K192178 · Imicryl Dis Malzemeleri Sanayi VE Ticaret AS · EBF · Dec 20, 2019 · Dental
Device Facts
Record ID
K192178
Device Name
R&D Series Nova Compo N
Applicant
Imicryl Dis Malzemeleri Sanayi VE Ticaret AS
Product Code
EBF · Dental
Decision Date
Dec 20, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Indications for Use
R&D Series Nova Compo N is a light-curing, radiopaque, nano-hybrid universal composite for anterior restorations. It is intended for professional use by below mentioned Indications for Use; - Restorations of deciduous teeth - Splinting of mobile teeth - Restoration in the posterior region (Classes I and II) - Anterior restorations (Classes III, IV) - Class V restorations (cervical caries, root erosion, wedge-shaped defects) - Extended fissure sealing in molars and premolars - Veneering of discoloured anterior teeth - Repair of composite and ceramic veneers
Device Story
R&D Series Nova Compo N is a light-curing, radiopaque, nano-hybrid universal composite material; used by dental professionals for restorative procedures. Input: light energy applied intra-orally to initiate polymerization. Transformation: high molecular weight resin matrix with prepolymerized fillers and nanofillers reduces polymerization shrinkage, improves handling, and enhances wear resistance/surface polish. Output: hardened dental restorative material. Clinical use: applied in clinic settings for cavity filling, splinting, and veneer repair. Benefit: provides durable, aesthetic, and radiopaque tooth restoration with low water absorption and high surface hardness.
Clinical Evidence
Bench testing only. Biocompatibility evaluated per ISO 10993-1 and ISO 7405 (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation). Physical performance tested per ISO 4049:2009, including compressive strength (360 MPa), flexural strength (130 MPa), depth of cure (>=2 mm), water sorption (21.2 µg/mm3), and radiopacity (350% Al).
Technological Characteristics
Nano-hybrid composite; resin matrix (Bis-GMA, UDMA, Bis-EMA, hydrophobic aromatic dimethacrylates); inorganic fillers (barium glass, ytterbium trifluoride); organic fillers (prepolymers); nano fillers. Light-cured (ISO 4049:2009 Type 1, Class 2, Group 1). Particle size 40 nm to 3 µm. Radiopaque. Biocompatible.
Indications for Use
Indicated for professional dental use in patients requiring restorations of deciduous teeth, splinting of mobile teeth, Class I-V restorations, extended fissure sealing in molars/premolars, veneering of discolored anterior teeth, and repair of composite/ceramic veneers.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 20, 2019
Imicryl Dis Malzemeleri Sanayi Ve Ticaret A.S. Husamettin Sonmez General Manager Fetih Mahallesi Mahir Sokak No:5/201 Konya, 42030 Karatay TURKEY
Re: K192178
Trade/Device Name: R&D Series Nova Compo N Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: Class II Product Code: EBF Dated: October 1, 2019 Received: October 7, 2019
Dear Husamettin Sonmez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K192178
Device Name
R&D Series Nova Compo N
Indications for Use (Describe)
R&D Series Nova Compo N is a light-curing, radiopaque, nano-hybrid universal composite for anterior restorations. It is intended for professional use by below mentioned Indications for Use;
- · Restorations of deciduous teeth
- · Splinting of mobile teeth
- · Restoration in the posterior region (Classes I and II)
- · Anterior restorations (Classes III, IV)
- · Class V restorations (cervical caries, root erosion, wedge-shaped defects)
- · Extended fissure sealing in molars and premolars
- · Veneering of discoloured anterior teeth
- · Repair of composite and ceramic veneers
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for IMICRYL. The logo consists of a blue circle with the letters "IMI" inside, followed by the word "IMICRYL" in blue, and the words "DIŞ MALZEMELERI SAN. VE TIC A.Ş." in black. The logo is simple and modern, and the use of blue gives it a sense of trustworthiness and reliability.
### K192178
### 510(k) Summary
# R&D SERIES NOVA COMPO N® Light Cure Nano Hybrid Composite
Date of Summary Preparation: July 17, 2019
Type of Submission: Traditional 510(k)
#### Submitter Information:
Company Name: IMICRYL DIS MALZEMELERI SANAYI VE TICARET A.S.
