K192079 · Eos Imaging · KPR · Nov 27, 2019 · Radiology
Device Facts
Record ID
K192079
Device Name
EOSedge
Applicant
Eos Imaging
Product Code
KPR · Radiology
Decision Date
Nov 27, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
EOSedge is intended for use in general radiographic exams and applications, excluding the evaluation of lung nodules and exams involving fluoroscopy, and mammography. EOSedge allows the radiographic acquisition of either one or two orthogonal X-ray images, for diagnostic purposes, in one single body or a reduced area of investigation of a patient, in the upright or seated position. The Micro Dose feature is indicated for assessing global skeletal deformities in follow-up pediatric exams.
Device Story
Digital radiography system; acquires one or two orthogonal X-ray images simultaneously via scanning patient over area of interest. System comprises acquisition workstation, gantry with electrical cabinet, and acquisition software. Two orthogonal acquisition chains (HV generators, X-ray tubes, collimators, solid-state photon counting detectors) mounted on C-shaped arms translating vertically. Used by trained medical personnel in clinical settings. Input: patient X-ray exposure; Output: diagnostic images stored locally, displayed on high-resolution monitor, transmitted via DICOM for archiving/printing. Micro Dose feature provides low-dose imaging for pediatric follow-ups. Healthcare providers use images for diagnostic assessment of skeletal structures. Benefits include simultaneous orthogonal imaging and reduced radiation dose options.
Clinical Evidence
Bench testing only. Performance and functional testing conducted per FDA's 'Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices' (2016) to verify specifications, dosing, and image quality. Software verification and validation testing performed. No clinical data presented.
Technological Characteristics
Digital radiography system; solid-state photon counting detectors (direct conversion); two orthogonal acquisition chains on C-shaped arms; HV generators; X-ray tubes; collimators. Dimensions: 2.58m x 2.58m x 2.706m. Weight: 2005 kg. Connectivity: DICOM compatible network. Standards: IEC 60601-1. Filtration: 1.7 mm Al at 75 kV plus 0.1 mm copper. Pixel depth: 17 bits; Pixel size: 100 μm; Resolution: 3.7 lp/mm; Dynamic range: >100 dB.
Indications for Use
Indicated for general radiographic exams (excluding lung nodules, fluoroscopy, angiography, mammography) in upright or seated patients. Micro Dose feature indicated for assessing global skeletal deformities in follow-up pediatric exams.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below.
November 27, 2019
EOS Imaging c/o Bernard Ismael Quality and Regulatory Affairs Director 10 rue Mercoeur Paris, 75011 FRANCE
Re: K192079
Trade/Device Name: EOSedge Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: KPR Dated: July 31, 2019 Received: August 2, 2019
Dear Bernard Ismael:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K192079
Device Name
EOSedge™
Indications for Use (Describe)
EOSedge is intended for use in general radiographic exams and applications, excluding the evaluation of lung nodules and exams involving fluoroscopy, and mammography. EOSedge allows the radiographic acquisition of either one or two orthogonal X-ray images, for diagnostic purposes, in one single body or a reduced area of investigation of a patient, in the upright or seated position.
The Micro Dose feature is indicated for assessing global skeletal deformities in follow-up pediatric exams.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> of Health Department and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
### EOS imaging's EOSedge System
EOS imaging 10 Rue Mercoeur F-75011, Paris FRANCE
Phone: Facsimile: + 33 1 55 25 60 61
Contact Person: Bernard ISMAEL, Quality and Requlatory Affairs Director
Date Prepared: July 31, 2019
| Name of Device: | EOSedge™ |
|-----------------------|---------------------------------------------|
| Common or Usual Name: | Digital Radiography System |
| Classification Name: | 21 C.F.R §892.1680; Stationary X-ray System |
| Regulatory Class: | Class II |
| Product Code: | KPR - System, X-ray, Stationary |
| Predicate Device: | EOS imaging's EOS System (K152788) |
### Device Description
The EOSedge system is a digital radiography system comprised of an acquisition workstation, a gantry including an electrical cabinet housing the system power and communication controls, and an acquisition software to obtain diagnostic images. Two sets of detectors and X-ray tubes are positioned orthogonally to generate frontal and lateral images simultaneously by scanning the patient over the area of interest. If desired, the Micro Dose feature enables image acquisition for assessing global skeletal deformities in follow-up exams. The diagnostic images are stored in a local database and are displayed on a highresolution medical-quality non-diagnostic image can be transmitted through a DICOM compatible digital network for printing and archiving.
