← Product Code [ODP](/productcode/ODP) · K192018

# Endoskeleton TA Interbody Fusion Device, Endoskeleton TAS and TAS Hyperlordotic Interbody Fusion Device, Endoskeleton TO Interbody Fusion Device, Endoskeleton TT Interbody Fusion Device, Endoskeleton TC Interbody Fusion Device, Endoskeleton TCS Interbody Fusion Device, Endoskeleton TL Interbody Fusion Device (K192018)

_Titan Spine, Inc. · ODP · Aug 16, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K192018

## Device Facts

- **Applicant:** Titan Spine, Inc.
- **Product Code:** [ODP](/productcode/ODP.md)
- **Decision Date:** Aug 16, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The ENDOSKELETON® TA Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the device may be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). ENDOSKELETON® TA is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate. The ENDOSKELETON® TAS Interbody Fusion device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received 6 months of nonoperative treatment prior to treatment with the device is a standalone system that is intended to be used with the bone screws provided and requres no additional supplementary fixation. The Device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate. Hyperlordotic Devices ≥16°: The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device (≥ 16°) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Patients should have received 6 months of non-operative treatment with the devices. The device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/ or demineralized allograft bone with bone marrow aspirate.The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device must be used with a posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine must be used with a posterior supplemental inxation that has been cleared by the FDA for use in the lumbar spine. The ENDOSKELETON® TC is indicated for use for anterior cervical interbody fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine at one disc level from C-3 to C-7. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received 6 weeks of non-operative treatment prior to treatment with the device. The ENDOSKELETON® TC is indicated to be used with supplemental fixation that has been cleared by the FDA for use in the cervical spine and be used with autograft bone and/ or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate. The ENDOSKELETON® TCS System is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment with the device. The device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate. The device is a stand-alone system when used with ENDOSKELETON® TCS integrated screws and when used without the integrated screws it requires additional supplemental fixation cleared for the cervical spine. The ENDOSKELETON® TL Interbody Fusion Device is indicated for use in spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. The device must be used with supplemental fixation systems that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate. The ENDOSKELETON® TO Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate. The ENDOSKELETON® TT Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non- operative treatment prior to treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate.

## Device Story

Endoskeleton® System consists of interbody and vertebral body fusion devices of varying sizes/geometries; manufactured from Ti6Al4V-ELI titanium alloy. Implants feature hollow centers for bone graft material (autograft, allograft, or demineralized allograft with bone marrow aspirate) and windows for graft visualization/bone formation. Superior/inferior surfaces utilize Chemtex® or nanoLOCK® (MMN™) surface treatments to enhance fixation. Used in spinal fusion procedures; implanted by surgeons. Modification expands indications to include use with demineralized allograft bone with bone marrow aspirate. No changes to design, materials, or surgical technique.

## Clinical Evidence

No clinical data or performance testing was performed for this submission; substantial equivalence is based on literature review and similarity to predicate devices.

## Technological Characteristics

Medical grade titanium alloy (Ti6Al4V-ELI) per ASTM F-136/ISO 5832-3 and ASTM F-3001. Features hollow geometry for bone graft, visualization windows, and Chemtex® or nanoLOCK® (MMN™) surface treatments. Implanted for >30 days. Sterilization via gamma irradiation.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Medtronic Sofamor Danek USA, Inc. ARTiC-L™ 3D and ARTiC-XL™ 3D Ti Spinal System ([K190959](/device/K190959.md))
- Endoskeleton System ([K183557](/device/K183557.md))
- Endoskeleton® TAS and TCS ([K173535](/device/K173535.md))
- Endoskeleton® TO ([K170399](/device/K170399.md))

## Reference Devices

- Endoskeleton TCS nanoLOCK ([K153122](/device/K153122.md))
- Endoskeleton® System ([K141953](/device/K141953.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out in a sans-serif font.

August 16, 2019

Titan Spine, Inc Christine Scifert 6140 West Executive Dr., Suite A Mequon, Wisconsin 53092

Re: K192018

Trade/Device Name: Endoskeleton TA Interbody Fusion Device, Endoskeleton TAS and TAS Hyperlordotic Interbody Fusion Device, Endoskeleton TO Interbody Fusion Device, Endoskeleton TT Interbody Fusion Device, Endoskeleton TC Interbody Fusion Device, Endoskeleton TCS Interbody Fusion Device, Endoskeleton TL Interbody Fusion Device

Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP, OVD, MAX, OVE Dated: July 26, 2019 Received: July 29, 2019

Dear Christine Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Brent Showalter -S

for Melissa Hall Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

#### K192018

#### Device Name

ENDOSKELETON® TA Interbody Fusion Device

#### Indications for Use (Describe)

The ENDOSKELETON® TA Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the device may be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). ENDOSKELETON® TA is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{3}------------------------------------------------

510(k) Number (if known)

K192018

#### Device Name

ENDOSKELETON® TAS Interbody Fusion device

#### Indications for Use (Describe)

The ENDOSKELETON® TAS Interbody Fusion device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received 6 months of nonoperative treatment prior to treatment with the device is a standalone system that is intended to be used with the bone screws provided and requres no additional supplementary fixation. The Device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate.

