K192249 · InMode , Ltd. · IPF · Dec 17, 2019 · Physical Medicine
Device Facts
Record ID
K192249
Device Name
InMode System with Tone Applicator
Applicant
InMode , Ltd.
Product Code
IPF · Physical Medicine
Decision Date
Dec 17, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The InMode System with Tone Applicator is used in EMS mode for: Prevention or retardation of disuse atrophy Maintaining or increasing range of motion Muscle re-education Relaxation of muscle spasms Increasing local blood circulation Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis And in TENS mode for: Symptomatic relief and management of chronic, intractable pain Post-surgical acute pain Post-traumatic acute pain
Device Story
InMode System with Tone Applicator; console-based electrical stimulator; provides EMS and TENS therapy. Inputs: user-selected parameters via LCD touch screen. Transformation: AC/DC power supply converts input to symmetrical biphasic rectangular electrical pulses; controlled by microprocessor. Outputs: electrical stimulation delivered via Tone Applicator electrodes to patient. Used in professional healthcare facilities; operated by clinicians. Output allows clinicians to manage pain or stimulate muscle tissue; benefits include atrophy prevention, pain relief, and improved circulation. Safety features include automatic overload/no-load trips, patient override, and automatic shut-off.
Clinical Evidence
Bench testing only. No clinical data. Bench testing validated output waveforms, unit characteristics, and design requirements. Biocompatibility assessment (cytotoxicity, acute systemic toxicity, sensitization) confirmed non-toxic and non-irritating.
Indicated for patients requiring EMS for muscle atrophy prevention, range of motion maintenance, muscle re-education, spasm relaxation, local blood circulation, or post-surgical calf stimulation; and TENS for chronic, intractable, post-surgical, or post-traumatic pain management.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 17, 2019
InMode Ltd. % Amit Goren Regulatory Manager A. Stein - Regulatory Affairs Consulting Ltd. 20 Hata'as Str., Suite 102 Kfar Saba, 4442520 Israel
Re: K192249
Trade/Device Name: InMode System with Tone Applicator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, GZJ Dated: September 16, 2019 Received: September 18, 2019
Dear Amit Goren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Vivek Pinto, Ph.D. Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192249
Device Name InMode System with Tone Applicator
Indications for Use (Describe)
The InMode System with Tone Applicator is used in EMS mode for:
Prevention or retardation of disuse atrophy
Maintaining or increasing range of motion
Muscle re-education
Relaxation of muscle spasms
Increasing local blood circulation
Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
And in TENS mode for: Symptomatic relief and management of chronic, intractable pain Post-surgical acute pain Post-traumatic acute pain
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
# THE INMODE SYSTEM WITH TONE APPLICATOR
# 510(k) Number K192249
# Applicant Name:
| Company Name: | InMode Ltd. |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Tabor Building, Shaar Yokneam<br>Yokneam 20692<br>Israel<br>Tel: +972-4-9097470<br>Fax: +972-4-9097471<br>E-mail: amit@asteinrac.com |
| Contact Person: | |
| Official Correspondent: | Amit Goren |
| Company Name: | A. Stein - Regulatory Affairs Consulting Lt |
| Address: | 20 Hata'as Str., Suite 102<br>Kfar Saba 4442520 Israel<br>Tel: +972-9-7670002<br>Fax: +972-9-7668534<br>E-mail: amit@asteinrac.com |
| Date Prepared: | December 11, 2019 |
| Trade Name: | The InMode System with Tone Applicator |
| Classification Name: | Powered Muscle Stimulator |
| Common Name: | Powered Muscle Stimulator |
| Regulation Number: | 890.5850 & 882.5890 |
| Product Code: | IPF & GZJ |
| Classification: | Class II Medical Device |
#### Predicate Device:
The InMode System with Tone Applicator is substantially equivalent to the following predicate device.
| Predicate | Manufacturer | 510(k) No. |
|------------------------------------|--------------|------------|
| Vectra Neo Clinical Therapy System | DJO, LLC | K132284 |
{4}------------------------------------------------
## Device Description:
The InMode System in combination with Tone Applicator (manufactured by InMode Ltd.), is a versatile device intended to employ EMS (Electrical Muscle Stimulation) and TENS (Transcutaneous Electrical Nerve Stimulation) technologies for various medical applications. The InMode System with Tone Applicator consists of an AC/DC power supply unit, controller and user interface including an LCD touch screen. The delivery of the electrical energy is controlled by a Start/Stop button positioned on the front panel.
