Canady Plasma SMART XL-1000 Electrosurgical Generator with Accessories
K192124 · Us Medical Innovations, LLC · GEI · Nov 21, 2019 · General, Plastic Surgery
Device Facts
Record ID
K192124
Device Name
Canady Plasma SMART XL-1000 Electrosurgical Generator with Accessories
Applicant
Us Medical Innovations, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 21, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories is intended to provide gas-enhanced coagulation during general surgery procedures. The Cana® SMART XL-1000™ Electrosurgical Generator with Accessories is designed for gas enhanced coagulation when used only with the Canady Plasma Electrosurgery Unit and compatible monopolar and/or bipolar handpieces. The Canady Electrosurgery Unit Generators are intended to cut and/or coagulate tissue when used with compatible monopolar and/or bipolar RF handpieces. The Canady Electrosurgery Unit Generators are capable of monopolar argon gas enhanced coagulation when used with Canady Plasma Coagulator and probes
Device Story
High-frequency (HF) electrosurgical generator; provides tissue cutting and coagulation via monopolar/bipolar RF energy and argon gas-enhanced plasma. Input: user-selected settings via touchscreen; electrical energy from wall power. Transformation: FPGA-controlled HF voltage applied to tissue via compatible handpieces; argon gas ionized to create plasma stream for coagulation. Output: plasma gas stream, RF energy for cutting/coagulation, visual/audio status/error alerts on LCD monitor. Used in clinical/surgical settings by physicians. Device features programmable modes (pediatrics, orthopedics, endoscopy, gynecology) and real-time display of power (watts), gas flow, and energy mode. Benefits: provides surgical flexibility for tissue removal and hemostasis; improved programmable software over predicate; integrated gas coagulation unit.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed product specifications meet requirements and are equivalent to the predicate device. Testing included electromagnetic compatibility (IEC 60601-1-2), electrical safety (IEC 60601-1, IEC 60601-2-2), and performance/usability testing (e.g., power output, flow rate, touchscreen responsiveness, alarm functionality).
Technological Characteristics
High-frequency (390 kHz) electrosurgical generator; FPGA-based control; LCD touchscreen interface. Monopolar and bipolar modes; argon gas-enhanced coagulation. Input voltage 100-240 VAC; convection cooling. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2. Integrated gas coagulation unit.
Indications for Use
Indicated for general surgery patients requiring tissue cutting and/or coagulation, including gas-enhanced coagulation, using compatible monopolar or bipolar RF handpieces.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Canady Plasma® Electrosurgical Unit SS Series with Accessories (K100669)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
November 21, 2019
US Medical Innovations, LLC % Timothy Joiner, Senior Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746
Re: K192124
Trade/Device Name: Canady Plasma SMART XL-1000 Electrosurgical Generator with Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: October 22, 2019 Received: October 28, 2019
Dear Timothy Joiner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192124
Device Name
Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories
#### Indications for Use (Describe)
The Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories is intended to provide gasenhanced coagulation during general surgery procedures. The Cana® SMART XL-1000™ Electrosurgical Generator with Accessories is designed for gas enhanced coagulation when used only with the Canady Plasma Electrosurgery Unit and compatible monopolar and/or bipolar handpieces.
The Canady Electrosurgery Unit Generators are intended to cut and/or coagulate tissue when used with compatible monopolar and/or bipolar RF handpieces. The Canady Electrosurgery Unit Generators are capable of monopolar argon gas enhanced coagulation when used with Canady Plasma Coagulator and probes
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Special 510(k) Summary
Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories
## 1. Submission Sponsor
US Medical Innovations, LLC 6930 Carroll Ave Suite 1000 Takoma Park MD, 20912 USA Contact: Jerome Canady Title: CEO
# 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327-9997 Contact: Timothy Joiner Title: Senior Consultant
### 3. Date Prepared
August 5, 2019
# 4. Device Identification
Trade/Proprietary Name: Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories Common/Usual Name: Electrosurgical Generator Classification Name: Electrosurgical cutting and coagulation device and accessories Regulation Number: 878.4400 Product Code: GEI Class: 2 Classification Panel: General and Plastic Surgery
### 5. Legally Marketed Predicate Device
Canady Plasma® Electrosurgical Unit SS Series with Accessories (K100669)
### 6. Indication for Use Statement
{4}------------------------------------------------
The Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories is intended to provide gas-enhanced coagulation during general surgery procedures. The Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories is designed for gas enhanced coagulation when used only with the Canady Plasma Electrosurgery Unit and compatible monopolar and/or bipolar handpieces.
