Halyard Purple Xtra and Purple Sterile, Low Dermatitis Potential, Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate
K192241 · O&M Haylard, Inc. · LZA · Mar 13, 2020 · General Hospital
Device Facts
Record ID
K192241
Device Name
Halyard Purple Xtra and Purple Sterile, Low Dermatitis Potential, Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate
Applicant
O&M Haylard, Inc.
Product Code
LZA · General Hospital
Decision Date
Mar 13, 2020
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The Halyard Purple Xtra and Purple Sterile, Low Dermatitis Potential, Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs and Fentanyl Citrate as per ASTM -D6978-05 : Arsenic Trioxide (1 mg/ml) No breakthrough up to 240 minutes Azacıtıdine (Vidaza) (25 mg/ml) No breakthrough up to 240 minutes Bendamustine (5 mg/ml) No breakthrough up to 240 minutes Bortezomib (Velcade) (1 mg/ml) No breakthrough up to 240 minutes Bleomycin sulfate (15 mg/ml) No breakthrough up to 240 minutes Busulfan (6 mg/ml) No breakthrough up to 240 minutes Carboplatin (10 mg/ml) No breakthrough up to 240 minutes Carfilzomib (2 mg/ml) No breakthrough up to 240 minutes Carmustine (3.3 mg/ml) permeation occurred at 169.8 minutes Cetuximab (Erbitux) (2 mg/ml) No breakthrough up to 240 minutes Cisplatin (1 mg/ml) No breakthrough up to 240 minutes Cladiibine (1 0 mg/ml) No breakthrough up to 240 minutes Cyclophosphamide (20 mg/ml) No breakthrough up to 240 minutes Cytarabıne HCL (100 mg/ml) No breakthrough up to 240 minutes Cytovene (10 mg/ml) No breakthrough up to 240 minutes Dacarbazine (10 mg/ml) No breakthrough up to 240 minutes Daunorubicin HCL (5 mg/ml) No breakthıough up to 240 minutes Decitabine (5 mg/ml) No breakthrough up to 240 minutes Docetaxel (10 mg/ml) No breakthrough up to 240 minutes Doxorubicin HCL (2 mg/ml) No breakthrough up to 240 minutes Ep11 ubicin (Ellence) (2 mg/ml) No breakthrough up to 240 minutes Etoposide (20 mg/ml) No breakthrough up to 240 minutes Fludarabıne (25 mg/ml) No breakthrough up to 240 minutes Fluorouracil (50 mg/ml) No breakthıough up to 240 minutes Fulvestrant (50 mg/ml) No breakthrough up to 240 minutes Gemcitabine (38 mg/ml) No breakthrough up to 240 minutes Idarubicin (1 mg/ml) No breakthrough up to 240 minutes Ifosfamide (50 mg/ml) No breakthrough up to 240 minutes Iiinotecan (20 mg/ml) No breakthrough up to 240 minutes Mechlorethamine HCL (1 mg/ml) No breakthrough up to 240 minutes Melphalan (5 mg/ml) No breakthrough up to 240 minutes Methotrexate (25 mg/ml) No breakthrough up to 240 minutes Mitomycin-C (0 5 mg/ml) No breakthrough up to 240 mınutes Mitoxantıone (2 mg/ml) No breakthrough up to 240 minutes Oxaliplatın (2 mg/ml) No breaktlı ough up to 240 minutes Paclitaxel (6 mg/ml) No breakthrough up to 240 minutes Paraplatın (10 mg/ml) No breakthrough up to 240 mınutes Pemetrexed (25 mg/ml) No breakthrough up to 240 minutes Pertuzumab (30 mg/ml) No breakthrough up to 240 mınutes Raltitlexed (0.5 mg/ml) No breakthrough up to 240 minutes Retrovii (10 mg/ml) No breakthrough up to 240 minutes Rituximab (10 mg/ml) No breakthrough up to 240 minutes Temsuolimus (25 mg/ml) No breakthrough up to 240 minutes Trastuzumab (21 mg/ml) No breakthrough up to 240 minutes ThioTEPA (10 mg/ml) No breakthrough up to 240 minutes Topotecan HCL (1 mg/ml) No breakthrough up to 240 minutes Triclosan (1 mg/ml) No breakthrough up to 240 minutes Trisenox (0.1 mg/ml) No breakthrough up to 240 minutes Vincrinstine Sulfate (1 mg/ml) No breakthrough up to 240 minutes Vinblastine (1 mg/ml) No breakthrough up to 240 minutes Vinorelbine (10 mg/ml) No breakthrough up to 240 minutes Zoledronic Acid (0.8 mg/ml) No breakthrough up to 240 minutes Fentanyl Citrate (100mcg/2ml) No breakthrough up to 240 minutes
Device Story
Disposable, sterile, purple-colored, chlorinated, nitrile, powder-free, textured-fingertip, ambidextrous patient examination gloves. Used by clinicians to prevent cross-contamination between examiner and patient. Tested for permeation resistance against chemotherapy drugs and Fentanyl Citrate per ASTM D6978-05. Provides barrier protection; reduces risk of exposure to hazardous drugs and opioids during clinical handling.
