K192102 · Aktina Medical Corporation · IYE · Sep 25, 2019 · Radiology
Device Facts
Record ID
K192102
Device Name
Thermoplastic Masks
Applicant
Aktina Medical Corporation
Product Code
IYE · Radiology
Decision Date
Sep 25, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
Aktina Thermoplastic Masks are medical devices used for patient positioning and immobilization during external beam radiation therapy procedures. Thermoplastic masks are intended for single patient use for the duration of the treatment.
Device Story
Device consists of low-temperature polycaprolactone thermoplastic sheets mounted on polycarbonate frames; used for patient immobilization during external beam radiation therapy. Material softens when immersed in 70-75°C water; user molds sheet over patient anatomy; material hardens at room temperature to maintain anatomical shape. Mask provides consistent immobilization across multiple treatment sessions. Used in clinical radiation therapy settings; operated by radiation therapy staff. Output is a custom-molded physical immobilization device; aids in precise patient positioning to ensure accurate radiation delivery; benefits patient by minimizing movement during treatment.
Clinical Evidence
No clinical testing performed. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1, -5, -10), rigidity, shrinkage, radiation attenuation, and risk management (ISO 14971) and usability engineering (IEC 62366-1) assessments.
Technological Characteristics
Materials: Polycaprolactone sheet, polycarbonate frame. Principle: Thermoplastic molding via heat (70-75°C). Form factor: Perforated sheets mounted to frames. Connectivity: None. Sterilization: Non-sterile. Software: None.
Indications for Use
Indicated for patient positioning and immobilization during external beam radiation therapy procedures.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
September 25, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Aktina Medical Corporation % Mr. Tony Spaccarotella Director, Quality Assurance / Regulatory Affairs 360 North Route 9W CONGERS NY 10920
Re: K192102
Trade/Device Name: Thermoplastic Mask Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE Dated: July 30, 2019 Received: August 5, 2019
Dear Mr. Spaccarotella:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Section 4
## Indications for Use
ng
No. of Pages Including Cover: 2
{3}------------------------------------------------
## Expiration Date: 06/30/2020 See PRA Statement below.
Form Approved: 0MB No. 0910-0120
510(k) Number (if known)
K192102
Device Name Thermoplastic Mask
## Indications for Use (Describe)
Aktina Thermoplastic Masks are medical devices used for patient positioning and immobilization during external beam radiation therapy procedures. Thermoplastic masks are intended for the duration of the treatment.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
D Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.●
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of infonnation unless it displays a currently valid OMB number."
{4}------------------------------------------------
Section 5
510(k) Summary
360 North Route 9W Congers, NY 10920 USA www.aktina.com 888.433.3380 Tel: 845.268.0101 Fax: 845.268.1700
ﻟﺘﻘﺪﻳ
:
{5}------------------------------------------------
### 510(k) Summary
K192102
### In Compliance with 21 CFR Section 807.92(c)
#### 1. General Provisions
| Device Trade Name: | Thermoplastic Mask |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Moldable Thermoplastic |
| Owner Name<br>and Address: | Aktina Medical Corporation<br>360 North Route 9 W<br>Congers, New York, 10920<br>Phone: 845-268-0101<br>Fax: 845-268-1700<br>Registration Number: 2436865 |
| Contact Person: | Tony Spaccarotella, Director, QA/RA |
#### 3. Date Prepared: July 1, 2019
#### 4. Classification
2.
This device is classified as a class II device according to 21 CFR 892.5050, "Medical chargedparticle radiation therapy system." The product code is IYE.
#### 5. Predicate Device
Embrace Thermoplastic, 510(k) No. K120335, manufactured by Bionix Development Corporation, 5154 Enterprise Blvd, Toledo, OH 43612 USA.
#### 6. Description
This is a Traditional 510(k) for the Aktina Medical Thermoplastic Mask. The low-temperature thermoplastic mask is made of Aquaplast polycaprolactone sheets. When immersed in water heated to 70-75°C [160 to 170°F], the material softens, allowing the user to shape or stretch the mask over the contours of the patient's anatomy while cooling. When the material reaches room temperature, the mask hardens into the anatomical shape. The mask can then be placed over the patient treatment area to consistently immobilize the area from treatment to treatment. Predicate device manufacturers have used polycaprolactone materials for this intended use for over 25 years. The mask also has a polycarbonate frame with holes for mounting to the user's existing patient support hardware.
360 North Route 9W Congers, NY 10920 USA www.aktina.com 888.433.3380 Tel: 845.268.0101 Fax: 845.268.1700
{6}------------------------------------------------
#### 7. Intended Use
Aktina Thermoplastic Masks are medical devices used for patient positioning and immobilization during external beam radiation therapy procedures. Thermoplastic masks are intended for single patient use for the duration of the treatment.
