MEDONE ULTRA Soft tissue programmable automatic biopsy system: biopsy system has been designed to be used for histological biopsy of soft tissues (such as breast, kidney, liver, prostate and various soft tissue masses). It is not intended for use in bone biopsy. MEDEXTRA lymph node biopsy system with rectangular shaped cannula and adjustable penetration depth: is intended for use in obtaining biopsies from lymph nodes and other superficial non-vascular lumps. It is not intended for use in bone, deep visceral structures and breast biopsy.
Device Story
Medax Soft Tissue Biopsy System consists of single-use, spring-loaded, guillotine-style biopsy devices (MEDONE ULTRA and MEDEXTRA). MEDONE ULTRA is a programmable automatic system; MEDEXTRA is a semi-automatic system with a rectangular cannula and adjustable penetration depth. Devices are used by clinicians in clinical settings to obtain soft tissue core samples for histological examination. Operation involves single-hand automatic activation to advance the needle/cannula into the target tissue. The mechanical action captures tissue within the sample notch. The output is a tissue core sample retrieved for pathology analysis, aiding in the diagnosis of soft tissue abnormalities. Benefits include minimally invasive tissue acquisition for diagnostic purposes.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical in vitro bench testing, including performance/functional testing (ISO 9626), biocompatibility (ISO 10993), bioburden/sterility testing (ISO 11737), and packaging validation (ISO 11607).
Technological Characteristics
Materials: AISI 304 stainless steel cannula and mandrel. Principle: Spring-loaded, guillotine-style mechanical biopsy. Sterilization: Ethylene Oxide (EO) per ISO 11135-1. Form factor: Disposable, single-use, hand-held. No software or electronic components.
Indications for Use
Indicated for individuals requiring histological biopsy of soft tissues, including breast, kidney, liver, prostate, spleen, lymph nodes, and superficial non-vascular lumps. Contraindicated for use in bone, deep visceral structures (for MEDEXTRA), and breast (for MEDEXTRA).
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Achieve® Programmable Automatic Biopsy Systems (K141552)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is in blue and consists of the letters "FDA" and the words "U.S. Food & Drug Administration".
April 2, 2020
MEDAX S.R.L. UNIPERSONALE Stefano Cavalieri Quality Assurance Manager Via R. Piva 1/A Poggio Rusco Mantova, Italy 46025
Re: K192101
Trade/Device Name: Medax Soft Tissue Biopsy System: MEDONE ULTRA and MEDEXTRA Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology-Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: February 21, 2020 Received: February 25, 2020
Dear Stefano Cavalieri:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192101
Device Name Medax Soft Tissue Biopsy System
#### Indications for Use (Describe)
MEDONE ULTRA Soft tissue programmable automatic biopsy system: biopsy system has been designed to be used for histological biopsy of soft tissues (such as breast, kidney, liver, prostate and various soft tissue masses). It is not intended for use in bone biopsy.
MEDEXTRA lymph node biopsy system with rectangular shaped cannula and adjustable penetration depth: is intended for use in obtaining biopsies from lymph nodes and other superficial non-vascular lumps. It is not intended for use in bone, deep visceral structures and breast biopsy.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MEDAX medical devices. The logo is in a red box with rounded corners. The word "MEDAX" is in large, white, sans-serif letters. Below the word "MEDAX" is the phrase "medical devices" in smaller, white, sans-serif letters.
## 510(K) SUMMARY, AS REQUIRED BY 21 CFR 807.92
| Submitter's Name | MEDAX S.R.L. UNIPERSONALE | | | | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| Address | Via R. Piva 1/A | | | | |
| | Poggio Rusco Mantova, ITALY 46025 | | | | |
| Establishment<br>Registration Number | 3007648417 | | | | |
| Summary Preparation<br>Date | March 30th 2020 | | | | |
| Contact Person | Stefano Cavalieri | | | | |
| | Quality Assurance Manager | | | | |
| Telephone Number | +39.0535.1813915 | | | | |
| Fax Number | +39.0535.1812744 | | | | |
| Medax Soft Tissue Biopsy System | | | | | |
| Name of the Device | Medax Soft Tissue Biopsy System | | | | |
| Common name of the<br>device | Medax Soft Tissue Biopsy System (MEDONE ULTRA, MEDEXTRA) | | | | |
| Classification Name<br>and class | Instrument, Biopsy | | | | |
| | Device Class: II | | | | |
| | Product Code: KNW | | | | |
| | Regulation Number 21 CFR 876.1075 | | | | |
| Performance Standard | - ISO 9626:2016 Stainless steel needle tubing for the manufacture of<br>medical devices -- Requirements and test methods<br>- ASTM F899-12b Standard Specification for Wrought Stainless<br>Steels for Surgical Instruments<br>- ISO 10993:2009 series and FDA Guidance on Use of International<br>Standard ISO 10993-1, "Biological evaluation of medical devices -<br>Part 1: Evaluation and testing within a risk management process ",<br>Date:06/16/16<br>- ISO 11607-1.2006 Packaging for terminally sterilized medical<br>devices -- Part 1: Requirements for materials, sterile barrier<br>systems and packaging systems<br>- ISO 11737:2006 Sterilization of medical devices --<br>Microbiological methods -- Part 1: Determination of a population<br>of microorganisms on products | | | | |
| DESCRIPTION of the<br>device: | Medax Soft Tissue Biopsy System portfolio is composed by single use<br>devices intended to obtain biopsy samples from soft tissue for histological | | | | |
| examinations. |
|------------------------------------------------------------------|
| MEDONE ULTRA Devices are available in different gauge dimensions |
| (identified by different colors) and needle length. |
Indications for Use MEDONE ULTRA Soft tissue programmable automatic biopsy system: biopsy system has been designed to be used for histological biopsy of soft tissues (such as breast, kidney, liver, prostate and various soft tissue masses). It is not intended for use in bone biopsy.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for MEDAX medical devices. The logo is in a red box with white text. The word "MEDAX" is in large, bold letters on the top line. Below that, the words "medical devices" are in smaller letters.