Clinical Evidence
Clinical evidence includes a 204-subject study using the Jordan-King modification of the Draize test to evaluate Type IV allergic contact sensitization. Results showed the device was non-irritating and induced no clinical evidence of residual chemical additives causing Type IV allergy. Non-clinical testing included ASTM D6319, D5151, D6124, D6978, and ISO 10993 (systemic toxicity, irritation, sensitization).
Technological Characteristics
Nitrile material; powder-free (ASTM D6124); textured fingertips; ambidextrous; sterile. Dimensions and physical properties (tensile strength, elongation) meet ASTM D6319. Biocompatibility per ISO 10993. Permeation resistance tested per ASTM D6978-05.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for resistance to permeation by specific chemotherapy drugs and Fentanyl Citrate.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol of the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
March 13, 2020
O & M Halyard, Inc. Steven Dowdley Associate Director of Regulatory Affairs 5405 Windward Parkway Alpharetta, Georgia 30004
Re: K192241
Trade/Device Name: Halyard Purple Xtra and Purple Sterile, Low Dermatitis Potential, Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, QDO Dated: February 10, 2020 Received: February 12, 2020
Dear Steven Dowdley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K192241 Device Name
The Halyard Puple Xtra Steile, Low Dermatitis Potential Nitule Powder-Free Exam Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citiate
#### Indications for Use (Describe)
The Halyad Purple Xtra Steile, Low Dermatitis Potental Nutile Powder-Fiee Exam Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citateis are disposable devices intended for medical purposes that is worn on the examinel's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs and Fentanyl Citrate as per ASTM -D6978-05 :
Arsenic Trioxide (1 mg/ml) No breakthrough up to 240 minutes Azacıtıdine (Vidaza) (25 mg/ml) No breakthrough up to 240 minutes Bendamustine (5 mg/ml) No breakthrough up to 240 minutes Bortezomib (Velcade) (1 mg/ml) No breakthrough up to 240 minutes Bleomycin sulfate (15 mg/ml) No breakthrough up to 240 minutes Busulfan (6 mg/ml) No breakthrough up to 240 minutes Carboplatin (10 mg/ml) No breakthrough up to 240 minutes Carfilzomib (2 mg/ml) No breakthrough up to 240 minutes Carmustine (3.3 mg/ml) permeation occurred at 169.8 minutes Cetuximab (Erbitux) (2 mg/ml) No breakthrough up to 240 minutes Cisplatin (1 mg/ml) No breakthrough up to 240 minutes Cladiibine (1 0 mg/ml) No breakthrough up to 240 minutes Cyclophosphamide (20 mg/ml) No breakthrough up to 240 minutes Cytarabıne HCL (100 mg/ml) No breakthrough up to 240 minutes Cytovene (10 mg/ml) No breakthrough up to 240 minutes Dacarbazine (10 mg/ml) No breakthrough up to 240 minutes Daunorubicin HCL (5 mg/ml) No breakthıough up to 240 minutes Decitabine (5 mg/ml) No breakthrough up to 240 minutes Docetaxel (10 mg/ml) No breakthrough up to 240 minutes Doxorubicin HCL (2 mg/ml) No breakthrough up to 240 minutes Ep11 ubicin (Ellence) (2 mg/ml) No breakthrough up to 240 minutes Etoposide (20 mg/ml) No breakthrough up to 240 minutes Fludarabıne (25 mg/ml) No breakthrough up to 240 minutes Fluorouracil (50 mg/ml) No breakthıough up to 240 minutes Fulvestrant (50 mg/ml) No breakthrough up to 240 minutes Gemcitabine (38 mg/ml) No breakthrough up to 240 minutes Idarubicin (1 mg/ml) No breakthrough up to 240 minutes Ifosfamide (50 mg/ml) No breakthrough up to 240 minutes Iiinotecan (20 mg/ml) No breakthrough up to 240 minutes Mechlorethamine HCL (1 mg/ml) No breakthrough up to 240 minutes Melphalan (5 mg/ml) No breakthrough up to 240 minutes Methotrexate (25 mg/ml) No breakthrough up to 240 minutes Mitomycin-C (0 5 mg/ml) No breakthrough up to 240 mınutes Mitoxantıone (2 mg/ml) No breakthrough up to 240 minutes Oxaliplatın (2 mg/ml) No breaktlı ough up to 240 minutes Paclitaxel (6 mg/ml) No breakthrough up to 240 minutes Paraplatın (10 mg/ml) No breakthrough up to 240 mınutes Pemetrexed (25 mg/ml) No breakthrough up to 240 minutes
