K192049 · Innovasis, Inc. · KWQ · Nov 8, 2019 · Orthopedic
Device Facts
Record ID
K192049
Device Name
Oryx Cervical Plate System
Applicant
Innovasis, Inc.
Product Code
KWQ · Orthopedic
Decision Date
Nov 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Innovasis Oryx Cervical Plate System is intended for use in Anterior Cervical Discectomy and Fusion (ACDF) procedures. Users of these products are limited to physicians trained in orthopedic surgery. Clinical locations include hospitals and surgery sites equipped to perform spinal surgery.
Device Story
Oryx Cervical Plate System is an anterior cervical fixation device consisting of titanium alloy plates and screws. Used by orthopedic surgeons in hospitals/surgical centers for spinal stabilization (C2-C7). Plates feature lordotic and transverse curvature for anatomical fit. Device provides mechanical support to the cervical spine to facilitate fusion in patients with degenerative, traumatic, or deformity-related conditions. Surgeon selects appropriate plate/screw sizes based on patient anatomy; implants are secured to the spine to provide rigid fixation. Benefits include stabilization of the cervical column, potentially reducing pain and restoring spinal alignment.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included static and dynamic compression, bending, and torsion per ASTM F1717, and axial pull-out, insertion/removal torque, and torsional yield strength per ASTM F543.
Technological Characteristics
Implant components: Ti6Al4V (ELI) titanium alloy (ASTM F136). Instruments/accessories: Surgical Stainless Steel (ASTM F899) and medical grade Silicone. Sterilization trays: Anodized 5052 Aluminum, Nylon, Stainless Steel, Polypropylene, and Radel (ASTM D6394 SP031). Design includes lordotic and transverse curvature plates. Mechanical fixation via screws.
Indications for Use
Indicated for anterior cervical fixation in patients requiring stabilization of the cervical spine from C2 to C7 for Degenerative Disc Disease (DDD), Spondylolisthesis, Trauma (fracture/dislocation), Spinal stenosis, Deformities (scoliosis, kyphosis, lordosis), Tumor, Pseudoarthrosis, and previous failed fusion.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Innovasis Opteryx Anterior Cervical Plate System (K061147)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 8, 2019
Innovasis, Inc. Marshall C. McCarty, RAC Director QA/RA 614 East 3900 South Salt Lake City, Utah 84107
Re: K192049
Trade/Device Name: Oryx™ Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: August 13, 2019 Received: August 14, 2019
Dear Mr. McCarty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192049
Device Name OryxTM Cervical Plate System
#### Indications for Use (Describe)
- The Innovasis Oryx Cervical Plate System is intended for use in anterior cervical fixation for the following indications: i. Degenerative Disc Disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc
- confirmed by history and radiographic studies),
- ii. Spondylolisthesis,
- iii. Trauma (i.e., fracture or dislocation)
- iv. Spinal stenosis.
- v. Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
- vi. Tumor,
- vii. Pseudoarthrosis: and
- viii. Previous failed fusion.
The Innovasis Oryx Cervical Plate System is indicated for stabilizing the cervical spine from C2 to C7.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Innovasis. The logo is black and features the word "INNOVASIS" in a bold, sans-serif font. Below the logo is the text "INNOVATE / INVOLVE / INVENT" in a smaller font. The logo is simple and modern, and the text below it suggests that the company is focused on innovation.
510(k)
#### 510(k) Summary Report: 5.0
# Oryx™ Cervical Plate System
| Company: | Innovasis, Inc. |
|----------|--------------------------|
| | 614 E. 3900 South |
| | Salt Lake City, UT 84107 |
Contact: Marshall McCarty Alternate: Michael Thomas Phone: (801) 261-2236 x8012 mmccarty@innovasis.com
(801) 261-2236 x8002 mthomas@innovasis.com
- Trade Name: Oryx™ Cervical Plate System
- Common Name: Anterior cervical fixation system
Regulation No.: 21CFR 888.3060 Spinal Intervertebral Body Fixation Orthosis Classification: Class 2 Product Code: KWQ Review Panel: Orthopedic Review Team: ISDT
- Primary Predicte: K030866 Synthes Anterior CSLP System
- Additional: K061147 Innovasis Opteryx Anterior Cervical Plate System
- Device Description: The Oryx Cervical Plate System consists of a number of plates and screws of varying lengths. The plates feature lordotic curvature and a transverse plane curvature for an anatomical fit. The set contains ancillary products and instruments. All implant components are machined from implant grade titanium alloy, Ti6Al4V (ELI) per ASTM F136.
- (Non-clinical)—Performance testing on plates per ASTM F1717 Performance Data: for Static and Dynamic Compression Bend and Static Torsion and ASTM F543 for bone screw Axial Pull-Out, Insertion and Removal Torque and Torsional Yield Strength indicates that the Oryx Cervical Plate System is substantially equivalent to the predicate device.
- Materials: The Oryx implants are machined from medical grade Ti6Al4V (ELI) per ASTM F136. The Orvx instruments/accessories are machined from Surgical Stainless Steel per ASTM F899 and some have medical grade Silicone handles.
{4}------------------------------------------------
| INNOVASIS<br>INNOVATE / INVOLVE / INVENT | Oryx™ Cervical Plate System | 510(k)<br>July 31, 2019<br>Page 11 of 107 |
|------------------------------------------|-----------------------------|-------------------------------------------|
|------------------------------------------|-----------------------------|-------------------------------------------|
The Oryx sterilization trays are comprised of Anodized 5052 Aluminum and have components made of Nylon. Stainless Steel. Polypropylene, and Radel per ASTM D6394 SP031.
The Oryx Cervical Plate System is intended for use in Anterior Intended Use: Cervical Discectomy and Fusion (ACDF) procedures. Users of these products are limited to physicians trained in orthopedic surgery. Clinical locations include hospitals and surgery sites equipped to perform spinal surgery.
Indications for Use: The Innovasis Oryx Cervical Plate System is intended for use in anterior cervical fixation for the following indications:
> i. Degenerative Disc Disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies).
ii. Spondvlolisthesis.
iii. Trauma (i.e., fracture or dislocation)
- iv. Spinal stenosis.
v. Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis).
vi. Tumor.
- vii. Pseudarthrosis; and
viii. Previous failed fusion.
The Innovasis Oryx Cervical Plate System is indicated for stabilizing the cervical spine from C2 to C7.
## Comparison of the Technological Characteristics with Predicate Device:
The Innovasis Oryx Cervical Plate System has been subjected to risk analysis, engineering analysis and testing to recognized standards and has been shown to be substantially equivalent to the Synthes Anterior CSLP System (K030866) predicate.
- . Design configurations are substantially equivalent.
- . Applied mechanical loads are substantially equivalent.
- . Materials used are substantially equivalent.
- Biocompatibility requirements have been demonstrated. ●
- Intended Use and Indications for Use are substantially equivalent.
- Conclusion: Performance data and rationales submitted herein demonstrate that the subject Oryx Cervical Plate System is substantially equivalent to the predicate device with regard to design, technological characteristics, materials, performance, intended use and indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.