← Product Code [DQO](/productcode/DQO) · K192122

# Trevo Trak 21 Microcatheter (K192122)

_Stryker · DQO · Nov 22, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K192122

## Device Facts

- **Applicant:** Stryker
- **Product Code:** [DQO](/productcode/DQO.md)
- **Decision Date:** Nov 22, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1200
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Microcatheter is indicated for use in the selective placement of devices and/or fluids, such as contrast media, into the peripheral, coronary, and neuro vasculature during diagnostic and/or therapeutic procedures.

## Device Story

Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter; includes radiopaque markers for fluoroscopic visualization; features hydrophilic coating to reduce friction. Used in peripheral, coronary, and neuro vasculature; operated by physicians during diagnostic or therapeutic procedures. Device advanced over guide wire to target location; facilitates delivery of interventional devices or fluids. Provided with rotating hemostasis valve. Benefits include precise navigation and delivery in tortuous anatomy.

## Clinical Evidence

Bench testing only. No clinical or animal studies conducted. Testing included dimensional verification, tensile strength, leak resistance (air/liquid), burst pressure, particulate characterization, coating integrity, torsion bond strength, kink resistance, chemical compatibility, tip flexibility, trackability, and simulated use in tortuous anatomical models. All tests met acceptance criteria.

## Technological Characteristics

Single-lumen, braided stainless steel shaft, variable stiffness. Materials: PTFE inner layer, polyolefin strain relief, polyurethane hub, platinum/iridium marker bands, acrylic adhesive. Hydrophilic coating. Dimensions: 2.4F/2.0F OD, 0.021" ID, 162 cm length. EO sterilization. Single-use. No software/algorithm.

## Regulatory Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

## Predicate Devices

- Trevo® Pro 18 Microcatheter ([K113260](/device/K113260.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.

November 22, 2019

Stryker Neurovascular Rhoda Santos Senior Principal Regulatory Affairs Specialist 47900 Bayside Parkway Fremont, California 94538

Re: K192122

Trade/Device Name: Trevo Trak™ 21 Microcatheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DQO, DQY Dated: November 7, 2019 Received: November 8, 2019

## Dear Rhoda Santos:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Xiaolin Zheng, Ph.D., M.S. Director (Acting) DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192122

Device Name Trevo Trak 21 Microcatheter

Indications for Use (Describe)

The Microcatheter is indicated for use in the selective placement of devices and/or fluids, such as contrast media, into the peripheral, coronary, and neuro vasculature during diagnostic and/or therapeutic procedures.

Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K192122

| Trade/Proprietary Name: | Trevo Trak™ 21 Microcatheter                                    |
|-------------------------|-----------------------------------------------------------------|
| Common Name:            | Catheter, intravascular, diagnostic                             |
| Regulation Name:        | Catheter, intravascular, diagnostic, 21 CFR 870.1200 – Class II |
| Product Code:           | DQO                                                             |
| Trade/Proprietary Name: | Trevo Trak™ 21 Microcatheter                                    |
| Common Name:            | Percutaneous Catheter                                           |
| Regulation Name:        | Percutaneous Catheter, 21CFR 870.1250 – Class II                |
| Product Code:           | DQY                                                             |
| Submitter:              | Stryker Neurovascular                                           |
|                         | 47900 Bayside Parkway                                           |
|                         | Fremont, CA 94538-6515                                          |
|                         | (FDA Registration Number: 3008853977)                           |
| Contact:                | Rhoda M. Santos                                                 |
|                         | Senior Principal Regulatory Affairs Specialist                  |
|                         | Phone: 510-413-2269                                             |
|                         | Fax: 510-413-2724                                               |
|                         | Email: rhoda.santos@stryker.com                                 |
| Date Prepared:          | November 20, 2019                                               |

## Legally Marketed Predicate Devices

| Name of Predicate Device       | Name of<br>Manufacturer  | 510(k)<br>Number |
|--------------------------------|--------------------------|------------------|
| Trevo® Pro 18<br>Microcatheter | Stryker<br>Neurovascular | K113260          |

# Device Description

The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization.

