Disposable Valve Sets

K192258 · Zhejiang Chuangxiang Medical Technology Co., Ltd. · ODC · Dec 30, 2019 · Gastroenterology, Urology

Device Facts

Record IDK192258
Device NameDisposable Valve Sets
ApplicantZhejiang Chuangxiang Medical Technology Co., Ltd.
Product CodeODC · Gastroenterology, Urology
Decision DateDec 30, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

Disposable Air/Water Valve This device intended to be fitted to an endoscope air/water channel to enable the operator to control inflow of medical gases and water, whilst preventing backflow during a GI Endoscopic procedure. Disposable Suction Valve The device intended to be fitted to an endoscope suction channel to enable the operator to control suction whilst preventing inflow of air during a GI Endoscopic procedure. Disposable Biopsy Valve This device is intended to be fitted to an endoscope biopsy port to enable access for/exchange of endoscopic devices while maintaining insufflation and minimizing leakage of bio material during an endoscopic procedure.

Device Story

Disposable Valve Sets consist of single-use air/water, suction, and biopsy valves; designed for compatibility with Olympus 140/160/180/190/240/260 series endoscopes. Valves manually fitted to endoscope channels/ports by clinicians during GI endoscopic procedures. Air/water valve controls gas/water inflow and prevents backflow; suction valve controls suction and prevents air inflow; biopsy valve enables instrument exchange while maintaining insufflation and minimizing bio-material leakage. Single-use design supports infection prevention protocols. Supplied sterile via EO sterilization. Output is mechanical control of endoscopic fluid/gas channels; assists clinicians in maintaining procedural environment and device access.

Clinical Evidence

Bench testing only. No clinical data. Testing included endoscope compatibility, vacuum/water/air flow, depression force, and leak tests. Biocompatibility testing (cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, pyrogen) performed per ISO 10993-1. Shelf-life validated per ASTM F1980-16; packaging per ISO 11607-1/2. EO residuals met ISO 10993-7.

Technological Characteristics

Single-use, sterile, EO-sterilized endoscopic valves. Materials evaluated for biocompatibility per ISO 10993-1. Mechanical operation via manual depression. Compatible with Olympus 140/160/180/190/240/260 series endoscopes. Shelf life: 3 years. Packaging: single tray in sealed pouch.