{3}------------------------------------------------
Pertuzumab (30 mg/ml) No breakthrough up to 240 mınutes Raltitlexed (0.5 mg/ml) No breakthrough up to 240 minutes Retrovii (10 mg/ml) No breakthrough up to 240 minutes Rituximab (10 mg/ml) No breakthrough up to 240 minutes Temsuolimus (25 mg/ml) No breakthrough up to 240 minutes Trastuzumab (21 mg/ml) No breakthrough up to 240 minutes ThioTEPA (10 mg/ml) No breakthrough up to 240 minutes Topotecan HCL (1 mg/ml) No breakthrough up to 240 minutes Triclosan (1 mg/ml) No breakthrough up to 240 minutes Trisenox (0.1 mg/ml) No breakthrough up to 240 minutes Vincrinstine Sulfate (1 mg/ml) No breakthrough up to 240 minutes Vinblastine (1 mg/ml) No breakthrough up to 240 minutes Vinorelbine (10 mg/ml) No breakthrough up to 240 minutes Zoledronic Acid (0.8 mg/ml) No breakthrough up to 240 minutes
Fentanyl Citrate (100mcg/2ml) No breakthrough up to 240 minutes
Type of Use (Select one or both, as applicable)
_ | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to
> Department of Health and Human Services Food and Drug Admınıstration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda hhs gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of ınformation unless it displays a currently valid OMB number "
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green design inside, followed by the word "HALYARD" in blue. The design inside the circle appears to be two curved shapes, one white and one green.
## 510(k) Summary – K192241
| Date Summary<br>was Prepared | March 12, 2020 |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter | O & M Halyard, Inc.<br>5405 Windward Parkway<br>Alpharetta, GA 30004 |
| Primary Contact for<br>this 510(k) Submission | Steven Dowdley, RAC<br>Tel: 678-451-8062<br>Email: steven.dowdley@hyh.com |
| Device Trade Name | Halyard Purple Xtra Sterile, Low Dermatitis Potential, Powder-Free Exam<br>Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate |
| | Halyard Purple Sterile, Low Dermatitis Potential, Powder-Free Exam Gloves<br>Tested for Use with Chemotherapy Drugs and Fentanyl Citrate |
| Device Common Name | Medical Exam Gloves |
| Device Product Code<br>and Classification Name | LZA<br>Class I, 21 CFR §880.6250<br>Patient Examination Glove |
| Subsequent Product Codes | LZC<br>Class I, 21 CFR §880.6250<br>Patient Examination Glove, Specialty<br>QDO<br>Class I, 21 CFR §880.6250<br>Fentanyl and other opioid protection glove |
| Predicate Device | K102032<br>KIMBERLY-CLARK PURPLE NITRILE XTRA* STERILE POWDER-FREE<br>EXAM GLOVE (CHEMOTHERAPY GLOVE) |
| Subject Device Description | Halyard Purple Xtra and Purple Sterile, Low Dermatitis Potential, Powder-<br>Free Exam Gloves Tested for Use with Chemotherapy Drugs and Fentanyl<br>Citrate are disposable, purple-colored, chlorinated, nitrile, powder-free,<br>textured fingertip, ambidextrous, sterile patient examination gloves.<br>The devices follow consensus standards:<br>· ASTM D5151-06 Standard Test Method for Detection of Holes in<br>Medical Gloves<br>· ASTM D6319-10 Standard Specification for Nitrile Examination Gloves<br>for Medical Applications<br>• ASTM D6124-06 Standard Test Method for Residual Powder on Medical<br>Gloves<br>· ASTM D6978-05 Standard Practice for Assessment of Resistance of<br>Medical Gloves to Permeation by Chemotherapy Drugs<br>· ISO 10993-11:2017, Biological evaluation of medical devices - Part 11:<br>Tests for Systemic Toxicity<br>· ISO 10993-10: 2010: Biological evaluation of medical devices - Part 10:<br>Tests for Irritation and Skin Sensitization |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green design inside, followed by the word "HALYARD" in blue, block letters. The design inside the circle appears to be two stylized shapes, one white and one green.