| MEDEXTRA lymph node biopsy system with rectangular shaped cannula<br>and adjustable penetration depth: is intended for use in obtaining biopsies<br>from lymph nodes and other superficial non-vascular lumps. It is not<br>intended for use in bone or deep visceral structures. | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|----------------------------------|---------------------------------------------------------|
| Comparison of<br>Technological<br>Characteristics | In vitro bench testing was performed to support a determination of<br>substantial equivalence (refer to performance testing below) between<br>Medax Soft Tissue Biopsy System portfolio and predicate devices.<br>The results of these tests provide reasonable assurance that proposed<br>devices have been designed and tested to assure conformance to the<br>requirements for its intended use and perform comparably to the existing<br>predicate devices. | | | |
| Performance Testing<br>(non-clinical) | In vitro bench tests were carried out, according to the requirements of<br>FDAs document Guidance for Industry and FDA Staff: Format for<br>Traditional and Abbreviated 510(k)s and applicable standards.<br>The following areas have been tested and/or evaluated:<br>- Performance and functional tests according to ISO 9626;<br>- Biocompatibility tests according to ISO 10993 series and FDA<br>- Guidance on Use of International Standard ISO 10993-1, "Biological<br>evaluation of medical devices - Part 1: Evaluation and testing within<br>a risk management process "<br>- Bioburden and Sterility tests;<br>- Validation of the EtO Sterilization process,<br>- Packaging validation,<br>- Labelling evaluation,<br>- EtO Residual, Ethylene Chlorohydrin and Ethylene Glycol according<br>to EN ISO 10993-7.<br>Results from these performances evaluation demonstrated that the Medax<br>Soft Tissue Biopsy System devices met the acceptance criteria defined in<br>the product specification and performed comparably to the predicate<br>device. | | | |
| SUBSTANTIAL<br>EQUIVALENCE: | Medax Soft Tissue Biopsy System devices are identical to the predicate<br>device in terms of intended use, indications for use and medical technique.<br>Equivalence has been identified as follows: | | | |
| | Medax Soft<br>Tissue Biopsy<br>System Device | Name | Predicate Device<br>Manufacturer | 510(k) ID |
| | MEDONE<br>ULTRA Soft<br>tissue<br>programmable<br>automatic<br>disposable<br>biopsy system | Achieve®<br>Programmable<br>Automatic<br>Biopsy<br>Systems | MERIT<br>MEDICAL | K141552,<br>cleared by<br>FDA on June<br>10, 2014 |
| | MEDEXTRA<br>lymph node<br>biopsy system<br>with rectangular<br>shaped cannula | Velox 2 | Medax s.r.l. | K181803,<br>cleared by<br>FDA on<br>October 18,<br>2018 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for "MEDAX medical devices". The word "MEDAX" is in large, bold, white letters on a red background. Below "MEDAX" are the words "medical devices" in smaller, white letters. The logo is simple and clean, and the red and white color scheme is eye-catching.