#### Technological Characteristics 8.
The Table below compares the technological characteristics of the Aktina Thermoplastic Mask to the Predicate Device:
| Item | Predicate Device, K120035<br>Embrace Thermoplastic<br>Bionix Development Corporation,<br>5154 Enterprise Blvd, Toledo, OH 43612<br>USA | This 510(k) Submission<br>Thermoplastic Mask<br>Aktina Medical Corp.,<br>360 N. Route 9W, Congers, NY 10920<br>USA | Equivalent or<br>Better for<br>Intended Use? |
|------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| 1. | Intended Use:<br>Immobilize patients receiving external beam<br>radiation therapy. | Intended Use:<br>Same as predicate. | Equivalent |
| 2. | Design:<br>Moldable plastic sheet used for patient<br>immobilization. Heat is applied to the sheet<br>to soften it and mold it to the shape of the<br>patient anatomy. The mask is used for<br>multiple treatments on a single patient. The<br>sheet is pre-mounted to a non-patient<br>contacting frame to interface with the user's<br>existing support hardware. | Design:<br>Same as predicate. | Equivalent |
| 3. | Components:<br>Low temperature thermoplastic sheet with<br>non-low temperature plastic mounting<br>frame. | Components:<br>Same as predicate. | Equivalent |
| 4. | Technology:<br>1. Perforated plastic sheets that soften for<br>molding when exposed to heat.<br><br>2. Fastened to patient support equipment. | Technology:<br>1. Same as predicate.<br><br>2. Same as predicate. | Equivalent |
360 North Route 9W Congers, NY 10920 USA www.aktina.com 888.433.3380 Tel: 845.268.0101 Fax: 845.268.1700
{7}------------------------------------------------
| 5. | Materials:<br>a. Plastic sheet:<br>Polycaprolactone | Materials:<br>a. Plastic sheet:<br>Same as predicate. | a. Equivalent |
|-----|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| | b. Frame:<br>"Non-low temperature thermoplastic" | b. Frame:<br>Polycarbonate | b. Equivalent |
| 6. | Biocompatibility:<br>Plastic sheet is biocompatible for skin<br>contact up to 24 hours. | Biocompatibility:<br>Same as predicate. | Equivalent |
| 7. | Sterility:<br>Non-sterile. | Sterility:<br>Same as predicate. | Equivalent |
| 8. | Melting Temperature:<br>160°F | Melting Temperature:<br>Same as predicate. | Equivalent |
| 9. | Rigidity:<br>Pressure (psi)<br>Deflection (inches)<br>@ 5.08 psi<br>0.011<br>@ 10.4 psi<br>0.052 | Rigidity:<br>Pressure (psi)<br>Deflection (inches)<br>@ 5.08 psi<br>0.011<br>@ 10.4 psi<br>0.019 | Equivalent<br>at 5.08 psi<br>Submitted device<br>has less deflection<br>at 10.4 psi |
| 10. | Shrinkage:<br>0.3% | Shrinkage:<br>0.3% | Equivalent |
| 11. | Sheet Thickness:<br>2.4mm | Sheet Thickness:<br>Same as predicate. | Equivalent |
| 12. | Sheet Type:<br>Perforated sheets. | Sheet Type:<br>Same as predicate. | Equivalent |
| 13. | Radiation Attenuation (X-ray):<br>Less than 2% | Radiation Attenuation (X-ray):<br>Same as predicate device. | Equivalent |
#### Performance Standards, Non-Clinical Testing, and Data 9.
The FDA under Section 514 of the Food, Drug and Cosmetic Act has not established performance standards for this product
Specification testing has been performed for this device to show that the verification, validation and safety requirements have been met regarding:
- All specified functional, performance, safety and labeling requirements, such as ● biocompatibility, melting temperature, shrinkage, rigidity, and radiation attenuation.
Congers, NY 10920 USA www.aktina.com 888.433.3380 Tel: 845.268.0101 Fax: 845.268.1700 360 North Route 9W
{8}------------------------------------------------
- o Assessment against ISO 14971 Risk Management requirements.
- Assessment against applicable sections of IEC Standard: 62366-1, Application of 0 usability engineering to medical devices.
This device does not contain software.
The testing has demonstrated substantial equivalence or better when compared to the predicate device.
#### 10. Clinical Testing
No clinical testing was performed in support of this pre-market submission.
#### 11. Biocompatibility
The patient contact component for this device is the low-temperature thermoplastic mask material, which contacts the skin surface for less than 24 hours. This material has been shown to be biocompatible for the intended use when tested against the requirements of ISO 10993-1, -5 and -10 for cytotoxicity, irritation, and sensitization.
#### 12. Conclusion Regarding of Substantial Equivalence
This device is similar in design, intended use, technological, physical and performance characteristics to the predicate device. No new issues of safety or effectiveness are introduced by using this device. Therefore, Aktina Medical Corp. believes that the Thermoplastic Mask is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.