K192226 · Philips Health Care · IYN · Sep 6, 2019 · Radiology
Device Facts
Record ID
K192226
Device Name
Lumify Diagnostic Ultrasound System
Applicant
Philips Health Care
Product Code
IYN · Radiology
Decision Date
Sep 6, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Philips Lumify Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined (B+Color), and M modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Fetal/Obstetric, Abdominal, Pediatric, Cephalic, Urology, Gynecological, Cardiac Fetal Echo, Small Organ, Musculoskeletal, Peripheral Vessel, Carotid, Cardiac. Lumify is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals.
Device Story
Mobile, software-controlled diagnostic ultrasound system; acquires and displays ultrasound data in B, Color Doppler, Combined, and M modes. System comprises COTS Android or iOS mobile device, Philips ultrasound app, USB-based transducers (C5-2, L12-4, S4-1), and Lumify Power Module (LPM). LPM enables connectivity to Apple iAP2 Lightning interface and provides battery power for iOS devices. Used by healthcare professionals in point-of-care settings. Output displayed on mobile device screen for clinical diagnostic assessment; aids in real-time visualization of anatomy and fluid flow.
Clinical Evidence
No clinical data. Substantial equivalence established through bench testing, verification of hardware/software components, and compliance with international standards (IEC 60601 series) and FDA ultrasound guidance.
Technological Characteristics
Mobile ultrasound system; COTS mobile device platform; USB-interfaced transducers; Lumify Power Module (LPM) for iOS connectivity/power. Complies with IEC 60601-2-37, IEC 60601-1-11, IEC 60601-1-12, and ISO 10993-1. Acoustic output limits: Ispta.3 ≤ 720 mW/cm2, MI < 1.9, TI < 6.0.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal/obstetric, abdominal, pediatric, cephalic, urological, gynecological, cardiac fetal echo, small organ, musculoskeletal, peripheral vessel, carotid, and cardiac applications. Intended for use by healthcare professionals in clinical environments.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
September 6, 2019
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
Philips Healthcare % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive. Suite 510k SAINT PAUL MN 55114
Re: K192226
Trade/Device Name: Lumify Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 15, 2019 Received: August 16, 2019
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192226
Device Name Lumify Diagnostic Ultrasound System
#### Indications for Use (Describe)
Philips Lumify Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined (B+Color), and M modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications:
Fetal/Obstetric, Abdominal, Pediatric, Cephalic, Urology, Gynecological, Cardiac Fetal Echo, Small Organ, Musculoskeletal, Peripheral Vessel, Carotid, Cardiac.
Lumify is a transportable ultrasound system intended for use in environments where healthcare professionals.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
# Lumify Diagnostic Ultrasound System
This summary of safety and effectiveness information is submitted in accordance with 21CFR §807.92
- 1. Submitter's name, address, telephone number, contact person. Philips Ultrasound, Inc. 22100 Bothell Everett Hwy Bothell, WA 98021-8431
| Contact person: | Paul Elias, Regulatory Affairs Specialist |
|--------------------|---------------------------------------------|
| Email: | Paul.Elias@philips.com |
| Tel: | 425-482-8396 |
| Fax: | 425-487-8666 |
| Secondary Contact: | Hebe Sun, Senior Regulatory Affairs Manager |
| Email: | Hebe.Sun@philips.com |
July 15, 2019 Date prepared:
- 2. Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
Common name: Diagnostic ultrasound system and transducers Proprietary name: Lumify Diagnostic Ultrasound System
These devices are classified as follows:
| Classification Description | 21 CFR Section | Product Code |
|------------------------------------------|----------------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | ITX |
As stated in 21 CFR, parts 892.1550, 892.1560, and 892.1570, each of these generic types of devices has been classified as Class II.
#### 3. Substantially Equivalent Devices
Philips Ultrasound believes the proposed Lumify Diagnostic Ultrasound System is substantially equivalent to the following currently marketed predicate device:
| Predicate Device | 510(k) |
|--------------------------|---------|
| Lumify Ultrasound System | K162549 |
{4}------------------------------------------------
## 4. Device Description
The Lumify Diagnostic Ultrasound System is a mobile, general purpose, softwarecontrolled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data in various modes of operation. Lumify is a portable system facilitating point of care ultrasound applications.