Company Address: Fetih Mahallesi Mahir Sokak No: 5/201 Zip Code: 42030 Karatay/Konya/TURKEY
Phone Number: 0090 444 71 30
Fax Number: 0090 332 355 76 19
Contact Person: Husamettin SONMEZ
Phone: +90 444 71 30 ext: 120
E-mail: sonmez@imicryl.com.tr
#### Secondary Contact Person: Muhammed Hulusi SONMEZ
Phone: +90 542 372 27 23
E-mail: hsonmez@imicryl.com.tr
### Device Information:
Trade Name: R&D SERIES NOVA COMPO N®
Common Name: Light Cure Nano Hybrid Composite
Product Code: EBF
Classification: Class II
Classification Name: Tooth Shade Resin Material
Regulation Number: 872.3690
Review Panel: Dental
#### Predicate Devices:
R&D Series Nova Compo N light cure nano hybrid composite is substantially equivalent to the following marketed products:
{4}------------------------------------------------
| COMPANY | DEVICE | 510(k) NUMBER | PRODUCT CODE |
|--------------------------------|---------------------------|---------------|--------------|
| Ivoclar Vivadent, Incorporated | Tetric Evoceram | K042819 | EBF |
| Kuraray Medical, Inc. | Clearfil Majesty Esthetic | K061860 | EBF |
### Indication for Use:
R&D Series Nova Compo N is a light-curing, radiopaque, nano-hybrid universal composite for anterior and posterior restorations. It is intended for professional use by below mentioned Indications for Use;
- . Restorations of deciduous teeth
- Splinting of mobile teeth ●
- Restoration in the posterior region (Classes I and II) ●
- Anterior restorations (Classes III, IV)
- . Class V restorations (cervical caries, root erosion, wedge-shaped defects)
- . Extended fissure sealing in molars and premolars
- Veneering of discoloured anterior teeth .
- Repair of composite and ceramic veneers .
### Device Description:
R&D Series Nova Compo N is a light-curing, radiopaque, nano hybrid composite for anterior and posterior restorations.
According to the applicable FDA recognized consensus standard: ISO 4049:2009 "Dentistry-Polymer Based Restorative Materials", this device is classified into the following:
- -Type 1: polymer-based filling and restorative materials;
-Class 2: Materials whose setting is effected by light;
-Group 1: Materials whose use requires the energy to be applied intra-orally.
R&D Series Nova Compo N light cure nano hybrid composite is designed with the high reduced shrinkage resin. The higher molecular weight of the resin results in less shrinkage, reduced aging and a slightly softer resin matrix. Additionally these resins impart a greater hydrophobicity and are less water absorption. Prepolymerized fillers provide low stress, non-sticky, easy handling and shaping formulation. High loading Prepolymerized fillers provide better wear resistance than traditional less loaded Prepolymerized fillers in other marketing composites. The nanofillers provide high polish, high wear, surface hardness and making the material smooth.
R&D Series Nova Compo N light cure nano hybrid composite was tested for biocompability and found to be biocompatible as a result of testing.
### Substantial Equivalence:
The applicant device has the same intended use as the 510(k) cleared predicates listed above.
Table 1 below shows a comparison of R&D Series Nova Compo N light cure nano hybrid composite and the predicates.
### Table 1
| DESCRIPTIVE<br>INFORMATION | NEW DEVICE/R&D SERIES<br>NOVA COMPO N<br>Light cure nano hybrid<br>composite | PREDICATE/TETRIC EVOCERAM | PREDICATE/CLEARFILL<br>MAJESTY ETSHETIC |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INDICATION FOR USE | Restorations of deciduous teeth Splinting of mobile teeth Restorations in the posterior region (classes I and II) Anterior restorations (classes III, IV) | Restorations in the posterior region (Classes I and II) Anterior restorations (Classes III, IV) Class V restorations (cervical caries, root erosion, wedge-shaped defects) Veneering of discoloured | Direct restorations for all cavity classes in anterior and posterior teeth Direct veneers Correction of tooth position and tooth shape (e.g. diastema closure, dwarfed tooth, etc.) |
| | Class V restorations (cervical caries, root erosion, wedge-shaped defects) Extended fissure sealing in molars and premolars Veneering of discoloured anterior teeth Repair of composite and ceramic veneers | anterior teeth Splinting of mobile teeth Repair of composite/ceramic veneers | Intraoral repairs of fractured restorations |
| COMPOSITION | | | |
| | The monomer matrix is<br>composed of<br>Bis-GMA, hydrophobic aromatic<br>dimethacrylates, UDMA,<br>Bis-EMA,<br>The fillers contain barium glass,<br>ytterbium trifluoride, nano filler<br>and prepolymer as an organic<br>filler and inorganic filler.<br>dl-Camphorquinone<br>additives,<br>initiators,<br>stabilizers<br>Accelerators<br>Pigments. | The monomer matrix is composed<br>of dimethacrylates.<br>The fillers contain barium glass,<br>ytterbium trifluoride, mixed oxide<br>and copolymers.<br>Additives,<br>Initiators,<br>Stabilizers<br>Pigments<br>Urethane dimethacrylates,<br>Bis-GMA,<br>Ytterbium trifluoride,<br>Ethyoxylated bisphenol A<br>dimethacrylates (Bis-EMA)<br>dl-Camphorquinone | Silanated barium glass filler.<br>Pre-polymerized organic filler<br>including nano filler<br>Bis-phenol A diglycidyl<br>methacrylate (Bis-GMA) .<br>Hydrophobic aromatic<br>dimethacrylate .<br>dl-Camphorquinone .<br>Initiators,<br>Accelerators<br>Pigments<br>YbF3( Ytterbium trifluoride) |
| PHYSICAL PROPERTIES | | | |
| Compressive<br>Strength (MPa) | 360 | 250 | 356 |
| Flexural Strength<br>(MPa) | 130 | 120 | 118 |
| Depth Of Cure<br>(mm) | >=2 | >1,5 | >1,5 |
| Particle Size<br>Range | 40 nm and 3 µm. | Particle Size Range: 40nm-<br>3micron | Particle Size Range: 20nm-<br>3micron |
| Water Sorption (<br>µg/mm3) | 21,2 | 21,2 | 25,3 |
| Water Solubility<br>( μg/mm3) | < 1,0 | < 1,0 | <1,5 |
| Elastic Modulus | 10 GPa | 10 GPa | 10 GPa |
| Radio-opacity<br>( mm of Al) | %350 | %400 | %180 |
| Working Time<br>(second) | 250 | 200 | 250 |
| Curing Time<br>(second) | 20 | 20 | 20 |
| Sensitivity to<br>Ambient Light<br>(second) | 250 | 200 | 250 |
| Color<br>stability | YES | YES | YES |
{5}------------------------------------------------
# Similarities
- Tetric Evo Ceram, Clearfil Majesty and Nova Compo N are light cured glass filled nano hybrid composites. All . devices have the same basic structure. The basic structures are glass fillers and methacrylate resins.