| Indications for Use | | The Halyard Purple Xtra and Purple Sterile, Low Dermatitis Potential, Powder-<br>Free Exam Gloves Tested for Use with Chemotherapy Drugs and Fentanyl<br>Citrate are disposable devices intended for medical purposes that is worn on<br>the examiner's hand to prevent contamination between patient and examiner.<br>These gloves were tested for use with the following chemotherapy drugs and<br>Fentanyl Citrate as per ASTM -D6978-05:<br>Arsenic Trioxide (1 mg/ml) No breakthrough up to 240 minutes<br>Azacitidine (Vidaza) (25 mg/ml) No breakthrough up to 240 minutes<br>Bendamustine (5 mg/ml) No breakthrough up to 240 minutes<br>Bortezomib (Velcade) (1 mg/ml) No breakthrough up to 240 minutes<br>Bleomycin sulfate (15 mg/ml) No breakthrough up to 240 minutes<br>Busulfan (6 mg/ml) No breakthrough up to 240 minutes<br>Carboplatin (10 mg/ml) No breakthrough up to 240 minutes<br>Carfilzomib (2 mg/ml) No breakthrough up to 240 minutes<br>Carmustine (3.3 mg/ml) permeation occurred at 169.8 minutes<br>Cetuximab (Erbitux) (2 mg/ml) No breakthrough up to 240 minutes<br>Cisplatin (1 mg/ml) No breakthrough up to 240 minutes<br>Cladribine (1.0 mg/ml) No breakthrough up to 240 minutes<br>Cyclophosphamide (20 mg/ml) No breakthrough up to 240 minutes<br>Cytarabine HCL (100 mg/ml) No breakthrough up to 240 minutes<br>Cytovene (10 mg/ml) No breakthrough up to 240 minutes<br>Dacarbazine (10 mg/ml) No breakthrough up to 240 minutes<br>Daunorubicin HCL (5 mg/ml) No breakthrough up to 240 minutes<br>Decitabine (5 mg/ml) No breakthrough up to 240 minutes<br>Docetaxel (10 mg/ml) No breakthrough up to 240 minutes<br>Doxorubicin HCL (2 mg/ml) No breakthrough up to 240 minutes<br>Epirubicin (Ellence) (2 mg/ml) No breakthrough up to 240 minutes<br>Etoposide (20 mg/ml) No breakthrough up to 240 minutes<br>Fludarabine (25 mg/ml) No breakthrough up to 240 minutes<br>Fluorouracil (50 mg/ml) No breakthrough up to 240 minutes<br>Fulvestrant (50 mg/ml) No breakthrough up to 240 minutes<br>Gemcitabine (38 mg/ml) No breakthrough up to 240 minutes<br>Idarubicin (1 mg/ml) No breakthrough up to 240 minutes<br>Ifosfamide (50 mg/ml) No breakthrough up to 240 minutes<br>Irinotecan (20 mg/ml) No breakthrough up to 240 minutes<br>Mechlorethamine HCL (1 mg/ml) No breakthrough up to 240 minutes<br>Melphalan (5 mg/ml) No breakthrough up to 240 minutes<br>Methotrexate (25 mg/ml) No breakthrough up to 240 minutes<br>Mitomycin-C (0.5 mg/ml) No breakthrough up to 240 minutes<br>Mitoxantrone (2 mg/ml) No breakthrough up to 240 minutes<br>Oxaliplatin (2 mg/ml) No breakthrough up to 240 minutes<br>Paclitaxel (6 mg/ml) No breakthrough up to 240 minutes<br>Paraplatin (10 mg/ml) No breakthrough up to 240 minutes<br>Pemetrexed (25 mg/ml) No breakthrough up to 240 minutes<br>Pertuzumab (30 mg/ml) No breakthrough up to 240 minutes<br>Raltitrexed (0.5 mg/ml) No breakthrough up to 240 minutes<br>Retrovir (10 mg/ml) No breakthrough up to 240 minutes<br>Rituximab (10 mg/ml) No breakthrough up to 240 minutes<br>Temsirolimus (25 mg/ml) No breakthrough up to 240 minutes<br>Trastuzumab (21 mg/ml) No breakthrough up to 240 minutes<br>ThioTEPA (10 mg/ml) No breakthrough up to 240 minutes<br>Topotecan HCL (1 mg/ml) No breakthrough up to 240 minutes<br>Triclosan (1 mg/ml) No breakthrough up to 240 minutes<br>Trisenox (0.1 mg/ml) No breakthrough up to 240 minutes<br>Vincrinstine Sulfate (1 mg/ml) No breakthrough up to 240 minutes<br>Vinblastine (1 mg/ml) No breakthrough up to 240 minutes<br>Vinorelbine (10 mg/ml) No breakthrough up to 240 minutes | Zoledronic Acid (0.8 mg/ml) No breakthrough up to 240 minutes |
|---------------------|------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| | Fentanyl Citrate (100mcg/2ml) No breakthrough up to 240 minutes. | | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Halyard. The logo consists of a blue circle with a white and green design inside, followed by the word "HALYARD" in blue, bold letters. The design inside the circle appears to be two curved shapes, one white and one green, resembling sails or stylized mountains.
