Ultraverse 014 and 018 PTA Balloon Dilatation Catheters
K192318 · Bard Peripheral Vascular, Inc. · LIT · Oct 3, 2019 · Cardiovascular
Device Facts
Record ID
K192318
Device Name
Ultraverse 014 and 018 PTA Balloon Dilatation Catheters
Applicant
Bard Peripheral Vascular, Inc.
Product Code
LIT · Cardiovascular
Decision Date
Oct 3, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Ultraverse® 014 and Ultraverse® 018 PTA Balloon Dilatation Catheters are recommended for use in percutaneous trasluminal angioplasty (PTA) of the renal, popliteal, tibial, femoral, and peroneal arteries. These catheters are not for use in coronary arteries.
Device Story
Over-the-wire PTA balloon dilatation catheter; distal balloon fixed to catheter tip; includes atraumatic tip and hydrophilic coating. Used in peripheral vascular interventions (renal, popliteal, tibial, femoral, peroneal arteries). Operated by physicians in clinical settings. Input: guidewire (0.014" for 014 platform; 0.014" or 0.018" for 018 platform). Mechanism: balloon inflation via luer lock hub to dilate stenosis. Features GeoAlign® Marking System (non-radiopaque ruler on shaft) for external measurement of intravascular advancement/retraction and geographic alignment of adjunctive therapies. Radiopaque markers delineate balloon working length. Output: mechanical dilation of arterial stenosis. Benefits: restores vessel patency in peripheral artery disease.
Clinical Evidence
Bench testing only. No clinical data. Testing included balloon burst pressure, fatigue, tensile strength, trackability, inflation/deflation, and GeoAlign® marking durability/legibility. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity, hemocompatibility).
Indicated for percutaneous transluminal angioplasty (PTA) of renal, popliteal, tibial, femoral, and peroneal arteries. Contraindicated for use in coronary arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters (K121856)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 3, 2019
Bard Peripheral Vascular, Inc. % Prithul Bom Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K192318
Trade/Device Name: Ultraverse 014 and 018 PTA Balloon Dilatation Catheters Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: LIT Dated: August 23, 2019 Received: August 26, 2019
Dear Mr. Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters
Indications for Use (Describe)
Ultraverse® 014 and Ultraverse® 018 PTA Balloon Dilatation Catheters are recommended for use in percutaneous trasluminal angioplasty (PTA) of the renal, popliteal, tibial, femoral, and peroneal arteries. These catheters are not for use in coronary arteries.
| | Type of Use (Select one or both, as applicable) |
|--|-------------------------------------------------|
|--|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Ultraverse® 014 and 018 PTA Balloon Dilatation Catheter 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
### Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>Tempe, Arizona 85281 |
|------------|-------------------------------------------------------------------------------|
| Phone: | 480-449-2587 |
| Fax: | 480-449-2546 |
| Contact: | Andrew Quach, Regulatory Affairs Associate |
| Date: | September 20, 2019 |
#### Subject Device Name:
| Device Trade Name: | Ultraverse® 014 and 018 PTA Balloon Dilatation<br>Catheters |
|-----------------------|-------------------------------------------------------------|
| Common or Usual Name: | Catheter, Angioplasty, Peripheral, Transluminal |
| Product Code: | LIT |
| Classification: | Class II |
| Review Panel: | Cardiovascular |
| Regulation Number: | 21 CFR 870.1250 |
### Predicate Device:
- Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters (K121856; cleared July 11, 2012)
#### Reference Device:
- . Ultraverse® 035 PTA Balloon Dilatation Catheters (K142261; cleared September 24, 2014)
{4}------------------------------------------------
### Device Description:
The Ultraverse® 014 and 018 PTA Balloon Dilatation Catheter is a small vessel balloon catheter consisting of an over the wire catheter with a balloon fixed at the distal tip. For all balloon lengths, radiopaque markers delineate the working length of the balloon and aid in balloon placement. For balloon lengths of 100 mm and greater, two radiopaque markers are positioned on the distal portion of the balloon and one radiopague marker is positioned on the proximal portion of the balloon to differentiate between the distal and proximal ends of the balloon. The catheter includes an atraumatic tip to facilitate advancement of the catheter to and through the stenosis as well as a hydrophilic coating. The coaxial lumen, over the wire catheter is compatible with an 0.014" guidewire for the 014 platform, and compatible with an 0.014" or 0.018″ guidewire on the 018 platform, and is available in 75, 90, 100, 150 and 200 cm working lengths. The proximal portion of the catheter includes a female luer lock hub connected to the inflation lumen, and a female luer-lock hub connected to the guidewire lumen. Packaged with every product is a profile reducing sheath that is positioned over the balloon for protection before use. A stylet is placed into the tip of the catheter to aid in rewrap/refolding of the balloon. This product is not manufactured with any latex.
