← Product Code [EIH](/productcode/EIH) · K192231

# Dental Glass Ceramics Blocks (K192231)

_Aidite (Qinhuangdao) Technology Co., Ltd. · EIH · Jul 8, 2020 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K192231

## Device Facts

- **Applicant:** Aidite (Qinhuangdao) Technology Co., Ltd.
- **Product Code:** [EIH](/productcode/EIH.md)
- **Decision Date:** Jul 8, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Dental Glass Ceramics Blocks are indicating all-ceramic restorations such as veneers, inlay/ onlay, partial crowns, anterior crowns, posterior crowns, using the hot press technique or CAD/CAM system.

## Device Story

Dental Glass Ceramics Blocks consist of pre-formed ceramic blanks composed of SiO2, Li2O, Al2O3, and inorganic pigments. Used in dental laboratories; processed by technicians via hot press or CAD/CAM milling systems to create final all-ceramic restorations (veneers, crowns, inlays/onlays). Final restorations are placed by dentists to restore tooth structure and function. Benefits include biocompatible, durable, and aesthetic tooth replacement.

## Clinical Evidence

No clinical data. Bench testing performed per ISO 6872:2015 to verify physical/chemical properties. Biocompatibility testing conducted per ISO 10993 series (cytotoxicity, oral mucosa irritation, sensitization, subacute/subchronic toxicity, genotoxicity) confirmed safety equivalent to predicate.

## Technological Characteristics

Ceramic blocks composed of SiO2, Li2O, K2O, Al2O3, and inorganic pigments. Form factor: blocks. Processing: hot press or CAD/CAM. Non-sterile. Meets ISO 6872:2015 standards for density. Biocompatible per ISO 10993.

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- Dental Lithium Disilicate Glass Ceramic Block(Up.Press Series and Up. CAD Series) ([K141727](/device/K141727.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 8, 2020

Aidite (Oinhuangdao) Technology Co., Ltd. % Christy Young Consultant Shenzhen Joyantech Consulting Co., Ltd NO. 55 Shizhou middle road , Nanshan District Shenzhen, Guangdong GD755, China

Re: K192231

Trade/Device Name: Dental Glass Ceramics Blocks Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: May 22, 2020 Received: June 8, 2020

Dear Christy Young:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K192231

Device Name Dental Glass Ceramics Blocks

Indications for Use (Describe)

Dental Glass Ceramics Blocks are indicating all-ceramic restorations such as veneers, inlay/ onlay, partial crowns, anterior crowns, posterior crowns, using the hot press technique or CAD/CAM system.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Product: Dental Glass Ceramics Block

# 510(k) Summary

K192231

#### 1. Contact Details

1.1 Applicant information

| Applicant Name | Aidite (Qinhuangdao) Technology Co., Ltd.                                               |
|----------------|-----------------------------------------------------------------------------------------|
| Address        | No.9 Dushan Road,Economic and Technological Development Zone,<br>Qinhuangdao City China |
| Phone No.      | +86 335 8587898                                                                         |
| Fax No.        | +86 335 8587198                                                                         |
| Contact person | Zhang Wei                                                                               |
| Date Prepared  | May 28, 2019                                                                            |
| Website        | www.zro2blocks.com                                                                      |

- 1.2 Submission Correspondent

|                         | Shenzhen Joyantech Consulting Co., Ltd                                                                           |
|-------------------------|------------------------------------------------------------------------------------------------------------------|
|                         | Room 1122, International Mayors Communication Centre, No. 55 Shizhou<br>middle road , Nanshan District, Shenzhen |
|                         | 卓远天成                                                                                                             |
| Phone No.               | +86-755-86069197                                                                                                 |
| Contact person          | Christy Young; Field Fu;                                                                                         |
| Contact person's e-mail | christy@cefda.com; cefda@foxmail.com                                                                             |
| Website                 | http://www.cefda.com                                                                                             |

#### Device information 2.

.

-

-

| Trade name          | Dental Glass Ceramics Blocks      |
|---------------------|-----------------------------------|
| Common name         | Dental Glass Ceramics             |
| Model               | /                                 |
| Classification      | II                                |
| Classification name | Porcelain Powder for Clinical Use |
| Product code        | EIH                               |
| Regulation No.      | 872.6660                          |

#### Legally Marketed Predicate Device 3.

| Trade Name    | Dental Lithium Disilicate Glass Ceramic Block(Up.Press Series and Up.<br>CAD Series) |
|---------------|--------------------------------------------------------------------------------------|
| 510(k) Number | K141727                                                                              |
| Product Code  | EIH                                                                                  |
| Manufacturer  | Liaoning Upcera Co., Ltd                                                             |

#### 4. Device Description

Dental Glass Ceramics Blocks are derived from dental porcelain powder that has been processed into their final net shapes. These blanks are then being further fabricated (using hot press or

{4}------------------------------------------------

CAD/CAM technologies) into all-ceramic restorations such as veneers, inlay/ onlay, partial crowns, anterior crowns, posterior crowns. The ceramics material is composed of SiO2, Li2O, Al2Os and other oxides. It also contains inorganic pigments to provide different shades on the product surface.