The System support the following components:
- . LCD display touch screen
- . Audio loudspeaker
- 48V AC/DC power supply .
- . Controller
- . Fans
The System operates while connected to the Tone Applicator.
Following are The InMode System with Tone Applicator specifications:
| Main Line Frequency (nominal): | 50-60 Hz |
|--------------------------------|------------------------------------------|
| Input Voltage (nominal): | 100-240 VAC |
| Input Current (rms) | 2A |
| Dimension: | |
| Console | 35cm W x 35cm D x 100cm H |
| [W x H x D] | [18.2" W x 18.2" D x 40" H] |
| Applicator | Tone Applicator |
| [L x D] | 12cm L x 10cm D [4.7" L x 4" D] |
| Weight Console: | 20 Kg (44 lbs) |
| Tone Applicator Weight: | 0.22 Kg [0.5 lbs.] |
| Platform modules | Converts AC input voltage (90-264Vac) to |
| AC/DC power supply | 48Vdc, 300W |
| Waveform | Symmetrical Biphasic |
| Shape | Rectangular |
| Intensity (output Voltage) | Up to 50 intensity level (=54 Vpeak) |
| Pulse Width | |
| Tone Applicator | 20 to 400 $ \mu $ S |
| Frequency | |
| Tone Applicator | 3 to 200 Hz |
{5}------------------------------------------------
## Intended Use/Indication for Use:
The InMode System with Tone Applicator is used in EMS mode for:
- 트 Prevention or retardation of disuse atrophy
- Maintaining or increasing range of motion ■
- Muscle re-education ■
- Relaxation of muscle spasms
- Increasing local blood circulation
- Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
And in TENS mode for:
- Symptomatic relief and management of chronic, intractable pain 트
- 트 Post-surgical acute pain
- Post-traumatic acute pain
{6}------------------------------------------------
#### Performance Standards:
The InMode System with Tone Applicator has been tested and complies with the following FDA recognized consensus standards:
- · IEC 60601-1: Medical Electrical Equipment Part 1: General Requirements For Basic Safety And Essential Performance ES60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005 Edition 3rd, MOD)
- · Appendix 6-Clause 3.2.2.1 and Clause 3.2.2.2 from ANSI/AAMI NS4:2013/(R)2017
- · IEC 60601-1-2: Medical electrical equipment; Part 1-2: Collateral Standard: Electromagnetic compatibility - Requirements and tests, Edition 4.0 (2014). Environment of intended uses: Professional Healthcare Facility Environment
- · IEC 60601-2-10 Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators (IEC 60601-2-10 Edition 2.1 2016-04) IEC 60601-2-10: 2012, AMD1:2016 for use in conjunction with IEC 60601-1:2005/AMD1:2016
- · IEC 60601-1-6 Medical Electrical Equipment Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability (60601-1-6 Edition 3.1 2013-10, AMD1:2013)
## Non-Clinical (Bench) Performance Data:
Bench testing was conducted to demonstrate that the InMode System with Tone Applicator performs as expected under anticipated conditions of use and to verify that the device performance meets the device design requirements. The device was tested for validation of output waveform, basic unit characteristics, and output specifications.
The bench testing results demonstrated that the device performs as expected under anticipated conditions of use
The biocompatibility of the Tone Applicator outer components (handle and electrodes)
{7}------------------------------------------------
was justified using a biocompatibility assessment performed on a family of related products and by scientific evidence. The following biocompatibility tests were performed as part of the biocompatibility assessment:
| Test | Test Summary | Conclusions |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Cytotoxicity Study Using the ISO<br>Elution Method | The test article extract<br>showed no evidence of<br>causing cell lysis or<br>toxicity | Non-toxic |
| ISO Acute Systemic Toxicity<br>Study in Mice | The test article extracts<br>showed no evidence of<br>causing delayed dermal<br>contact sensitization in the<br>guinea<br>pig. | Non-toxic |
| ISO Guinea Pig Maximization<br>Sensitization Test | There was no mortality or<br>evidence of systemic<br>toxicity from the extracts<br>injected into mice | No irritation/<br>sensitization |
# Pre-Clinical (Animal) Performance Data:
Non-Applicable.