The GeoAlign® Marking System is a non-radiopaque ruler on the catheter shaft measured from the distal tip. The GeoAlign® markings are designated on the catheter shaft by 1cm increment bands with an accuracy within ±1mm. The distance from the distal catheter tip is labeled in 10cm increments. Thicker bands denote the midway point (5cm) between the labeled distances. The GeoAlign® Marking System is designed to be used as a tool to externally measure the intravascular advancement and/or retraction of the catheter. This can provide an intravascular reference regarding the location of the distal tip of the catheter or an approximate intravascular length measurement between two points. The GeoAlign® Marking System may also facilitate geographic alignment of an adjunctive therapy that includes the same GeoAlign® Marking System.
| Attribute | Ultraverse® 014 and 018 PTA Balloon Dilatation Catheter |
|----------------------------|---------------------------------------------------------|
| Balloon Diameter (mm) | 1.5, 2, 2.5, 3, 3.5, 4, 5, 6, 7, 8, 9 |
| Balloon Length (mm) | 20, 40, 60, 80, 100, 120, 150, 200, 220, 250, 300 |
| Catheter Shaft Length (cm) | 75, 90, 100, 130, 150, 200 |
| .014" Platform Introducer | 4F, 5F |
| Sheath Compatibility | |
| .018" Platform Introducer | 4F, 5F, 6F |
| Sheath Compatibility | |
{5}------------------------------------------------
## Indications for Use of Device:
The Ultraverse® 014 and Ultraverse ® 018 PTA Balloon Dilatation Catheters are recommended for use in percutaneous transluminal angioplasty (PTA) of the renal, popliteal, tibial, femoral and peroneal arteries. These catheters are not for use in coronary arteries.
# Comparison of Indications for Use to Predicate Device:
The indications for use statement for the subject device, the Ultraverse® 014 and 018 PTA Balloon Dilatation Catheter, are the same as compared to the predicate device. Therefore, the subject device, the Ultraverse® 014 and 018 PTA Balloon Dilatation Catheter, is substantially equivalent to the predicate device.
# Technological Comparison to Predicate Devices:
The subject Ultraverse® 014 and 018 PTA Dilatation Balloon Catheters have the following similarities to the Ultraverse® 014 and 018 PTA Balloon Catheters predicate device (K121856 - cleared on July 11, 2012):
- Same intended use .
- Same indications for use
- Same target population ●
- Same operating principle ●
- . Same fundamental scientific technology
- . Same sterility assurance level
- . Same method of sterilization
- Similar packaging materials and configuration ●
- . Same shelf life
The subject device, Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters has the following differences when compared to the predicate device:
- Increased radiopaque marker bands lengths and additional marker band . added to the distal end of the balloon for balloon lengths 100mm or longer
- . GeoAlign® Marking System added along the shaft of the catheter
- Opaque colorant used instead of clear shaft color .
{6}------------------------------------------------
- Strain relief on Ultraverse® 014 platform changed to blue color to . differentiate product
- . 0.014" guidewire compatibility for Ultraverse®018 platform
- Extension of product size offering .
- . Extended 4Fr compatibility range
#### Performance Data:
To demonstrate substantial equivalence of the subject device to the predicate device, its technological characteristics and performance criteria were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, the following in vitro tests were performed on the subject device:
- Catheter Shaft Length .
- Balloon Working Length .
- Marker Band Alignment
- Balloon OD at OP .
- Balloon Rated Burst Pressure, Leak, . & Burst Mode
- Crossing Profile ●
- . Sheath Compatibility
- . Shaft Outer Diameter
- Balloon Compliance / Distensibility ●
- . Fatigue
- . Hub to Shaft Tensile
- Guidewire Compatibility .
- Inflation .
- . Deflation
- Balloon to Shaft Tensile .
- Catheter Elongation .
- . Flushability
- GeoAlign® Marking Positions
- GeoAlign® Marking Durability .
- GeoAlign® Marking Legibility .
- Trackability ●
- . Reinsertion
- Kink .
The results from these tests demonstrate that the technological characteristics and performance criteria of the Ultraverse® 014 & 018 PTA Balloon Dilatation Catheters are substantially equivalent to the predicate device, and that it can perform in a manner equivalent to devices currently on the market for the same intended use.
{7}------------------------------------------------
The following tests performed for the predicate device are leveraged as applicable tests for the subject device for components which remained unchanged from the predicate device:
- Radiopacity .
- . Tip Taper / Tip Length
- . Stylet / Refold
- . Hub Torsion / Hub Stress
- Device Compatibility .
- . Packaging Integrity
- Packaging Visual Inspection o
- Dye Penetration o
- Pouch Seal Strength o
# Biocompatibility:
To demonstrate substantial equivalence of the subject device, the Ultraverse® 014 and 018 PTA Balloon Dilatation Catheter to the predicate device, the following biocompatibility testing was performed in accordance ISO 10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process.
- . Cytotoxicity
- Sensitization .
- Intracutaneous Reactivity .
- . Acute Systemic Toxicity
- Material Mediated Pyrogenicity .
- . Hemocompatibility (Hemolysis, Complement Activation, Thrombogenicity)
- Chemical Characterization .
The results from these tests demonstrate that the subject device, the Ultraverse® 014 & 018 PTA Balloon Dilatation Catheters, is comparable to the predicate device and that it is considered safe and biocompatible for its intended use.
# Conclusions:
The subject device, the Ultraverse® 014 & 018 PTA Balloon Dilatation Catheters, met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The Ultraverse® 014 & 018 PTA Balloon Dilatation Catheters is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.