This femoral stem is for total hip or hemi-hip arthroplasty and is indicated for the following conditions: · Noninflammatory degenerative joint disease (NID), e. g., avascular necrosis, osteoarthritis and inflammatory joint disease (IJD), e. g., rheumatoid arthritis. · Failed previous hip surgery (not THA) where pain, deformity or dysfunction persists. · Optional use in revision: in some medical conditions (e. g., early revision when healthy and good bone stock exists) the surgeon may opt to use primary implants in a revision procedure. This stem is for uncemented use only.
Device Story
Modular femoral hip stem for total or hemi-hip arthroplasty; uncemented fixation. Proximal Titanium Vacuum Plasma Spray coating; distal rough-blasted surface for primary stability. Curved design with trapezoidal cross-section; available in multiple sizes and offset options with CCD angles of 140°, 137°, 129°, and 127°. Implanted by orthopedic surgeons in clinical settings. Provides structural support for hip joint replacement; restores joint function and reduces pain in patients with degenerative or inflammatory joint disease.
Clinical Evidence
No clinical data required. Non-clinical performance testing included packaging integrity verification per ISO 11607-1:2009 and ISO 11607-2:2006, and verification of label/IFU legibility.
Technological Characteristics
Modular femoral stem; Titanium Vacuum Plasma Spray proximal coating; rough-blasted distal surface. Curved geometry with trapezoidal cross-section. Available in multiple sizes/offsets with CCD angles of 140°, 137°, 129°, 127°. Uncemented fixation. Sterilization and manufacturing processes identical to predicate.
Indications for Use
Indicated for patients requiring total or hemi-hip arthroplasty due to noninflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis), inflammatory joint disease (e.g., rheumatoid arthritis), or failed previous non-THA hip surgery. May be used in revision procedures when healthy bone stock exists. Uncemented use only.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
December 11, 2019
Zimmer GmbH Tobias Moller Regulatory Affairs Specialist Sulzer Allee 8 WINTERTHUR CH 8404
Re: K192236
Trade/Device Name: Fitmore Hip Stem Regulation Number: 21 CFR 888.3350 Regulation Name: Hip joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JDI, KWL, KWY, KWZ, LWJ, LZO Dated: October 2, 2019 Received: October 7, 2019
Dear Tobias Moller:
This letter corrects our substantially equivalent letter of November 5, 2019.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vesa Vuniqi Date: 2019.12.11 -5
Vesa Vuniqi Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192236
Device Name Fitmore® Hip Stem
## Indications for Use (Describe)
This femoral stem is for total hip or hemi-hip arthroplasty and is indicated for the following conditions:
· Noninflammatory degenerative joint disease (NID), e. g., avascular necrosis, osteoarthritis and inflammatory joint disease (IJD), e. g., rheumatoid arthritis.
· Failed previous hip surgery (not THA) where pain, deformity or dysfunction persists.
· Optional use in revision: in some medical conditions (e. g., early revision when healthy and good bone stock exists) the surgeon may opt to use primary implants in a revision procedure.
This stem is for uncemented use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Zimmer Biomet. The logo consists of a blue circle with a stylized "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. Below the company name is the tagline "Your progress. Our promise." in a smaller font size.
