# AcQRef Introducer Sheath (K192106)

_Acutus Medical, Inc. · DYB · Sep 3, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K192106

## Device Facts

- **Applicant:** Acutus Medical, Inc.
- **Product Code:** [DYB](/productcode/DYB.md)
- **Decision Date:** Sep 3, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Acutus Medical AcQRef Introducer Sheath is indicated for use in percutaneous procedures to facilitate venous access from the lower extremities for introduction of catheters and other devices, and may be used to sense intravenous signals.

## Device Story

AcQRef Introducer Sheath is a sterile, single-use venous access device; consists of straight shaft with lumen, hemostasis valve, flush port, single electrode, and extension cable. Used in clinical settings by physicians to facilitate percutaneous access from lower extremities; allows introduction of catheters/devices. Electrode senses intravenous signals; transmits data to compatible AcQMap High Resolution Imaging and Mapping System. Simplification from predicate (1 electrode vs 4; 1 cable vs 2) maintains fundamental technology. Provides venous access and signal sensing capabilities to assist in electrophysiology procedures.

## Clinical Evidence

Bench testing only. Verification and validation studies performed to confirm device performance against pre-determined acceptance criteria. No clinical data presented.

## Technological Characteristics

7Fr introducer sheath, 30 cm length. Materials: Pebax 72D SA01 (sheath), HDPE (dilator), Platinum/Iridium (electrode). Features: hemostasis valve, flush port, locking dilator. Connectivity: wired electrode connection to external mapping system. Sterilization: EtO. Shelf life: 24 months.

## Regulatory Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

## Predicate Devices

- AcQRef Introducer Sheath ([K171557](/device/K171557.md))

## Reference Devices

- AcQMap High Resolution Imaging and Mapping System ([K181577](/device/K181577.md))
- AcQMap High Resolution Imaging and Mapping System ([K190131](/device/K190131.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 3, 2019

Acutus Medical, Inc. Serena Sanginthirath Senior Regulatory Affairs Specialist 2210 Faraday Ave., Suite 100 Carlsbad, California 92008

Re: K192106

Trade/Device Name: AcQRef Introducer Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: August 1, 2019 Received: August 5, 2019

Dear Serena Sanginthirath:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark Fellman Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192106

Device Name AcQRef® Introducer Sheath

Indications for Use (Describe)

The Acutus Medical AcQRef Introducer Sheath is indicated for use in percutaneous procedures to facilitate venous access from the lower extremities for introduction of catheters and other devices, and may be used to sense intravenous signals.

Type of Use (Select one or both, as applicable)

| <div style="display:flex; align-items:center;"> <div style="margin-right:5px;">☑</div> <div>Residential Use (Part 21 CFR 601 Subpart D)</div> </div>      |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <div style="margin-right:5px;">☐</div> <div>Study That Supports Use of a Combination Product</div> </div> |

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) Notification K192106

#### General Information [807.92(a)(1)]

Date Prepared: August 2, 2019

Applicant:

Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008 USA Phone: 1-442-232-6080 Fax: 1-442-232-6081

#### Contact Person:

Serena Sanginthirath Senior Regulatory Affairs Specialist Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008 USA Phone: 1-442-232-6178 FAX: 1-442-232-6081 Email: Serena.Sanginthirath@acutus.com

## DEVICE INFORMATION [807.92(a)(2)]

#### Trade Name:

AcQRef® Introducer Sheath

#### Generic/Common Name:

Catheter Introducer

#### Classification:

Class II, 21 CFR §870.1340

#### Product Code:

DYB

{4}------------------------------------------------

### PREDICATE DEVICE(S) [807.92(a)(3)]

AcQRef Introducer Sheath (K171557), manufactured by Acutus Medical, Inc.

### DEVICE DESCRIPTION [807.92(a)(4)]

The Acutus Medical AcQRef Introducer Sheath is a sterile, single use, venous access device that consists of a straight shaft with lumen, hemostasis valve, flush port, electrode, and attached extension cable for electrode connectivity

The AcQRef Introducer Sheath consists of the following components:

- One (1) 7Fr Introducer Sheath with electrode and connection cable ●
- One (1) 7Fr Vessel Dilator .
- One (1) Guidewire .038 in, J-tip .

