The Aerodentis System is indicated for use in movement and alignment of teeth during orthodontic treatment of malocclusion.
Device Story
Aerodentis System moves teeth via pulsatile mechanical force; intended for nighttime use (8-14 hours/day). System comprises custom-made, latex-free plastic dental mouthpiece and Personal Treatment Controller (PTC). PTC delivers compressed air to mouthpiece balloon via polyurethane tube; inflation provides pressure to malpositioned teeth. Used by patients at home under dental practitioner prescription. Practitioner uses software to develop patient-specific treatment plan and control delivery. Output is physical tooth movement; allows patient to remain appliance-free during daytime hours. Benefits include effective alignment with reduced daily wear time compared to constant-force aligners.
Clinical Evidence
Single-center controlled trial (n=43; 28 Aerodentis, 15 Invisalign). Primary endpoint: Little's Irregularity Index (LII) ≤ 1.5. Results: All subjects completing treatment achieved LII ≤ 1.5. No adverse events or clinically significant root resorption reported. Aerodentis demonstrated statistically significantly higher rates of tooth movement when normalized for actual hours of device use.
Technological Characteristics
System includes custom latex-free plastic mouthpiece and PTC. Operates via pneumatic inflation of an internal balloon to apply pulsatile mechanical force. Materials tested per ISO 10993-3, -5, and -10. Electrical safety per IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11. Software-controlled treatment delivery.
Indications for Use
Indicated for movement and alignment of permanent teeth in adults and pediatric patients with full permanent dentition. Suitable for Class 1 and 2 malocclusions, including crowding, proclination, retroclination, tipping, uprighting of anterior teeth, and rotation of incisors.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
May 6, 2020
Dror Orthodesign % Janice Hogan Partner Hogan Lovelle US LLP 1735 Market Street, Suite 2300 Philadelphia, Pennsylvania 19103
Re: K192069
Trade/Device Name: Aerodentis System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: April 10, 2020 Received: April 10, 2020
Dear Ms. Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K192069
Device Name
Aerodentis System
Indications for Use (Describe)
The Aerodentis System is indicated for use in movement of teeth during orthodontic treatment of malocclusion.
Type of Use (Select one or both, as applicable)区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Page 1 of 3
# K192069 510(k) SUMMARY Aerodentis System
#### Submitter
Dror Orthodesign, Inc. 7 Hartom St., Mount Hotzvim Jerusalem, Israel 9777507
Phone: +972 74 700 6700 Facsimile: +972 74 700 6767
Contact Person: Michael Nadav
Date Prepared: May 4, 2020
#### Name of Device:
Aerodentis System
#### Common or Usual Name:
Sequential Aligner
#### Classification Name:
872.5470 Orthodontic plastic bracket
#### Regulatory Class:
Class II
#### Product Code:
NXC
#### Primary Predicate:
K181739 Align Technology, Inc.'s Invisalign System
#### Reference Device:
K130643 Orthoaccel Technologies, Inc.'s Acceledent
#### Device Description:
The Aerodentis System is an orthodontic system intended to adjust the patient's teeth through movement from their initial position to the desired position as a result of pulsatile mechanical force. The device is designed to achieve orthodontic tooth movement over a period of approximately 10
{4}------------------------------------------------
hours per day (typically at night). The device is suitable for patients with full complement of permanent teeth in need of upper, lower or combined jaw treatments.
The system is composed of two primary components:
- . A custom-made, latex-free, plastic dental mouthpiece is inserted and worn by the patient according to a customized treatment prescription specified by the dental practitioner.
- . The Personal Treatment Controller (PTC), which provides compressed air to the mouthpiece via a polyurethane tube
# Indications for Use:
The Aerodentis System is indicated for use in movement and alignment of teeth during orthodontic treatment of malocclusion.