The Canady Electrosurgery Unit Generators are intended to cut and/or coagulate tissue when used with compatible monopolar and/or bipolar RF handpieces. The Canady Electrosurgery Unit Generators are capable of monopolar argon gas enhanced coagulation when used with Canady Plasma Coagulator and probes.
# 7. Device Description
The Canady Plasma® SMART XL-1000™ Electrosurgical Generator (CPSXEG) with Accessories is a High Frequency (HF) Electrosurgery Unit (ESU) used in combination with electrosurgical probes. The CPSXEG combines (HF) voltage to electrically enhanced plasma gas (i.e. Argon) to produce a plasma gas stream. Current density upon arrival at the tissue surface from Canady Plasma Probes causes coagulation of the tissue. The CPSXEG is also a monopolar and bipolar RF surgical device for cutting and coagulation. It is used for tissue removal and coagulation in accordance with the operative procedure performed by the physician. The unit provides various cutting, coagulation, hybrid plasma, and argon plasma modes to meet the physician's flexibility performing the specific procedure with the device
The modified Canady Plasma Smart XL-1000™ ESU has the same intended use as the predicate SS-601/Argon 4 Canady Plasma® Electrosurgical Unit as well as the same performance specifications. The modified device has the same protective circuits as the predicate ESU which displays the equipment output error on the monitor screen with a graphic symbol so that the user can determine the cause of the error. The packaging, labeling (i.e. user manual) are the same.
The Canady Plasma® SMART XL-1000™ Electrosurgical Generator has a touch screen color monitor that provides the user with quick start, on screen tutorial and user manual as well as operational and settings. Various cutting, coagulation, bipolar, and plasma modes are the predicate unit, but the modified device has improved programmable software. When turned on, an APPS screen is displayed on the monitor screen. The user can decide which medical application is started by touching the particular application (i.e Pediatrics, orthopedics, endoscopy, gynecology). Upon activation of the device audio and visual error system malfunctions or user errors can be observed by the user. Once the device is activated, power in watts, gas flow rate, gas volume, PPM, and energy mode are displayed on the monitor screen.
# 8. Substantial Equivalence Discussion
The following table compares the Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories to the predicate device with respect to intended use, principles of operation, technological characteristics, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.
# Table 5A – Comparison of Characteristics
{5}------------------------------------------------
| Trade Name | Canady Plasma Electrosurgery<br>Unit<br>(Predicate Device) | Canady Plasma SMART<br>Electrosurgical Generator XL-<br>1000 | Significant<br>Difference |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Manufacturer | USMI | USMI | N/A |
| 510(k) Number | K100669 | N/A | N/A |
| Model | SS-601MCa | XL-1000 | N/A |
| CPC Used With | Canady Plasma Coagulator<br>Argon 4 | CPC is consolidated into the<br>XL-1000 | N/A |
| Intended Use | Cut and/or coagulate tissue | Cut and/or coagulate tissue | None |
| Generator Type | High Frequency | High Frequency | None |
| Frequency | 390 kHz | 390 kHz | None |
| Peak to Peak Voltage<br>(rated Ω) | 4kV (1000 Ω) | 4kV (1000 Ω) | None |
| Main CPU Type | FPGA | FPGA | None |
| Display Type | LED with backlight | LCD with backlight | Minor |
| Monopolar | Yes, insulated | Yes, insulated | None |
| Bipolar | Yes | Yes | None |
| Handswitch | Yes | Yes | None |
| Footswitch | Yes, independent switches for<br>monopolar and bipolar | Yes, independent switches for<br>monopolar and bipolar | None |
| Monopolar Modes | Cut (Pure, Blend I,II,III, Pure<br>High Cut, Blend High Cut I,II), E-<br>Cut<br>Coagulation (Spray, Fulgurate<br>High, Fulgurate Low, Desiccate<br>1,II,III) | Cut (Pure, Blend I,II,III, Pure<br>High Cut, Blend High Cut I,II), E-<br>Cut<br>Coagulation (Spray, Fulgurate<br>High, Fulgurate Low, Desiccate<br>I,II,III) | None |
| Bipolar Modes | Cut, Coagulate (Micro, Default,<br>Macro) | Cut, Coagulate (Micro, Default,<br>Macro) | None |
| Volume Control | Yes | Yes | None |
| Defeatable-Inaudible<br>Volume Control | No | No | None |