Hyperlordotic Devices ≥16°: The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device (≥ 16°) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Patients should have received 6 months of non-operative treatment with the devices. The device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/ or demineralized allograft bone with bone marrow aspirate.The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device must be used with a posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine must be used with a posterior supplemental inxation that has been cleared by the FDA for use in the lumbar spine.

| Type of Use (Select one or both, as applicable)                                                           |                                                         |
|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| <div><span style="text-decoration: overline;">区</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{4}------------------------------------------------

510(k) Number (if known)

K192018

Device Name ENDOSKELETON® TC

#### Indications for Use (Describe)

The ENDOSKELETON® TC is indicated for use for anterior cervical interbody fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine at one disc level from C-3 to C-7. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received 6 weeks of non-operative treatment prior to treatment with the device. The ENDOSKELETON® TC is indicated to be used with supplemental fixation that has been cleared by the FDA for use in the cervical spine and be used with autograft bone and/ or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.

| Type of Use (Select one or both, as applicable)                                               |                                         |
|-----------------------------------------------------------------------------------------------|-----------------------------------------|
| <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
| Over-The-Counter Use (21 CFR 801 Subpart C)                                                   | <span style="font-size: 20px;">☐</span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{5}------------------------------------------------

510(k) Number (if known)

K192018

Device Name ENDOSKELETON® TCS

#### Indications for Use (Describe)

The ENDOSKELETON® TCS System is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment with the device. The device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate. The device is a stand-alone system when used with ENDOSKELETON® TCS integrated screws and when used without the integrated screws it requires additional supplemental fixation cleared for the cervical spine.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{6}------------------------------------------------

510(k) Number (if known)

K192018

Device Name ENDOSKELETON® TL

#### Indications for Use (Describe)

The ENDOSKELETON® TL Interbody Fusion Device is indicated for use in spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. The device must be used with supplemental fixation systems that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D)  | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{7}------------------------------------------------

510(k) Number (if known)

K192018

Device Name ENDOSKELETON® TO

#### Indications for Use (Describe)

The ENDOSKELETON® TO Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate.

Type of Use (*Select one or both, as applicable*)> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{8}------------------------------------------------

510(k) Number (if known)

K192018

Device Name ENDOSKELETON® TT

#### Indications for Use (Describe)

The ENDOSKELETON® TT Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non- operative treatment prior to treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{9}------------------------------------------------

| Company:                       | Titan Spine, Inc.<br>6140 West Executive Drive, Suite A<br>Mequon, WI 53092, USA                                                                                                                                                                                                                                                          |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration: | 3006340236                                                                                                                                                                                                                                                                                                                                |
| Primary Contact:               | Christine Scifert<br>Phone: 901-831-8053                                                                                                                                                                                                                                                                                                  |
| Company Contact:               | Jane Rodd<br>Phone: 866-822-7800<br>Fax: 262-242-7802                                                                                                                                                                                                                                                                                     |
| Trade Name:                    | Endoskeleton® TA Interbody Fusion Device<br>Endoskeleton® TAS and TAS Hyperlordotic Interbody Fusion Devices<br>Endoskeleton® TO Interbody Fusion Device<br>Endoskeleton® TT Interbody Fusion Device<br>Endoskeleton® TC Interbody Fusion Device<br>Endoskeleton® TCS Interbody Fusion Device<br>Endoskeleton® TL Interbody Fusion Device |
| Common Name:                   | Intervertebral Fusion Device With Integrated Fixation, Lumbar<br>Intervertebral Fusion Device With Bone Graft, Cervical<br>Intervertebral Fusion Device With Integrated Fixation, Cervical<br>Intervertebral Fusion Device With Bone Graft, Lumbar<br>Spinal Vertebral Body Replacement Device                                            |
| Classification:                | Class II                                                                                                                                                                                                                                                                                                                                  |
| Regulation Number:             | 21 CFR 888.3080 (Intervertebral body fusion device)                                                                                                                                                                                                                                                                                       |
| Panel:                         | 87- Orthopedic                                                                                                                                                                                                                                                                                                                            |
| Product Code:                  | ODP, OVD, OVE, MAX                                                                                                                                                                                                                                                                                                                        |
| Predicate Devices:             | Primary Predicate:<br>Medtronic Sofamor Danek USA, Inc. ARTiC-L™ 3D and ARTiC-XL™<br>3D Ti Spinal System – K190959                                                                                                                                                                                                                        |
| Additional Predicates:         | Endoskeleton System – K183557<br>Endoskeleton® TAS and TCS - K173535<br>Endoskeleton® TO - K170399                                                                                                                                                                                                                                        |