{4}------------------------------------------------

## Indications for Use

The Microcatheter is indicated for use in the selective placement of devices and/or fluids, such as contrast media, into the peripheral, coronary, and neuro vasculature during diagnostic and/or therapeutic procedures.

# Technological Characteristics and Product Feature Comparison

Stryker Neurovascular has demonstrated the Trevo Trak™ 21 Microcatheter is substantially equivalent to the Predicate device, Trevo Pro18 Microcatheter (K113260) based on the same or similar materials, similar design, and the same fundamental operating principles. A comparison of the Subject device with the Predicate device is summarized in

Table 1 below.

| Detail                                                                   | Subject Device (K192122)                                                                                                                                                                                                                                              | Predicate Device (K113260)                                                                                                                                                                                                                                                    |
|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                                             | Stryker Neurovascular                                                                                                                                                                                                                                                 | Stryker Neurovascular                                                                                                                                                                                                                                                         |
| 510(k) Number                                                            | K192122                                                                                                                                                                                                                                                               | K113260                                                                                                                                                                                                                                                                       |
| Device Trade Name                                                        | Trevo TrakTM 21<br>Microcatheter                                                                                                                                                                                                                                      | Trevo® Pro 18 Microcatheter                                                                                                                                                                                                                                                   |
| Regulation Number                                                        | 21 CFR 870.1250                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                          |
| Regulation Name                                                          | Percutaneous Catheter                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                          |
| Classification                                                           | II                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                          |
| Product Code                                                             | DQY                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                          |
| Intended<br>Use/Indication for<br>Use                                    | The Microcatheter is indicated<br>for use in the selective placement<br>of devices and/or fluids, such as<br>contrast media, into the<br>peripheral, coronary, and neuro<br>vasculature during diagnostic<br>and/or therapeutic procedures.                           | The Microcatheter is indicated<br>for use in the selective placement<br>of fluids and/or other devices or<br>agents into the peripheral,<br>coronary, and neuro vasculature<br>during diagnostic and/or<br>therapeutic procedures.                                            |
| Device Description                                                       | The Microcatheter is a single-<br>lumen, braided shaft, variable<br>stiffness catheter with radiopaque<br>marker(s) on the distal end and a<br>luer hub on the proximal end.<br>The catheter shaft has a<br>hydrophilic coating to reduce<br>friction during use. The | The Microcatheter is a single-<br>lumen, braided shaft, variable<br>stiffness catheter with radiopaque<br>marker(s) on the distal end and a<br>luer hub on the proximal end.<br>The catheter shaft has a<br>hydrophilic coating on the distal 100cm to reduce friction during |
| Detail                                                                   | Subject Device (K192122)                                                                                                                                                                                                                                              | Predicate Device (K113260)                                                                                                                                                                                                                                                    |
|                                                                          | radiopaque shaft and distal<br>marker(s) facilitate fluoroscopic<br>visualization.<br>Device dimensions and<br>configuration are shown on the<br>product label. A rotating<br>hemostasis valve with side-arm<br>adapter is provided with each<br>microcatheter.       | use. The radiopaque shaft and<br>distal marker(s) facilitate<br>fluoroscopic visualization.<br>Device dimensions and<br>configuration are shown on the<br>product<br>label. A rotating hemostasis<br>valve with side-arm adapter is<br>provided with each<br>microcatheter.   |
| Accessory Devices<br>Provided (not in<br>direct contact with<br>patient) | Rotating Hemostasis Valve<br>(RHV) packaged within device                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                          |
| Outer Jacket                                                             | Polymeric microcatheter                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                          |
| Shaft Braid                                                              | Stainless Steel                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                          |
| Strain Relief                                                            | Polyolefin                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                          |
| Inner Layer                                                              | PTFE                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                          |
| Catheter Hub                                                             | Polyurethane                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                          |
| Marker Band                                                              | Platinum/Iridium                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                          |
| Adhesive                                                                 | Acrylic (Acrylated Urethane)                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                          |
| Outer Jacket Coating                                                     | Hydrophilic Coating                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                          |
| Labeled Shaft Outer<br>Diameter                                          | 2.4F/2.0F                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                          |
| Labeled Shaft Inner<br>Diameter                                          | .021"                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                          |
| Effective Length                                                         | 162 cm                                                                                                                                                                                                                                                                | 150 cm                                                                                                                                                                                                                                                                        |
| Packaging Materials<br>and Configuration                                 | HDPE Packaging Hoop,<br>Tyvek/Film Pouch, SBS Carton                                                                                                                                                                                                                  | HDPE Packaging Hoop, HDPE<br>mounting card, Tyvek/Film<br>Pouch, HDPE Tubing Clips, SBS<br>Carton                                                                                                                                                                             |
| Sterilization Method                                                     | EO Sterilization                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                          |
| How Supplied                                                             | Single Use/Sterile                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                          |
| Detail                                                                   | Subject Device (K192122)                                                                                                                                                                                                                                              | Predicate Device (K113260)                                                                                                                                                                                                                                                    |
| Principles of<br>Operation                                               | The device is advanced into the<br>vasculature over an appropriately<br>sized guide wire. Once the<br>microcatheter is inserted, the<br>catheter can be advanced through<br>the vasculature to the desired<br>location.                                               | Same                                                                                                                                                                                                                                                                          |

