STERRAD VELOCITY Biological Indicator/Process Challenge Device and Reader
K192025 · Advanced Sterilization Products (Asp) · FRC · Jan 23, 2020 · General Hospital
Device Facts
Record ID
K192025
Device Name
STERRAD VELOCITY Biological Indicator/Process Challenge Device and Reader
Applicant
Advanced Sterilization Products (Asp)
Product Code
FRC · General Hospital
Decision Date
Jan 23, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
STERRAD VELOCITY® Biological Indicator/Process Challenge Device, in conjunction with the STERRAD VELOCITY® Reader, is intended to be used as a standard method for frequent monitoring and periodic testing of the following STERRAD Sterilization Systems: - STERRAD® 100NX (STANDARD, FLEX, EXPRESS, and DUO Cycles) with and without ALLClear® Technology - STERRAD NX® (STANDARD and ADVANCED Cycles) with and without ALLClear® Technology - STERRAD® 100S
Device Story
The STERRAD VELOCITY BI/PCD is a self-contained biological indicator containing Geobacillus stearothermophilus spores and nutrient medium with an alpha-glucosidase enzyme system. Used with the STERRAD VELOCITY Reader, it monitors hydrogen peroxide gas plasma sterilization. During sterilization, hydrogen peroxide exposure reduces enzyme activity. The Reader incubates the BI at 57°C, uses UV light to excite the fluorescent byproduct (4-MU) of the enzymatic reaction, and detects fluorescence via photodiode. The device provides a final result in 15 minutes. A positive result indicates sterilization failure; a negative result indicates successful inactivation. The system includes a barcode scanner and network connectivity for record-keeping. It is used in clinical settings to ensure sterilization equipment performance, directly impacting patient safety by confirming the sterility of processed instruments.
Clinical Evidence
No clinical data was generated. Substantial equivalence is supported by non-clinical bench testing, including hydrogen peroxide dose-response verification, performance qualification for periodic testing, reduced incubation time verification, holding time verification, system-level error checks, and operational vibration testing. All tests met acceptance criteria.
Technological Characteristics
Self-contained biological indicator with Geobacillus stearothermophilus (ATCC 7953) spores on glass fiber carrier. Reader uses UV LED excitation, optical filters, and photodiode detection. 8-well incubation system maintained at 57 ± 2°C. Connectivity via barcode scanner and network. Power: 100-240V AC. Compliance: UL/IEC 61010-1, FCC Part 15 Subpart B. Software algorithm modified for 15-minute readout.
Indications for Use
Indicated for frequent monitoring and periodic testing of STERRAD 100NX, STERRAD NX, and STERRAD 100S sterilization systems. Used by healthcare personnel to verify sterilization cycle efficacy.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
January 23, 2020
Advanced Sterilization Products (ASP) Elsie Kim Sr. Regulatory Affairs Program Lead 33 Technology Drive Irvine, California 92618
Re: K192025
Trade/Device Name: STERRAD VELOCITY Biological Indicator/Process Challenge Device and Reader Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: December 20, 2019 Received: December 23, 2019
Dear Elsie Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Sreekanth Gutala, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192025
#### Device Name
STERRAD VELOCITY® Biological Indicator/Process Challenge Device and Reader
Indications for Use (Describe)
STERRAD VELOCITY® Biological Indicator/Process Challenge Device, in conjunction with the STERRAD VELOCITY Reader, is intended to be used as a standard method for frequent monitoring and periodic testing of the following STERRAD Sterilization Systems:
- · STERRAD® 100NX (STANDARD, FLEX, EXPRESS, and DUO Cycles) with and without ALLClear® Technology
- STERRAD NX® (STANDARD and ADVANCED Cycles) with and without ALLClear® Technology
- · STERRAD® 100S
| | . | Comprehensive and consisted and the commended to the commendent of the commendent of the commendent of concession of consistences and | STATUTION CONSULTION CONSULTION CONSULTION COLLEGION COLLECTION COLLECTION COLLECTION COLLECTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION |
|--|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Type of Use (Select one or both, as applicable) | | |
Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the letters 'ASP' in a stylized font. The letters are connected and enclosed within a rounded rectangle shape. There is a registered trademark symbol in the upper right corner of the logo. The logo is a light blue color.
