For use in assisted reproduction procedures to ablate or thin the zona pellucida of an embryo to facilitate assisted hatching or recovery of cells for pre-implantation genetic diagnosis (blastomeres). The device can also be used on blastocyst stage embryos for biopsy of trophectoderm cells for preimplantation diagnosis procedures, and blastocyst collapse prior to vitrification procedures.
Device Story
Assisted reproduction laser systems (LaserShot M, NaviLase) for embryo manipulation. LaserShot M: static system; user moves microscope stage to align embryo with laser. NaviLase: includes motion module/controller for fixed and dynamic modes; dynamic mode allows laser targeting within field of view and automated firing along user-defined paths. Components: 1.48 µm infrared diode laser, microscope adapter, mirror block, 25X objective, USB camera, EyeWare software, target pointer, footswitch. Used in clinical embryology labs by trained personnel. Software controls laser parameters and firing to minimize localized heating. Output: visual display of embryo with laser targeting; ablation of zona pellucida. Benefits: facilitates assisted hatching, biopsy, and vitrification preparation.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing: laser performance validation (targeting accuracy, automated ablation modes), electrical safety, EMC, and software/cybersecurity documentation.
Technological Characteristics
1.48 µm infrared diode laser (Class 1M). Components: microscope adapter, mirror block, 25X objective, USB camera, motion module (NaviLase only). Connectivity: USB. Software: EyeWare for laser control and imaging. Standards: IEC 60601-1 (electrical safety), IEC 61326-1 (EMC).
Indications for Use
Indicated for assisted reproduction procedures including zona pellucida ablation/thinning for assisted hatching or blastomere biopsy, trophectoderm biopsy for preimplantation diagnosis, and blastocyst collapse prior to vitrification.
Regulatory Classification
Identification
The assisted reproduction laser system is a device that images, targets, and controls the power and pulse duration of a laser beam used to ablate a small tangential hole in, or to thin, the zona pellucida of an embryo for assisted hatching or other assisted reproduction procedures.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Assisted Reproduction Laser Systems.” See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
March 25, 2020
Vitrolife GmbH Susanne Schweitzer Manager QA/QC/RA Dr .- Pauling-Str. 9 Bruckberg, Bavaria 84079 GERMANY
K192008 Re:
> Trade/Device Name: NaviLase, LaserShot M Regulation Number: 21 CFR 884.6200 Regulation Name: Assisted Reproduction Laser System Regulatory Class: II Product Code: MRX Dated: February 19, 2020 Received: February 24, 2020
Dear Susanne Schweitzer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192008
Device Name NaviLase, LaserShot M
Indications for Use (Describe)
For use in assisted reproduction procedures to ablate or thin the zona pellucida of an embryo to facilitate assisted hatching or recovery of cells for pre-implantation genetic diagnosis (blastomeres). The device can also be used on blastocyst stage embryos for biopsy of trophectoderm cells for preimplantation diagnosis procedures, and blastocyst collapse prior to vitrification procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K192008
#### Submitter
Company Name Address
#### Contact person
Name Company Address
Phone FAX Email
Date Prepared
## Subject device
Trade name Common Name Regulation Name Requlation Number Product code Requlatory Class
## Predicate device
510(k) number Device name Manufacturer
Vitrolife GmbH Dr .- Pauling-Str. 9 84079 Bruckberg Germany
Susanne Schweitzer Vitrolife GmbH Dr.-Pauling-Str. 9 84079 Bruckberg Germany +49 (0) 8765 / 939 90-19 +49 (0) 8765 / 939 9070 SSchweitzer@vitrolife.com
March 22, 2020
NaviLase, LaserShot M Assisted Reproduction Laser System Assisted Reproduction Laser System 884.6200 MRX (System, Assisted Reproduction Laser) =
K141434 Saturn 5™ Laser Systems Research Instruments Ltd
The predicate device has not been subject to a design-related recall
## Device Description
This submission includes two assisted reproduction laser systems, the LaserShot M and the NaviLase. The LaserShot M is a static laser system where the embryo is moved to the laser by moving the stage of the microscope. The LaserShot M components include a laser module (1.48 um infrared diode laser, class 1M), laser adapter to connect to the microscope, mirror block, laser objective (25X), USB camera, EyeWare Software, and an optional target pointer and footswitch. The NaviLase includes the same components as the LaserShot M. but also includes a motion module and a motion controller. These additional components allow the NaviLase to work in two modes, fixed and dynamic. In fixed mode, the NaviLase operates similarly to the LaserShot M. In dynamic mode, the laser can be directed to any location within the field of view. In addition, dynamic mode allows automated laser firing along a user-defined path (straight line, arc, etc.). The EveWare controls the device components, operates the laser, and takes images and records videos of embryo undergoing laser procedures. The software determines how many holes are drilled along this line and controls/restricts laser firing parameters to
{4}------------------------------------------------
minimize localized heating of the embryo during these procedures. These laser devices have been designed to be fitted to compatible inverted microscopes.