## Table 1. Product Feature Comparison of Subject Device to Predicate Device

{5}------------------------------------------------

{6}------------------------------------------------

## Testing Summary

## Performance Data – Bench Testing

The results of design verification and design validation testing conducted on the Trevo Trak 21 Microcatheter demonstrates that it performs as designed, is suitable for the indication for use, and is substantially equivalent to the legally marketed Predicate device. The design verification and design validation bench testing are summarized in Table 2 below.

| Test                                    | Test Method Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Conclusions                                                            |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Dimensional<br>Verification             | Purpose: To describe the procedure and technique<br>of making dimensional measurements using<br>various measurement equipment.<br><br>Method: Verify dimensions using specified<br>measurement tool. Record measurements.                                                                                                                                                                                                                                                                                                                                                                 | Dimensional<br>verification met<br>acceptance criteria.                |
| Tensile Strength                        | Purpose: To determine tensile strength of<br>catheters based on EN ISO 10555-1.<br><br>Method: Identify catheter joints, including<br>catheter to hub junction, and prepare sample for<br>test. Use tensile tester to apply tensile load to<br>sample and identify peak tensile force. Record<br>results.                                                                                                                                                                                                                                                                                 | Tensile strength<br>met acceptance<br>criteria.                        |
| Air Leak Resistance                     | Purpose: To determine whether the device meets<br>the freedom from leakage-air aspiration<br>requirement of 4.7.2 of EN ISO 10555-1 and<br>4.2.2 of EN 1707.<br><br>Method: Connect test hub sample to a partially<br>filled syringe. With the nozzle of the syringe<br>pointing down towards the ground, withdraw the<br>plunger to the 10cc mark. Hold for 15 seconds                                                                                                                                                                                                                   | Air leak resistance<br>of catheter met<br>acceptance criteria.         |
| Test                                    | Test Method Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Conclusions                                                            |
|                                         | and examine the water in the syringe for the<br>formation of air bubbles. Record results.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                        |
| Liquid Leak<br>Resistance               | Purpose: To determine whether the device meets<br>the freedom from leakage-liquid leak requirement<br>4.7.1 of EN ISO 10555-1 and 4.2.1 of EN 1707.                                                                                                                                                                                                                                                                                                                                                                                                                                       | Liquid leak<br>resistance of<br>catheter met                           |
|                                         | Method: Connect test hub sample to fixture and<br>flush with water to expel air. Occlude distal tip.<br>Apply pressure of 300 kPa to 320 kPa and<br>maintain pressure for 30s. Visually inspect<br>catheter/hub joint and catheter shaft for leaks.<br>Record results.                                                                                                                                                                                                                                                                                                                    | acceptance criteria.                                                   |
| Burst pressure                          | Purpose: To determine if the catheter can resist its<br>rated burst pressure without causing leakage or<br>damage to the catheter. This test is performed in<br>accordance with EN ISO 10555-1:2013, Annex F.                                                                                                                                                                                                                                                                                                                                                                             | Burst pressure met<br>acceptance criteria.                             |