# K192025 510(k) Summary
# Advanced Sterilization Products, Inc. STERRAD VELOCITY® Biological Indicator/Process Challenge Device and Reader
This summary of 510(k) information is being submitted per 21 CFR 807.92.
#### General Information
| Submitter Name: | Advanced Sterilization Products |
|-----------------|-----------------------------------------------------------------------------------------------------|
| Address | 33 Technology Drive,<br>Irvine, CA 92618 |
| Contact Person: | Elsie Kim<br>Sr. Regulatory Affairs Program Lead<br>Phone: 949-932-4234<br>Email: elsie.kim@asp.com |
Date Prepared: January 22, 2020
#### Device Name
| Proprietary Name: | STERRAD VELOCITY® Biological Indicator/Process Challenge Device and<br>Reader |
|----------------------|-------------------------------------------------------------------------------|
| Common Name: | Biological Indicator |
| Classification Name: | Biological Sterilization Process Indicator |
| Device Class: | Class II |
| Product Code: | FRC |
| CFR Section: | 21 CFR 880.2800 |
#### Predicate Device
Predicate Device: STERRAD VELOCITY Biological Indicator, K182404 cleared December 27, 2018
#### Description
#### STERRAD VELOCITY Biological Indicator/Process Challenge Device
The STERRAD VELOCITY Biological Indicator (BI) /Process Challenge Device (PCD) is a selfcontained biological indicator, used in conjunction with the STERRAD VELOCITY Reader, that is intended for frequent monitoring and periodic testing of the STERRAD Sterilization Cycles, using rapid readout technology that provides a final fluorescence result in 15 minutes at the incubation temperature of 57 ± 2ºC.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image features a logo with a teal background and white lettering. The logo includes a triangle shape with the letters 'ASP' to the right of the triangle. A registered trademark symbol is located in the upper right corner of the logo.
The STERRAD VELOCITY BI/PCD can also be determined as growth-positive or growth-negative via an optional visual pH-based color change result (using bromocresol purple) if used for frequent monitoring purposes. When using this method, the biological indicator must be cultured in an incubator at 55-60°C for 5 to 7 days to get a final visual result.
The STERRAD VELOCITY BI/PCD consists of a glass fiber disc containing a minimum of 1 x 100 Geobacillus stearothermophilus (ATCC 7953) spores and a glass ampoule containing nutrient growth medium and non-fluorescent substrate, as well as a vial, cap, cap label, insert, and chemical indicator. The spore disc, growth media ampoule, and insert are contained in a clear plastic vial with a vented cap. The cap is designed with sterilant ingress openings which allow for penetration of hydrogen peroxide vapor into the vial during the sterilization process. The chemical indicator (CI), placed on the top of the cap, is a Type 1 process indicator that changes color from red/pink to yellow with some red/orange/brown dots when exposed to hydrogen peroxide.
The STERRAD VELOCITY BI/PCD has the same a-glucosidase enzyme system for the fundamental scientific technology as the predicate device cleared under K182404. The a-glucosidase enzyme, which is generated naturally during growth of G. stearothermophilus and released during spore germination, hydrolyzes the bond between the glucose and 4-methylumbelliferyl (4-MU) moieties of 4methylumbelliferyl a-D-glucopyranoside (α-MUG). In the combined state, α-MUG is not fluorescent. Once the bond between the glucose and 4-MU is hydrolyzed, the 4-MU component becomes fluorescent when excited with UV light. Therefore, the a-glucosidase enzyme in its active state can be detected by measuring the fluorescence produced by the enzymatic hydrolysis of a-MUG.
