The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone. Indications for the Cephalomedullary nails include: • Compound and simple shaft fractures • Proximal, metaphyseal and distal shaft fractures • Segmental fractures • Comminuted fractures • Fractures involving osteopenic and osteoporotic bone • Pathological fractures • Fractures with bone loss • Pseudoarthrosis, non-union, mal-union and delayed union • Periprosthetic fractures • Surgically created defects such as osteotomies • Intertrochanteric and subtrochanteric fractures
Device Story
Zimmer Natural Nail System Cephalomedullary Nails are intramedullary fixation devices for femur fracture stabilization. System comprises nails, screws, and nail caps; nails secured by screws passing through nail holes; set screw with polyethylene peg provides guided rotational stability for lag screw. Components available in various lengths/diameters; color-coded for identification. Used by orthopedic surgeons in clinical/OR settings to restore pre-injured bone shape. Provides temporary fixation to facilitate healing. No software or electronic components.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials: Tivanium Ti-6Al-4V Alloy (nails, caps, screws) and Polyethylene (set screw pegs). Intramedullary fixation rod design. Mechanical stabilization via screw-in-nail fixation. Color-coded component system. Sterilization processes consistent with predicate devices.
Indications for Use
Indicated for patients requiring temporary fracture fixation and stabilization of the femur, including compound, simple, segmental, comminuted, pathological, periprosthetic, intertrochanteric, and subtrochanteric fractures; fractures with bone loss; fractures in osteopenic/osteoporotic bone; and conditions including pseudoarthrosis, non-union, mal-union, delayed union, and osteotomies.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
Zimmer Natural Nail System Cephalomedullary Nails (K091566)
Zimmer Natural Nail System Cephalomedullary Femoral Nails - Asia Short (K120715)
Submission Summary (Full Text)
{0}------------------------------------------------
October 11, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Zimmer GmbH Danielle Madureira Senior Specialist, Regulatory Affairs Sulzerallee 8 CH-8404 Winterthur Switzerland
Re: K192312
Trade/Device Name: Zimmer Natural Nail System Cephalomedullary Nails Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: August 21, 2019 Received: August 26, 2019
Dear Danielle Madureira:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Stereotaxic. Trauma and Restorative Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K192312
Device Name
ZIMMER® NATURAL NAIL® SYSTEM CEPHALOMEDULLARY NAIL
Indications for Use (Describe)
The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone. Indications for the Cephalomedullary nails include:
- · Compound and simple shaft fractures
- · Proximal, metaphyseal and distal shaft fractures
- Segmental fractures
- Comminuted fractures
- · Fractures involving osteopenic and osteoporotic bone
- · Pathological fractures
- · Fractures with bone loss
- · Pseudoarthrosis, non-union, mal-union and delayed union
- · Periprosthetic fractures
- · Surgically created defects such as osteotomies
- · Intertrochanteric and subtrochanteric fractures
Type of Use (Select one or both, as applicable)
|X | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Zimmer Biomet logo. The logo consists of a blue circle with a stylized "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. The color of the text is a dark gray.
## K192312
# 510(k) Summary
| Sponsor: | Zimmer GmbH<br>Sulzerallee 8, P.O. Box<br>8404 Winterthur, Switzerland |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Danielle Jannuzzi Madureira<br>Senior Specialist, Regulatory Affairs<br>Telephone: +41 58 854 82 60<br>Fax: +41 52 244 86 58 |
| Date: | October 21, 2019 |
| Trade Name: | ZIMMER® NATURAL NAIL® SYSTEM CEPHALOMEDULLARY NAIL |
| Classification Product Code : | HSB - Intramedullary Fixation Rod |
| Device Classification Name: | Rod, fixation, intramedullary and accessories |
| Regulation Number / Description: | 21 CFR § 888.3020 Intramedullary fixation rod |
| Predicate Device: | Zimmer Natural Nail System Cephalomedullary Nails, manufactured<br>by Zimmer GmbH, K091566, cleared October 2009 and its line<br>extension Zimmer Natural Nail System Cephalomedullary Femoral<br>Nails - Asia Short, manufactured by Zimmer GmbH, K120715,<br>cleared October 2012 |
| Device Description: | The Zimmer® Natural Nail® System Cephalomedullary Nails are<br>temporary fixation intramedullary nails designed for fracture fixation<br>and stabilization of the femur. They restore the shape of preinjured<br>bone and are available in a variety of lengths and diameters to meet<br>assorted anatomical needs. Nail caps are available to protect the nail<br>threads from tissue ingrowth and extend the nail length if necessary.<br>Each of the intramedullary nails is secured by a series of screws that<br>pass through holes in the nail. A set screw (with a polyethylene peg)<br>allows guided rotational stability of the lag screw. Nails, nail caps and<br>screws are made from Tivanium Ti-6Al-4V Alloy. Set screws are made<br>from Tivanium Ti-6Al-4V Alloy and Polyethylene (PE).<br>Selected components of the Zimmer Natural Nail System are color<br>coded to aid in identifying which components should be used<br>together. The package label for each of the nails indicates which<br>screw sizes should be used with that nail. A screw case that holds<br>unused screws intended for resterilization is available and contains<br>designated sections for each screw size. Each section contains the<br>color name that matches the color/color name on the label of the<br>appropriate screw size for that section. Refer to the color coding chart<br>and/or surgical technique for more detailed instructions on the use of<br>Zimmer Natural Nail System components. |
| Intended Use: | The Zimmer Natural Nail System is intended for temporary fracture<br>fixation and stabilization of the bone.<br>Indications for the Cephalomedullary nails include:<br>• Compound and simple shaft fractures<br>• Proximal, metaphyseal and distal shaft fractures<br>• Segmental fractures<br>• Comminuted fractures<br>• Fractures involving osteopenic and osteoporotic bone<br>• Pathological fractures<br>• Fractures with bone loss<br>• Pseudoarthrosis, non-union, mal-union and delayed union<br>• Periprosthetic fractures<br>• Surgically created defects such as osteotomies<br>• Intertrochanteric and subtrochanteric fractures |
| Comparison to Predicate Device: | The intended use of the modified devices, as described in its labeling,<br>has not changed as a result of the modifications proposed in the<br>present submission. Zimmer GmbH is seeking clearance for certain<br>system specific Class II instruments - these instruments that have<br>previously considered Class I exempt and correction of classification<br>to Class II is proposed within present submission. |
| Performance Data (Nonclinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions:<br>Class II: Amendment of Design Controls with verification of<br>mechanical integrity and resistance.<br>Clinical Performance and Conclusions:<br>Clinical data and conclusions were not needed for this device. |
| Conclusion: | The subject devices have the same intended use and similar<br>indications for use as the predicate devices. The subject devices use<br>the same operating principle, incorporate the same basic design and<br>labeling and are manufactured and sterilized using the same materials<br>and processes as the predicate devices.<br>Except for the modifications described in this submission the subject<br>devices are identical to the predicate devices, and the performance<br>data and analyses demonstrate that:<br>• any differences do not raise new questions of safety and<br>effectiveness as established with performance testing; and |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Zimmer Biomet. The logo consists of a blue circle with a stylized "Z" inside, followed by the text "ZIMMER BIOMET" in a gray sans-serif font. The logo is simple and modern, with a focus on the company name.
- the subject devices are at least as safe and effective as the • legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.