TidalPort-AP Implantable Apheresis Vascular Access Port
K192291 · Norfolk Medical Products, Inc. · PTD · Aug 20, 2020 · General Hospital
Device Facts
Record ID
K192291
Device Name
TidalPort-AP Implantable Apheresis Vascular Access Port
Applicant
Norfolk Medical Products, Inc.
Product Code
PTD · General Hospital
Decision Date
Aug 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TidalPort-AP™ Implantable Apheresis Vascular Access Port is indicated for therapies requiring repeated access to the vascular system. The port system can be used for long-term therapeutic apheresis, withdrawal of blood, and infusion of medications, I.V. fluids, parenteral nutrition solutions, blood, and blood products. The TidalPort-AP™ Implantable Apheresis Vascular Access Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 mL/s.
Device Story
TidalPort-AP is a totally implantable, titanium-based vascular access port with a silicone septum and 9.6F radiopaque polyurethane catheter. Designed for long-term therapeutic apheresis, blood withdrawal, and infusion. Implanted subcutaneously with catheter tip placed in lower 1/3 of superior vena cava. Accessed percutaneously via perpendicular insertion of 16G or 18G high-flow non-coring needles for apheresis, or Huber point needles for routine infusion. Provides power injection capability for contrast media. Used in clinical settings by healthcare providers. Benefits include reliable repeated vascular access without daily external catheter care or repeated venipuncture. System includes port, catheter, and securement components; provided sterile via EtO.
Clinical Evidence
Bench testing only. Performance evaluations included catheter-to-port connection, tensile strength, radiopacity, gravity flow rate, burst pressure, corrosion resistance, septum puncture/leak, clearance volume, power injection, simulated apheresis, recirculation, hemolysis, MRI safety, and packaging/ship testing. Biocompatibility testing performed per ISO 10993-1.
Technological Characteristics
Titanium port body with silicone septum; 9.6F radiopaque polyurethane catheter. Perpendicular access via non-coring needle. Sterilized via EtO. Materials comply with ASTM F67 (titanium), ASTM F640 (radiopacity). Standards: EN ISO 10555-1/6, ISO 10993-1/7, ISO 11135, ISO 14971, ISO 11607-1.
Indications for Use
Indicated for patients requiring long-term repeated vascular access for therapeutic apheresis, blood withdrawal, and infusion of medications, IV fluids, parenteral nutrition, blood, and blood products. Also indicated for power injection of contrast media at a maximum rate of 5 mL/s.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
PowerFlow™ Implantable Apheresis IV Port with 9.6 Fr. ChronoFlex Catheter (K163001)
Norfolk Medical SportPort™ (now TidalPort™) Family of Vascular Access Ports (K112713)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
August 20, 2020
Norfolk Medical Products, Inc. Natan Pheil Product Development Manager 7350 N. Ridgeway Avenue Skokie, Illinois 60076
Re: K192291
Trade/Device Name: TidalPort-AP Implantable Apheresis Vascular Access Port Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port And Catheter Regulatory Class: Class II Product Code: PTD Dated: July 17, 2020 Received: July 20, 2020
Dear Natan Pheil:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192291
Device Name
TidalPort-APTM Implantable Apheresis Vascular Access Port
Indications for Use (Describe)
The TidalPort-AP™ Implantable Apheresis Vascular Access Port is indicated for therapies requiring repeated access to the vascular system. The port system can be used for long-term therapeutic apheresis, withdrawal of blood, and infusion of medications, I.V. fluids, parenteral nutrition solutions, blood, and blood products.
The TidalPort-AP™ Implantable Apheresis Vascular Access Port is indicated for power injection of contrast media. For power injection of contrast media, the maximium recommended infusion rate is 5 mL/s.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Norfolk Medical. The logo consists of the word "Norfolk" in large, bold font, followed by a red heart-shaped symbol, and then the word "Medical" in the same font as "Norfolk". Below the logo is the address "7350 N. Ridgeway • Skokie, IL, USA • 847-674-7075".
