OSW Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
Device Story
The OSW Aligner System consists of custom-made, thin, clear, removable polyurethane orthodontic appliances and proprietary 3D software. The system is used by dental practitioners to treat malocclusion as an alternative to conventional wire/bracket appliances. Input data consists of PVS impressions or digital intraoral scans of the patient's teeth. The 3D software generates a digital model of the current state, creates a provisional final treated state, and interpolates intermediate tooth positions. The practitioner reviews, edits, and approves the treatment plan via the software. Upon approval, the system generates digital files for rapid prototyping machines to create physical models; aligners are then thermoformed over these models. The final series of aligners is shipped to the practitioner and dispensed to the patient, who wears them sequentially to move teeth in small increments, typically changing every two weeks. The device benefits patients by providing a removable, aesthetic orthodontic treatment option.
Clinical Evidence
No clinical testing was performed. The device relies on the well-established clinical performance of sequential aligners (product code NXC) cleared since 1998 and literature to demonstrate substantial equivalence.
Technological Characteristics
Material: 0.03" thick thermoformed polyurethane. Form factor: Removable clear aligners. Manufacturing: Digital model creation from PVS/intraoral scans, software-based treatment planning, rapid prototyping of physical models, and thermoforming. Software: Proprietary 3D planning software. Biocompatibility: Tested per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (acute systemic toxicity).
Indications for Use
Indicated for patients requiring orthodontic treatment for the alignment of teeth due to malocclusion.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 23, 2019
OSW Manufacturing, LLC % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K191990
Trade/Device Name: OSW Aligner System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NXC Dated: July 24, 2019 Received: July 25, 2019
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191990
Device Name OSW Aligner System
Indications for Use (Describe)
OSW Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary (K191990)
| 510(k) Owner | OSW Manufacturing, LLC<br>35778 Elk Lane,<br>Murrieta, CA 92563<br>Phone: 408.621.1152 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------|
| Contact person | Ian Kitching<br>OSW Manufacturing, LLC<br>35778 Elk Lane,<br>Murrieta, CA 92563<br>Phone: 408.621.1152<br>ian@oswaligner.com |
| Date Prepared | August 23, 2019 |
| Product Code: | |
| Common Name | Orthodontic Plastic Brackets |
| Trade Name | OSW Aligner System |
| Classification Name | aligner, sequential |
| Regulation | 872.5470 |
| Class | Class II |
| Panel | Dental |
| Product Code | NXC |
| Primary Predicate | K171674<br>Angel Align System |
#### Description
The OSW Aligner System consists of a series of doctor-prescribed, thin, clear plastic removable orthodontic appliances (aligners) and proprietary 3D software. The aligners progressively move the patient's teeth in small increments from their original state to a final treated state.
The OSW Aligner System is intended as an alternative to conventional wire/bracket technology and fixed appliances for the treatment of patients with malocclusion.
The 3D software generates the model of a provisional treated state and then creates a series of models that represent intermediate teeth states. The dental practitioner reviews these models to view, edit or approve the orthodontic treatment plan. The dental practitioner has the option to
{4}------------------------------------------------
reject or request modifications to the setup prior to approving it for aligner fabrication. Once the doctor approves the setup, the series of custom-made aligners are manufactured, packaged, and shipped to the dental practitioner to be dispensed to the patient for treatment.