#### Clinical Performance Data:
Non-Applicable.
# Substantial Equivalence:
The below table summarizes the main comparison aspects between the InMode System with Tone Applicators and the proposed predicate device.
| Characteristic | Subject Device | Predicate |
|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) file No. | K192249 | K132284 |
| Device Name | InMode System with<br>Tone Applicator | Vectra Neo<br>Clinical Therapy<br>System |
| Manufacturer | InMode Ltd. | DJO, LLC |
| Class, Product Code. | Class II<br>IPF GZJ | Class II<br>IPF, GZJ<br>IMG, HCC, GZI |
| Characteristic | Subject Device | Predicate |
| 510(k) file No. | K192249 | K132284 |
| Device Name | InMode System with<br>Tone Applicator | Vectra Neo<br>Clinical Therapy<br>System |
| Manufacturer | InMode Ltd. | DJO, LLC<br>LIH, ILY |
| Design: | The InMode System with<br>Tone Applicator consists<br>of an AC/DC power<br>supply unit, controller and<br>user interface including an<br>LCD touch screen. The<br>delivery of the electrical<br>energy is controlled by a<br>Start/Stop button<br>positioned on the front<br>panel.<br>The System support the<br>following components:<br>• LCD display touch<br>screen<br>• Audio loudspeaker<br>• 48V AC/DC power<br>supply<br>• Controller<br>• Fans<br>The System operates<br>while connected to the<br>Tone Applicator. | The Vectra Neo<br>Vlinical Therapy<br>System consists of<br>an AC/DC power<br>supply unit,<br>controller and user<br>interface including<br>color display. The<br>delivery of the<br>electrical energy is<br>controlled by a<br>Start/Stop button<br>positioned on the<br>front panel. |
| Mechanism of Action | Muscle contraction by<br>electrical pulsing. | Idem |
| Components<br>Console | The InMode System<br>consists of the following<br>components:<br>• Console, including a<br>power supply unit,<br>controller and user<br>interface including an<br>LCD touch screen.<br>• Tone Applicator<br>connected to the<br>console via a cable. | Vectra Neo Clinical<br>Therapy System<br>consists of the<br>following<br>components:<br>• Console,<br>including a power<br>supply unit,<br>controller and<br>user interface<br>including an LCD<br>touch screen.<br>• Several types of<br>applicators for<br>different |
| Characteristic | Subject Device | Predicate |
| 510(k) file No. | K192249 | K132284 |
| Device Name | InMode System with<br>Tone Applicator | Vectra Neo<br>Clinical Therapy<br>System |
| Manufacturer | InMode Ltd. | DJO, LLC |
| | | indications (EMS,<br>Ultrasound, EMG<br>& Laser)<br>connected to the<br>console via a<br>cable. |
| Dimension<br>Console<br>[W x H x D] | 35cm W x 35cm D<br>x 100cm H<br>[18.2" W x 18.2" D x<br>40" H] | not publicly<br>available |
| Applicator<br>[L x D] | Tone Applicator<br>12cm L x 10cm D<br>[4.7" L x 4" D] | |
| Weight Console | 20.0 Kg [44 lbs.] | not publicly<br>available |
| Weight applicator | Tone: 0.22 Kg [0.5 lbs.] | |
| Performance<br>Specifications:<br>Components<br>Console | Main Line Frequency<br>(nominal)<br>50-60Hz<br>Input Voltage (nominal)<br>100-240VAC<br>Input Current (rms)<br>2A | not publicly<br>available |
| Method of Line Current Isolation | Independent transformer<br>isolated | not publicly<br>available |
| Electrical Type | Type BF | not publicly<br>available |
| Patient Leakage Current - Normal<br>Condition (μΑ) | <100uA patient leakage | not publicly<br>available |
