Dragonfly OpStar™ Imaging Catheter, AptiVue™ Software version E.5.1
K192019 · ABBOTT MEDICAL · DQO · Nov 8, 2019 · Cardiovascular
Device Facts
Record ID
K192019
Device Name
Dragonfly OpStar™ Imaging Catheter, AptiVue™ Software version E.5.1
Applicant
ABBOTT MEDICAL
Product Code
DQO · Cardiovascular
Decision Date
Nov 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The Dragonfly OpStar™ imaging catheter with OCT imaging system is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Dragonfly OpStar Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The Dragonfly OpStar Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure. The AptiVue™ E-series software is intended for use only with compatible OPTIS™ imaging systems. The OPTIS imaging system with a compatible Dragonfly™ imaging catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The compatible Dragonfly™ imaging catheters are intended for use in vessels 2.0 to 3.5 mm in diameter. The compatible Dragonfly™ imaging catheters are not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure. The OPTIS imaging system is intended for use in the catheterization and related cardiovascular specialty laboratories and will further compute and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. The physician may use the acquired physiological parameters, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
Device Story
Dragonfly OpStar is a sterile, single-use intravascular catheter for coronary artery imaging. Device features external sheath and internal rotating fiber optic core; core emits near-infrared light and receives reflections. Driven by stainless-steel torque wire; pulled back through sheath window by Drive-motor and Optical Controller (DOC). System processes light data to construct OCT images. Used in catheterization labs by physicians. AptiVue software (vE.5.1) controls imaging engine; collects/stores pressure waveform data to compute Fractional Flow Reserve (FFR) and Resting Full-cycle Ratio (RFR). Physician uses OCT images and physiological parameters (pressure drop across lesion) to assess coronary lesion severity and guide therapeutic intervention decisions.
Clinical Evidence
No clinical testing provided. Substantial equivalence supported by bench testing and pre-clinical animal testing comparing the subject device to the predicate.
Technological Characteristics
Intravascular catheter with external sheath and internal rotating fiber optic core. Stainless-steel torque wire drive. Near-infrared light sensing. Software-controlled OCT imaging and physiological parameter computation (FFR, RFR). Sterile, single-use.
Indications for Use
Indicated for patients who are candidates for transluminal interventional procedures requiring coronary artery imaging. Intended for vessels 2.0 to 3.5 mm in diameter. Contraindicated for use in the left main coronary artery or target vessels with previous bypass procedures.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
ILUMIEN OPTIS, OPTIS Integrated, OPTIS Mobile with AptiVue™ Software version E.5 (K183320)
Submission Summary (Full Text)
{0}------------------------------------------------
November 8, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with an abstract design. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG" are written in blue, with the word "ADMINISTRATION" written in a smaller font size below.
Abbott Medical Richard Demello Senior Regulatory Affairs Specialist 4 Robbins Road Westford, Massachusetts 01886
## Re: K192019
Trade/Device Name: Dragonfly OpStar™ Imaging Catheter, AptiVue™ Software version E.5.1 Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DQO, NQQ Dated: October 4, 2019 Received: October 7, 2019
## Dear Richard Demello:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Nicole Gillette Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K192019
#### Device Name
Dragonfly OpStar™ Imaging Catheter, AptiVue™ Software version E.5.1
#### Indications for Use (Describe)
The Dragonfly OpStar™ imaging catheter with OCT imaging system is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Dragonfly OpStar Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The Dragonfly OpStar Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.
The AptiVue™ E-series software is intended for use only with compatible OPTIS™ imaging systems. The OPTIS imaging system with a compatible Dragonfly™ imaging catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The compatible Dragonfly™ imaging catheters are intended for use in vessels 2.0 to 3.5 mm in diameter. The compatible Dragonfly™ imaging catheters are not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.