#### Intended Use / Indications for Use
EOSedge is intended for use in general radiographic exams and applications, excluding the evaluation of lung nodules and exams involving fluoroscopy, angiography, and mammography. EOSedge allows the radiographic acquisition of either one or two orthogonal X-ray images, for diagnostic purposes, in one single scan, of the whole body or a reduced area of investigation of a patient, in the upright or seated position.
The Micro Dose feature is indicated for assessing global skeletal deformities in follow-up pediatric exams.
#### Performance Data
EOSedge is designed and has been certified to conform to IEC 60601-1 and collateral standards. Software verification and validation testing was also conducted. Additional performance and functional
{4}------------------------------------------------
testing have confirmed the equivalent performance of EOSedge compared to the cleared predicate EOS System. This included bench testing to confirm appropriate dosing and image quality. Bench performance testing were conducted based on FDA's Guidance for the Submission of 510/k)'s for Solid-State X-ray Imaging Devices (September 1, 2016), to verify that EOSedge performs according to specifications and is as safe and effective as the predicate device.
## Substantial Equivalence
EOSedge has the same intended use and similar indications, principles of operation, and technological characteristics as the cleared predicate device. EOSedge is an updated version of the cleared EOS System (K152788) with the main modification being a change in the type of detector used, namely, solid state photon counting detectors rather than gaseous detectors. Although the design of the equipment is being updated to accommodate the new detector, there are no major changes to the tube assembly and beam-limiting features. Both systems have two sets of detectors and X-ray tubes positioned orthogonally to generate frontal and lateral images simultaneously by scanning the patient over the area of interest. The two orthogonal acquisition chains consist of HV generators, X-ray tubes, collimators and detectors, positioned on C-shaped arms translating along a vertical axis. The software for the two systems are also fundamentally the same.
A substantial equivalence table summarizing the similarities and differences between EQSedge and its predicate device is provided below.
| | EOS imaging's EOSedge | EOS imaging's cleared EOS System<br>(K152788) |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>Indications for Use | General X-ray imaging system<br><br>EOSedge is intended for use in general<br>radiographic exams and applications,<br>excluding the evaluation of lung nodules and<br>exams involving fluoroscopy, angiography,<br>and mammography. EOSedge allows the<br>radiographic acquisition of either one or two<br>orthogonal X-ray images, for diagnostic<br>purposes, in one single scan, of the whole<br>body or a reduced area of investigation of a<br>patient, in the upright or seated position.<br><br>The Micro Dose feature is indicated for<br>assessing global skeletal deformities in<br>follow-up pediatric examinations. | General X-ray imaging system<br><br>EOS is intended for use in general<br>radiographic examinations and applications,<br>excluding the evaluation of lung nodules and<br>examinations involving fluoroscopy,<br>angiography, and mammography. EOS<br>allows the radiographic acquisition of either<br>one or two orthogonal X-ray images for<br>diagnostic purposes, in one single scan, of<br>the whole body or a reduced area of<br>investigation of a patient in the upright or<br>seated position.<br><br>The Micro Dose feature is indicated for<br>imaging with a patient entrance dose of 10<br>to 90 µGy for assessing global skeletal<br>deformities in follow-up pediatric<br>examinations. Micro Dose is not indicated<br>for focal skeletal abnormalities and/or other<br>pediatric abnormalities. Micro Dose is not<br>indicated for use in patients with a Body<br>Mass Index over 30. |
| Contraindications | EOSedge is not designed to perform<br>fluoroscopy, angiography or mammography<br>exams.<br><br>The Micro Dose feature is not designed for<br>morphological analysis of bone structures and<br>lesions.<br><br>The Micro Dose feature is not indicated:<br>• for analysis of spine static in the | This system was not designed to perform<br>mammograms, angiograms or fluoroscopy<br>examinations.<br><br>The Micro Dose functionality is not designed<br>for morphological analysis of bone<br>structures and lesions. It is not indicated for<br>analysis of spine static in the presence of<br>fusion material or for analysis of the bone- |
| | EOS imaging's EOSedge | EOS imaging's cleared EOS System<br>(K152788) |
