K192202 · World Class Technology Corporation · NJM · Nov 12, 2019 · Dental
Device Facts
Record ID
K192202
Device Name
Clear21
Applicant
World Class Technology Corporation
Product Code
NJM · Dental
Decision Date
Nov 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Clear21™ orthodontic ceramic bracket system is intended to aid in the movement of teeth during orthodontic treatment.
Device Story
Clear21 is an aesthetic passive self-ligating orthodontic bracket system. Brackets are bonded directly to the facial surface of maxillary teeth (second bicuspid to second bicuspid). System consists of a ceramic body, ceramic door, and nitinol spring. Clinician inserts an archwire into the bracket slot and closes the door, which secures the wire and eliminates the need for elastomeric or steel ligatures. The archwire provides the force to move teeth according to the clinician's treatment plan. Used in orthodontic clinics by trained dental professionals. Benefits include aesthetic appearance of ceramic materials combined with the functional performance of traditional self-ligating brackets.
Clinical Evidence
No clinical data. Evidence is based on non-clinical bench testing, including biocompatibility and functional performance comparisons (bond strength, ligation mechanism, structural integrity) against the predicate and reference devices.
Indicated for patients requiring orthodontic treatment to aid in tooth movement. Intended for use by dental professionals trained in orthodontic treatment.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
Predicate Devices
Damon 4Clear (currently marketed as Damon Clear and Damon Clear2) (K081415)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text.
November 12, 2019
World Class Technology Corporation % Alyssa Thomas Principal Consultant Allegiance Regulatory Consulting LLC 16642 SW Lansford Ct. Beaverton, Oregon 97007
#### Re: K192202
Trade/Device Name: Clear21 Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NJM Dated: August 9, 2019 Received: August 14, 2019
#### Dear Alyssa Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known) K192202
Device Name
Clear21TM
#### Indications for Use (Describe)
The Clear21™ orthodontic ceramic bracket system is intended to aid in the movement of teeth during orthodontic treatment.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:12px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This summary of 510(k) information is being submitted in accordance with the requirements of SMDA (Safe Medical Devices Act) 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K192202
## 5.1. Submitter Contact Information
| Name: | World Class Technology Corporation |
|----------------|---------------------------------------------------|
| Address: | 1300 NE Alpha Drive<br>McMinnville, OR 97128, USA |
| Phone: | 503-472-8320 |
| Fax: | 503-435-2432 |
| Contact: | Alan Kozlowski |
| Date Prepared: | November 11, 2019 |
## 5.2. Device Identification
| Trade Name: | Clear21™ |
|----------------------|-----------------------------|
| Common Name: | Orthodontic Ceramic Bracket |
| Classification Name: | Orthodontic Plastic Bracket |
| Regulatory Class: | 872.5470 |
| Product Code: | NJM |
# 5.3. Legally Marketed Predicate Device(s)
Predicate Device: K081415 Ormco Corporation's Damon 4Clear (currently marketed as "Damon Clear" and "Damon Clear2")
Reference Device: K173440 Ortho Organizers, Incorporated/Henry Schein Orthodontics' Carriere SLX 3D Clear
## 5.4. Device Description
The Clear21™ orthodontic ceramic bracket system from World Class Technology Corporation is a system of aesthetic passive self-ligating brackets for fixed appliance orthodontic treatment. The brackets are bonded directly to patient's teeth of the
{4}------------------------------------------------
maxillary arch. The self-ligating mechanism (i.e. the door) eliminates the need for elastomeric and steel ligatures. The system is intended for use with clinician chosen orthodontic auxiliary devices.
# 5.5. Intended Use/Indications for Use
The Clear21™ orthodontic ceramic bracket system is intended to aid in the movement of teeth during orthodontic treatment.
# 5.6. Summary of Technological Characteristics
The Clear21 bracket system includes brackets designed to provide the aesthetic benefit of clear ceramic materials with the function of traditional metal brackets. The brackets are available for the upper maxillary teeth from the second bicuspid to the second bicuspid, i.e. 5 to 5. The bracket body and door are ceramic injection molded. The threecomponent bracket assembly includes the bracket body, door and a spring. The Clear21 bracket is similar in materials, design, manufacturing processes and principles of operation as the predicate device (Damon Clear2). There are no significant differences between the Clear21 device and the predicate. A table comparing the key features of the subject device and the predicate device is provided below (Table 5-1Table 5-1).