# Technological Characteristics Comparison Table
| | Subject Device | Predicate Device | Comparison |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | K192241 | K102032 | |
| FDA Product Code | LZA, LZC, QDO | LZC, LZA | Same |
| FDA Classification | Class I | Class I | Same |
| Regulation Number | 880.6250 | 880.6250 | Same |
| Common Name | Medical Exam Glove | Medical Exam Glove | Same |
| Device Trade Name | Halyard Purple Xtra and<br>Purple Sterile, Low<br>Dermatitis Potential,<br>Powder-Free Exam<br>Gloves Tested for Use<br>with Chemotherapy<br>Drugs and Fentanyl<br>Citrate | Kimberly Clark Purple Xtra<br>and Purple Sterile, Powder-<br>Free Exam Gloves Tested for<br>Use with Chemotherapy<br>Drugs | Same |
| Intended Use | The Halyard Purple Xtra<br>and Purple Sterile, Low<br>Dermtatis Potential,<br>Powder-Free Exam<br>Gloves Tested for Use<br>with Chemotherapy<br>Drugs and Fentanyl<br>Citrates are disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hand to prevent<br>contamination between<br>patient and examiner.<br>These gloves were<br>tested for use with<br>chemotherapy drugs<br>listed on the label. | Kimberly Clark Purple Xtra<br>and Purple Sterile, Powder-<br>Free Exam Gloves Tested for<br>Use with Chemotherapy<br>Drugs are disposable device<br>intended for medical purposes<br>that is worn on the examiner's<br>hand to prevent<br>contamination between<br>patient and examiner.<br>These gloves were tested for<br>use with chemotherapy drugs<br>listed on the label. | Different |
| Technological<br>Characteristics | The glove is a colored,<br>nitrile, powder-free,<br>textured fingertip,<br>ambidextrous, patient<br>examination glove. | The glove is a colored, nitrile,<br>powder-free, textured<br>fingertip, ambidextrous,<br>patient examination glove. | Same |
| Sizes of gloves | XS, S, M, L, XL | XS, S, M, L, XL | Same |
| Texture | Textured fingertips | Textured fingertips | Same |
| Sterility | Sterile | Sterile | Same |
| Biocompatibility | Based ISO 10993<br>Biological evaluation of<br>Medical devices - Test<br>for Systemic Injection,<br>the test article was<br>considered non-toxic.<br>Meets the acceptance<br>criteria. | Based ISO 10993 Biological<br>evaluation of Medical devices<br>- Test for Systemic Injection,<br>the test article was<br>considered non-toxic. Meets<br>the acceptance criteria. | Same |
| | Based on ISO 10993-<br>Biological evaluation of<br>Medical Devices – Test<br>for Skin Irritation, the<br>device extracts were not<br>found to cause a<br>systemic response in the<br>animal model. Meets the<br>acceptance criteria. | Based on ISO 10993-<br>Biological evaluation of<br>Medical Devices - Test for<br>Skin Irritation, the device<br>extracts were not found to<br>cause a systemic response in<br>the animal model. Meets the<br>acceptance criteria. | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green design inside, followed by the word "HALYARD" in blue, sans-serif font. The design inside the circle appears to be two curved shapes, one white and one green, creating a stylized image.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Halyard. The logo consists of a blue circle with a white and green design inside, followed by the word "HALYARD" in blue, sans-serif font. The design inside the circle appears to be two stylized shapes, one white and one green, resembling sails or stylized mountains.