| Patient Leakage Current - Single<br>Fault Condition (μΑ) | <300uA line leakage | not publicly<br>available |
| Number of Output Modes | 2 | not publicly<br>available |
| Number of Output Channels | 2 | not publicly<br>available |
| Characteristic | Subject Device | Predicate |
| 510(k) file No.<br>Device Name | K192249<br>InMode System with<br>Tone Applicator | K132284<br>Vectra Neo<br>Clinical Therapy<br>System |
| Manufacturer | InMode Ltd. | DJO, LLC |
| Synchronous or Alternating | See Output Specifications<br>Below | not publicly<br>available |
| Method of Channel Isolation | Through transformers and<br>isolators | not publicly<br>available |
| Regulated Current or Regulated<br>Voltage (output signals only) | Regulated voltage on all<br>channels<br>With current limit | not publicly<br>available |
| Software/Firmware/Microprocessor<br>Control | Yes | not publicly<br>available |
| Automatic Overload Trip | Yes | not publicly<br>available |
| Automatic No-Load Trip | Yes | not publicly<br>available |
| Automatic Shut Off | Yes, On/off switch | not publicly<br>available |
| Patient Override Control | Yes | not publicly<br>available |
| Indicator Display | Yes | not publicly<br>available |
| On/Off Status | Yes | not publicly<br>available |
| Battery | No battery | not publicly<br>available |
| Voltage/Current level | Yes, voltage levels | not publicly<br>available |
| Timer Range (Minutes) | 0-60 [minutes] | not publicly<br>available |
| Compliance with 21 CFR 890.5850<br>(IPF) | Yes | not publicly<br>available |
| Compliance with 21 CFR 882.5890<br>(GZJ) | Yes | not publicly<br>available |
| Applicator Name | Tone Applicator | Vectra Neo Clinical<br>Therapy System<br>Applicator |
| Characteristic | Subject Device | Predicate |
| 510(k) file No. | K192249 | K132284 |
| Device Name | InMode System with<br>Tone Applicator | Vectra Neo<br>Clinical Therapy<br>System |
| Manufacturer | InMode Ltd. | DJO, LLC |
| EMS Output Mode | | |
| Output Specifications:<br>Waveform | Symmetrical Biphasic<br>Waveform | not publicly<br>available |
| Pulse Shape | Rectangular | not publicly<br>available |
| Maximum Output Voltage (± 10%) | 56V @500Ω | not publicly<br>available |
| | 56V @2 kΩ | not publicly<br>available |
| | 56V @10kΩ | not publicly<br>available |
| Maximum Output Current (± 10%) | 112 mA @ 500 Ω | not publicly<br>available |
| | 28 mA @ 2 kΩ | not publicly<br>available |
| | 5.6 mA @ 10 kΩ | not publicly<br>available |
| Pulse Width (μsec.) - The output<br>active positive pulse width | 20 to 400 [µs] | not publicly<br>available |
| Frequency (Hz) | 3 to 200 [Hz] | not publicly<br>available |
| Net Charge @ 500 ohms<br>[µC/pulse] | 0 [μC] @ 500Ω | not publicly<br>available |
| Maximum Phase Charge [µC] | 44.8 [µC] @ 500Ω | not publicly<br>available |
| Maximum Current Density<br>[mA/cm2] | 1 [mA/cm2]<br>Surface = 12cm² | not publicly<br>available |
| Maximum Power Density<br>[mW/cm2] | 55[mW/cm2] @500Ω | not publicly<br>available |
| Burst Mode (i.e., pulse trains) | Yes: | not publicly<br>available |
| a. Pulses per burst | a. 3 - 200 | |
| b. Bursts per second | b. 1 | |
| c. Burst duration (seconds) | c. 1-60 sec | |
| d. Duty Cycle [Line (b) x Line (c)] | d. Time on / off | |
| On Time (sec.) | 1 - 60 [sec] | not publicly<br>available…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.