|                                         | Method: Prepare test sample. Apply specified<br>burst pressure (greater than the rated burst<br>pressure) to test sample and maintain pressure for<br>at least 2 seconds. Observe sample for any<br>leakage or damage. Record results.                                                                                                                                                                                                                                                                                                                                                    |                                                                        |
| Particulate<br>characterization         | Purpose: To characterize acute phase particulate<br>generated as a result of product specific simulated<br>use steps.                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Particulate<br>generation was<br>acceptable.                           |
|                                         | Method: Measure the total number of particulates<br>generated during simulated use. Report in each of<br>four size ranges ≥10μm, ≥25μm, ≥50μm, and<br>≥100μm.                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                        |
|                                         | If >100μm are observed, complete particle count<br>analysis of bin sizes ≥200μm, ≥500μm and<br>≥1000μm. Record results.                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                        |
| Coating Integrity                       | Purpose: Coated surface shall be visually<br>characterized before and after simulated use with<br>sufficient magnification to identify any defects<br>which could be associated with loss of coating<br>under anticipated clinical use conditions.                                                                                                                                                                                                                                                                                                                                        | Coating integrity<br>was acceptable.                                   |
|                                         | Method: Remove test sample from packaging and<br>inspect for damage. Visually inspect external<br>surface of test sample using minimum 10x<br>magnification for surface defects. Record results.                                                                                                                                                                                                                                                                                                                                                                                          |                                                                        |
| Test                                    | Test Method Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Conclusions                                                            |
| Torsion Bond<br>Strength                | Purpose: To measure the strength of a catheter<br>shaft when torque is applied.<br><br>Method: Prepare test sample and insert into<br>torsional bond strength test fixture with tortuous<br>path model. Apply torque to proximal end of<br>catheter while distal end is fixed. Record<br>maximum observed number of 360-degree<br>rotations.                                                                                                                                                                                                                                              | Torsional Bond<br>Strength met<br>acceptance criteria.                 |
| Kink Resistance                         | Purpose: To measure the kink radius of a catheter<br>shafts.<br><br>Method: Prepare test sample. Position test sample<br>in a loop and place within a vice ensuring loop<br>apex is within the vice. Slowly close the vice until<br>the test sample kinks. Measure distance between<br>vice grips and record result.                                                                                                                                                                                                                                                                      | Kink resistance met<br>acceptance criteria.                            |
| Chemical<br>Compatibility               | Purpose: To determine visual and dimensional<br>integrity of catheter following exposure to saline<br>and non-ionic contrast liquids.<br><br>Method: Prepare sample for test. Flush sample<br>with water for 1 minute and then measure OD and<br>ID of test sample. Insert mandrel and take images<br>of test sample for initial characterization. Dip test<br>sample and mandrel into specified chemicals and<br>hold for 1 minute. Remove test sample, measure<br>OD, ID and cross-sectional area. Take images of<br>test sample for post-exposure characterization.<br>Record results. | Chemical<br>compatibility met<br>acceptance criteria.                  |
| Tip Flexibility                         | Purpose: To measure the maximum force required<br>to deflect test sample by 45°.<br><br>Method: Prepare sample for test. Use Instron and<br>appropriate fixtures to deflect test sample while<br>measuring force. Record maximum force<br>achieved.                                                                                                                                                                                                                                                                                                                                       | Tip Flexibility met<br>acceptance criteria.                            |
