K192330 · Covidien · GDW · Jan 23, 2020 · General, Plastic Surgery
Device Facts
Record ID
K192330
Device Name
EEA Circular Stapler with Tri-Staple Technology
Applicant
Covidien
Product Code
GDW · General, Plastic Surgery
Decision Date
Jan 23, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EEA™ circular stapler with Tri-Staple™ technology has application throughout the alimentary tract for the creation of end-to-end, end-to-side and side-to-side anastomoses in both open and laparoscopic surgeries.
Device Story
Manual, single-use surgical stapler; used in open or laparoscopic alimentary tract surgeries; creates circular anastomoses via intraluminal resection. Device features Tri-Staple technology with three height-progressive titanium staple rows (4.0mm, 4.5mm, 5.0mm) for extra-thick tissue. Operator approximates tissue, squeezes handle to fire; device provides audible/tactile feedback upon completion. Includes Tilt-Top anvil and blunt-tipped trocar for insertion. Output is a stapled, resected tissue connection. Benefits include consistent staple formation and hemostasis in thick tissue applications.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing (visual inspection, force testing, staple formation, leak/burst tests), ex-vivo firing, and in-vivo animal studies (atraumatic tissue test, hemostasis, chronic survival study). Human factors/usability validated per IEC 62366.
Indicated for patients undergoing open or laparoscopic surgery of the alimentary tract requiring end-to-end, end-to-side, or side-to-side anastomoses.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
EEA™ Circular Stapler with Tri-Staple™ Technology (Purple) (K172361)
EEA™ Circular Stapler with DST Series™ Technology (K062850)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 23, 2020
Covidien Frank Gianelli Senior Regulatory Affairs Specialist 60 Middletown Avenue North Haven, Connecticut 06473
Re: K192330
Trade/Device Name: EEA Circular Stapler with Tri-Staple Technology Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: December 16, 2019 Received: December 19, 2019
## Dear Mr. Gianelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Dnug Administration
# Indications for Use
510(k) Number (if known)
K192330
Device Name
EEATM Circular Stapler with Tri-Staple™ Technology
Indications for Use (Describe)
The EEATM circular stapler with Tri-Staple™ technology has application throughout the alimentary tract for the creation of end-to-end, end-to-side and side-to-sicle anastomoses in both open and laparoscopic surgeries.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: underline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17) Page 1 of 1
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
#### Date Prepared:
January 20, 2020
#### Submitter:
Sarah Tang Covidien Rooms 501, 502, 601, 602, No.3 building No.2388 Chen Hang Road Min Hang District, Shanghai, 201114, China
### Contact:
Frank Gianelli (U.S. Agent) on behalf of Sarah Tanq Covidien 60 Middletown Avenue North Haven, CT 06473, USA Senior Regulatory Affairs Specialist Telephone: (203) 492-5352 Fax: (203) 492-5029 frank.j.gianelli@medtronic.com Email:
#### Name of Device:
Proprietary/Trade Name: EEATM Circular Stapler with Tri-Staple™ Technology (Black) Model Numbers: TRIEEA28XT, TRIEEA31XT, TRIEEA33XT, TRIEEAXL28XT, TRIEEAXL31XT, TRIEEAXL33XT Classification Name: Staple, Implantable Requlations Number: 21 CFR 878.4750 Product Codes: GDW, GAG FDA Panel Number: 79 Device Class: Class II Review Panel: General and Plastic Surgery Common Name: Surgical Stapler
#### Predicate Devices:
Primary Predicate Device: Proprietary/Trade Name: 510(k) Number: Classification Name: Regulations Number: Product Codes: FDA Panel Number: Device Class: Review Panel: Common Name:
Additional Predicate Device: Proprietary/Trade Name: 510(k) Number: Classification Name:
EEA™ Circular Stapler with Tri-Staple™ Technology (Purple) K172361 Staple, Implantable 21 CFR 878.4750 GDW, GAG 79 Class II General and Plastic Surgery Surgical Stapler
EEA™ Circular Stapler with DST Series™ Technology K062850 Staple, Implantable
{4}------------------------------------------------
Regulations Number: Product Codes: FDA Panel Number: Device Class: Review Panel: Common Name:
21 CFR 878.4750 GDW, GAG 79 Class II General and Plastic Surgery Surgical Stapler
#### Device Description:
The EEA™ circular stapler with Tri-Staple™ technology (Black) places a circular, triple staggered row of titanium staples and resects the excess tissue, creating a circular anastomosis as an endto-end, end-to-side and side-to-side anastomoses in both open and laparoscopic surgeries. The instrument is activated by squeezing the handle firmly as far as it will go. The subject circular stapler will be offered for an extra thickness range, which is identified by the black staple guide. Staplers with extra thick staple size deploy three height progressive rows of 4.0 mm, 4.5 mm and 5.0 mm titanium staples. The subject circular staplers are available in 3 lumen sizes 28, 31. 33mm and 2 shaft lengths a standard 22 cm shaft and an XL 35 cm shaft. The Tilt-Top™ anvil is available on all staplers. A blunt tipped anvil trocar accessory is provided to assist in introducing the anvil into the surgical field.