| Standard | Results<br>Subject Devices | Results<br>Predicate Devices K102032 | Remarks |
|--------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|---------------------------------------------------------------------|---------|
| ASTM D6978-05<br>Standard<br>Practice for<br>Assessment of<br>Resistance of<br>Medical Gloves<br>to Permeation by<br>Chemotherapy<br>Drugs | Arsenic Trioxide (1 mg/ml) No<br>breakthrough up to 240 minutes | Bleomycin sulfate (15 mg/ml) No<br>breakthrough up to 240 minutes | Similar |
| | Azacitidine (Vidaza) (25 mg/ml) No<br>breakthrough up to 240 minutes | Busulfan (6 mg/ml) No breakthrough up to<br>240 minutes | |
| | Bendamustine (5 mg/ml) No<br>breakthrough up to 240 minutes | Carboplatin (10 mg/ml) No breakthrough up<br>to 240 minutes | |
| | Bortezomib (Velcade) (1 mg/ml) No<br>breakthrough up to 240 minutes | Carmustine (3.3 mg/ml) permeation<br>occurred at 48.6 minutes | |
| | Bleomycin sulfate (15 mg/ml) No<br>breakthrough up to 240 minutes | Cyclophosphamide (20 mg/ml) No<br>breakthrough up to 240 minutes | |
| | Busulfan (6 mg/ml) No breakthrough up<br>to 240 minutes | Cytarabine HCL (100 mg/ml) No<br>breakthrough up to 240 minutes | |
| | Carboplatin (10 mg/ml) No breakthrough<br>up to 240 minutes | Dacarbazine (10 mg/ml) No breakthrough<br>up to 240 minutes | |
| | Carfilzomib (2 mg/ml) No breakthrough<br>up to 240 minutes | Daunorubicin HCL (5 mg/ml) No<br>breakthrough up to 240 minutes | |
| | Carmustine (3.3 mg/ml) permeation<br>occurred at 169.8 minutes | Docetaxel (10 mg/ml) No breakthrough up<br>to 240 minutes | |
| | Cetuximab (Erbitux) (2 mg/ml) No<br>breakthrough up to 240 minutes | Doxorubicin HCL (2 mg/ml) No<br>breakthrough up to 240 minutes | |
| | Cisplatin (1 mg/ml) No breakthrough up<br>to 240 minutes | Epirubicin (Ellence) (2 mg/ml) No<br>breakthrough up to 240 minutes | |
| | Cladribine (1.0 mg/ml) No breakthrough<br>up to 240 minutes | Etoposide (20 mg/ml) No breakthrough up<br>to 240 minutes | |
| | Cyclophosphamide (20 mg/ml) No<br>breakthrough up to 240 minutes | Fludarabine (25 mg/ml) No breakthrough up<br>to 240 minutes | |
| | Cytarabine HCL (100 mg/ml) No<br>breakthrough up to 240 minutes | Fluorouracil (50 mg/ml) No breakthrough up<br>to 240 minutes | |
| | Cytovene (10 mg/ml) No breakthrough<br>up to 240 minutes | Gemcitabine (38 mg/ml) No breakthrough<br>up to 240 minutes | |
| | Dacarbazine (10 mg/ml) No<br>breakthrough up to 240 minutes | Idarubicin (1 mg/ml) No breakthrough up to<br>240 minutes | |
| | Daunorubicin HCL (5 mg/ml) No<br>breakthrough up to 240 minutes | Ifosfamide (50 mg/ml) No breakthrough up<br>to 240 minutes | |
| | Decitabine (5 mg/ml) No breakthrough<br>up to 240 minutes | Irinotecan (20 mg/ml) No breakthrough up<br>to 240 minutes | |
| | Docetaxel (10 mg/ml) No breakthrough<br>up to 240 minutes | Mechlorethamine HCL (1 mg/ml) No<br>breakthrough up to 240 minutes | |
{9}------------------------------------------------
# HALYARD
| Doxorubicin HCL (2 mg/ml) No breakthrough up to 240 minutes | Melphalan (5 mg/ml) No breakthrough up to 240 minutes |
|---------------------------------------------------------------------|------------------------------------------------------------------|
| Epirubicin (Ellence) (2 mg/ml) No breakthrough up to 240 minutes | Methotrexate (25 mg/ml) No breakthrough up to 240 minutes |
| Etoposide (20 mg/ml) No breakthrough up to 240 minutes | Mitomycin (0.5 mg/ml) No breakthrough up to 240 minutes |