| | Clear21<br>(Subject Device) | Damon Clear / Damon<br>Clear2<br>(K081415) |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer /<br>Marketing Company | World Class Technology<br>Corporation | Ormco Corporation |
| Intended Use /<br>Indication for Use | The Clear21™ orthodontic<br>ceramic bracket system is<br>intended to aid in the<br>movement of teeth during<br>orthodontic treatment. | The Damon 4Clear* is a<br>ceramic bracket system<br>intended to aid in the<br>movement of patient teeth<br>during orthodontic<br>treatment.<br>* Name at the time of submission |
| Clinical Application | Orthodontic Treatment | Orthodontic Treatment |
| Clinical Setting / Site<br>of Use | Prescription Device for<br>Orthodontic Clinic Use | Prescription Device for<br>Orthodontic Clinic Use |
| | Clear21 | Damon Clear / Damon<br>Clear2<br>(K081415) |
| | (Subject Device) | |
| Target User | Dental professionals trained<br>in orthodontic treatment | Dental professionals trained<br>in orthodontic treatment |
| | Technology Characteristics | |
| Device Materials | Polycrystalline Alumina &<br>Nitinol | Polycrystalline Alumina &<br>Nitinol |
| Device Design | Three-component<br>construction comprised of a<br>ceramic door & body with a<br>nitinol spring element. The<br>spring provides force for the<br>ligating mechanism | Three-component<br>construction comprised of a<br>ceramic door & body with a<br>nitinol spring element. The<br>spring provides force for the<br>ligating mechanism |
| Bracket Base Design | Base designed for direct<br>bonding to the facial surface<br>of the teeth. The bracket<br>bonding pad/base includes<br>mechanical retention<br>features that are formed in<br>the ceramic material | Base designed for direct<br>bonding to the facial surface<br>of the teeth. The bracket<br>bonding pad/base utilizes<br>mechanical retention<br>features that are laser cut<br>into the ceramic material |
| Bracket Hook Feature | Select brackets include an<br>integrated hook for<br>accessory attachment | Select brackets include an<br>integrated hook for<br>accessory attachment |
| Ligation Method | Self-Ligating | Self-Ligating |
| Bracket Identification | Color-coded Dots | Color-coded Dots |
| Manufacturing Method | - Ceramic Injection Molded<br>- Processed Pad/Base<br>- Spring Cut from Flat<br>Stock<br>- Component Assembly | - Ceramic Injection Molded<br>- Laser-Etch Pad/Base<br>- Spring Formed into a<br>Tubular Split Pin<br>- Component Assembly |
| Functional Dimension (Prescription) | | |
| In-Out | 0.0265" - 0.0385" | 0.03" - 0.040" |
| Torque | -11 deg. through +12 deg. | -11 deg. through +22 deg. |
| Angulation | 2 to 9 deg. | 2 to 9 deg. |
| Rotation | None of the brackets in the<br>system incorporate rotation | None of the brackets in the<br>system incorporate rotation |
| | Clear21 | Damon Clear / Damon<br>Clear2 |
| | (Subject Device) | (K081415) |
| Principles of Operation | | |
| Mode of Use | An archwire, selected by the<br>clinician, is inserted into the<br>archwire slots in the<br>brackets and the ligating<br>door is closed, securing the<br>archwire in the slot. The<br>archwires provide the forces<br>needed to move the teeth in<br>accordance with the dental<br>professional's goals and<br>technique. | An archwire, selected by the<br>clinician, is inserted into the<br>archwire slots in the<br>brackets and the ligating<br>door is closed, securing the<br>archwire in the slot. The<br>archwires provide the forces<br>needed to move the teeth in<br>accordance with the dental<br>professional's goals and<br>technique. |
| Application | Bracket pad/base is directly<br>bonded to the facial surface<br>of the tooth using typical<br>orthodontic adhesives and<br>bonding techniques | Bracket pad/base is directly<br>bonded to the facial surface<br>of the tooth using typical<br>orthodontic adhesives and<br>bonding techniques |
Table 5-1 Subject & Predicate Device Comparison
{5}------------------------------------------------
{6}------------------------------------------------
# 5.7. Performance Data
Non-Clinical Testing - The application includes detailed discussions of nonclinical testing. In all instances, the Clear21 bracket device preformed as intended. Biocompatibility testing was also performed, demonstrating that the bracket materials and processing are acceptable based on the contact category for the Clear21 brackets. Functional testing was done to compare the Clear21 bracket's performance to that of its predicate and reference devices. A reference device, the Carriere SLX bracket, was used to compare the bond strength performance of the Clear21 bracket. The predicate device incorporates a unique proprietary bond base and the reference device uses a bond base design similar to the Clear21 and is more consistent with typical bond base designs. The ligation mechanism, structural integrity and bond strength of the Clear21 bracket are similar to the predicate and reference devices, as demonstrated through the results of functional testing. The Clear21 bracket produces similar results for ligating door open/close force range values, door pull-out, rotational strength, hook strength and bond strength. The nonclinical evaluations provide evidence of the Clear21 bracket's safety and effectiveness when used as intended.
{7}------------------------------------------------
Clinical Testing - Clinical testing is not included in this submission
## 5.8. Substantial Equivalence
The Clear21 brackets are as safe and effective as the predicate, the Damon Clear2. The Clear21 bracket system has the same indications for use as the predicate device. Clear21 has similar technological characteristics and principles of operation as its predicate device. The minor technological differences between the Clear21 brackets and its predicate device raise no new or different issues of safety or effectiveness, as evident by the comparison testing. The performance data demonstrates that the Clear21 brackets are as safe and effective as the Damon Clear/Clear2 brackets. Thus, the Clear21 is substantially equivalent.
## 5.9. Conclusions
Conclusions drawn from the comparison to the predicate device and the nonclinical testing demonstrate that the Clear21 ceramic bracket system is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.