The resultant fluorescent by-product (4-MU), is detected by the Reader and the fluorescent signal is used to determine the positive or negative result of the biological indicator. The measured enzyme activity is reduced upon exposure to hydrogen peroxide. As the enzyme activity is directly correlated with the spore outgrowth, the reduction of the enzyme activity below a certain level indicates that all spores have been inactivated. The level of the fluorescence response is determined using the algorithm developed for the STERRAD VELOCITY BI/PCD and is used to distinguish between the positive and negative responses.
#### STERRAD VELOCITY READER
The STERRAD VELOCITY Reader is designed to automatically read the STERRAD VELOCITY BI/PCD to obtain the final fluorescence result in 15 minutes at the incubation temperature of 57 ± 2℃. The STERRAD VELOCITY Reader utilizes the fluorometric assay method to detect the enzyme activity from the BI and the fluorescence emitted from the BI is converted into a voltage. This voltage reading is then used by the fluorescence algorithm in the Reader to determine the final fluorescence result.
There are eight individual BI incubation wells in the STERRAD VELOCITY Reader. Its heater system is designed to maintain the biological indicators at 57 ± 2℃ to promote the outgrowth of the indicator organisms. Each well contains an ultraviolet light source that excites fluorescence in the growth medium, and a photodetector to detect that fluorescence.
The STERRAD VELOCITY Reader features a touch screen for an effective user interface. Directly under each well is a well number illuminated by a well status indicator light. Three colors (white, green, and red) and two states (off and solid line) are used for the indicator light on the touch screen to show the status of the BI processing. The Reader has a thermoplastic exterior which makes it easy to clean and maintain. A built-in barcode scanner coupled with network connectivity makes maintaining sterilization records easy.
The STERRAD VELOCITY Reader of the subject device has the same hardware and uses same fundamental scientific technology as the predicate device cleared under K182404. Only the algorithm for
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters ASP in a stylized font. The letters are white and are set against a blue-green background. The logo is enclosed in a rounded rectangle, and there is a small registered trademark symbol in the upper right corner of the rectangle.
fluorescence reading has been modified in the subject device to reduce the fluorescence readout time from 30 minutes to 15 minutes.
### Intended Use/Indications for Use
The STERRAD VELOCITY® Biological Indicator/Process Challenge Device, in conjunction with the STERRAD VELOCITY® Reader, is intended to be used as a standard method for frequent monitoring and periodic testing of the following STERRAD Sterilization Systems:
- STERRAD® 100NX (STANDARD, FLEX, EXPRESS, and DUO Cycles) with and without . ALLClear® Technology
- STERRAD NX® (STANDARD and ADVANCED Cycles) with and without ALLClear ● Technology
- STERRAD 100S ●
The intended use of the modified device, as described in the labeling, has not changed as a result of the modification which is the subject of this submission.
The name of the device is changed from STERRAD VELOCITY Biological Indicator to STERRAD VELOCITY Biological Indicator/Process Challenge Device (or BIPCD) as the periodic testing indication was cleared for the predicate device under K182404.
#### Comparison of Technological Characteristics with Predicate Device
The modified device and predicate device have the same technological characteristics based on their use in the monitoring of hydrogen peroxide gas plasma sterilization processes. Refer to the following table for a comparison of the modified device and predicate device characteristics, showing the similarities and differences.