| Fludarabine (25 mg/ml) No breakthrough up to 240 minutes | Mitoxantrone (2 mg/ml) No breakthrough up to 240 minutes |
| Fluorouracil (50 mg/ml) No breakthrough up to 240 minutes | Paclitaxel (6 mg/ml) No breakthrough up to 240 minutes |
| Fulvestrant (50 mg/ml) No breakthrough up to 240 minutes | Paraplatin (10 mg/ml) No breakthrough up to 240 minutes |
| Gemcitabine (38 mg/ml) No breakthrough up to 240 minutes | Rituximab (10 mg/ml) No breakthrough up to 240 minutes |
| Idarubicin (1 mg/ml) No breakthrough up to 240 minutes | ThioTEPA (10 mg/ml) No breakthrough up to 240 minutes |
| Ifosfamide (50 mg/ml) No breakthrough up to 240 minutes | Trisenox (0.1 mg/ml) No breakthrough up to 240 minutes |
| Irinotecan (20 mg/ml) No breakthrough up to 240 minutes | Vincrinstine Sulfate (1 mg/ml) No breakthrough up to 240 minutes |
| Mechlorethamine HCL (1 mg/ml) No breakthrough up to 240 minutes | |
| Melphalan (5 mg/ml) No breakthrough up to 240 minutes | |
| Methotrexate (25 mg/ml) No breakthrough up to 240 minutes | |
| Mitomycin-C (0.5 mg/ml) No breakthrough up to 240 minutes | |
| Mitoxantrone (2 mg/ml) No breakthrough up to 240 minutes | |
| Oxaliplatin (2 mg/ml) No breakthrough up to 240 minutes | |
| Paclitaxel (6 mg/ml) No breakthrough up to 240 minutes | |
| Paraplatin (10 mg/ml) No breakthrough up to 240 minutes | |
| Pemetrexed (25 mg/ml) No breakthrough up to 240 minutes | |
| Pertuzumab (30 mg/ml) No breakthrough up to 240 minutes | |
| Raltitrexed (0.5 mg/ml) No breakthrough up to 240 minutes | |
| Retrovir (10 mg/ml) No breakthrough up to 240 minutes | |
| Rituximab (10 mg/ml) No breakthrough<br>up to 240 minutes | |
| Temsirolimus (25 mg/ml) No<br>breakthrough up to 240 minutes | |
| Trastuzumab (21 mg/ml) No<br>breakthrough up to 240 minutes | |
| ThioTEPA (10 mg/ml) No breakthrough<br>up to 240 minutes | |
| Topotecan HCL (1 mg/ml) No<br>breakthrough up to 240 minutes | |
| Triclosan (1 mg/ml) No breakthrough up<br>to 240 minutes | |
| Trisenox (0.1 mg/ml) No breakthrough<br>up to 240 minutes | |
| Vincrinstine Sulfate (1 mg/ml) No<br>breakthrough up to 240 minutes | |
| Vinblastine (1 mg/ml) No breakthrough<br>up to 240 minutes | |
| Vinorelbine (10 mg/ml) No breakthrough<br>up to 240 minutes | |
| Zoledronic Acid (0.8 mg/ml) No<br>breakthrough up to 240 minutes | |
{10}------------------------------------------------
## HALYARD 11
| Performance Data for Fentanyl Citrate Injection | | | |
|-------------------------------------------------|------------------------------------------------------------------------------------|--------------------------------------|-----------|
| Standard | Results<br>Subject Devices | Results<br>Predicate Devices K102032 | Remarks |
| ASTM D 6978 | Fentanyl Citrate Injection, (100mcg<br>/2mL). No breakthrough up to 240<br>minutes | Not tested | Different |
| ASTM D5151-06<br>Standard Test<br>Method for<br>Detection of<br>Holes in Medical<br>Gloves | Testing of the subject device shows<br>it meets the 2.5% AQL requirement<br>in the standards for leakage. The<br>device meets the acceptance criteria<br>of the standard. | Testing of the subject device shows it<br>meets the 2.5% AQL requirement in<br>the standards for leakage. The device<br>meets the acceptance criteria of the<br>standard. | Same |
|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green design inside, followed by the word "HALYARD" in blue. The design inside the circle appears to be two curved shapes, one white and one green, resembling stylized sails or mountains.