The electrode is connected to the yellow cable and connector. The AcQRef Introducer Sheath is compatible for use with the AcQMap® High Resolution Imaging and Mapping System Models 900000, cleared under 510(k) K181577 and 900100, cleared under 510(k) K190131.

#### INDICATIONS FOR USE [807.92(a)(5)]

The Acutus Medical AcQRef Introducer Sheath is indicated for use in percutaneous procedures to facilitate venous access from the lower extremities for introduction of catheters and other devices, and may be used to sense intravenous signals.

{5}------------------------------------------------

| Substantial Equivalence Table                                     |                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                    |
|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature                                                           | Subject Device<br>AcQRef Introducer<br>Sheath<br>TBD                                                                                                                                                                                                                          | Predicate Device<br>AcQRef Introducer<br>Sheath<br>(K171557)                                                                                                                                                                                                                  | Analysis of<br>Differences                                                                                                                                                                                         |
| Classification                                                    | 21 CFR § 870.1340                                                                                                                                                                                                                                                             | 21 CFR § 870.1340                                                                                                                                                                                                                                                             | Identical                                                                                                                                                                                                          |
| Product Code                                                      | DYB                                                                                                                                                                                                                                                                           | DYB                                                                                                                                                                                                                                                                           | Identical                                                                                                                                                                                                          |
| Indications for<br>Use                                            | The Acutus Medical<br>AcQRef Introducer Sheath<br>is indicated for use in<br>percutaneous procedures<br>to facilitate venous access<br>from the lower extremities,<br>for introduction of catheters<br>and other devices, and<br>may be used to sense<br>intravenous signals. | The Acutus Medical<br>AcQRef Introducer Sheath<br>is indicated for use in<br>percutaneous procedures<br>to facilitate venous access<br>from the lower extremities,<br>for introduction of catheters<br>and other devices, and<br>may be used to sense<br>intravenous signals. | Identical                                                                                                                                                                                                          |
| Intended Use                                                      | Catheter delivery system<br>including electrodes for<br>intravenous signal sensing<br>when used with compatible<br>mapping system.                                                                                                                                            | Catheter delivery system<br>including electrodes for<br>intravenous signal sensing<br>when used with compatible<br>mapping system.                                                                                                                                            | Identical                                                                                                                                                                                                          |
| Device Inner<br>Diameter (ID)<br>(French)                         | 7F                                                                                                                                                                                                                                                                            | 7F                                                                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                          |
| Device Length                                                     | 30 cm                                                                                                                                                                                                                                                                         | 30 cm                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                          |
| Shaft Material                                                    | Pebax 72D SA01, BaSO4<br>Loaded                                                                                                                                                                                                                                               | Pebax 72D SA01, BaSO4<br>Loaded                                                                                                                                                                                                                                               | Identical                                                                                                                                                                                                          |
| Deflection (Y/N)                                                  | No                                                                                                                                                                                                                                                                            | No                                                                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                          |
| Side Port for<br>Flush                                            | Yes                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                           | Identical                                                                                                                                                                                                          |
| Guide wire<br>Compatibility                                       | 0.038 in.                                                                                                                                                                                                                                                                     | 0.038 in.                                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                                          |
| Substantial Equivalence Table (continued)                         |                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                    |
| Feature                                                           | Subject Device<br>AcQRef Introducer<br>Sheath<br>TBD                                                                                                                                                                                                                          | Predicate Device<br>AcQRef Introducer<br>Sheath<br>(K171557)                                                                                                                                                                                                                  | Analysis of<br>Differences                                                                                                                                                                                         |
| Dilator, Locking<br>Feature                                       | Yes<br>Locking feature                                                                                                                                                                                                                                                        | No<br>Locking feature                                                                                                                                                                                                                                                         | The subject device has<br>a lock-in feature. This<br>change does not raise<br>new questions of safety<br>and efficacy.<br>The Dilator functions the<br>same and has the same<br>purpose as the cleared<br>dilator. |