#### Indications for Use
OSW Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
#### Technological Characteristics
The predicate and the OSW Aligner System were compared in the following areas and found to have similar technological characteristics and to be equivalent:
{5}------------------------------------------------
| Device Comparison Table | | |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | OSW Aligner System | K171674 |
| Indications for<br>Use Statement | The OSW Aligner System is indicated<br>for the alignment of teeth during<br>orthodontic treatment of<br>malocclusion. | The Angel Align System is<br>indicated for the alignment of<br>teeth during orthodontic treatment<br>of malocclusion. |
| 3-D Software<br>Description | The OSW Manufacturing 3-D<br>Software uses a scan of a PVS<br>impression or a digital scan (which<br>represents an untreated state) to<br>generate the image of a final,<br>provisional treated state and then<br>interprets a series of images that<br>represent intermediate teeth states.<br>The dental practitioner then reviews<br>these images and has the option to<br>reject or request modifications to the<br>set-up prior to approving it for<br>aligner fabrication. Once the dental<br>practitioner approves the treatment<br>plan, the software converts the files<br>to produce the series of custom-<br>made aligners | The Angel 3-D Software uses a<br>scan of a PVS impression or a<br>digital scan (which represents an<br>untreated state) to generate the<br>image of a final, treated state and<br>then interprets a series of images<br>that represent intermediate teeth<br>states. The dental practitioner then<br>reviews these images and has the<br>option to reject or request<br>modifications to the set-up prior to<br>approving it for aligner fabrication.<br>Once the dental practitioner<br>approves the treatment plan, the<br>software converts the files to<br>produce the series of custom-made<br>aligners |
| Mode of<br>Operation for 3-D<br>Software | OSW Aligner System 3-D software<br>performs the following operations:<br>Produce 3D-model file of the<br>PVS impression or digital<br>scan. Identifies the individual<br>teeth that will require<br>treatment (i.e.<br>repositioning). Creates a treatment plan (i.e.<br>3-D models that represent<br>the treatment plan). The<br>treating dental practitioner<br>reviews these images using<br>OSWDP software and has<br>the option to reject or<br>request modifications to the<br>set-up prior to approval. | Angel Align System 3-D software<br>performs the following operations:<br>Produce 3D-model file of<br>the PVS impression or<br>digital scan. Identifies the individual<br>teeth that will require<br>treatment (i.e.<br>repositioning). Creates a treatment plan<br>(i.e. 3-D models that<br>represent the treatment<br>plan). The treating dental<br>practitioner reviews these<br>images using iOrtho<br>software and has the<br>option to reject or request<br>modifications to the set-up<br>prior to approval. |
| Material | 0.03" thick, thermoformed<br>polyurethane | 0.03" thick, thermoformed<br>polyurethane |
| Mode of Use | Each appliance is worn by the patient as determined by the dental practitioner, generally 2 weeks prior to being replaced by the next aligner in sequence. | Each appliance is worn by the patient as determined by the dental practitioner, generally 2 weeks prior to being replaced by the next aligner in sequence. |
| Description of Appliance Application | Removable | Removable |
| Manufacturing Method | A digital model of the patient's teeth is created from either scanning a PVS impression or directly from an intraoral scan of the patient's teeth. From the digital model, following a dental practitioner's prescription, the software generates model transforms describing the provisional final, treated state and then interpolates a series of model transforms that represent intermediate states of alignment. The resulting computer "setups" relay this information to rapid prototyping machines that produce physical positive models. The aligners are produced by thermoforming on each physical model to fabricate the sequence of aligners. | A digital model of the patient's teeth is created from either CT scanning a PVS impression or directly from an intraoral scan of the patient's teeth. From the digital model, following a dental practitioner's prescription, the software generates model transforms describing the final, treated state and then interpolates a series of model transforms that represent intermediate states of alignment. The resulting computer "setups" relay this information to rapid prototyping machines that produce physical positive models. The aligners are produced by thermoforming on each physical model to fabricate the sequence of aligners. |
## Device Comparison Table
{6}------------------------------------------------
### Technological Characteristic Differences
There are no major technological characteristic differences when comparing the subject device and the predicate device, with only differences in location of manufacturing. Non-clinical performance testing mitigated any concerns regarding this difference.
### Non-Clinical Performance Testing
The following non-clinical performance tests were conducted:
| Report Name | Standard | Result |
|--------------------------------------------------------------|--------------------|--------|
| OSW OrthoSetup Software Verification Report | N/A (FDA Guidance) | PASS |
| OSWDP System and Software Verification Report | N/A (FDA Guidance) | PASS |
| Biocompatibility testing of device - Cytotoxicity | ISO 10993-5 | PASS |
| Biocompatibility testing of device - Irritation | ISO 10993-10 | PASS |
| Biocompatibility testing of device - Sensitization | ISO 10993-10 | PASS |
| Biocompatibility testing of device - Acute Systemic Toxicity | ISO 10993-11 | PASS |
{7}------------------------------------------------
## Clinical Testing
The performance of sequential aligners in the clinical environment has been well established since the first aligners were cleared by the FDA under product code NXC in 1998. Therefore, clinical testing was not necessary to demonstrate substantial equivalence of the OSW Aligner System to the predicate device.
## Bench Testing
Bench testing of the aligners has not been performed due to the difficulty in evaluating this type of dental device in a laboratory setting. There is sufficient information available from the scientific literature and from the predicate device to demonstrate substantial equivalence of the OSW Aligner System to the predicate device.
# Substantial Equivalence Conclusion
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Based on the comparison and analysis above, the OSW Aligner System is determined to be substantially equivalent to the referenced primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.