| ASTM D6124-06<br>Standard Test<br>Method for<br>Residual Powder<br>on Medical<br>Gloves | Residual powder on the subject<br>device is an average of 0.4<br>mg/glove within the powder-free limit<br>of < 2 mg maximum powder per<br>glove and meets the acceptance<br>criteria for powder- free. | Residual powder on the subject device<br>is an average of 0.4 mg/glove within<br>the powder-free limit of < 2 mg<br>maximum powder per glove and meets<br>the acceptance criteria for powder-<br>free. | Same |
|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| ASTM D6319-10<br>Standard<br>Specification for<br>Nitrile<br>Examination<br>Gloves for<br>Medical<br>Applications | The physical dimensions of the<br>subject device are within the limits of<br>the standard and the physical<br>properties of the subject device met<br>the requirements for tensile strength<br>before and after aging. The subject<br>device also met the requirement for<br>elongation before and after aging. | The physical dimensions of the subject<br>device are within the limits of the<br>standard and the physical properties of<br>the subject device met the<br>requirements for tensile strength<br>before and after aging. The subject<br>device also met the requirement for<br>elongation before and after aging. | Same |
| ISO 10993<br>Biological<br>evaluation of<br>medical devices | Meets acceptance criteria | Meets acceptance criteria | Same |
## Non- Clinical Testing Summary
## PERFORMANCE CHARACTERISTICS OF THE SUBJECT DEVICE:
| Brief description<br>of non-clinical<br>tests: | Test | Standard | Acceptance Criteria | Results |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Dimensions | ASTM D 6319 | Length $\geq$ 230mm<br>Palm width Size<br>Xtra-Small: 60 – 80 mm<br>Small: 70 – 90 mm<br>Medium: 85 – 105 mm<br>Large: 100 – 120 mm<br>Xtra-Large: 110-130 mm<br>Finger Thickness<br>$\geq$ 0.050 mm<br>Palm Thickness<br>$\geq$ 0.050 mm<br>Cuff Thickness<br>0.10-0.13 mm (Halyard) | The physical<br>dimensions of the<br>subject device are<br>within the limits of the<br>standard. |
| | Physical Properties | ASTM D 6319 | AQL 4.0<br>Before<br>Tensile Strength: $\geq$ 14 MPa<br>Ultimate elongation: $\geq$ 500%<br>After<br>Tensile Strength: $\geq$ 14 MPa<br>Ultimate elongation: $\geq$ 400% | Physical properties<br>of the subject device<br>meet the<br>requirements for<br>tensile strength and<br>elongation in<br>the standard. |
| | Freedom from Pinholes | ASTM D 6319<br>ASTM D 5151 | AQL 2.5%<br>No leakage | Testing of the<br>subject device<br>shows it meets the<br>AQL requirement in<br>the<br>standards for leakage. |
| | Powder-Free | ASTM D 6124<br>ASTM D 6319 | $≤$ 2 mg / glove | Residual powder on<br>the subject device is<br>within the powder-<br>free limits prescribed<br>in the<br>standards. |
| | ISO Indirect Irritation<br>Study | ISO 10993, Part 10 | Primary Irritation Index<br>$\le$ 2.0 | Under the conditions of<br>the<br>study the device is<br>not an irritant. |
| | ISO Systemic Toxicity<br>Study | ISO 10993, Part 11 | No animals treated with test<br>extracts exhibit greater<br>reaction than control<br>animals | No evidence of<br>systemic toxicity. |
| | ISO Dermal<br>Sensitization | ISO 10993, Part 10 | Grade < 1 | Under the conditions<br>of the study the<br>device is not a<br>sensitizer. |
| | Standard Practice for<br>Assessment of<br>Resistance of Medical<br>Gloves to Permeation<br>by Chemotherapy<br>Drugs | ASTM D6978-05 | No breakthrough was detected<br>for up to 240 minutes | Acceptance criteria:<br>No signs of<br>breakthrough for the<br>subject device after 4<br>hours for 51<br>chemotherapy drugs.<br>Carmustine showed<br>no signs of<br>breakthrough until<br>169.8 minutes.<br>No signs of<br>breakthrough after 4<br>hours for Fentanyl<br>Citrate Injection<br>(100 mcg/2ml) |
| Brief description<br>of clinical tests: | A 204 subject study was completed to evaluate whether the level of residual chemical additives in the<br>subject device induced Type IV allergic contact sensitization by repetitive applications to the skin of<br>normal healthy human volunteers using the Jordan-King modification of the Draize test as<br>recommended by the FDA.<br>Under the conditions of the study, the subject device was nonirritating and showed no clinical<br>evidence of residual chemical additives that may induce Type IV allergy in human subject | | | |
| Conclusions: | The conclusions drawn from the nonclinical and clinical tests demonstrate that the subject devices (the<br>Halyard Purple Xtra and Purple Sterile, Low Dermatitis Potential Powder-Free Exam Gloves Tested for<br>Use with Chemotherapy Drugs and Fentanyl Citrate) are as safe, as effective, and performs as well as<br>or better than the legally marketed devices cleared under K102032. | | | |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green design inside, followed by the word "HALYARD" in blue, bold letters. The design inside the circle appears to be an abstract representation of a sail or a similar shape.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.