The EEA™ Circular Stapler with Tri-Staple™ Technology (Black) is manufactured with patient contact materials (stainless steel, titanium) that are utilized within the predicate devices (K172361 and K062850).
The EEA™ Circular Stapler with Tri-Staple™ Technology (Black) is a manual device which is utilized by approximating tissue, instrumentation activation by squeezing the handle firmly as far as it will go, audible and tactile firing indicator upon completion of the staple firing and removal of the device from the surgical field in the same manner as the predicate devices (K172361 and K062850).
The EEA™ Circular Stapler with Tri-Staple™ Technology (Black) is a single-use device that is packaged and sterilized via ETO (Ethylene Oxide) with a 5-year shelf life, which is the same as the predicate devices (K172361 and K062850).
#### Indications for Use:
The EEA™ Circular Stapler with Tri-Staple™ Technology (Black) has application throughout the alimentary tract for the creation of end-to-end, end-to-side anastomoses in both open and laparoscopic surgeries.
#### Technoloqical Characteristics:
The subject new devices EEA™ circular stapler with Tri-Staple™ technology (Black) are substantially equivalent to primary predicate device K172361 EEA™ Circular Stapler with Tri-Staple™ Technology (Purple) and the additional predicate device K062850 EEA™ circular stapler with DST Series™ technology regarding the fundamental stapling technologies employed, intended use and indications for use. All of them are single-use manual circular staplers that have application throughout the alimentary tract for the creation of end-to-end, end-to-side and side-toside anastomoses in both open and laparoscopic surgeries.
The Tri-Staple™ technology used in the subject devices are fundamentally the same as the primary predicate device K172361 with the exception that the subject devices deploy 3 height progressive rows of 4.0 mm, 4.5 mm and 5.0 mm titanium staples for use in extra thick tissue applications and is identified by the black staple quide. Conversely, the primary predicate devices
{5}------------------------------------------------
deploy 3 height progressive rows of 3.0 mm, 3.5 mm and 4.0 mm titanium staples for use in medium/thick tissue applications and is identified by the purple staple quide. That's why an additional predicate offering for an extra thick tissue thickness is introduced as control device in performance testing.
The additional predicate device K062850 EEA™ circular stapler with DST Series™ technology deploy 2 staggered rows of 4.8mm staples, for use in extra thickness range similar to the subject device. A series of comparison testing has been conducted between the subject new devices and the additional predicate devices and is provided in this submission.
#### Substantial Equivalent:
The subject devices have the same intended use and indications for use as the predicate devices.
They are similar in fundamental scientific technology in that they are all sterile, single used, handheld, manual surgical instruments equipped with titanium staples intended to be used during open or laparoscopic surgical procedures of the alimentary tract, to create anastomoses (end-to-end, end-to-side, or side-to-side) via intraluminal (within the lumen) resection. The subject and predicate devices are similar in design, materials and are sterilized via ethylene oxide.
The below table further summarizes the similarities and differences between the subject and predicate devices.
| | Subject Device | Primary Predicate Device<br>(K172361) | Additional Predicate Device (K062850) |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Features | EEATM Circular Stapler with<br>Tri-StapleTM Technology<br>(Black) | EEATM Circular Stapler with<br>Tri-StapleTM Technology<br>(Purple) | EEATM Circular Stapler with<br>DST SeriesTM Technology |
| Manufacturer | Covidien | Covidien | Covidien |
| Constructional | Image: EEA Circular Stapler with Tri-Staple Technology (Black) | Image: EEA Circular Stapler with Tri-Staple Technology (Purple) | Image: EEA Circular Stapler with DST Series Technology |
| Indications for Use | The EEATM Circular Stapler<br>with Tri-StapleTM Technology<br>has application throughout<br>the alimentary tract for the<br>creation of end-to-end, end-<br>to-side and side-to-side<br>anastomoses in both open<br>and laparoscopic surgeries. | The EEATM Circular Stapler<br>with Tri-StapleTM Technology<br>has application throughout<br>the alimentary tract for the<br>creation of end-to-end, end-<br>to-side and side-to-side<br>anastomoses in both open<br>and laparoscopic surgeries. | The EEATM Circular Stapler<br>with DST SeriesTM<br>Technology has application<br>throughout the alimentary<br>tract for the creation of end-<br>to-end, end-to-side and side-<br>to-side anastomoses in both<br>open and laparoscopic<br>surgeries. |
| Operation Method | Manual | Manual | Manual |
| Anatomical Site | Alimentary tract | Alimentary tract | Alimentary tract |
| Surgical Approach | Open and laparoscopic | Open and laparoscopic | Open and laparoscopic |
| Staple Rows | 3 staggered rows of staples<br>with different staple height in<br>each staple row | 3 staggered rows of staples<br>with different staple height in<br>each staple row | 2 staggered rows of staples<br>with same staple height in<br>each staple row |
{6}------------------------------------------------