| Dilator Material                                                  | HDPE                                                                                                                                                                                                                                                                          | Polypropylene                                                                                                                                                                                                                                                                 | All devices use<br>biocompatible polymer<br>material.                                                                                                                                                              |
| Electrodes                                                        | Yes (1x 90/10 Pt-Ir)<br>Sensing Only                                                                                                                                                                                                                                          | Yes (4x 90/10 Pt-Ir)<br>Sensing Only                                                                                                                                                                                                                                          | The subject device is<br>simplified It has only 1<br>of the original 4<br>electrodes. This<br>change does not raise<br>any new questions of<br>safety or efficacy.                                                 |
| Electrode<br>Material                                             | Platinum / iridium                                                                                                                                                                                                                                                            | Platinum / iridium                                                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                          |
| Connection to<br>Recording<br>System or<br>Mapping System         | Yes                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                           | Identical                                                                                                                                                                                                          |
| Radiopaque<br>Elements                                            | Yes                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                           | Identical                                                                                                                                                                                                          |
| Anatomical<br>location                                            | Peripheral (Venous)                                                                                                                                                                                                                                                           | Peripheral (Vascular)                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                          |
| Package<br>Contents                                               | Introducer sheath,<br>Guidewire, Dilator                                                                                                                                                                                                                                      | Introducer sheath,<br>Guidewire, Dilator                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                          |
| Sterilization<br>Method                                           | EtO                                                                                                                                                                                                                                                                           | EtO                                                                                                                                                                                                                                                                           | Identical                                                                                                                                                                                                          |
| Sheath<br>Hemostasis<br>Control                                   | Yes                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                           | Identical                                                                                                                                                                                                          |
| Biocompatible<br>Blood, Body and<br>Fluid Contacting<br>Materials | Yes                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                           | Identical                                                                                                                                                                                                          |
| Shelf Life                                                        | 24 months                                                                                                                                                                                                                                                                     | 24 months                                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                                          |
| Substantial Equivalence Table (continued)                         |                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                    |
| Feature                                                           | Subject Device<br>AcQRef Introducer<br>Sheath<br>TBD                                                                                                                                                                                                                          | Predicate Device<br>AcQRef Introducer<br>Sheath<br>(K171557)                                                                                                                                                                                                                  | Analysis of<br>Differences                                                                                                                                                                                         |
| Connector Cable                                                   | 1 (yellow)                                                                                                                                                                                                                                                                    | 2 (1 yellow and 1 blue)                                                                                                                                                                                                                                                       | The subject device has<br>been simplified. This<br>change does not raise<br>new questions of safety<br>and efficacy.                                                                                               |

## Comparison of Technological Characteristics with the Predicate Devices [807.92(A)(6)]

{6}------------------------------------------------

# 5 I 0(K) Summary

{7}------------------------------------------------

## 510(K) Summary

#### SUBSTANTIAL EQUIVALENCE

The indication for use of the subject device is substantially equivalent to the predicate. Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. The assessment of non-clinical performance data demonstrates that the modified AcQRef® Introducer Sheath does not change the fundamental scientific technology of the device.

### PERFORMANCE DATA [807.92(B)]

All necessary bench testing was conducted on the modified AcQRef Introducer Sheath to support a determination of substantial equivalence to the predicate device. Verification and validation studies demonstrated that the modified device met the pre-determined acceptance criteria. The results passed to support a determination of substantial equivalence.

## CONCLUSION [807.92(B)(3)]

The AcQRef® Introducer Sheath is substantially equivalent in its intended use, technological characteristics and performance to the previously cleared AcQReet® Introducer Sheath (K171557). The subject device has the same fundamental technological characteristics, principles of operation, specifications and is biocompatible to perform per its intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K192106](https://fda-staging.innolitics.com/device/K192106)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).