The OPTIS imaging system is intended for use in the catheterization and related cardiovascular specialty laboratories and will further compute and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. The physician may use the acquired physiological parameters, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <span> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY 5.0
| 510(k) Summary<br>Per 21 CFR §807.92 | |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K192019 |
| Date Prepared | October 4, 2019 |
| Submitter Name<br>& Address | Abbott Medical<br>4 Robbins Road<br>Westford, MA, 01886 |
| Contact Person | Richard DeMello<br>978-577-3504 |
| Alternative<br>Contact Person | Jose Marquez<br>978-577-3578 |
| Proprietary /<br>Trade Name | Dragonfly OpStar™ Imaging Catheter, AptiVue™ Software version E.5.1 |
| Common / Usual<br>Name | Diagnostic Imaging Catheter |
| Product<br>Classification | Product Code: DQO, NQQ |
| Product<br>Regulation<br>Number | 21 CFR 870.1200 |
| Predicate Device | OPTIS Integrated System, Dragonfly OPTIS™ Imaging Catheter (K141769), cleared<br>18 August 2014 |
| Reference Device | ILUMIEN OPTIS, OPTIS Integrated, OPTIS Mobile with AptiVue™ Software version<br>E.5 (K183320), cleared April 2, 2019 |
| Device<br>Description | The Dragonfly OpStar™ Imaging Catheter is a sterile, single-use intravascular catheter<br>consisting of a catheter body external sheath and an internal rotating fiber optic<br>imaging core. The external sheath serves two primary functions: 1) to facilitate<br>placement of the device into the coronary artery and 2) to cover and protect the internal<br>rotating fiber optic imaging core. The inner rotating fiber optic imaging core emits<br>near infrared light to tissues and receives reflected light. It is driven by a stainless-steel<br>torque wire visible under fluoroscopy and pulled back through the window tube of the<br>external sheath by the Drive-motor and Optical Controller (DOC). The emitted and<br>returned reflected light are combined and processed by the OPTIS System software to<br>construct an OCT image. The patient is never exposed to moving parts as the external<br>sheath completely covers the rotating imaging core.<br>The AptiVue™ Software version E.5.1 controls the Optical Coherence Tomography<br>(OCT) imaging engine to collect and store pressure wave form data for computing,<br>Fractional Flow Reserve (FFR), and Resting Full-cycle Ratio (RFR) during procedures<br>and to provide images of the coronary arteries in patients who are candidates for<br>transluminal interventional procedures. FFR, Pd/Pa at rest, and RFR physiological<br>waveforms measured by the system are used to assess the severity of a coronary lesion<br>by measuring the pressure drop across the lesion (distal vs proximal pressure). |
{4}------------------------------------------------
| Indications for<br>Use / Intended<br>Use | The Dragonfly OpStar™ Imaging Catheter with OCT imaging system is intended for<br>the imaging of coronary arteries and is indicated in patients who are candidates for<br>transluminal interventional procedures. The Dragonfly OpStar Imaging Catheter is<br>intended for use in vessels 2.0 to 3.5 mm in diameter. The Dragonfly OpStar Imaging<br>Catheter is not intended for use in the left main coronary artery or in a target vessel<br>which has undergone a previous bypass procedure.<br><br>AptiVue™ E-series software:<br>The AptiVue™ E-series software is intended for use only with compatible OPTIS™<br>imaging systems.The OPTIS imaging system with a compatible Dragonfly™ imaging<br>catheter is intended for the imaging of coronary arteries and is indicated in patients who<br>are candidates for transluminal interventional procedures. The compatible Dragonfly™<br>imaging catheters are intended for use in vessels 2.0 to 3.5 mm in diameter. The<br>compatible Dragonfly™ imaging catheters are not intended for use in the left main<br>coronary artery or in a target vessel which has undergone a previous bypass procedure.<br>The OPTIS imaging system is intended for use in the catheterization and related<br>cardiovascular specialty laboratories and will further compute and display various<br>physiological parameters based on the output from one or more electrodes, transducers,<br>or measuring devices. The physician may use the acquired physiological parameters,<br>along with knowledge of patient history, medical expertise and clinical judgment to<br>determine if therapeutic intervention is indicated. | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison of<br>Subject to<br>Predicate and<br>Reference<br>Devices | The Dragonfly OpStar Imaging Catheter is equivalent to the predicate Dragonfly<br>OPTIS Imaging Catheter (K141769) in terms of intended use, indications for use,<br>operational characteristics, fundamental design, and technological characteristics.<br>Changes to technological characteristics of the device do not raise new questions of<br>safety or effectiveness. | | |
| | Feature | Predicate Device:<br>Dragonfly OPTIS™ Imaging<br>Catheter (K141769) | Proposed Device:<br>Dragonfly OpStar™ Imaging<br>Catheter |
| | Intended Use | The Dragonfly OPTIS™<br>Imaging Catheter with OCT<br>Imaging System is intended for<br>the Visualization and Imaging<br>of Coronary Arteries during an | The Dragonfly OpStar™ Imaging<br>Catheter with OCT Imaging<br>System is intended for the<br>Visualization and Imaging of<br>Coronary Arteries during an |
| | | | |