| Distal Tip Shape                        | Purpose: To verify that the catheter tip is smooth,<br>rounded, tapered or similarly finished in order to<br>minimize trauma to vessels during use per EN<br>ISO 10555-1.<br><br>Method: Visually inspect distal tip to verify distal<br>tip end is smooth, rounded, tapered or similarly<br>finished. Record results.                                                                                                                                                                                                                                                                    | Distal Tip Shape<br>met acceptance<br>criteria.                        |
| Test                                    | Test Method Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Conclusions                                                            |
| Surface Condition                       | Purpose: To verify the external surface of the<br>catheter is free from extraneous matter, process or<br>surface defects along the shaft.<br>Method: Visually inspect external surface of test<br>sample using minimum 2.5x magnification for<br>surface defects. Record results.                                                                                                                                                                                                                                                                                                         | Surface condition<br>met acceptance<br>criteria.                       |
| Distal Shaft<br>Trackability            | Purpose: To evaluate the force required to track<br>the catheter through a tortuous model as well as<br>evaluate the lubricity and durability of the<br>catheter.<br>Method: Prepare sample for test. Use track tester<br>to measure the track force of the test sample<br>through a tortuous model using a tri-axial<br>approach. Record the peak force during advance<br>tracking at 6th cycle.                                                                                                                                                                                         | Distal shaft<br>trackability met<br>acceptance criteria.               |
| Luer Testing per EN<br>1707             | Purpose: To determine if microcatheter hub and<br>accessory meet requirements 4.1 through 4.7 of<br>EN 1707.<br>Method: Test methods conforms with section 5 of<br>EN 1707.                                                                                                                                                                                                                                                                                                                                                                                                               | Luer testing meets<br>acceptance criteria.                             |
| <i>In-vitro</i> Simulated<br>Use Study  | Purpose: To evaluate interventional device<br>delivery, durability and integrity of the Subject<br>device as a general access and delivery catheter in<br>a tortuous anatomical model with multiple<br>physician users.<br>Method: Perform a simulated interventional<br>procedure by tracking the device to the target site<br>for interventional device delivery using a<br>neurovascular model that replicates the tortuosity,<br>diameter and location of the arteries in the<br>neurovasculature. Devices were visually<br>inspected post testing.                                   | All test samples met<br>acceptance criteria.                           |
| Compatibility with<br>Retrieval devices | Purpose: To ensure the microcatheter shall be<br>able to resheath a retriever in good condition<br>without a captured thrombus after deployment of<br>the entire shaped section into the vessel at least 2<br>times without functional impact and integrity to<br>the tip.                                                                                                                                                                                                                                                                                                                | Compatibility with<br>Retrieval devices<br>met acceptance<br>criteria. |
| Test                                    | Test Method Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Conclusions                                                            |
|                                         | Method: Entire shaped section is deployed into<br>the vessel at least 2 times and inspected for<br>functional impact and integrity of the tip.<br>Integrity is defined as no missing material or tears<br>at the tip. Functional impact is defined as inability<br>to resheath the shaped section. A deployment is<br>defined as the entire shaped section of the<br>retriever is deployed into the vessel.                                                                                                                                                                               |                                                                        |
| Product Integrity Post<br>Removal       | Purpose: To determine damage to product<br>following removal of the catheter and its<br>accessories from its packaging.<br><br>Method: Prepare product and remove from<br>packaging per DFU, followed by<br><br>-inspecting for kinks, stretches or crushing of the<br>catheter shaft.<br><br>Self-evident inspection:<br>-No damage on RHV.                                                                                                                                                                                                                                              | Product integrity<br>post removal met<br>acceptance criteria.          |