| Sponsor: | Zimmer GmbH |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Sulzerallee 8, P.O. Box |
| | 8404 Winterthur, Switzerland |
| Contact Person: | Tobias Möller |
| | Specialist, Regulatory Affairs |
| | Telephone: +41 58 854 89 94 |
| | Fax: +41 52 244 86 58 |
| Date: | October 18, 2019 |
| Trade Name: | Fitmore® Hip Stem |
| Common or Usual Name: | Hip Prosthesis |
| Classification Product Code: | JDI, KWL, KWY, KWZ, LWJ, LZO |
| Device Classification Name: | Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular<br>Component) |
| Regulation Number / Description: | 21 CFR § 888.3350 - Hip joint metal/polymer semi-constrained<br>cemented prosthesis |
| | 21 CFR § 888.3360 - Hip joint femoral (hemi-hip) metallic<br>cemented or uncemented prosthesis |
| | 21 CFR § 888.3390 – Hip joint femoral (hemi-hip) metal/polymer<br>cemented or uncemented prosthesis |
| | 21 CFR § 888.3310 - Hip joint metal/polymer constrained<br>cemented or uncemented prosthesis |
| | 21 CFR § 888.3360 – Hip joint femoral (hemi-hip) metallic<br>cemented or uncemented prosthesis |
| | 21 CFR § 888.3353 - Hip joint metal/ceramic/polymer semi-<br>constrained cemented or nonporous uncemented prosthesis |
| Predicate Device: | Fitmore® Hip Stem, manufactured by Zimmer GmbH, K071723,<br>cleared March 7, 2008 |
| Device Description: | The Fitmore® Hip Stem is a modular femoral stem intended for<br>total or hemi-hip arthroplasty. The component is an uncemented<br>stem which is coated proximally with Titanium Vacuum Plasma<br>Spray and rough-blasted distally for primary stability. The stem<br>design is curved and features a trapezoidal cross-section. |
| | The 12/14 femoral stem is available in multiple sizes, consisting of<br>three stem families. The different family designs were developed<br>in order to meet various patients' anatomic needs. As the<br>anatomic offset varies considerably between individuals, the<br>Fitmore® Hip Stem offers a wide range of offset options. Each<br>family is designed with a CCD angle of 140°, 137°, 129° and 127°. |
| Indications for Use: | This femoral stem is for total hip or hemi-hip arthroplasty and is<br>indicated for the following conditions:<br>• Noninflammatory degenerative joint disease (NIDJD), e. g.,<br>avascular necrosis, osteoarthritis and inflammatory joint<br>disease (IJD), e. g., rheumatoid arthritis.<br>• Failed previous hip surgery (not THA) where pain, deformity or<br>dysfunction persists.<br>• Optional use in revision: in some medical conditions (e. g.,<br>early revision when healthy and good bone stock exists) the<br>surgeon may opt to use primary implants in a revision<br>procedure.<br>This stem is for uncemented use only. |
| Comparison to Predicate Device: | The intended use of the proposed device is identical to those of<br>the previously cleared predicate device. The amount of indications<br>for use have been reduced and reworded for more clarity<br>compared to the previously cleared device. The proposed<br>changes do not alter the fundamental scientific technology shared<br>by both the proposed and predicate devices. Zimmer GmbH is<br>furthermore seeking clearance for minor modifications that have<br>been implemented since the last clearance. |
| Performance Data (Nonclinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions:<br>Packaging performance testing was to verify that packaging<br>configuration maintains integrity of the sterile barrier system up to<br>the point of use and provides adequate protection to the product<br>through the hazards of sterilization, handling, distribution and<br>storage according to ISO 11607-1:2009 and ISO 11607-2:2006.<br>Performance testing further verified that labels and IFU remain<br>intact and legible. Packaging Configuration testing was conducted<br>by representative worst-case products.<br>Clinical Performance and Conclusions:<br>Clinical data and conclusions were not needed for this device. |
| Conclusion | The subject device has the same intended use and similar<br>indications for use as the predicate device. The subject device<br>uses the same operating principle, incorporates the same basic<br>design and labeling and is manufactured and sterilized using the<br>same materials and processes as the predicate device. Except for<br>the modifications described in this submission the subject device<br>is identical to the predicate device and the performance data and<br>analyses demonstrate that:<br>• any differences do not raise new questions of safety and<br>effectiveness as established with performance testing; and<br>• the subject devices are at least as safe and effective as the<br>legally marketed predicate devices. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for Zimmer Biomet. The logo consists of a blue circle with a white "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. Below the company name is the tagline "Your progress. Our promise." in a smaller font size.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.