## Summary of Technological Characteristics:
| | Dror Orthodesign<br>Aerodentis System<br>(K192069) | Align Technology, Inc.<br>Invisalign System<br>(K181739) | Orthoaccel<br>Technologies, Inc.<br>Acceledent Aura<br>(K130643) |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Aerodentis System is<br>indicated for use in<br>movement and alignment of<br>permanent teeth during<br>orthodontic treatment of<br>malocclusion. | The Invisalign System is<br>indicated for the<br>orthodontic treatment of<br>malocclusion. | AcceleDent® Aura is<br>intended for use<br>during orthodontic<br>treatment. It is used<br>in conjunction with<br>orthodontic<br>appliances such as<br>braces and helps<br>facilitate minor<br>anterior tooth<br>movement. |
| User<br>Population | The Aerodentis System is<br>suitable for adults and<br>pediatric patients with full<br>complement of permanent<br>teeth. It is suitable for Class<br>1 and 2 malocclusions,<br>including crowding,<br>proclination and<br>retroclination. The following<br>movements are supported:<br>tipping and uprighting of<br>anterior teeth, and rotation of<br>incisors. | The Invisalign System is<br>suitable for adults and<br>pediatric patients with full<br>complement of permanent<br>teeth. It is suitable for Class 1<br>and 2 malocclusions,<br>including crowding,<br>proclination and<br>retroclination. The following<br>movements are supported:<br>tipping and uprighting of<br>anterior teeth and rotation of<br>incisors. | AcceleDent® Aura is an<br>orthodontic accessory<br>for the treatment of<br>tooth malocclusion. It is<br>used as an adjunctive<br>therapy for patients<br>with orthodontic<br>appliances such as<br>braces to help facilitate<br>tooth movement. |
| Components | Mouthpiece and<br>Personalized Treatment<br>Controller | Incremental sets of<br>aligners | Activator,<br>Mouthpiece and<br>Travel Case |
{5}------------------------------------------------
| Principles of<br>Operation | Inflation of the balloon within<br>the Mouthpiece provides<br>pressure to the malpositioned<br>teeth to move them into the<br>desired position over time. | Incremental set of aligners<br>provides pressure to the<br>malpositioned teeth to move<br>them into the desired position<br>over time. | Uses soft pulse<br>technology and cyclic<br>forces to accelerate the<br>movement of teeth |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Pressure<br>Application | Pulsatile | Constant | Pulsatile |
| Daily Treatment<br>Length | 8 - 14 hr/day | 24 hr/day | 20 minutes per day<br>during Orthodontic<br>treatment |
| Software | Yes; used to develop a<br>patient-specific plan and<br>control treatment delivery. | Yes; used to develop a<br>patient specific plan. | Not Applicable |
## Performance Data:
The following testing was performed to establish equivalence:
- . Balloon leak and fatigue testing
- Aerodentis System force analysis .
- . FEA to support use life
- . Biocompatibility testing per ISO 10993-3, -5 and -10
- . Software validation per FDA guidance
- IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11 testing
A single center, controlled, clinical trial was conducted comparing the investigational Aerodentis System to the Invisalign control. 28 subjects were enrolled in the investigational arm and 15 in the control. 7 subjects dropped from the study (5 from the investigational arm, and 2 from the control). Subjects were followed until the average score of the mandibular and maxillary teeth achieved a Little's Irregularity Index (LII) of 1.5 or less or until 15 months of treatment, whichever came first. Safety was assessed by recording adverse events and radiographic assessment of root resorption. Efficacy was assessed by achieving an average LII score of 1.5 or less.
There were no adverse events reported throughout the study for either treatment arm. In addition, no clinically significant root resorption was observed for either group. Thus, both devices have excellent safety profiles. All subjects that completed treatment achieved an average LII score of 1.5 or less. While the time to treatment until success was longer for the subject device, once accounting for actual hours of device use, the Aerodentis System provided statistically significantly higher rates of tooth movement. This allows the user to achieve tooth alignment while being free of all appliances during normal working hours.
# Conclusion:
The Aerodentis System is substantially equivalent to the Invisalign System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.