The VistaScan™ USB Ultrasound Imaging device is intended for diagnostic ultrasound imaging in B mode for the following applications: Fetal/obstetric Gynecology Abdominal Pediatric Small organ Musculo-skeletal (conventional) Musculo-skeletal (superficial) Urology Pelvic floor Neuro-muscular Peripheral vessel Indications for use vary by probe type. The VistaScan™ USB Ultrasound Imaging System is indicated for Rx (prescription) use and intended for use by appropriately trained healthcare professionals such as physicians and ultrasound technologists. The VistaScan™ USB Ultrasound Imaging System is intended for use in healthcare facilities such as hospitals, clinics and facilities where ultrasound testing is performed.
Device Story
VistaScan™ is a portable, B-mode ultrasound imaging system consisting of handheld probes (non-array, curved, and flat linear array) and an Android-based application. Probes contain an ultrasound generator/receiver, analog-to-digital converter, microcontroller, and USB 2.0 interface. The system connects to a Samsung Galaxy S4 tablet via USB. Operators use the application to control transmit power, image settings, and Time Gain Compensation (TGC), and to perform 2D measurements. The device is used by physicians and technologists in hospitals and clinics to visualize internal structures for diagnostic purposes. It provides high-resolution, high-penetration imaging to assist in clinical decision-making across various anatomical applications. The system includes automated safety/performance checks (initialization, hardware check, probe ID verification) prior to use.
Clinical Evidence
No clinical data. Bench testing only. Evaluation included acoustic output, clinical measurement accuracy, system sensitivity, thermal/mechanical/electrical safety, biocompatibility of patient-contacting materials, and software verification against FDA-recognized standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37).
Technological Characteristics
Portable B-mode ultrasound system; non-array, curved, and flat linear array transducers (2.5-10.0 MHz). USB 2.0 connectivity. Android-based software platform. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-37 (ultrasonic diagnostic equipment). Non-sterile. Mechanical and solid-state transducer types.
Indications for Use
Indicated for diagnostic ultrasound imaging in B-mode for fetal/obstetric, gynecological, abdominal, pediatric, small organ, musculoskeletal (conventional/superficial), urological, pelvic floor, neuromuscular, and peripheral vessel applications. Intended for prescription use by trained healthcare professionals (physicians, ultrasound technologists) in clinical settings.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 23, 2020
Emagine Solutions Technology, LLC % Ms. Michelle Lott Senior RA & QA Consultant Lean RAQA, LLC 12602 North Summerwind Drive MARANA AZ 85658
# Re: K192170
Trade/Device Name: VistaScan™ USB Ultrasound Imaging System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO, ITX Dated: December 18, 2019 Received: December 23, 2019
Dear Ms. Lott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192170
#### Device Name
VistaScan™ USB Ultrasound Imaging System
Indications for Use (Describe) The VistaScan™ USB Ultrasound Imaging System is intended for diagnostic ultrasound imaging in B mode for the following applications: Fetal/obstetric Gynecology Abdominal Pediatric Small organ Musculo-skeletal (conventional) Musculo-skeletal (superficial) Urology Pelvic floor Neuro-muscular Peripheral vessel
Indications for use vary by probe type. The VistaScan™ USB Ultrasound Imaging System is indicated for Rx (prescription) use and intended for use by appropriately trained healthcare professionals such as physicians and ultrasound technologists. The VistaScan™ USB Ultrasound Imaging System is intended for use in healthcare facilities such as hospitals, clinics and facilities where ultrasound testing is performed.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| <span style="white-space: nowrap;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="white-space: nowrap;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K192170
# 510(k) Summary
# GENERAL INFORMATION
This 510(k) Summary is submitted in accordance with 21 CFR 807, Section 807.92.
### Submitted by:
| Owner's Name:<br>Address:<br>Contact:<br>Title:<br>Tel:<br>Fax:<br>Email: | Emagine Solutions Technology, LLC<br>9040 S. Rita Road, Suite 1270 Tucson, AZ 85747<br>Courtney Williams<br>CEO<br>+1 (520) 485-9905<br>None<br>courtney.williams@emaginesolutionstech.com |
|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | |
| Name:<br>Title:<br>Tel:<br>Email: | Michelle Lott<br>Senior RA & QA Consultant at Lean RAQA, LLC<br>+1 (520) 275 9838<br>michelle@leanraqa.com |
| Date Prepared: | January 15, 2020 |
| Trade Name: | VistaScan™ USB Ultrasound Imaging System |
| Common Name: | Diagnostic Ultrasound System and Accessories |
| Classification Name:<br>Classification: | Ultrasonic pulsed echo imaging system, Diagnostic Ultrasound<br>Transducer<br>Class II |
| Product Code: | IYO, ITX |
| Regulation Number: | 892.1560, 892.1570 |
| Predicate Device(s): | Interson USB Ultrasound System, Interson Corporation.<br>(K163443) (Primary Predicate device) |
| | Interson USB Ultrasound Probe System, Interson Corporation.<br>(K070907) |
| Reference Device: | Philips Lumify<br>(K152899) |
{4}------------------------------------------------
### DEVICE DESCRIPTION:
The VistaScan™ USB Ultrasound Imaging System is a self-contained portable single-mode and multiple-application ultrasound imaging system. The system comprises a series of handheld probes containing an ultrasound generator/receiver, analog to digital converter, microcontroller, control logic, USB 2.0 interface, and software-based controls offering B operating mode, parameter controls, and recording functions. The selection of transducers offered with the system permit a wide range of clinical applications including Fetal/Obstetric, Abdominal, Pediatric, Small Organ, Muscular-skeletal, Urology, Gynecology, Pelvic Floor, Neuromuscular, and Peripheral Vessel.
