The MEDRON Vessel Dilator is intended for short-term vascular access for percutaneous catheterization.
Device Story
MEDRON Vessel Dilator is a radiopaque high-density polyethylene (HDPE) shaft with an overmolded HDPE hub; available in straight or stepped models; diameters 6F to 17.5F. Device facilitates percutaneous vascular access by enlarging vessel openings for catheter placement. Used in clinical settings by healthcare providers. Operates by mechanical dilation of the vessel entry site. Benefits include enabling safe, short-term vascular access for subsequent catheterization procedures.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility verified per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility, complement activation). Functional testing included dimensional verification, hub workmanship, luer taper compliance (ANSI/HIMA MD70.1, ISO 594-2), tensile strength (ISO 11070), tip integrity, and guidewire compatibility.
Technological Characteristics
Materials: High Density Polyethylene (HDPE) body and hub. Dimensions: 6F-17.5F O.D., 8.40-8.66 in length. Guidewire compatibility: 0.018" and 0.038". Sterilization: Ethylene Oxide (EO). Mechanical dilation principle. Standards: ISO 10993-1, ISO 11070, ANSI/HIMA MD70.1, ISO 594-2.
Indications for Use
Indicated for percutaneous entry into a vessel to enlarge the opening for catheter placement in a vein or artery.
Regulatory Classification
Identification
A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 6, 2020
MEDRON, LLC David Kujawa QARA Manager 1518 S Gladiola Street Salt Lake City, UT 84104
Re: K192283
Trade/Device Name: MEDRON Vessel Dilator Regulation Number: 21 CFR 870.1310 Regulation Name: Vessel Dilator For Percutaneous Catheterization Regulatory Class: Class II Product Code: DRE Dated: April 1, 2020 Received: April 2, 2020
Dear David Kujawa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192283
Device Name MEDRON Vessel Dilator
Indications for Use (Describe)
The MEDRON Vessel Dilator is designed for percutaneous entry into a vessel in order to enlarge the opening of the vessel for the placement of a catheter in a vein or artery.
Type of Use (Select one or both, as applicable)
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|-----------------------------------------------------------------------------|
| <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Vessel Dilator
# 510(k) Summary
| 510(k) Sponsor: | MEDRON, LLC<br>1518 S Gladiola Street<br>Salt Lake City, UT 84104 |
|-----------------|-------------------------------------------------------------------|
| Contact Person: | David Kujawa<br>Manager, QARA<br>Tel: 801-974-3010 |
E-mail: mike.kujawa@flexan.com
Date Prepared: March 31, 2020
Prepared by: Ryan O'Callaghan, MS, RAC Phil Triolo and Associates LC Tel: 801.699.9846 Fax: 801.328.2399 E-mail: ryano@philt.com
| Trade Name: | MEDRON Vessel Dilator |
|----------------------|-------------------------------------------------------------------|
| Common Name: | Vessel dilator |
| Classification Name: | Vessel dilator for percutaneous catheterization (21 CFR 870.1310) |
| Product Code: | DRE |
| Predicate Device: | Medcomp Vessel Dilator (K162389) |
## 1. Device Description
The MEDRON Vessel Dilator is a dilator consisting of a radiopaque HDPE (high density polyethylene) shaft with an overmolded HDPE Hub. The dilator comes in a variety of diameters ranging from 6F to 17.5F. It is also available in two models, a straight model and a stepped model. The straight model has a consistent diameter along the length of the shaft until it tapers at the distal tip. In the stepped model, the distal segment of the dilator shaft has a smaller diameter than the proximal segment as well as a taper at the distal tip.
{4}------------------------------------------------
#### 2. Intended Use
The MEDRON Vessel Dilator is intended for short-term vascular access for percutaneous catheterization.
The intended use of the subject device is identical to the intended use of the predicate device.
## 3. Indications for Use
The MEDRON Vessel Dilator is designed for percutaneous entry into a vessel in order to enlarge the opening of the vessel for the placement of a catheter in a vein or artery. The subject device and predicate device have identical indications for use statements.
## 4. Technological Characteristics
As shown in the table below, the technological characteristics of the subject device, the MEDRON Vessel Dilator, are substantially equivalent to those of the predicate device, the Medcomp Vessel Dilator.
| Comparison between Subject & Predicate Device Technological Characteristics | | | |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|------------|
| Characteristic | Predicate Device<br>Medcomp Vessel Dilator<br>(K162389) | Subject Device<br>MEDRON Vessel Dilator | Comparison |
| Dimensions | O.D.: 4F to 24F<br>Length: 6", 8"<br>Guidewire Sizes: 0.025",<br>0.035", 0.038" | O.D.: 6F - 17.5F<br>Length: 8.40 in - 8.66 in<br>Guidewire Sizes: 0.018",<br>0.038" | Equivalent |
| Materials:<br>Dilator Body | High Density Polyethylene | High Density Polyethylene | Same |
| Materials:<br>Dilator Hub | High Density Polyethylene | High Density Polyethylene | Same |
{5}------------------------------------------------
| Comparison between Subject & Predicate Device Technological Characteristics | | | |
|-----------------------------------------------------------------------------|---------------------------------------------------------|-----------------------------------------|------------|
| Characteristic | Predicate Device<br>Medcomp Vessel Dilator<br>(K162389) | Subject Device<br>MEDRON Vessel Dilator | Comparison |
| Sterilization Method | EO | EO | Same |
Results of tests performed on the subject demonstrate that it meets the requirements of relevant standards. Further, any differences in technological characteristics of the MEDRON Vessel Dilator when compared with predicate device characteristics do not raise different questions of safety and effectiveness.