| | Predicate Device | Subject Device |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Characteristics | STERRAD VELOCITY BI<br>(K182404) and Reader | STERRAD VELOCITY BI/PCD and<br>Reader with 15-minute fluorescence<br>readout |
| Intended Use | Monitoring of hydrogen peroxide gas plasma<br>sterilization processes | Same |
| Indications for Use | Used as a standard method for frequent<br>monitoring and periodic testing of the following<br>STERRAD Sterilization Systems:<br>• STERRAD 100NX (STANDARD, FLEX,<br>EXPRESS, and DUO Cycles) with and<br>without ALLClear Technology<br>• STERRAD NX (STANDARD and<br>ADVANCED Cycles) with and without<br>ALLClear Technology<br>• STERRAD 100S | Same |
| Organism<br>(Spore, Species,<br>Strain) | <i>Geobacillus stearothermophilus</i><br>ATCC 7953 | Same |
| | Predicate Device | Subject Device |
| Device<br>Characteristics | STERRAD VELOCITY BI<br>(K182404) and Reader | STERRAD VELOCITY BI/PCD and<br>Reader with 15-minute fluorescence<br>readout |
| Viable Spore<br>Population | $\ge$ 1x106 CFU/BI | Same |
| Carrier Material | Glass Fiber | Same |
| Device Design | Self-contained biological indicator | Same |
| Resistance<br>Characteristics | Tested at 5 mg/L hydrogen peroxide:<br>D-value 5mg/L: ≥ 1 second, using two D-value<br>methods (Survivor Curve and Fraction<br>Negative) per ISO 11138-1.<br>Equal to or greater than the most difficult item<br>routinely processed (biological model). | Same |
| Rapid Readout<br>Technology | The α-glucosidase enzyme system is generated<br>naturally during growth of Geobacillus<br>stearothermophilus. The α-glucosidase enzyme<br>in its active state is detected by measuring the<br>fluorescence produced by the enzymatic<br>hydrolysis of a non-fluorescent substrate. The<br>resultant fluorescent by-product is detected in the<br>reader. The fluorescence signal is used to<br>determine a positive or negative result. | Same |
| Incubation<br>Temperature | 57 +/- 2°C | Same |
| Reduced<br>Incubation Time | 30-minute fluorescence results read in<br>STERRAD VELOCITY Reader and optional<br>visual pH color change results at 5 to 7 days. | 15-minute fluorescence results read in<br>STERRAD VELOCITY Reader and<br>optional visual pH color change results at 5<br>to 7 days. |
| Holding Time | Within 2 hours after sterilization cycle | Same |
| Carrier growth<br>inhibition/media<br>growth promotion | No bacteriostatic effects that inhibit the growth<br>of the indicator microorganism | Same |
| Chemical Indicator | Type 1 process indicator changes color from<br>red/pink to yellow or yellow with some<br>red/orange/brown dots to indicate exposure to<br>hydrogen peroxide. | Same |
| Shelf Life | 9 months | Same |
| Method of<br>Fluorescence<br>Detection | UV LED, optical filters, with sensing by<br>photodiode | Same |
| Indicator of<br>Adequate<br>Sterilization Cycle | "Negative" displayed on LCD Display | Same |
| Indicators of<br>Possible<br>Sterilization Cycle<br>Failure | "Positive" displayed on LCD Display, Audible<br>Alarm | Same |
| Incubation Wells | 8 incubation/reader wells | Same |
| | Predicate Device | Subject Device |
| Device | STERRAD VELOCITY BI | STERRAD VELOCITY BI/PCD and<br>Reader with 15-minute fluorescence |
| Characteristics | (K182404) and Reader | readout |
| Voltage Range | 100-240 Volts AC (12 Volt DC conversion for<br>internal circuitry) | Same |
| Product Safety | UL/IEC 61010-1 | Same |
| EMC Compliance | FCC Part 15, Subpart B, Class A | Same |
#### Technological Characteristic Comparison Table
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with the letters "ASP" in a stylized font. The "A" is represented by a triangle, and the "S" and "P" are connected with curved lines. The logo is enclosed in a rounded rectangle with a light blue fill. There is a registered trademark symbol in the upper right corner of the logo.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image features a logo with the letters 'ASP' in a stylized, interconnected design. The letters are white against a rounded, light blue-green background. A registered trademark symbol is visible in the upper right corner of the logo.
The modified device and its predicate device have the same intended use and indications for use.
