DICOM format.
- C. Technological Characteristics of the Device as compared to Predicate Device:
| | Predicate Device<br>TomTec-Image Arena<br>TomTec Imaging<br>Systems GmbH<br>(K132544) | Subject Device<br>ASCEND Image Viewer<br>Ascend HIT | Substantially<br>Equivalent |
|-----------------------------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------------|-----------------------------|
| Software only device | Yes | Yes | Yes |
| Web accessible | Yes | Yes | Yes |
| Display Images | still images and image<br>sequences | Yes | Yes |
| Select Images for closer<br>examination | Yes | Yes | Yes |
| Echocardiogram<br>analysis and review<br>capability | Manual only | Manual | Yes |
| Data Source | Multiple modalities | Ultrasound | Yes |
| DICOM compliant | Yes | Yes | Yes |
| Ultrasound mode<br>utilized for echo<br>analysis | B mode | 2D, Color Doppler, M-<br>mode, Spectral Doppler<br>images. | Yes |
| Viewing mode – 2D | Yes | Yes | Yes |
| Viewing mode - 3D | Yes | No | No |
{6}------------------------------------------------
| General Cardiac<br>Measurements | | Distance LV distance Angle Area Elipse Disks Time Velocity Heart rate Acceleration PHT VTI PIRI Slope | Linear (distance)<br>Angle<br>Area<br>Volume (method of disks)<br>Time<br>Velocity<br>VTI<br>PHT | Yes – Different features<br>from predicate have<br>been verified and<br>validation to assure<br>that there is no impact<br>on safety and<br>effectiveness. |
|---------------------------------------|-------------------|-------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display format | | AVI, BMP, JPEG<br>or DCM export | Yes | Yes |
| Minimum Hardware Requirements | | | | |
| Operates on off the<br>shelf hardware | Yes | | Yes | Yes |
| OS | Microsoft Windows | | Microsoft Windows 7<br>and above. | Yes |
# D. Brief Discussion of Test Results Submitted:
# Bench Testing
The test plan covers all aspects of the functionality, exercising every user interface, all menus and submenus, as well as the measurement tools.
The software was tested according to the company's Design Control process. Risks were analyzed according to ISO 14971 to find risks in the non-acceptable area. No risks have been identified in the non-acceptable area. Hence, no measures are considered essential performance characteristics. The software is also verified for compliance with the DICOM standard.
## Clinical Testing
{7}------------------------------------------------
ASCEND Image Viewer was tested in US city-hospital environments throughout the development cycle. Clinical testers consisted of resident cardiologists, cardiology fellows, and sonographers. The clinical test results showed ASCEND Image Viewer performed effectively its intended use. Where the functionality of ASCEND Image Viewer intersected with the predicate device, it was found to be substantially equivalent. No late-arising risks or other safety concerns were recorded during clinical testing.
- E. Conclusions from Test Results:
The Graphic User Interface conforms to the ASCEND Image Viewer functional specification.
The comparison ASCEND Image Viewer with the predicate devices shows that the ASCEND Image Viewer has similar functionality, intended use, technological characteristics, and typical users as the predicate devices. Verification and validation activities ensure that the design specifications are met, and that ASCEND Image Viewer does not introduce new issues concerning safety and effectiveness. Hence ASCEND Image Viewer is substantial equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.