| | Subject Device | Primary Predicate Device<br>(K172361) | Secondary Predicate<br>Device (K062850) |
|----------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------|--------------------------------------------------------|
| Features | EEATM Circular Stapler with<br>Tri-StapleTM Technology<br>(Black) | EEATM Circular Stapler with<br>Tri-StapleTM Technology<br>(Purple) | EEATM Circular Stapler with<br>DST SeriesTM Technology |
| Staple Guide<br>Color | Black | Purple | Green |
| Staple Size<br>(open leg height) | 4.0mm, 4.5mm 5.0mm | 3.0mm, 3.5mm, 4.0mm | 4.8mm, 4.8mm |
| Staple Material | Titanium per ASTM F67<br>Grade I | Titanium per ASTM F67<br>Grade I | Titanium per ASTM F67<br>Grade I |
| Lumen Sizes | 28mm/31mm/33mm | 28mm/31mm/33mm | 28mm/31mm/33mm |
| Stapler Length | Standard length: 22cm<br>Extra length XL: 35cm | Standard length: 22cm<br>Extra length XL: 35cm | Standard length: 22cm<br>Extra length XL: 35cm |
| Audible Feedback | Yes | Yes | Yes |
| Anvil Head | 3 staggered rows of anvil<br>bucket, lipless design | 3 staggered rows of anvil<br>bucket, lipless design | 2 staggered rows of anvil<br>bucket, lipped design |
| Safety Lever | Red color | Red color | White color |
| Key Parts | Knife: Stainless Steel | Knife: Stainless Steel | Knife: Stainless Steel |
| Materials | Anvil: Stainless Steel | Anvil: Stainless Steel | Anvil: Stainless Steel |
| | Tube: Aluminum | Tube: Aluminum | Tube: Aluminum |
| Biocompatibility | Evaluated per ISO 10993-1<br>series and FDA 2016<br>biocompatibility guidance | Evaluated per ISO 10993-1<br>series and FDA 2016<br>biocompatibility guidance | Evaluated per ISO 10993-1<br>series |
| Single Use | Yes | Yes | Yes |
| Disposable | Yes | Yes | Yes |
| Sterile | Ethylene oxide | Ethylene oxide | Ethylene oxide |
| Shelf Life | 5 years | 5 years | 5 years |
#### Tests performed to evaluate and compare technological and performance characteristics:
Non-clinical performance data - the following testing has been performed to demonstrate substantial equivalence to the predicate devices.
- 1. Performance Test _ In-Vitro
- . Visual inspection on product, packaging and instruction for use
- . Safety lock release force test
- Staple formation on test media .
- Anvil attach force test
- . Anvil detach force test
- . Clamping force test
- . Unclamping force test
- Firing force test ●
The acceptance criteria of each test within the In-Vitro section has been satisfied and results deemed acceptable.
- 2. Performance Test _ Ex-Vivo
- EX-Vivo Firings .
- Knife cut evaluation .
- . Leak / Burst test
{7}------------------------------------------------
All units evaluated vielded comparable or better results in accordance with the acceptance criteria for Ex-Vivo test, which demonstrate the substantial equivalence of the subject devices as compared to the predicate device.
- 3. Performance Test _ In-Vivo Study
- Atraumatic tissue test ●
- Hemostasis test ●
- Staple formation on tissues
- . Chronic survival study
All units evaluated yielded comparable or better results in accordance with the acceptance criteria for In-Vivo acute test, which demonstrate the substantial equivalence of the subject devices as compared to the predicate device.
The chronic animal study performed to evaluate the performance of the subject device and predicate device shows no differences in healing metrics or anastomotic index.
- 4. Performance Test Human Factors/Usability per IEC 62366 and FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices".
- Human factors evaluation was conducted on the subject device and the predicate . devices. This testing report has been included in the predicate submission (K172361).
- . The usability testing validated the performance of the device meets identified user needs and intended use; validated potential use errors identified which have been mitigated through the product design or instruction; validate training and instruction via the proposed device Instruction for Use on device operation.
- The results of the testing demonstrate the substantial equivalence of the subject ● devices as compared to the predicate device.
- 5. Biocompatibility tests per ISO 10993-1 and FDA guidance "Use of international Standard ISO 10993-1" issued June 16, 2016
- Cytotoxicity test o
- Sensitization .
- Intracutaneous irritation
- . Acute system toxicity
- . Pyrogenicity
The biocompatibility test results support the subject devices are biocompatible.
- 6. Sterilization assessment per ISO 11135
- Overkill method used for validation ●
- ETO residual test
The sterilization validation demonstrates the sterilization cycle used for the single use devices can effectively achieve a minimum Sterility Assurance Level (SAL)of 10-6.
- 7. Stability/Shelf-life studies
- Aging test for 5 years shelf life ●
- . Product and packaging functional test
The stability study results support a 5-year shelf life of the single use devices.
{8}------------------------------------------------
- 8. Clinical performance data No clinical study is deemed necessary since the substantial equivalence has been sufficiently demonstrated by non-clinical studies.
#### Conclusion:
Based upon the intended use, technological characteristics and supporting performance data summarized above, we concluded that the new subject devices EEA™ Circular Stapler with Tri-Staple™ Technology are substantially equivalent to the legally-marketed predicate devices K172361 and K062850.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.