| Indications for Use | Intended for the imaging of<br>coronary arteries and is<br>indicated in patients who are<br>candidates for transluminal<br>interventional procedures. The<br>Dragonfly OPTIS Imaging<br>Catheter is intended for use in<br>vessels 2.0 to 3.5 mm in<br>diameter. The Dragonfly OPTIS<br>Imaging Catheter is not<br>intended for use in the left main<br>coronary artery or in a target<br>vessel which has undergone a<br>previous bypass procedure. | Intended for the imaging of<br>coronary arteries and is indicated<br>in patients who are candidates for<br>transluminal interventional<br>procedures. The Dragonfly<br>OpStar Imaging Catheter is<br>intended for use in vessels 2.0 to<br>3.5 mm in<br>diameter. The Dragonfly OpStar<br>Imaging Catheter is not intended<br>for use in the left main coronary<br>artery or in a target vessel which<br>has undergone a previous bypass<br>procedure. | |
| Design<br>Modifications | N/A | Modifications are made for<br>improved shaft support,<br>deliverability, and reliability.<br>These changes have been<br>thoroughly assessed through<br>head-to-head bench and animal<br>testing, which conclude these<br>modifications do not change the<br>safety or effectiveness of the<br>Dragonfly OpStar Imaging<br>Catheter. | |
| AptiVue™ Version E.5.1 software is equivalent to the reference AptiVue™ Version<br>E.5 Software (K183320) for intended use, operational characteristics, and fundamental<br>design of the device. Changes to technological characteristics of the device do not raise<br>different questions of safety or effectiveness. | | | |
| Feature | AptiVue™ Version E.5<br>Software (Reference) | AptiVue™ Version E.5.1<br>Software (Proposed) | |
| Intended Use | The AptiVue™ E-series<br>software is intended for use only<br>with compatible OPTISTM<br>imaging systems. OPTISTM<br>imaging systems are intended<br>for use in the catheterization and<br>related cardiovascular<br>specialty laboratories. | The AptiVue™ E-series software<br>is intended for use only with<br>compatible OPTIST™ imaging<br>systems. OPTIST™ imaging<br>systems are intended for use in<br>the catheterization and related<br>cardiovascular specialty<br>laboratories. | |
| Indications for<br>Use | The AptiVue™ E series<br>software is intended to be used<br>only with compatible OPTISTM<br>imaging systems.<br>The OPTIS imaging system<br>with a compatible Dragonfly™<br>imaging catheter is intended for<br>the imaging of coronary arteries | The AptiVue™ E series software<br>is intended to be used only with<br>compatible OPTIS™ imaging<br>systems.<br>The OPTIS imaging system with<br>a compatible Dragonfly™<br>imaging catheter is intended for<br>the imaging of coronary arteries | |
| | are candidates for transluminal interventional procedures. The compatible DragonflyTM imaging catheters are intended for use in vessels 2.0 to 3.5 mm in diameter. The compatible DragonflyTM imaging catheters are not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure. The OPTIS imaging system is intended for use in the catheterization and related cardiovascular specialty laboratories and will further compute and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. The physician may use the acquired physiological parameters, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated. | are candidates for transluminal interventional procedures. The compatible DragonflyTM imaging catheters are intended for use in vessels 2.0 to 3.5 mm in diameter. The compatible DragonflyTM imaging catheters are not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure. The OPTIS imaging system is intended for use in the catheterization and related cardiovascular specialty laboratories and will further compute and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. The physician may use the acquired physiological parameters, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated. | |
| Measurement &<br>Display<br>Features | OCT recordings, FFR, Pd/Pa at rest, and RFR physiological waveforms | OCT recordings, FFR, Pd/Pa at rest, and RFR physiological waveforms | |
| Design<br>Modifications | N/A | Modifications to the AptiVue software version E.5.1 have been made to include recognition of the Dragonfly OpStar catheter, improvements to the Angio Co-registration feature, minor configuration changes and upgrade support. Software design verification and validation testing has been performed which concludes these modifications do | |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
| Summary on<br>Non-Clinical<br>Testing | Design verification and validation was performed on the Dragonfly OPSTAR Imaging<br>Catheter as compared to the predicate Dragonfly OPTIS Imaging catheter in<br>compliance with internal design control procedures, which included bench testing and<br>pre-clinical animal testing. The testing involved head-to-head comparison between the<br>proposed and predicate devices for modifications to technological characteristics. The<br>results of this testing conclude the Dragonfly OPSTAR Imaging Catheter is<br>substantially equivalent to the Dragonfly OPTIS Imaging Catheter.<br>Verification and Validation testing was completed for the AptiVue software version<br>E.5.1 to demonstrate safety and effectiveness and ensure that the subject device<br>performs as intended. Software design verification concludes these modifications do<br>not raise different questions of safety or effectiveness. |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of<br>Clinical Testing | No clinical testing is provided in this pre-market notification. |
| Statement of<br>Equivalence | The Dragonfly OpStar Imaging Catheter is equivalent to the predicate Dragonfly<br>OPTIS Imaging Catheter (K141769) in terms of operational and technological features,<br>hardware components, operational use and target population. Changes to technological<br>characteristics of the device do not raise new questions of safety or effectiveness. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.