The Lumify Diagnostic Ultrasound System includes:
- A commercial off-the-shelf (COTS) Android or iOS mobile device ●
- Philips Ultrasound software running as an app (Android or iOS) on the COTS device ●
- The C5-2 Curved array USB transducer
- The L12-4 Linear array USB transducer
- The S4-1 Sector array USB transducer
- The Lumify Power Module (LPM) to convert the USB interface used on the family of Lumify transducers to Apple's iAP2 Lightning interface standard used on iPhones and iPads, and to provide battery power to the transducers when using an iOS mobile device
# 5. Indications for Use
Philips Lumify Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined (B+Color), and M modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications:
Fetal/Obstetric, Abdominal, Pediatric, Cephalic, Urology, Gynecological, Cardiac Fetal Echo, Small Organ, Musculoskeletal, Peripheral Vessel, Carotid, Cardiac.
Lumify is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals.
# 6. Technological Comparison to Predicate Devices
The Lumify Diagnostic Ultrasound System employs the same fundamental scientific technology as the currently marketed predicate Lumify Ultrasound System (K162549). The transducers, scanning modes, and indications for use are identical. The primary differences are the implementation of the Lumify software application for iOS and the addition of the Lumify Power Module (LPM). The LPM converts the USB interface used on the family of Lumify transducers to Apple's iAP2 Lightning interface, and powers transducers when using an iOS mobile device.
# 7. Safety Considerations and Nonclinical Performance Testing
As a Track 3 ultrasound device, the Philips Lumify Diagnostic Ultrasound System is designed to comply with the acoustic output display requirements of IEC 60601-2-37 Ed 2.1 (Particular requirements for the basic safety and essential performance of ultrasonic medical and monitoring equipment). The Lumify Diagnostic Ultrasound System complies with the referenced standard as well as the FDA ultrasound guidance document,
{5}------------------------------------------------
"Marketing Clearance of Diagnostic Ultrasound Systems and Transducers," issued on June 27, 2019.
System acoustic output limits are identical to the currently marketed predicate Lumify Ultrasound System (K162549):
- Ispta.3 ≤ 720 MW/cm2 트
- 트 MI < 1.9
- . TI < 6.0
To ensure safety and effectiveness, the Lumify Diagnostic Ultrasound System has been tested to, and is compliant with, the following standards:
- AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012. C1:2009/(R)2012 ■ And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment -Part 1: General Requirements For Basic Safety And Essential Performance
- IEC 60601-2-37: Medical electrical equipment. Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment, 2015
- I IEC 60601-1-11:2015: Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 60601-1-12:2015: Medical Electrical Equipment Part 1-12: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment.
- IEC 60601-1-2 Medical Electrical Equipment - Part 1-2, General Requirements for Basic Safety and Essential Performance - Collateral Standard Electromagnetic Compatibility, 2014
- AAMI ANSI ISO 10993-1:2009/(R)2013 Biological Evaluation of Medical l Devices – Part 1: Evaluation and Testing within a Risk Management Process
Quality assurance measures applied to the system design and development included risk analysis, product specifications, design reviews, and verification testing.
Test results supported a determination of substantial equivalence to the predicate device and demonstrated that the Lumify Diagnostic Ultrasound System:
- complies with the aforementioned international and FDA-recognized . consensus standards and FDA ultrasound guidance document, and
- . meets the acceptance criteria and is adequate for its intended use.
{6}------------------------------------------------
#### 8. Clinical Data
The Lumify Diagnostic Ultrasound System did not require clinical studies because substantial equivalence to the currently marketed predicate Lumify Ultrasound System (K162549) has been established.
## 9. Conclusion
The hardware and software components enabling the Lumify Diagnostic Ultrasound System expansion from Android to support iOS mobile devices have been successfully verified. The differences between the Lumify Diagnostic Ultrasound System and the predicate Lumify Ultrasound System (K162549) do not raise new questions of safety and/or effectiveness. With identical transducers, scanning modes, and indications for use, the Lumify Diagnostic Ultrasound System is substantially equivalent to the currently marketed predicate Lumify Ultrasound System (K162549).
# 514 Performance Standards
There are no Sec. 514 performance standards for the Lumify Diagnostic Ultrasound System.
## Prescription Status
The Lumify Diagnostic Ultrasound System is a prescription device. The prescription device statement appears in the labeling.
#### Sterilization Sites
Not applicable. No components are supplied sterile.
# Track
The Lumify Diagnostic Ultrasound System is a Track 3 system.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.