Indications for Use

Indicated for use in GI endoscopic procedures to control air/water inflow, suction, and biopsy port access/insufflation maintenance in patients undergoing endoscopy.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration". December 30, 2019 Zhejiang Chuangxiang Medical Technology Co., LTD. Lucius Long RA Manager 301B, No.22, XinYan Road Hanzhou, Zhejiang 311100 CHINA Re: K192258 > Trade/Device Name: Disposable Valve Sets Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODC, OCX Dated: August 21, 2019 Received: November 26, 2019 Dear Lucius Long: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general control's provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Shanil P. Haugen, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K192258 Device Name Disposable Valve Sets #### Indications for Use (Describe) Disposable Valve Sets includes either one air/water and suction valve or one air/water, suction and biopsy valve. #### Disposable Air/Water Valve This device intended to be fitted to an endoscope air/water channel to enable the operator to control inflow of medical gases and water, whilst preventing backflow during a GI Endoscopic procedure. Disposable Suction Valve The device intended to be fitted to an endoscope suction channel to enable the operator to control suction whilst preventing inflow of air during a GI Endoscopic procedure. Disposable Biopsy Valve This device is intended to be fitted to an endoscope biopsy port to enable access for/exchange of endoscopic devices while maintaining insufflation and minimizing leakage of bio material during an endoscopic procedure. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows a logo for a company called MEDNOVA. The logo consists of a geometric shape in shades of green and blue, along with the company name in blue and green text. The text "MEDNOVA" is in a sans-serif font, with "MED" in blue and "NOVA" in green. Below the logo is the text "Zhejiang Chuangxiang Medical Technolo", which is likely the full name of the company. ng Medical Technology Co.,Ltd Page: ## Section 8 510(k) Summary( 21CFR 807.92) ## 1. Submitter's information Name: Zhejiang Chuangxiang Medical Technology Co., LTD. Address: 301B, No.22, XinYan Road Yuhang Economic And Technological Development Zone Hangzhou Zhejiang China Contact person: Lucius.Long Telephone: 86-571-89167088 Fax: 86-571-89167086 2. Date of Submission Dec. 24, 2019 ## 3. Device Trade/Device Name: Disposable Valve Sets Regulation name: Endoscope and accessories Regulation class: II Regulation number:876.1500 Panel: Gastroenterology/Urology Product code: ODC, OCX ## 4. Predicative device 4.1) 510(k) Number: K181509, Device Name: Endorate™ Valve Sets ## 5. Device description Disposable Valve Sets includes either one air/water and suction valve or one air/water, suction and biopsy valve, So we divided into two models for these devices. One suction valve, air/water valve and biopsy valve means model B, one suction valve and air/water valve means model A. These devices are supplied sterile. These valves easily incorporate into infection prevention policies as a single use item, and these valves are suitable for Olympus 140/160/190/240/260 series endoscopes. The following table shows the details information on the model. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains a logo with a geometric shape on the left and text on the right. The geometric shape is a hexagon with a smaller hexagon inside, and it is filled with a gradient of blue and green. To the right of the shape, there is the text "MEDNOVA" in blue and green, with the word "MED" in blue and the word "NOVA" in green. Above the text "MEDNOVA", there are three Chinese characters in blue. ## g Medical Technology Co.,Ltd 2 / 7 Page: | Model | Specification | ENDOSCOPIC COMPATIBILITY | Sterilization<br>(Y/N) | |-------|---------------|--------------------------|------------------------| | A | MD-A-AW01 | Olympus | Y | | B | MD-B-AW02 | 140/160/180/190/240/260 | Y | ### 6. Indications for use Disposable Valve Sets includes either one air/water and suction valve or one air/water, suction and biopsy valve. ### Disposable Air/Water Valve This device intended to be fitted to an edoscope air/water channel to enable the operator to control inflow of medical gases and water, whilst preventing backflow during a GI Endoscopic procedure. ### Disposable Suction Valve The device intended to be fitted to an endoscope suction channel to enable the operator to control suction whilst preventing inflow of air during a GI Endoscopic procedure. #### Disposable Biopsy Valve This device is intended to be fitted to an endoscope biopsy port to enable access for/exchange of endoscopic devices while maintaining insufflation and minimizing leakage of bio material during an endoscopic procedure. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image contains a logo for a company called MEDNOVA. The logo consists of a hexagon shape with a white outline and a gradient fill of green and blue. To the right of the hexagon is the company name, with "MED" in blue and "NOVA" in green. Above the company name are three blue Chinese characters. # Zhejiang Chuangxiang Medical Technology Co.,Ltd Page: # 7. Technological Characteristics The following tables are summaries of the proposed device's technological characteristics as compared to the predicate devices. Table 7.1 comparison of Disposable Air/Water Valve | Item | Proposed device | Predicate device | Comparison to<br>Predicate Devices | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------| | Device name | Disposable Air/Water Valve | Andorate Disposable Air/Water valve | / | | 510(k) number | / | K181509 | / | | Manufacturer | Zhejiang Chuangxiang<br>Medical Technology Co., LTD. | Smardata Suzhou Co., Ltd | / | | Product Code | ODC | ODC | Same | | Regulation No. | 876.1500 | 876.1500 | Same | | Class | II | II | Same | | Indications for Use | This device intended to be<br>fitted to an edoscope<br>air/water channel to enable<br>the operator to control inflow<br>of medical gases and water,<br>whilst preventing backflow<br>during a GI Endoscopic<br>procedure. | The ENDORATE™ Disposable<br>Air/Water valve is intended to be<br>used control the air/water<br>function of an endoscope during a<br>GI Endoscopic procedure. | Substantial<br>Equivalence | | Supplied Sterile | Sterile | Sterile | Same | | Sterilization method | EO | EO | Same | | Compatibility | Olympus®<br>140/160/180/190/240/260<br>series endoscopes | Olympus®<br>140/160/180/190/240/260 series<br>Endoscope | Same | | Single Use | Yes | Yes | Same | | Packaging | Suction, air/water and/or<br>biopsy valves are housed in a<br>single tray and packaged in a<br>sealed pouched | Suction, air/water, biopsy valves are<br>housed in a single tray and<br>packaged in a sealed pouched | Substantial<br>Equivalence | | Shelf Life | Three years | Three years | Substantial<br>Equivalence | {6}------------------------------------------------ Image /page/6/Picture/1 description: The image contains a logo for Zhejiang Chuangxiang Medical Technology Co., Ltd. The logo features a stylized hexagon shape in green and blue, with the text "MEDNOVA" underneath. The company name is written in a larger font below the logo. Page: | Item | Proposed device | Predicate device | Comparison to<br>Predicate Devices | |-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------| | Device name | Disposable Suction Valve | Endorate™<br>Disposable<br>Suction Valve | / | | 510(k) number | / | K181509 | / | | Manufacturer | Zhejiang Chuangxiang<br>Medical Technology Co.,<br>LTD. | Smardata Suzhou Co., Ltd | / | | Product Code | ODC | ODC | Same | | Regulation No. | 876.1500 | 876.1500 | Same | | Class | II | II | Same | | Indications for<br>Use | The device intended to be<br>fitted to an endoscope<br>suction channel to enable<br>the operator to control<br>suction whilst preventing<br>inflow of air during a GI<br>Endoscopic procedure. | The ENDORATE™ Disposable<br>Suction valve is intended to<br>be used control the suction<br>function of an endoscope<br>during a GI Endoscopic<br>procedure. | Substantial<br>Equivalence | | Supplied Sterile | Sterile | Sterile | Same | | Sterilization<br>method | EO | EO | Same | | Compatibility | Olympus®<br>140/160/180/190/240/260<br>series Endoscope | Olympus®<br>140/160/180/190/240/260<br>series Endoscope | Same | | Single Use | Yes | Yes | Same | | Packaging | Suction, air/water and/or<br>biopsy valves are housed in a<br>single tray and packaged in a<br>sealed pouched | Suction, air/water, biopsy<br>valves are housed in a single<br>tray and packaged in a sealed<br>pouched | Substantial<br>Equivalence | | Shelf Life | Three years | Three years | Substantial<br>Equivalence | # Table 7.2 comparison of Disposable Suction Valve {7}------------------------------------------------ Image /page/7/Picture/1 description: The image contains the logo for Zhejiang Chuangxiang Medical Technology Co., Ltd. The logo features a stylized hexagon shape in green and blue gradient, along with the company name in both Chinese and English. The Chinese characters are larger and in blue, while the English name "MEDNOVA" is in green. Below the logo, the full company name "Zhejiang Chuangxiang Medical Technology Co.,Ltd" is written in black text. Page: | Item | Proposed device | Predicate device | Comparison<br>to Predicate<br>Devices | | |-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|----------------------------| | Device name | Disposable