| | presence of fusion material or for<br>analysis of the bone-implant interface<br>in the case of focal skeletal anomalies<br>•<br>and/or other pediatric abnormalities<br>•<br>for use in patients with a Body Mass<br>Index over 30 | implant interface. |
| User Population | Trained medical personnel | Trained medical personnel |
| Technological<br>Characteristics | Digital radiography system in which two sets<br>of detectors and X-ray tubes are positioned<br>orthogonally to generate frontal and lateral<br>images simultaneously by scanning the<br>patient over the area of interest. The 2<br>orthogonal acquisition chains consist of HV<br>generators, X-ray tubes, collimators and<br>detectors, positioned on C-shaped arms<br>translating along a vertical axis. | Digital radiography system in which two sets<br>of detectors and X-ray tubes are positioned<br>orthogonally to generate frontal and lateral<br>images simultaneously by scanning the<br>patient over the area of interest. The 2<br>orthogonal acquisition chains consist of HV<br>generators, X-ray tubes, collimators and<br>detectors, positioned on C-shaped arms<br>translating along a vertical axis. |
| Dimensions (I x w x<br>h) | 2.58 m x 2.58 m x 2.706 m<br>(8.5 ft x 8.5 ft x 8.9 ft) | 2 m x 2 m x 2.7 m<br>(7.9 ft x 7.9 ft x 10.44 ft) |
| Weight | 2 005 kg (4 420 lb) | 1 623 kg (3579 lb) |
| Accessories | Laser safety barriers<br>Motorized lifting platform<br>Gantry console<br>Laser positioning system<br>Access and stabilization bars<br>Bar code reader | Mechanical platform<br>Laser positioning system<br>Access and stabilization bars |
| Principles of<br>Operation | Tube preheating<br>Selection of the patient, the acquisition<br>planes, and the anatomical area, as well as<br>patient positioning<br>Selection of either automatic mode or manual<br>mode<br>Selection of patient morphotype<br>Image preview<br>Display of the exposure area, configuration of<br>the reference planes, and verification and<br>validation of the acquisition parameters<br>Image analysis and export of the exam | Tube preheating<br>Daily detector calibration<br>Selection of the patient, the acquisition<br>planes, and the anatomical area, as well as<br>patient positioning<br>Selection of patient morphotype<br>Image preview<br>Display of the exposure area, configuration<br>of the reference planes, and verification and<br>validation of the acquisition parameters<br>Display of unprocessed images and<br>reprocessing with various options<br>Image analysis and export of the exam<br>Printing |
| Permanent<br>minimum total | 1.7 mm Al at 75 kV | 1.5 mm Al at 75 kV |
| filtration<br>(Al equivalent) | Additional filtrations:<br>• 0.1 mm copper thickness (used for<br>images with or without Micro-Dose) | Additional filtrations:<br>• 0.1 mm Cu copper thickness (used for<br>images with or without Micro-Dose) |
| | EOS imaging's EOSedge | EOS imaging's cleared EOS System<br>(K152788) |
| | ◦ 0.5 mm copper thickness (used for the<br>Scout View) | ◦ 1 mm aluminum thickness (used for<br>large patient with LF and power > 28<br>kW) |
| Detectors | Direct conversion device - solid state<br>detector | Direct conversion device - gaseous detector |
| Pixel Depth | 17 bits (> 131 000 grey levels) | 16 bits (> 65 000 grey levels) |
| Pixel Size | 100 μm | 254 μm |
| Resolution | 3.7 lp/mm | 1.6 - 1.7 lp/mm |
| Typical Dynamic<br>Range | > 100 dB | > 90 dB |
| Software<br>Specifications | Patient information management<br>Image acquisition<br>Display images<br>Send exam to the PACS<br>Maintenance management<br>Access management to the system<br>System acquisition configuration. | Patient information management<br>Image acquisition<br>Display images<br>Send exam to the PACS<br>Maintenance management<br>Access management to the system<br>System acquisition configuration. |
| Focal Spot | Large focal spot:<br>0.6 x 1.3 at 120 kV - 100 mA | Large focal spot:<br>0.6 x 1.3 at 120 kV - 100 mA<br>Small focal spot:<br>0.4 x 0.7 at 120 kV - 50 mA<br>Use of the two focal spot sizes enabled |
| Linear Scanning<br>Speed (cm/s) | From 4.1 to 32.5 | From 3.8 to 30.5 |
| Average<br>Acquisition Time | 8 seconds for a spine and 15 seconds for an<br>entire body | 5 to 10 seconds for a spine and 20 seconds<br>for an entire body |
## Table 1: EOS IMAGING'S EOSEDGE - SUBSTANTIAL EQUIVALENCE CHART
{5}------------------------------------------------
{6}------------------------------------------------
# Conclusion
EOSedge has the same intended use and similar indications for use, technological characteristics, and principles of operation as its predicate device, the EOS System (K152788). The minor differences in indications for use and technology between EOSedge and its predicate device raise no new questions of safety or effectiveness. Performance data further demonstrates that EOSedge is as safe and effective as the cleared EOS System (K152788). Thus, EOSedge is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.