{10}------------------------------------------------

Endoskeleton® System (TA IBD and VBR, TO, TT, TAS, TC, TL) - K141953 Endoskeleton® TAS and TCS - K173535 Endoskeleton® TO – K170399 Endoskeleton TCS nanoLOCK - K153122 K141953

#### Device Description:

The current Endoskeleton® System is an interbody and vertebral body family comprised of a variety of sizes and geometries to accommodate various patient anatomy and pathology. All implantable components are manufactured from medical grade titanium alloy (Ti6Al4V-ELI). The center of the implant is hollow and is to be filled with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate. The design incorporates "windows" through the implant to permit visualization of the graft material and, over time, formation of new bone. The superior and inferior surfaces include either the Chemtex® surface treatment or nanoLOCK® surface treatment (MMNTM) designed to improve fixation to the adjacent bone. The nanoLOCK® surface technology (MMN™) provides a microscopic roughened surface with nano-scale features. This submission seeks to expand the indications of these devices to include use with demineralized allograft bone with bone marrow aspirate.

The predicate Endoskeleton® System (K173535. K170399. K153122. and K141953) is provided either non-sterile or sterile via gamma irradiation.

#### Indication for Use:

The ENDOSKELETON® TA Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. The device may be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). ENDOSKELETON® TA is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.

The ENDOSKELETON® TAS Interbody Fusion device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Patients should have received 6 months of non-operative treatment with the devices. The device is a standalone system that is intended to be used with the bone screws provided and requires no additional supplementary fixation. The Device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.

Hyperlordotic Devices ≥16°: The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device (≥ 16°) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc 

{11}------------------------------------------------

confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received 6 months of nonoperative treatment prior to treatment with the device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device must be used with a posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine must be used with a posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine.

The ENDOSKELETON® TO Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow asprate.

The ENDOSKELETON® TT Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non- operative treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.

The ENDOSKELETON® TC is indicated for use for anterior cervical interbody fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine at one disc level from C-3 to C-7. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received 6 weeks of non-operative treatment with the device. The ENDOSKELETON® TC is indicated to be used with supplemental fixation that has been cleared by the FDA for use in the cervical spine and be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.

The ENDOSKELETON® TCS System is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment with the device. The device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow asprate. The device is a stand-alone system when used with ENDOSKELETON® TCS integrated screws and when used without the integrated screws it requires additional supplemental fixation cleared for the cervical spine.

{12}------------------------------------------------

The ENDOSKELETON® TL Interbody Fusion Device is indicated for use in spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.

### Substantial Equivalence:

Titan Spine has 14 previous clearances related to the predicate Endoskeleton® systems. The most recent clearances for each sub-family are provided below for the primary and secondary predicates:

| Primary Predicate:    | Medtronic Sofamor Danek USA, Inc. ARTiC-L             TM 3D and ARTiC-XLTM<br>3D Ti Spinal System – K190959 |
|-----------------------|-------------------------------------------------------------------------------------------------------------|
| Secondary Predicates: | Endoskeleton® TAS and TCS – K173535                                                                         |
|                       | Endoskeleton® TO – K170399                                                                                  |
|                       | Endoskeleton® System (TA IBD and VBR, TO, TT, TAS, TC, TL) – K141953                                        |
|                       | Endoskeleton TCS nanoLOCK – K153122                                                                         |

The modification for this Special 510(k) relates only to indications for use. The intended use, material, surgical technique, surface treatment, sterility, and design of the subject devices are the same as predicate devices. The only change that is the subject of this submission is to change the indication for use with demineralized allograft bone with bone marrow aspirate.

The purpose of this Special 510(k) is to modify the original indications of the Ensoskeleton® System. This submission seeks to clarify that demineralized HCT/Ps (e.g. Grafton™ DBM DBF) combined with bone marrow aspirate may be used as an alternative to autograft and/or allograft bone comprised of cancellous and/or corticocancellous bone. The review of the literature concluded that there were no additional risks due to the modification of the indications of these devices to include use with demineralized allograft bone with bone marrow aspirate and that these devices are substantially equivalent to the predicate devices.

### Technological Characteristics:

The modification for this Special 510(k) relates only to the indications for use. The intended use, material, surgical technique, surface treatment, sterility, and design of the subject devices are the same as predicate devices.

#### Performance Testing:

No performance testing was required or performed, as this modification for this Special 510(k) relates only to the indications for use.

{13}------------------------------------------------

#### Biocompatibility:

There are no changes to the Endoskeleton® System materials. The devices are patient contacting and are implanted for durations greater than 30 days. All implant components of the Endoskeleton® devices are made from medical grade titanium (Ti6A14V-ELI) alloy per ASTM F-136/ISO 5832-3 and ASTM F-3001.

#### Conclusion:

Based upon the clinical literature contained in this submission and the similarities of the subject and predicate devices, the subject Endoskeleton® system is substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K192018](https://fda-staging.innolitics.com/device/K192018)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).