| Power Setting Shown | Yes | Yes | None |
| Cooling Type | Convection | Convection | None |
| Input Voltage, VAC | 100 – 240 | 100 - 240 | None |
| Meets IEC 60601-1 | Yes | Yes | None |
| Meets IEC 60601-1-2 | Yes | Yes | None |
| Meets IEC 60601-2-2 | Yes | Yes | None |
# 9. Non-Clinical Performance Data
Internal verification and validation testing confirm that product specifications of the Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories are met. These are equivalent to those of the unmodified predicate device. The testing/measurement of the output parameters of the Canady Plasma® SMART XL-1000™ Electrosurgical Generator (CPSEG-XL-1000) was performed to demonstrate that the CPSEG-XL-1000 performed as specified and in accordance with the specifications of the predicate device USMI SS-601/Argon 4. The results support that the device's design changes do not affect the safe and effective use of the device as compared to the unmodified predicate device.
{6}------------------------------------------------
US Medical Innovations, LLC established design and development quality system procedures to direct and control hardware and software design and development activities. The following procedures are: Development plan, design, Design input / Software (as applicable) Risk Management/ Risk Analysis, Design Review, Validation Master Sheet, Verification Master Sheet, Design Change Master Sheet, Design Transfer Checklist and Release Documentation. The activities provide a framework for final project and management approval for the necessary controls on the design and development of new products.
The Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories has the same principles of operation and technological characteristics as the predicate device in the previous 510k (K100669).
Electromagnetic Compatibility, Electrical Safety, and Performance Testing of the Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories included the following testing:
| EM / ES | Performance | Usability |
|---------------------------------------------------------------------------|-------------------------|-------------------------------------------------------|
| Conducted Emissions - Voltage | Crest Factor Comparison | Operation Modes Selection Test |
| Radiated Emissions | Frequency | Power Adjustment Test |
| Harmonic Current Emission | Peak to Peak Voltage | Flow Rate Adjustment Test |
| Voltage Fluctuations / Flicker | Monopolar Cut Mode | Standby Mode System Activation Test |
| Radiated Electromagnetic Field<br>Immunity | Maximum Watts | Gas Tank Pressure Value Indication<br>Test |
| Immunity to Proximity Fields from RF<br>Wireless Communications Equipment | Maximum Voltage | Audio / Visual Warning / Notification<br>Test |
| Surge Immunity | Bipolar Cut Mode | User Profile Selection Test |
| Conducted RF Immunity | Maximum Watts | User Profile Editing Test |
| Magnetic Immunity | Maximum Voltage | Font Size Visibility Test |
| Voltage Dips Interruption and<br>Variations | | Icon Size Visibility Test |
| | | Button Response Test |
| | | Font / Icon Color Code Test |
| | | Power Button Increment / Decrement<br>Speed Test |
| | | Power Mode Icon Graphic Visibility<br>Test |
| | | Patient Pad Connectivity Notification /<br>Alarm Test |
| | | Touchscreen EMF Immunity Test |
| | | LCD Brightness Visibility Test |
| | | Audio Volume Adjustment Test |
| | | Multitouch Prevention Test |
| | | Glove Test |
| | | Liquid Spill Test |
| | | Touchscreen LED Test |
| | | Touchscreen Impact Test |
| | | Chemical Resistance Test |
### 10. Statement of Substantial Equivalence
The Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories has the same intended use as the predicate device, and the same or similar technological characteristics. The modifications in technological characteristics do not raise new or different questions of safety and effectiveness. Performance testing in accordance with standards has demonstrated the Canady Plasma® SMART XL-
{7}------------------------------------------------
1000™ Electrosurgical Generator with Accessories is as safe and effective as the predicate device. Therefore, the Canady Plasma® SMART XL-1000™ Electrosurgical Generator with Accessories is substantially equivalent to the Canady Plasma® Electrosurgical Unit SS Series with Accessories.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.