The sole difference between the predicate and modified device relates to the decreased time to read, associated with an update to the Reader to include a 15-minute fluorescence readout algorithm. The modified and predicate devices have identical technological characteristics.
#### Summarv of Non-Clinical Testing
Verification testing was conducted in support of the modification to the STERRAD VELOCITY BIPCD and Reader that is the subject of this submission; all testing yielded passing results. This testing is summarized in the following table.
| Performance Testing | Description | Acceptance Criteria | Pass/Fail |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Hydrogen Peroxide Dose<br>Response and Sterilization<br>Verification | The fluorescence results collected<br>for previously submitted dose<br>response testing were reanalyzed<br>using the modified algorithm. No<br>additional sterilization cycles were<br>performed. | General trend of increasing<br>number of sterile BIs (growth and<br>fluorescence) with increasing<br>hydrogen peroxide injection<br>volume. All BIs negative for<br>fluorescence and growth at full<br>cycle. | Pass |
| Design evaluation and<br>Performance Qualification<br>for Periodic Testing | The fluorescence results collected<br>for the previously submitted dose<br>response testing were reanalyzed<br>using the modified algorithm and<br>compared with the biological model<br>data. No additional sterilization<br>cycles were performed. | Fluorescence results demonstrate<br>equal or greater resistance to the<br>biological model by comparison<br>of BI complete inactivation points.<br>The BI shall demonstrate all<br>fluorescence-negative results in<br>full cycle | Pass |
| Verification of Reduced<br>Incubation Time | The fluorescence data collected<br>for the previously submitted<br>reduced incubation time<br>verification study were<br>reanalyzed using the modified<br>algorithm and compared to the 7-<br>day incubation spore growth<br>results from the same study. No<br>additional sterilization cycles<br>were performed. | BI fluorescence-positive for<br>greater than 97.0% of all growth-<br>positive BIs. | Pass |
#### Summary of Non-Clinical Testing
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo with the letters "ASP" in a stylized font. The letters are white and are set against a light blue background. The logo is enclosed in a rounded rectangle shape. There is a registered trademark symbol in the upper right corner of the logo.
| Performance Testing | Description | Acceptance Criteria | Pass/Fail |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Verification of BI Holding<br>Time | The fluorescence data<br>collected for the previously<br>submitted holding time<br>verification study were<br>reanalyzed using the<br>modified algorithm. No<br>additional sterilization<br>cycles were performed. | The number of fluorescence-<br>positive BIs after 2 hours of<br>holding time is not less than the<br>number of fluorescence-positive<br>BIs read immediately after<br>exposure to H2O2. | Pass |
| System Level Error Check | This study verified that the<br>STERRAD VELOCITY Reader<br>with the modified software<br>algorithm meets the system level<br>error check performance<br>requirements as defined in the<br>product design<br>specification and system<br>requirements for the<br>STERRAD VELOCITY<br>Reader. | The reader shall accurately<br>interrogate the BI or provide an<br>error message to alert the user<br>when the BI is subject to rotational<br>movement or handling. | Pass |
| Operational Vibration | This study verified that the<br>STERRAD VELOCITY Reader<br>with the proposed algorithm<br>continues to meet the<br>requirements for operational<br>vibration as defined in the<br>product design specification and<br>system requirements for the<br>STERRAD VELOCITY Reader. | The reader shall accurately<br>interrogate the BI when subjected<br>to random vibration per C-S 1-<br>9711-002 in the 3 orthogonal axes<br>for the entire incubation cycle. | Pass |
| Software Verification and<br>Validation | STERRAD VELOCITY Reader<br>Software Verification and<br>Validation, including<br>Cybersecurity Testing | The modified software meets<br>specifications and functions<br>correctly for the product's<br>intended use. | Pass |
## Clinical Data
No clinical data was generated in support of this Premarket Notification.
## Conclusions
The conclusions drawn from the performance tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device STERRAD VELOCITY Biological Indicator (K182404).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.