## Indications for Use
For use in assisted reproduction procedures to ablate or thin the zona pellucida of an embryo to facilitate assisted hatching or recovery of cells for pre-implantation genetic diagnosis (blastomeres). The device can also be used on blastocyst stage embryos for biopsy of trophectoderm cells for preimplantation diagnosis procedures, and blastocyst collapse prior to vitrification procedures.
| | Subject Device<br>LaserShot<br>M/NaviLase<br>K181776 | Predicate Device<br>Saturn 5<br>K141434 | Comparison |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | For use in assisted<br>reproduction<br>procedures to ablate or<br>thin the zona pellucida<br>of an embryo to<br>facilitate assisted<br>hatching or recovery of<br>cells for pre-<br>implantation genetic<br>diagnosis<br>(blastomeres). The<br>device can also be<br>used on blastocyst<br>stage embryos for<br>biopsy of<br>trophectoderm cells for<br>preimplantation<br>diagnosis procedures,<br>and blastocyst collapse<br>prior to vitrification<br>procedures. | For use in assisted<br>reproduction procedures<br>to ablate or thin the zona<br>pellucida of an oocyte or<br>embryo to facilitate<br>assisted hatching or<br>recovery of cells for pre-<br>implantation genetic<br>diagnosis (blastomeres).<br>The device can also be<br>used on blastocyst stage<br>embryos for biopsy of<br>trophectoderm cells for<br>pre-implantation<br>diagnosis procedures,<br>and blastocyst collapse<br>prior to vitrification<br>procedures | Different: The<br>indications for use for<br>the subject and<br>predicate devices<br>differ, as the subject<br>device is to be used to<br>ablate or thin the zona<br>pellucida of an embryo,<br>while the predicate is<br>to be used for the<br>same uses for oocytes<br>or embryos. This<br>difference does not<br>represent a new<br>intended use.<br>Therefore, the indented<br>uses are the same. |
| System<br>configuration | · Laser unit<br>· Target pointer<br>· 25X objective<br>· EyeWare software<br>· Installation adapters<br>· Mirror module<br>· Motion control unit<br>(NaviLase Only)<br>· Digital camera<br>· Foot switch | · Control unit (laser<br>unit)<br>· 40X objective<br>· Mirror module<br>· Motor module (Active<br>version only)<br>· RI Viewer software<br>· Installation adapters<br>· Control unit<br>· Camera<br>· Foot switch | Different: The subject<br>and predicate devices<br>have similar<br>components, but<br>differences exist (e.g.,<br>objective magnification,<br>target pointer, etc.).<br>These differences do<br>not raise different<br>questions of Safety<br>and Effectiveness<br>(S&E). |
| Modes of<br>Action | LaserShot M - Fixed<br>laser system where the<br>user moves the sample<br>to the target area.<br>NaviLase - Operates in<br>fixed (similar to<br>LaserShot M) and<br>dynamic modes. The<br>dynamic mode allows<br>the user to move the<br>laser target to the<br>sample within the field<br>of view on the display<br>screen. | Fixed mode and active<br>mode. Active mode<br>allows the user to move<br>the laser within the field<br>of view. | Same |
| Laser Details | 1.48 µm diode laser<br>Class 1M laser | 1.48 µm diode laser<br>Class 1 laser | Different: The class of<br>lasers are different.<br>This difference does<br>not raise different<br>questions of S&E. |
| Energy<br>delivered | 140 mW and 200 mW | 400 mW | Different: The subject<br>device has a lower<br>energy output. This<br>difference does not<br>raise different<br>questions of S&E. |
| Maximum<br>Pulse Length | 10 ms | 2ms | Different: The subject<br>device has a longer<br>maximum pulse length<br>than the predicate<br>device. This difference<br>does not raise different<br>questions of S&E. |
| Multi-Pulse<br>Mode | Yes - NaviLase<br>dynamic mode allows<br>user to ablate a series<br>of holes along a set,<br>pre-determined path. | Yes - active version<br>allows user to ablate a<br>series of holes along a<br>set, pre-determined path. | Same |
| Hole Size<br>Indicator | Yes | Yes | Same |
| Laser firing<br>mechanism | Mouse or foot pedal | Mouse or foot pedal | Same |
| Microscope<br>Compatibility: | Compatible with various<br>models of inverted<br>microscopes | Compatible with various<br>models of inverted<br>microscopes | Same |
| Hole Size<br>Indication: | Yes | Yes | Same |
## Substantial Equivalence Comparison
{5}------------------------------------------------
As noted in the table above, there are differences in the indications for use statements and technological features of the subject and predicate devices. However, as stated in the table, the
{6}------------------------------------------------
differences in indications for use do not represent a new intended use, and the differences in technological features do not raise different questions of safety and effectiveness.
#### Performance Testing
The following performance data were provided in support of the substantial equivalence determination:
- . Reprocessing: Validation testing conducted in accordance with the 2015 FDA guidance document, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
- Electrical Safety and Electromagnetic Compatibility Testing:
- Electrical Safety Testing: IEC 60601-1:2005/AMD1:2012 o
- EMC Testing: IEC 61326-1:2012 O
- Software and Cybersecurity:
- Software documentation in accordance with the 2005 FDA quidance document o "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" to support device software with a minor level of concern.
- Cybersecurity information in accordance with the 2014 FDA quidance document o "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices."
- Laser performance testing: Testing was performed in accordance with recommendations ● in Section 7 of FDA's 2004 quidance document "Assisted Reproduction Laser Systems – Class II Special Controls Guidance Document for Industry and FDA Staff." Additional testing included the following:
- Compatibility of the devices with different microscope systems. O
- Target pointer alignment validation to demonstrate that the center of the target o pointer marker was aligned with the center of the ablation laser beam.
- Validation of NaviLase dynamic targeting within the field of view. O
- Validation of the NaviLase dynamic targeting system's ability to complete o automated ablations under specific treatment modes (e.g., hole, hatching, thinning, and trophectoderm).
# Conclusion
The results of the testing described above demonstrate that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.