The initial operational settings for each transducer are preprogrammed in the system. User-customized parameter settings for each transducer may be set by the operator within a specific range which is controlled by the transducer hardwareffirmware and stored for recall as needed via the system control panel. Customization includes transmit power, images controls selection, and Time Gain Compensation (TGC). Controls are also provided to select display format and to utilize the Cine function.
The VistaScan™ USB Ultrasound Imaging System is a B-Mode scanner and supports a wide variety of applications. It is an ultrasound scanner providing high resolution and high penetration performance. Probes are supported in frequencies from 2.5 MHz to 10.0 MHz. The product family uses non-array, curved, and flat linear array transducers and includes two General Purpose (GP) probes and a Small Parts (SP) probe. The probes can be connected to a USB 2.0 port (micro USB and USB-C).
System Components:
- Android™ based device operating VistaScan™ application .
- Transducer GP-3.5 ●
- Transducer GP-C01 ●
- . Transducer SP-L01
### Supported Hardware
The VistaScan™ USB Ultrasound Imaging System has been verified and validated on the Samsung Galaxy S4 Tablet using the Android software Operating System (OS). Device safety and performance cannot be assured when using other computing hardware which utilizes the Android OS. Computing hardware other than the Samsung Galaxy S4 Tablet using the Android software Operating System (OS) is not supported and should not be used with the VistaScan USB Ultrasound Imaging System.
### Automated Safety/Performance Checks
The VistaScan™ USB Ultrasound Imaging System includes several automated safety/performance checks to verify that the USB ultrasound probe is functioning properly prior to use. These automated performance checks include:
- System Initialize
- . Hardware Initialize
- Hardware Check ●
- Probe ID Code Read ●
- . Setup Control
{5}------------------------------------------------
# INTENDED USE / INDICATIONS FOR USE:
The VistaScan™ USB Ultrasound Imaging device is intended for diagnostic ultrasound imaging in B mode for the following applications:
- . Fetal/obstetric
- Gynecology ●
- Abdominal ●
- Pediatric ●
- Small Organ ●
- Musculo-skeletal (conventional) ●
- Musculo-skeletal (superficial)
- Urology
- Pelvic Floor ●
- Neuro-muscular
- . Peripheral Vessel
Indications for use vary by probe type. The VistaScan™ USB Ultrasound Imaging System is indicated for Rx (prescription) use and intended for use by appropriately trained healthcare professionals such as physicians and ultrasound technologists. The VistaScan™ USB Ultrasound Imaging System is intended for use in healthcare facilities such as hospitals, clinics and facilities where ultrasound testing is performed.
### PRODUCT MODELS:
| Model | Track | Common Applications |
|-----------------------------|-------|-------------------------------------------------------------------------------------------------|
| Transducer Model GP 3.5 MHz | 1 | Fetal, Abdominal, Small Organ |
| Transducer Model GP-C01 | 3 | Fetal/Obstetric, Abdominal, Pelvic Floor, Urology, Musculoskeletal, Small Organs, Neuromuscular |
| Transducer Model SP-L01 | 3 | Pediatric, Small Organs, Peripheral Vessel, Pelvic Floor |
Table 1: Common Applications of Transducer Models
# LABELING AND TECHNOLOGICAL CHARACTERISTICS COMPARISON:
Both the proposed VistaScan™ USB Ultrasound Imaging System and the predicate devices are provided nonsterile. They use similar packaging systems except for adapted product documentation (instructions for use in package labeling). The proposed VistaScan™ device has an identical intended use for B mode imaging, and similar technology characteristics to the currently marketed Interson USB Ultrasound Probe System (K070907, K163443). Technological differences are limited to the software operating system. The VistaScan™ device uses a commercially available electronic mobile device with an Android-based operating system, while the predicate Interson USB Ultrasound Probe System (K070907, K163443) device uses a proprietary Windows-based imaging system. All other technological features are substantially equivalent to the Interson USB Ultrasound Probe System (K070907, K163443). The reference device was selected as a point of comparison for the software operating system. Both the
{6}------------------------------------------------
VistaScan™ device and the reference device, Philips Lumify (K152899), use an Android based operating system.