#### 5. Intended Use/Indication for Use

Dental Glass Ceramics Blocks are indicating all-ceramic restorations such as veneers, inlay/onlay, partial crowns, anterior crowns, posterior crowns, using the hot press technique or CAD/CAM system.

| Item                                                           | Proposed Device:<br>Dental Glass Ceramics Blocks                                                                                                                                                                                           | Predicate Device:<br>Dental Lithium Disilicate<br>Glass Ceramic<br>Block(Up.Press Series and<br>Up. CAD Series)<br>(K141727)                                                                                                                                                                            | Comments                |
|----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Product Code                                                   | EIH                                                                                                                                                                                                                                        | EIH                                                                                                                                                                                                                                                                                                     | Same                    |
| Indications for<br>Use                                         | Dental Glass Ceramics Blocks<br>are indicated for fabricating<br>all-ceramic restorations such as<br>veneers, inlay/onlay, partial<br>crowns, anterior crowns,<br>posterior crowns, using the hot<br>press technique or CAD/CAM<br>system. | Dental Lithium Disilicate Glass<br>Ceramic Blocks (Up. Press<br>Series and Up. CAD Series) are<br>indicated for fabricating<br>allceramic restorations such as<br>veneers, inlay/onlay, partial<br>crowns, anterior crowns,<br>posterior crowns, using the hot<br>press technique or CAD/CAM<br>system. | Same                    |
| Materials                                                      | SiO2, Li2O, K2O, Al2O3 and other<br>oxides                                                                                                                                                                                                 | SiO2, Li2O, K2O, P2O5, Al2O3,<br>B2O3, and other oxides                                                                                                                                                                                                                                                 | Comparable(<br>Issue 1) |
| Processing at<br>Dental lab                                    | Hot Press (Up. Press Series)<br>CAD/CAM (Up.CAD Series)                                                                                                                                                                                    | Hot Press (Up. Press Series)<br>CAD/CAM (Up.CAD Series)                                                                                                                                                                                                                                                 | Same                    |
| Geometry                                                       | Blocks                                                                                                                                                                                                                                     | Blocks, disc and rod                                                                                                                                                                                                                                                                                    | Similar                 |
| Dimension                                                      | Various                                                                                                                                                                                                                                    | Various                                                                                                                                                                                                                                                                                                 | Same                    |
| Single use                                                     | Yes                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                                                                     | Same                    |
| Available color                                                | Various                                                                                                                                                                                                                                    | Various                                                                                                                                                                                                                                                                                                 | Same                    |
| Sterile                                                        | Non-sterile                                                                                                                                                                                                                                | Non-sterile                                                                                                                                                                                                                                                                                             | Same                    |
| Radioactivity<br>(Bq · g-1)                                    |                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                         |                         |
| Density(g/cm³)                                                 | Meet the requirements of ISO<br>6872:2015                                                                                                                                                                                                  | Meet the requirements of ISO<br>6872:2008                                                                                                                                                                                                                                                               | Same                    |
| Flexural<br>Strength(MPa)                                      |                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                         |                         |
| Coefficient of<br>Thermal<br>Expansion (K-1)                   |                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                         |                         |
| Glass Transition<br>Temperature (° C)                          |                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                         |                         |
| Cytotoxicity (ISO<br>10993-5:2009)                             | No cytotoxicity effect                                                                                                                                                                                                                     | No cytotoxicity effect                                                                                                                                                                                                                                                                                  | Same                    |
| Irritation Oral<br>Mucosa Irritation<br>(ISO<br>10993-10:2010) | Not a primary oral mucosa<br>irritant under the conditions of<br>the study                                                                                                                                                                 | Not a primary oral mucosa<br>irritant under the conditions of<br>the study                                                                                                                                                                                                                              | Same                    |
| Sensitization(ISO<br>10993-10:2010)                            | Not a sensitizer under the<br>conditions of the study                                                                                                                                                                                      | Not a sensitizer under the<br>conditions of the study                                                                                                                                                                                                                                                   | Same                    |
| Subacute and<br>Subchronic<br>Toxicity(ISO<br>10993-11:2006)   | No subacute and subchronic<br>toxic effects observed                                                                                                                                                                                       | No subacute and subchronic<br>toxic effects observed                                                                                                                                                                                                                                                    | Same                    |
| Genotoxicity<br>( ISO10993-3:2003)                             | No genotoxic effects observed                                                                                                                                                                                                              | No genotoxic effects observed                                                                                                                                                                                                                                                                           | Same                    |

#### Substantial Equivalence Comparison 6.

{5}------------------------------------------------

## Product: Dental Glass Ceramics Block

### lssue 1:

The subject device and the predicate devices might difference in compositions but all the devices have SiO2, Li2O, K2O and Al2O3 as major components. Despite this difference, the test results per ISO 6872 shows that the subject device is substantially equivalent to the predicate device in physical and chemical properties and meets the necessary requirements.

#### 7. Non-clinical Testing

Bench testing was performed per ISO 6872:2015 and internal procedures to ensure that the Dental Glass Ceramics Blocks met its specifications. All tests were verified to meet acceptance criteria.

Biocompatibility testing was performed to verify the equivalent safety of the materials that are used.

#### 8. Clinical testing

Clinical testing was not performed for the proposed device.

#### 9. Conclusion

It has been shown in this 510(k) submission that Dental Glass Ceramics Blocks and its predicate devices have the similar indications for use, similar biocompatibility, and similar performance.

The difference between Dental Glass Ceramics Blocks and its predicate device do not raise any

{6}------------------------------------------------

Product: Dental Glass Ceramics Block

question regarding its safety and effectiveness.

Dental Glass Ceramics Blocks, as designed and manufactured, is as safe and effective as its predicated device, and therefore is substantially equivalent as its predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K192231](https://fda-staging.innolitics.com/device/K192231)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).