Biopsy Valve | Endorate™ Disposable<br>biopsy Valve | / | | | 510(k) number | / | K181509 | / | | | Manufacturer | Zhejiang Chuangxiang<br>Medical Technology Co., LTD. | Smardata Suzhou Co.,<br>Ltd | / | | | Product Code | OCX | OCX | Same | | | Regulation No. | 876.1500 | 876.1500 | Same | | | Class | II | II | Same | | | Indications for<br>Use | This device is intended to be<br>fitted to an endoscope biopsy<br>port to enable access<br>for/exchange of endoscopic<br>devices while maintaining<br>insufflation and minimizing<br>leakage of bio material during<br>an endoscopic procedure. | The Endorate™<br>Disposable biopsy valve is<br>intended to cover the<br>endoscope biopsy port<br>during an endoscopy<br>procedure. In addition, the<br>valve provides access for<br>endoscopic device passage<br>and exchange, helps<br>maintain insufflation and<br>minimizes leakage of<br>biomaterial from the biopsy<br>port throughout the<br>endoscopic procedure. | Substantial<br>Equivalence | | | Supplied Sterile | Sterile | Sterile | Same | | | Sterilization<br>method | EO | EO | Same | | | Compatibility | Olympus<br>140/160/180/190/240/260<br>series endoscopes | Olympus® / Fujinon®<br>and Pentax ® Endoscopes | Substantial<br>Equivalence | | | Single Use | Yes | Yes | Same | | | Packaging | Suction, air/water and/or biopsy<br>valves are housed in a single<br>tray and packaged in a sealed<br>pouched | Suction, Air/Water, biopsy<br>valves are housed in a single<br>tray and packaged in a<br>sealed pouch.<br>Non Sterile Biopsy valve is | Substantial<br>Equivalence | | | | | | Page: 6 / 7 | | | | | also sold individually<br>packaged in a sealed pouch.<br>Biopsy valve is either in blue<br>or black depending on the<br>availability. | | | | Shelf Life | Three years | Three years | | Substantial<br>Equivalence | ## Table 7.3 comparison of Disposable Biopsy Valve {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for MEDNOVA. The logo consists of a hexagon shape on the left, with the text "MEDNOVA" written in blue and green on the bottom. To the right of the hexagon and text is Chinese characters in blue. ## iang Medical Technology Co.,Ltd ## 8. Performance data In-vitro Testing has been performed and all components, subassemblies, and/or full devices met the required specifications for the completed tests. A summary of the test results has been provided for: ## Disposable Suction Valve - 1. Endoscope Compatibility - 2. Vacuum Leak Test - 3. Suction Flow Test - 4. Depression Force Test ## Disposable Air/Water Valve - 1. Endoscope Compatibility - 2. Water Flow Test - 3. Air Flow Test - 4. Leak Test - 5. Depression Force Test - Disposable Biopsy Valve - 1. Endoscope Compatibility - 2. Leak Test - 3. Squeegee Leak Test The EO residual was measured after sterilization of the device to meet the criteria defined in ISO 10993-7 Second Edition 2008-10-15. The EO sterilization cycle has a Sterility Assurance Level (SAL) of 10 °. The shelf-life for three years had been validated in accelerated testing according to ASTM F1980-16 (2016) and the requirements on packaging for terminally sterilized medical device per ISO 11607-1 First Edition 2006-04-15 and ISO 11607-2 First Edition 2006-04-15 are also met. The testing successfully demonstrated essential performance is achieved before and after the shelf life test. Biocompatibility testing was performed in accordance with the FDA Guidance," Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued on June 16, 2016. The cytotoxicity, sensitization and intracutaneous irritation test were performed to demonstrate the biocompatibility of the device. {9}------------------------------------------------ K192258 page 7 of 7 Image /page/9/Picture/1 description: The image contains the logo for Zhejiang Chuangxiang Medical Technology Co., Ltd. The logo features a stylized hexagon shape in green and blue, with the text "MEDNOVA" underneath. To the right of the logo are Chinese characters, and below the logo is the company name in English. The text is clear and legible. Page: ## 9. Non-Clinical Testing: The device has undergone both bench testing of performance and laboratory biocompatibility testing for cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity and pyrogen test in accordance with 21 CFR, Part 58. The Disposable Valve Sets results in no safety or efficacy concerns regarding biocompatibility or performance. Likewise, in conformance with 21 CFR 807.92(b)(3), the device performs as well as the predicate in all testing performed. ## 10. Conclusions Chuangxiang medical has demonstrated that the proposed device Disposable Valve Sets is substantially equivalent to Smardata Suzhou Co., Ltd. currently marketed Endorate™ Valve Sets (K181509).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...