## K192291
## 510(k) Summary 21 CFR 807.92
# TidalPort-AP™ Implantable Apheresis Vascular Access Port
August 17, 2020
#### Submitter/Sponsor Information 5.1
| Submitter Name:<br>FDA Establishment | Norfolk Medical Products, Inc. |
|--------------------------------------|------------------------------------------|
| Registration Number: | 1450392 |
| Address: | 7350 N. Ridgeway<br>Skokie, IL 60076 |
| Telephone Number: | (847) 674-7075 |
| Fax Number: | (847) 674-7066 |
| Contact Person: | Natan Pheil, Product Development Manager |
#### 5.2 Device Name
5.3
| Trade Name: | TidalPort-APT™ Implantable Apheresis Vascular<br>Access Port |
|-------------------------|------------------------------------------------------------------------------|
| Common Name: | Subcutaneous implanted apheresis port |
| Classification Panel: | 80 General Hospital |
| Regulation: | 21 CFR 880.5965 |
| Regulation Name: | Subcutaneous, Implanted, Intravascular Infusion Port and<br>Catheter |
| Class: | Class II |
| Product Code: | PTD |
| Predicate Device | |
| Device Name: | PowerFlow™ Implantable Apheresis IV Port with<br>9.6 Fr. ChronoFlex Catheter |
| Premarket Notification: | K163001 |
| Product Code: | PTD |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Norfolk Medical. The logo consists of the word "Norfolk" in large, bold letters, followed by a red heart symbol, and then the word "Medical" in the same font as "Norfolk". Below the logo is the address "7350 N. Ridgeway • Skokie, IL, USA • 847-674-7075".
| Predicate Device | |
|-------------------------|-----------------------------------------------------------------------------|
| Device Name: | Norfolk Medical SportPort™ (now TidalPort™) Family of Vascular Access Ports |
| Premarket Notification: | K112713 |
| Product Code: | LJT |
#### 5.4 Device Description
The TidalPort-AP™ has one port size (Standard) and one catheter size (9.6F Polyurethane). It is designed to provide repeated access to the vascular system without the need for repeated venipuncture or the daily care of an external catheter. The TidalPort-APTM is available as a standard profile totally implantable, titanium port-based design and is accessed perpendicularly to the skin like a typical, conventional access port. For the purpose of apheresis, high-flow procedures, it is accessed with using the FDA Cleared 16G or 18G high-flow, non-coring needle Tidal High-Flow Non-coring Needle (K151341).
The TidalPort-APTM can also be used for routine vascular access infusion or withdrawal using Huber point needle. For power injection infusion procedures, the subject device can be accessed with a power injection rated needle to create a power-injectable system.
The design of the TidalPort-AP™ utilizes a spherical reservoir and an elongated radius contoured septum to achieve the design purpose of creating a port with a smaller reservoir and less clearance volume. The TidalPort-APTM comes with a number of kit components to aid in the implantation procedure and/or access of the device once implanted. The TidalPort-AP™ and necessary kit components are provided sterile (EtO).
The overall implanted port system consists of three primary components: the titanium port body with a silicone septum, an attachable radiopaque polyurethane catheter, and a catheter lock which secures the catheter to the port body stem. Once implanted, the method of accessing the subject TidalPort-APTM device is the exact same as the predicate TidalPort™ device. After the implanted device has been identified and access is prepped per institutional policy, the user palpates the uniquely shaped port. Once the port is palpated, providing the location of the septum, the 16G or 18G high-flow needle (K151341) is inserted into the reservoir for apheresis procedure use. After the reservoir floor is reached, the stylet is unlocked and pulled back slightly, and the needle is once again advanced forward until contacting the reservoir floor again. The stylet is then completely removed, leaving the hollow cannula with luer lock fitting in place. After stylet removal, the cannula is attached to the appropriate extension set and secured for the necessary infusion or withdrawal procedure.
The kit components provided to aid in the implantation procedure and/or access of the device once implanted include:
Implantation Placement Kit:
- 1 x TidalPort-APTM Implantable Apheresis Vascular Access Port .
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Norfolk Medical. The logo consists of the word "Norfolk" in a serif font, followed by a red heart-shaped symbol with a white outline and a small drop at the bottom. To the right of the heart is the word "Medical" in the same serif font as "Norfolk."
7350 N. Ridgeway · Skokie, IL, USA · 847-674-7075
- 1 x Radiopaque Polyurethane Catheter (9.6F) .