#### 5. Non-Clinical Performance Tests
### Biocompatibility
Biocompatibility of the MEDRON Vessel Dilator has been verified in accordance with ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and FDA's Guidance for Industry and Food and Drug Administration Staff Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued June 16, 2016.
The results of the following biological and toxicological safety evaluations verified the biocompatibility of the subject device when tested as an external communicating, blood contact, limited duration (≤24 hours) device:
- Cytotoxicity;
- . Sensitization;
- . Irritation/Intracutaneous Reactivity;
- . Acute Systemic Toxicity;
- . Materials-Mediated Pyrogenicity;
- . Hemocompatibility (Hemolysis by Direct Contact and Extract); and
- . Complement Activation (SC5b-9).
Results of these tests are summarized in the following table.
{6}------------------------------------------------
## Vessel Dilator
| Summaries of Biocompatibility Tests Conducted to Support this Premarket<br>Notification | | | |
|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|--|
| Test | Results | Conclusions | |
| Cytotoxicity<br>[L-929 MEM Elution] | Cell culture treated with test<br>sample exhibited no<br>reactivity (Grade 0). | Non-cytotoxic | |
| Sensitization<br>[Maximization<br>(Magnusson-Kligman)] | Challenge sites treated with<br>test sample exhibited no<br>erythema or edema (Grade<br>0). | Negative for dermal<br>sensitization | |
| Irritation<br>[Intracutaneous Toxicity<br>(ISO)] | The mean test score in 0.9%<br>Normal Saline extract was 0,<br>and in Sesame Oil was 0.5. | Non-irritating | |
| Systemic Toxicity (Acute)<br>[Systemic Injection (ISO)] | No study animals were<br>observed with abnormal<br>clinical signs indicative of<br>toxicity during the 72-hour<br>test period. | Non-toxic | |
| Systemic Toxicity (Acute)<br>[Material Mediated<br>Pyrogen in a Rabbit<br>Model] | Temperature increases for<br>the all test animals did not<br>exceed the acceptable test<br>limit for maximum individual<br>temperature rise. | Non-pyrogenic | |
| Hemocompatibility<br>[Hemolysis, direct contact<br>- device/material (human<br>blood)] | The difference between the<br>hemolytic indices of the test<br>article and the negative<br>control was 0.00%. | Non-hemolytic | |
| Summaries of Biocompatibility Tests Conducted to Support this Premarket<br>Notification | | | |
| Test | Results | Conclusions | |
| Hemocompatibility<br>[Hemolysis, Extract -<br>device/material (human<br>blood)] | The difference between the<br>hemolytic indices of the test<br>article and the negative<br>control was 0.08%. | Non-hemolytic | |
| Complement Activation<br>[SC5b-9] | The activation caused by the<br>test article was similar to the<br>activation caused by the<br>negative control after 30 and<br>60 minutes exposure and<br>lower than the activation<br>caused by the negative<br>control after 90 minutes<br>exposure. | Negative for complement<br>system activation | |
{7}------------------------------------------------
## Vessel Dilator
## Functional Testing
The following table summarizes the functional tests performed on unaged (t=0) and 1year real time aged samples and results obtained to demonstrate substantial equivalence to the predicate device.
| Summaries of Functional Tests Conducted to Support this Premarket Notification | | | |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--|
| Test | Test Method Summary | Results | |
| Dimensional<br>Verification | Dimensional inspection per<br>engineering drawings | All devices met dimensional<br>specifications. | |
| Hub<br>Workmanship | Visual inspection of hub for<br>French size identification as well<br>as extraneous matter and process<br>or surface defects | All devices met acceptance criteria<br>for size identification and were free<br>of extraneous matter and process or<br>surface defects | |
{8}------------------------------------------------
### Summaries of Functional Tests Conducted to Support this Premarket Notification Test Test Method Summary Results Inspection for luer taper in All samples met the criteria accordance with ANSI/HIMA specified in ANSI/HIMA MD70.1-Luer Taper MD70.1-1983 and ISO 594-2 1983 and ISO 594-2 All devices met minimum force Tensile testing at hub and shaft Tensile breakage requirements specified in ioint per ISO 11070 Strength ISO 11070. Visual inspection of dilator tips All device tips were free of Tip deformation, discoloration or deformation, discoloration and Workmanship process or surface defects process or surface defects Compatibility with standard All devices met acceptance criteria guidewires evaluated by passing as the guidewire passed through the Guidewire Test guidewire through the dilator dilators from hub to tip without drag from hub end to tip
## 6. Conclusion
MEDRON, LLC has presented information in this premarket notification supporting its contention that the MEDRON Vessel Dilator is substantially equivalent with respect to technological characteristics and indications for use to the predicate device.
### Vessel Dilator
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.