- . Tetric Evo Ceram and Nova Compo N contain the same resin matrix is UDMA (Urethane Dimethacrylate), Bis-GMA, Bis-EMA (Ethoxylated bisphenol-A dimethacrylate) and dl-comphoroquione. Clearfil Majesty and Nova Compo N have similar resin matrix. The similar resins are Bis-gma, hydrophotic aromatic dimethacrylate, dl-comphoroquione.
{6}------------------------------------------------
- . All devices have inorganic (ba-glass etc.) and organic fillers (prepolymers, copolymers). Tetric Evo Ceram, Nova Compo N and Clearfil Majesty are containing organic fillers are called prepolymer in Nova Compo N, Clearfil Majesty and copolymer in Tetric Evo Ceram. Basically prepolymer and copolymers mean the same material. Prepolymers are being used in composite formulations to reduce the polymerization shrinkage and improve the handling properties. Prepolymer and copolymers: The composite resins and fillers are mixed, moulded, polymerized and grind to small particles. As a result sometimes the material is called prepolymer. Copolymer is a prepolymer. Because prepolymer is produced from different base materials.
- All devices contain ba-glass, ytterbium triflouride for radiopacity as inorganic filler.
- . Nova Compo N and Clearfil Majesty also contain nano filler.
- The usage areas and purposes of the products are the same.
- Physical Property data similar to the predicate device. ●
### Differences
- Tetric Evo Ceram contains mixed oxide and Nova Compo N contains nano filler. The aim of both mixed oxide and nano filler materials are improve the surface gloss, increase mechanical properties and to fill the voids of bigger particles.
- Clearfil Majesty does not contain UDMA and Bis-EMA but Tetric Evo Ceram and Nova Compo N contain these monomer. But all devices have similar resin ratio. In Clearfil Majesty Bis-gma and hydrophobic aromatic dimethacrylate are used instead of UDMA and Bis-EMA. But it is not significant because they are both used extensively in Dental composites, have no substantial history of adverse events.
#### Non-Clinical Performance Testing:
#### Biocompatibility Testing:
In accordance with ISO 10993-1 (Biological Assessment Medical Devices-Part1: Evaluation and Testing) and ISO 7405 (Dentistry - Biocompatibility of Medical Devices used in Dentistry) standards, biocompatibility was evaluated for R & D Series Nova Compo N light cure nano hybrid composite.
The biocompatibility data for R & D Series Nova Compo N light cure nano hybrid composite are given in the table below.
| TEST NAME | RESULT |
|----------------------------|----------------------------------------------|
| In vitro cytotoxicity | Does not have cytotoxic potential |
| Sensitization | Does not cause sensitivity |
| Irritation | Does not cause irritation |
| Acute systemic toxicity | Does not have acute systemic toxic effect |
| Subacute systemic toxicity | Does not have subacute systemic toxic effect |
| Genotoxicity | Does not have genotoxic potential |
| Implantation | Irritant effect was not found |
#### Physical Properties:
In-vitro bench tests were performed on the R&D Series Nova Compo N light cure nano hybrid composite according to the requirements in ISO 4049: 2009 (Dentistry - Polymer-based restorative).
Bench tests included in support of the substantial equivalence of R&D Series Nova Compo N light cure nano hybrid composite are:
- Compressive Strength
- Flexural Strength
- •Depth of Cure
- Particle Size Range
- Water Sorption
- Water Solubility
- Elastic Modulus
- Radiopacity
- Working Time
- Curing Time
- Sensitivity to Ambient Light
- Color Stability
The results from bench tests included in this premarket notification support the substantial equivalence of the R&D Series Nova Compo N light cure nano hybrid composite.
{7}------------------------------------------------
# Conclusion:
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, IMICRYL DIS MALZEMELERI SANAYI VE TICARET A.S. concludes that the R&D SERIES NOVA COMPO N light cure nano hybrid composite is safe, effective and substantially equivalent to the predicate devices as described herein. It does not introduce new indications for use, has similar technological characteristics and does not introduce new potential hazards or risks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.