- 2 x Catheter Securement Boot .
- 1 x Syringe, 5cc with Luer Lock .
- 1 x Barbed Malleable Tunneler .
- . 1 x Vein Pick
- 1 x Blunt Flushing Cannula .
- . 1 x Needleless Injection Site
- 1 x 22G Straight Huber Point Needle w/Luer Lock Connector .
- . 1 x Safety Scalpel with #11 Blade
Standard Introducer Kit (optional, pouched separately):
- 1 x Syringe, 10cc with Luer Slip .
- 1 x 18G x 2.75" Introducer Needle with Echogenic Tip .
- . 1 x J-Tip Guidewire with Straightener, 0.035-inch OD x 45 cm length
- 1 x Standard Introducer, Peel-Apart Sheath, with VesselDilator o
Valved Introducer Kit (optional, pouched separately):
- 1 x Syringe, 10cc with Luer Slip .
- 1 x 18G x 2.75" Introducer Needle, Echogenic ●
- 1 x J-Tip Guidewire with Straightener, 0.035-inch OD x 45 cm length .
- . 1 x Valved Introducer, Peel-Apart Sheath, with VesselDilator
- Tidal High-Flow Needle (pouched separately):
- o 1 x 16G x 1" Tidal High-Flow Non-coring Needle
#### న్.న Indications for Use Statement
The TidalPort-AP™ Implantable Apheresis Vascular Access Port is indicated for therapies requiring repeated access to the vascular system. The port system can be used for long-term therapeutic apheresis, withdrawal of blood, and infusion of medications, I.V. fluids, parenteral nutrition solutions, blood, and blood products.
The TidalPort-AP™ Implantable Apheresis Vascular Access Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 mL/s.
#### 5.6 Technological Characteristics Summary
The proposed device is substantially equivalent to Bard's PowerFlow Implantable Apheresis Port cleared under 510(k) premarket notification K163001 and Norfolk Medical's TidalPort cleared under 510(k) premarket notification K112713. The subject device is designed to be accessed with a 16G or 18G High-Flow, Non-coring Needle (K151341) for apheresis procedures and has a 9.6F catheter intended to provide optimal flow rates for patients requiring therapeutic apheresis.
An Implantable Port System with a silicone septum for access is the primary technological principle for both the subject and predicate devices. Both port systems have the same primary components: a titanium port w/silicone septum, a radiopaque polyurethane catheter securement mechanism that secures the catheter to the port pin.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Norfolk Medical. The logo consists of the words "Norfolk Medical" in a serif font, with a red heart-shaped symbol in between the two words. Below the logo is the address "7350 N. Ridgeway • Skokie, IL, USA" and the phone number "847-674-7075".
# Technological Characteristics:
| Feature | Subject Device<br>(TidalPort-AP™) | Predicate<br>(PowerFlow) | Predicate<br>(SportPort) | Comparison –<br>same/different |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| 510(k) Number | K192291 | K163001 | K112713 | |
| Indications For Use | The TidalPort-<br>AP™<br>Implantable<br>Apheresis<br>Vascular Access<br>Port is indicated<br>for therapies<br>requiring<br>repeated access<br>to the vascular<br>system. The port<br>system can be<br>used for long-<br>term therapeutic<br>apheresis,<br>withdrawal of<br>blood, and<br>infusion of<br>medications, I.V.<br>fluids, parenteral<br>nutrition<br>solutions, blood,<br>and blood<br>products.<br>The TidalPort-<br>AP™<br>Implantable<br>Apheresis<br>Vascular Access<br>Port is indicated<br>for power<br>injection of<br>contrast media.<br>For power<br>injection of<br>contrast media,<br>the maximum<br>recommended<br>infusion rate is 5<br>mL/s. | The Bard<br>PowerFlow™<br>Implantable Apheresis<br>IV Port is indicated<br>for patient therapies<br>requiring repeated<br>access to the<br>vascular system. The<br>port system can be<br>used for long-term<br>therapeutic apheresis,<br>withdrawal of blood,<br>and infusion of<br>medications, I.V.<br>fluids, parenteral<br>nutrition solutions,<br>blood and blood<br>products.