| Device<br>Features | Subject Device:<br>VistaScan ™ USB<br>Ultrasound<br>Imaging System | Primary Predicate:<br>Interson Ultrasound<br>System: K163443 | Secondary Predicate:<br>Interson Ultrasound<br>System K070907 | Reference Device:<br>Philips Lumify K152899 |
|-------------------------|-----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Intended Use | Diagnostic ultrasound<br>imaging in B mode. | Diagnostic ultrasound<br>imaging in B, color<br>Doppler and Combined<br>(B + Color)<br>modes. | Diagnostic ultrasound<br>imaging in B mode (all<br>transducers), A mode<br>(ophthalmic) | Diagnostic ultrasound<br>imaging in B, color<br>Doppler, and Combined<br>(B + Color)<br>modes. |
| Indications for<br>Use | Indicated for diagnostic<br>ultrasound imaging in<br>specified applications | Indicated for diagnostic<br>ultrasound imaging and<br>fluid flow analysis in<br>specified applications | Indicated for diagnostic<br>ultrasound imaging in<br>specified applications | Indicated for diagnostic<br>ultrasound imaging and<br>fluid flow analysis in<br>specified applications |
| Array<br>Geometry | Non-Array<br>Curved and linear | Curved and linear | Non-Array | Curved and linear |
| Mechanics | Mechanical<br>Solid State | Solid State | Mechanical | Solid State |
| Software<br>platform | Commercial off-the-shelf<br>operating system<br>(Android) | Commercial off-the-shelf<br>operating system<br>(Windows) | Commercial off-the-shelf<br>operating system<br>(Windows) | Commercial off-the-<br>shelf operating system<br>(Android) |
| Software<br>control | Standalone | Standalone | Standalone | Standalone |
| Measurement<br>function | 2D measurement | 2D measurement and<br>area measurement | 2D measurement tool | 2D measurement tool |
| Wireless<br>networking | Not supported | Not supported | Not supported | Supported |
| Connector | USB | USB | USB | USB |
#### Table 2: Predicate Comparison Table
# NON-CLINICAL PERFORMANCE TESTING
The VistaScan™ USB Ultrasound Imaging System was generally evaluated against relevant requirements in Marketing Clearance of Diagnostic Ultrasound Systems and Transducers: Guidance for Industry and Food and Drug Administration Staff, June 27, 2019.
- Acoustic Output ●
- Clinical measurement accuracy and system sensitivity
- Thermal, mechanical, and electrical safety ●
- Patient -contacting materials ●
- Cleaning, disinfection, sterilization, and pyrogenicity
- Software ●
{7}------------------------------------------------
Verification testing was conducted against predetermined acceptance criteria and recognized standards1 to show that the subject device performs functions equivalent to the predicate device. VistaScan™ successfully passed verification testing. The results support substantial equivalence to the predicate device and demonstrate that is safe and effective for its intended use.
# Electrical Safety / Electromagnetic Compatibility
Evaluation per standard IEC 60601 1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) and IEC/EN 60601-1-2:2014 was performed for use of the transducer with a specific computer model (HP Netbook). Use of alternate USB 2.0 compatible computer hardware requires verification by the end user. Further information is provided in the Instructions for Use.
# Standards Conformity
The VistaScan™ USB Ultrasound Imaging System references the following FDA recognized standards:
| Standard Number/FDA Recognition# | Standard Name |
|--------------------------------------|-----------------------------------------------------------|
| IEC 60601 1: 2005 + CORR. 1 (2006) + | Medical electrical equipment-Part 1: General |
| CORR. 2 (2007), | requirements for basic safety and essential performance |
| FDA Recognition#: 19-4 | |
| IEC/EN 60601-1-2:2014 4th Ed. | Medical electrical equipment -Part 1-2: General |
| FDA Recognition#: 19-8 | requirements for basic safety and essential performance . |
| | Collateral Standard: Electromagnetic disturbances - |
| | Requirements and tests |
| IEC 60601-2-37:2007 | Medical electrical equipment -Part 2-37: Particular |
| FDA Recognition#: 12-293 | requirements for the basic safety and essential |
| | performance of ultrasonic medical diagnostic and |
| | monitoring equipment |
# ANIMAL AND CLINICAL TESTING
VistaScan™ USB Ultrasound Imaging System introduces no new indication for use, modes, features, or technologies relative to the predicate devices that require animal or clinical safety and effectiveness of ultrasound systems with these characteristics are well accepted for both predicate and subject devices.
### CONCLUSION OF COMPARISON
The VistaScan™ USB Ultrasound Imaging System device and its predicate device are technologically The devices are identical in their use of probes, and similar in design, materials, and similar. technological characteristics. The packaging and labeling has been designed to provide sufficient information to the user to ensure the safe and effective use of the proposed device. Furthermore, the proposed device is comparable to the reference device which was cleared for the same intended use for B mode operation, the same indication for use, and utilizes the same operating platform. The proposed device does not raise substantial new questions of safety or effectiveness. Therefore, the proposed device (VistaScan™ USB Ultrasound Imaging System) is determined to be substantially equivalent to the predicate device.
<sup>1</sup> Electrical Safety and EMC standards are applied by 'right of reference'.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.