## Table 2. Performance Data - Bench Testing

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

## Performance Data - Animal Study, Clinical Study

No animal study or clinical study was conducted as bench testing was determined sufficient for verification and validation purposes.

## Shelf Life Testing

The labeled shelf life for the Trevo Trak 21 Microcatheter is two years. Shelf life testing (product and packaging) and Distribution Shipping Challenge Conditioning and testing were performed on the Subject device, and the results met established criteria.

## Sterilization

The Trevo Trak 21 Microcatheter and accessory (RHV) are sterilized with 100% Ethylene Oxide. The Trevo Trak 21 Microcatheter and accessory are provided sterility assurance level (SAL) of 10 6 has been demonstrated. The Trevo Trak 21 Microcatheter and accessory meet EO residuals per EN ISO 10993-7 for a limited contact delivery system - externally communicating. The Trevo Trak 21 Microcatheter and accessory are for single use only.

{11}------------------------------------------------

## Biocompatibility

The Trevo Trak 21 Microcatheter was assessed for biocompatibility in accordance with EN ISO 10993-1, "Biological evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process". The Subject device is considered an externally communicating medical device with circulating blood contact for less than 24 hours. The Trevo Trak 21 Microcatheter passed all recommended biocompatibility testing. An overview of the biocompatibility testing is summarized below.

| Test            |                                | Acceptance Criteria | Conclusion                                                                                                                         |                                                                                                                                       |
|-----------------|--------------------------------|---------------------|------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| ISO<br>10993-4  | Hemolysis                      | Direct Contact      | Hemolysis is $\le$ 5%.                                                                                                             | Pass                                                                                                                                  |
|                 |                                | Extraction          | Non-hemolytic<br>(0.0%)                                                                                                            | Pass                                                                                                                                  |
|                 | Thrombosis                     | Thromboresistance   | Thrombosis is acceptable for clinical<br>application.                                                                              | Pass                                                                                                                                  |
|                 |                                |                     | Complement<br>Activation                                                                                                           | Potential activator<br>( $p$ < 0.05 versus negative and activated<br>controls, ~4% of positive control,<br>~173% of negative control) |
|                 |                                |                     | Coagulation                                                                                                                        | Slight potential enhancement of<br>coagulation system<br>(17% faster clotting versus negative<br>control)                             |
| ISO<br>10993-5  | Cytotoxicity                   |                     | No less than or equal to 50% cell<br>viability.                                                                                    | Pass                                                                                                                                  |
| ISO<br>10993-10 | Sensitization                  |                     | Not a sensitizer.                                                                                                                  | Pass                                                                                                                                  |
|                 | Irritation                     |                     | No significant irritation.                                                                                                         | Pass                                                                                                                                  |
| ISO<br>10993-11 | Material Mediated Pyrogenicity |                     | No febrile reaction greater than 0.5°.                                                                                             | Pass                                                                                                                                  |
|                 | Acute Systemic Toxicity        |                     | No mortality or evidence of systemic<br>toxicity.                                                                                  | Pass                                                                                                                                  |
| ISO<br>10993-18 | Physicochemical<br>(USP <661>) |                     | Non-volatile Residue $\le$ 15 mg<br>Residue on Ignition $\le$ 5 mg<br>Heavy Metals $\le$ 1 ppm<br>Buffering Capacity $\le$ 10.0 mL | Pass                                                                                                                                  |
|                 | FTIR                           |                     | No unexpected readings.                                                                                                            | Pass                                                                                                                                  |
|                 | Latex<br>(LEAP - ASTM D6499)   |                     | No detectable latex.                                                                                                               | Pass                                                                                                                                  |

| Table 3. Overview of Biocompatibility Studies Performed on the Subject Device |
|-------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------|

# Conclusion

Stryker Neurovascular has demonstrated the Trevo Trak 21 Microcatheter is substantially equivalent to the Predicate device (K113260) based on same intended use / indications for use, same or similar materials, same fundamental design, and the same operating principles. The conclusions drawn from risk assessment and the bench testing results summarized above demonstrate that the benefits of the device outweigh any residual risks when used in accordance with device Directions for Use. Stryker Neurovascular has demonstrated that the Trevo Trak 21 Microcatheter is substantially equivalent to the legally marketed Predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K192122](https://fda-staging.innolitics.com/device/K192122)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).