<br>The Bard<br>PowerFlow™<br>Implantable Apheresis<br>IV Port is indicated<br>for power injection of<br>contrast media. For<br>power<br>injection of contrast<br>media, the maximum<br>recommended<br>infusion rate is 5<br>mL/s. | The SportPort™ is<br>indicated for use<br>when the patient<br>requires the<br>following: repeated<br>access to the<br>vascular system for<br>injections, infusion<br>of drugs,<br>administration of<br>blood or blood<br>products, and/or<br>withdrawal of<br>blood as part of the<br>therapy regimen.<br>When used with a<br>power injectable<br>needle infusion set,<br>the SportPort™ is<br>indicated for power<br>injection of contrast<br>media. For power<br>injection of contrast<br>media, the<br>maximum<br>recommended<br>infusion rate is<br>5ml/s with a 19 or<br>20 gauge non-<br>coring power<br>injectable needle<br>and 2ml/s with a 22<br>gauge non-coring<br>power injectable<br>needle. | 1. Same except for<br>device name<br>2. Different |
| Port Catheter Size | 9.6F Polyurethane | 9.6F Polyurethane | 9.6F Polyurethane | Same |
| Port Access Device | 22, 20, 19, 18, or<br>16-Gauge Needle | 16- or 14-Gauge IV<br>Catheter | 22, 20, 19, 18, or<br>16-Gauge Needle | 1. Different<br>2. Same |
| Septum | Silicone Septum | Single Layer Valve | Silicone Septum | 1. Different<br>2. Same |
| 7350 N. Ridgeway • Skokie, IL, USA • 847-674-7075 | | | | |
| Port Entry | Perpendicular<br>with non-coring<br>needle | Angle entry funnel<br>with an introducer<br>needle stop feature<br>for port system<br>access | Perpendicular with<br>non-coring needle | 1. Different<br>2. Same |
| Long-term Duration<br>of Use | Repeated access to<br>vascular system | Repeated access to<br>vascular system | Repeated access to<br>vascular system | Same |
| Subcutaneous<br>Implantation | Tunneled and<br>inserted into blood<br>vessel | Tunneled and<br>inserted into blood<br>vessel | Tunneled and<br>inserted into blood<br>vessel | Same |
| Catheter Insertion<br>Site | External jugular,<br>internal jugular, or<br>subclavian veins | External jugular,<br>internal jugular, or<br>subclavian veins | External jugular,<br>internal jugular, or<br>subclavian veins | Same |
| Catheter Tip<br>Placement | Lower 1/3 of<br>superior vena cava | Lower 1/3 of<br>superior vena cava | Lower 1/3 of<br>superior vena cava | Same |
| Catheter Tip | Open ended tip | Open ended tip | Open ended tip | Same |
| Port Body Materials | Titanium and<br>Silicone | Titanium Covered<br>with Silicone | Titanium and<br>Silicone | Same |
| Method of<br>Sterilization | EO Gas | EO Gas | EO Gas | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Norfolk Medical. The logo consists of the word "Norfolk" in a serif font, followed by a stylized red heart shape. To the right of the heart is the word "Medical" in the same serif font as "Norfolk". The heart has a small "TM" symbol at the bottom.
7350 N. Ridgeway • Skokie, IL, USA • 847-674-7075
The Indication For Use statements between the TidalPort-AP™ and PowerFlow are the same. It differs from the SportPort™ in that the TidalPort-AP™ adds an indication for therapeutic apheresis, and as such, is limited to one port size and one catheter size.
The differences between the TidalPort-AP™ and the Predicate (PowerFlow) relate to the Port Access Device, Septum, and Port Entry. The differences do not impact the safety and effectiveness of the Port Access Device, 16-Gauge and 18-Gauge needles are used for apheresis today. The silicone septum creates a selfsealing unit and is easy to identify. Finally, the port entry, being perpendicular to the skin, is well-established which makes the TidalPort-AP™ access technique familiar to clinicians. The performance tests conducted below demonstrate that the differences do not raise new or different questions of safety or effectiveness.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Norfolk Medical. The logo consists of the words "Norfolk" and "Medical" in a serif font, with a red heart-shaped symbol in between. Below the logo is the address "7350 N. Ridgeway • Skokie, IL, USA" and the phone number "847-674-7075".