Rev.04 – March 30th 2020
### Table 1 - Comparison of the Medax Soft tissue Biopsy System to the predicate Soft tissue Biopsy System devices.
K192101
| | Subject Device | Predicate device |
|-----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Medax MEDONE ULTRA | Achieve® Programmable Automatic |
| | Programmable Automatic disposable | Biopsy Systems |
| | Biopsy System | (K141552) |
| Regulation Number | 21 CFR §876.1075 | Same as current device |
| Device Description | Disposable programmable automatic<br>spring-loaded guillotine style biopsy<br>system for histological biopsy on soft<br>tissue. | Disposable programmable automatic<br>spring-loaded guillotine style biopsy<br>system for histological biopsy on soft<br>tissue. |
| Indication for Use | MEDONE ULTRA intended for use<br>in obtaining core biopsy samples<br>from soft tissue such as kidney, liver,<br>prostrate, spleen, lymph nodes, and<br>various soft tissue masses. Not<br>intended for use in bone | intended for use in obtaining core<br>biopsy samples from soft tissue such<br>as kidney, liver, prostrate, spleen,<br>lymph nodes, and various soft tissue<br>masses. Not intended for use in bone |
| Target Population | Individuals requiring biopsy for<br>sampling of soft tissue abnormalities | Same as current device |
| Mechanics of<br>Operation | Single hand automatic activation | Same as current device |
| Model Available<br>(Needle, Cannula(S),<br>And Stylet Size:<br>Diameter, Gauge and<br>Length) | Needle cannula from 14G to 20G<br>Length from 60 mm to 200 mm | Needle cannula from 12G to 20G<br>Length from 60 mm to 200 mm |
| Patient/Tissue<br>Contact Materials | Cannula and mandrel are made out of<br>AISI 304 stainless steel.<br>Only Stainless steel is in direct<br>surgical contact with all soft tissues<br>of the patient. | Cannula and mandrel are made out of<br>stainless steel.<br>Only Stainless steel is in direct<br>surgical contact with all soft tissues of<br>the patient. |
| Biocompatibility<br>Requirements | Biocompatible according to ISO<br>10993 applicable parts | Biocompatible according to ISO 10993<br>applicable parts |
| Description of The<br>Sterilization Method | Sterilized by Ethylene Oxide (EO | Sterilized by Ethylene Oxide (EO) |
The lengths of the predicate are exactly equivalent to the lengths range of the proposed device. The predicate device has a larger number of versions in term of diameters of the proposed devices fall within the range of the predicate devices and therefore the substantially equivalence is demonstrated.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a logo for MEDAX medical devices. The logo is red with white text. The word "MEDAX" is in large, bold letters on the top line. Below that, the words "medical devices" are in smaller letters.
| | Subject Device<br>MEDEXTRA | Predicate device<br>VELOX 2 (K181803) |
|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number | 21 CFR §876.1075 | Same |
| Device Description | Disposable semi-automatic spring-<br>loaded guillotine style biopsy system<br>with adjustable penetration depth and<br>rectangular shaped cannula for<br>histological biopsy on lymph nodes<br>and other superficial non-vascular<br>lumps. | Disposable semi-automatic spring-<br>loaded guillotine style biopsy system<br>with adjustable penetration depth for<br>histological biopsy on soft tissue. |
| Indication for Use | MEDEXTRA is intended for use in<br>obtaining biopsies from soft tissues<br>such as lymph nodes and other<br>superficial non-vascular lumps.<br>It is not intended for use in bone,<br>deep visceral structures and breast<br>biopsy. | The device is intended for use in<br>obtaining biopsies from soft tissues<br>such as liver, kidney, prostate, breast,<br>spleen, lymph nodes and various soft<br>tissue masses. It is not intended for use<br>in bone biopsy. |
| Target Population | Individuals requiring biopsy for<br>sampling of soft tissue abnormalities | Individuals requiring biopsy for<br>sampling of soft tissue abnormalities |
| Needle Advancement/<br>Penetration Depth | From 12 or 22 mm | From 15 mm or 22 mm |
| Sample Notch Size | Max 20 mm | Max 20 mm |
| Mechanics of<br>Operation | Single hand automatic activation | Single hand automatic activation |
| Model Available<br>(Needle, Cannula(S),<br>And Stylet Size:<br>Diameter, Gauge and<br>Length) | Needle cannula 4,6mm x 2.10mm<br>Single length 100mm | Needle cannula from 14G (2.1mm) to<br>20G (0.8mm)<br>Length from 80 mm to 300 mm |
| Patient/Tissue<br>Contact Materials | Cannula and mandrel are made of<br>stainless steel AISI 304. Only<br>Stainless steel AISI 304 is in direct<br>surgical contact with all soft tissues<br>of the patient. | Cannula and mandrel are made of<br>stainless steel AISI 304. Only Stainless<br>steel AISI 304 is in direct surgical<br>contact with all soft tissues of the<br>patient. |
| Biocompatibility<br>Requirements | Biocompatible according to ISO<br>10993 applicable parts | Biocompatible according to ISO 10993<br>applicable parts |
| Description of The<br>Sterilization Method | Sterilized by Ethylene Oxide (EO) In<br>accordance with ISO 11135-1 | Sterilized by Ethylene Oxide (EO) In<br>accordance with ISO 11135-1 |
The rectangular shape of the subject device Medextra cannula has equivalent functionality if compared with the rounded shape of the predicate cannula.
This difference does not affect the safety and effectiveness of the proposed device.
The predicate device has a larger number of versions in term of length of the proposed devices falls within the range of the predicate devices and therefore the substantially equivalent is demonstrated.
### CONCLUSION
Based on the safety and performance testing, technological characteristics and the indications for use, the devices proposed Medax Soft Tissue Biopsy System, have been demonstrated to be appropriate for their intended use and are considered substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.