#### 5.7 Discussion of Non-clinical Tests
Norfolk Medical develops product specifications based on design input and risk analysis activities related to the intended use of the product. These product specifications are used to create appropriate design verification tests with reference/guidance to established standards (listed below). For this application, the tests conducted were based on the following standards, specifically in reference to septum puncture / port leak information:
- FDA Guidance on 510(k) Submissions for Implanted Infusion Ports, dated October 1990 was ● used as a reference for the performance testing.
- FDA Guidance on Implanted Blood Access Devices for Hemodialysis
- FDA Guidance Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Device Labeled as Sterile
- FDA Guidance Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance Environment
- FDA Guidance Use of International Standard ISO 10993, Biological evaluation of medical ● devices - Part 1: Evaluation and testing within a risk management process
| • | EN ISO 10555-1:2013 | Sterile, single-use intravascular catheters - Part 1: General<br>requirements |
|---|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • | EN ISO 10555-6:2015 | Intravascular catheters - Sterile, single use catheters - Part 6:<br>Subcutaneous implanted ports |
| • | BS EN ISO 10993-1:2010 | Biological evaluation of medical devices -- Part 1: Evaluation and<br>testing within a risk management process |
| • | BS EN ISO 10993-7:2008 | Biological evaluation of medical devices -- Part 7: Ethylene oxide<br>sterilization residuals |
| • | BS ENISO 11135:2014 | Sterilization of health care products - Ethylene oxide - Part 1:<br>Requirements for development, validation and routine control<br>of a sterilization process for medical devices |
| • | BS EN ISO 14971:2012 | The application of risk management to medical devices |
| • | ISO 11607-1:2010 | Packaging for terminally sterilized medical devices Part 1:<br>Materials, sterile barrier systems and packaging systems |
| • | BS ENISO 14937:2009 | Sterilization of health care products - General requirements for<br>characterization of a sterilizing agent and the development,<br>validation, and routine control of a sterilization process for medica<br>devices |
| • | ASTM F67-13(2017) | Standard Specification for Unalloyed Titanium, for Surgical<br>Implant Applications |
| • | ASTM F86-13 | Standard practice for surface preparation and marking of metallic<br>surgical implants |
| • | ASTM F640-12 | Standard test methods for determining radiopacity for medical use |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Norfolk Medical. The logo consists of the words "Norfolk Medical" in a serif font, with a red heart-shaped symbol in between the two words. Below the logo is the address "7350 N. Ridgeway · Skokie, IL, USA · 847-674-7075".
### Performance Data:
The following bench tests were performed to evaluate the performance of the TidalPort-APTM Implantable Apheresis Vascular Access Port:
- 1. Catheter to Port Connection
- 2. Catheter Tensile Strength
- 3. Radiopacity
- 4. Gravity Flow Rate
- 5. Port System Burst under Power Injection
- 6. Stem Tensile Strength
- 7. Corrosion Resistance
- 8. Septum Puncture / Port Leak (w/both Huber point and 2-part large bore needles)
- 9. Port Clearance Volume
- 10. Power Injection / Multi-Power Injection
- 11. Simulated Apheresis Testing- comparison to predicate (PowerFlow)
- 12. Recirculation, Forward and Reverse Flow- comparison to predicate (PowerFlow)
- 13. Hemolysis, Forward and Reverse Flow- comparison to predicate (PowerFlow)
- 14. MRI Safety Testing
- 15. Packaging Ship Testing
- 16. Pyrogenicity Testing
### Biocompatibility:
The biocompatibility assessment was completed and appropriate endpoints conducted in compliance to ISO 10993-1:2009, Biological Evaluation of Medical Devices Part 1 - Evaluation and Testing within a Risk Management Process and the FDA Guidance for Industry - Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing".
#### 5.8 Conclusion
The TidalPort-AP™ Implantable Apheresis Vascular Access Port has met all predetermined acceptance criteria of design verification evaluations through testing examination. Based on the FDA's decision tree, it is concluded through performance testing that the TidalPort-APTM Implantable Apheresis Vascular Access Port is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.