K191976 · Piper Access, LLC · FMI · Nov 13, 2019 · General Hospital
Device Facts
Record ID
K191976
Device Name
Piper GO-IO Intraosseous Infusion System
Applicant
Piper Access, LLC
Product Code
FMI · General Hospital
Decision Date
Nov 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic, Pediatric, 3rd-Party Reviewed
Indications for Use
The Piper GO-IO® is intended to provide clinicians and emergency personnel with access to the intraosseous space.
Device Story
Piper GO-IO Intraosseous Infusion System provides emergency intraosseous (IO) access for fluid resuscitation. System includes hypodermic needle set, powered or manual driver, extension set, and securement dressing. Powered driver uses rechargeable lithium battery to assist needle insertion through bone cortex; manual driver uses integrated stylet handle. Stylet features passive safety mechanism for sharps injury prevention. Used by clinicians/emergency personnel in urgent/emergent settings. Device facilitates rapid vascular access when IV access is unattainable. Benefits include reliable, time-critical delivery of fluids/medications to patient circulation.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included needle mechanical properties (ISO 9626, ISO 7864), sharps injury prevention (ISO 23908), electrical safety (IEC 60601-1 series), and biocompatibility (ISO 10993).
Indicated for intraosseous access in proximal tibia, distal tibia, and humeral head of adult and pediatric patients, and distal femur in pediatric patients when IV access is difficult or impossible in emergent, urgent, or medically necessary cases for up to 24 hours.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 13, 2019
Piper Access, LLC % Mark Job Official Correspondent Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K191976
Trade/Device Name: Piper GO-IO Intraosseous Infusion System Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: FMI Dated: September 5, 2019 Received: September 17, 2019
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K191976
Device Name Piper GO-IO® Intraosseous Infusion System
#### Indications for Use (Describe)
The Piper GO-IO® Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K191976
### 510(k) Summary 21 CFR 807.92(a)
# Applicant
| Name: | Piper Access, LLC |
|----------------------------|-----------------------------------------------------------|
| Address: | 3981 South 700 East Suite #15<br>Salt Lake City, UT 84107 |
| Ph: | 801-210-2886 |
| Manufacturer Contact: | Jay Muse, President and CEO |
| Email: | Jay.muse@piperaccess.com |
| Application Correspondent: | Jacob Lee |
| Email: | Jacob.Lee@bd.com |
| Date prepared: | November 08, 2019 |
### Subject Device
| Trade Name: | Piper GO-IO® Intraosseous Infusion System |
|----------------------|-------------------------------------------|
| Common Name: | Intraosseous Infusion System |
| Classification Name: | Hypodermic single lumen needle |
| Class: | II |
| Regulation: | 21 CFR 880.5570 |
| Product Code: | FMI |
| Panel: | General Hospital |
# Predicate Device
| 510(k): | K141117 (Clearance Date July 8, 2014) |
|----------------------|---------------------------------------|
| Trade Name: | EZ-IO® Intraosseous Infusion System |
| Manufacturer: | Arrow/Teleflex Medical |
| Common Name: | Intraosseous Infusion System |
| Classification Name: | Hypodermic single lumen needle |
| Class: | II |
| Regulation: | 21CFR 880.5570 |
| Product Code: | FMI |
| Panel: | General Hospital |
# Reason for Submission
This is a new device.
{4}------------------------------------------------
### Subject Device Description
The Piper GO-IO® Intraosseous Infusion System provides clinicians and emergency personnel with access to the intraosseous space for resuscitation and lifesaving fluid delivery for up to 24 hours. The Piper GO-IO® Intraosseous Infusion System consists of the following:
- a single use hypodermic needle (with needle safety cap), ●
- a powered or manual driver to assist with needle insertion, ●
- an extension set, and;
- an adhesive-backed securement dressing.
For insertions using the powered driver, the hypodermic needle hub that mates with a stylet connected to a drive adapter hub. The drive adapter hub includes a magnetic insert that attaches to the powered driver prior to needle insertion. The Piper GO-IO® Powered Driver is a hand-held, batterypowered device with a rechargeable lithium battery used to assist in the insertion of the subject device needle through the bone cortex. The assembly of the hypodermic needle and stylet with connected drive adapter hub is referred to as the needle set.
For insertions using the manual driver, the needle and the needle hub mate with a stylet in the same way as the needle set that is used with the powered driver, except the stylet is integrated into the handle of the manual driver instead of a drive adaptor hub (i.e. the manual driver needle assembly does not include a drive adapter hub).
The stylet was designed to include a passive safety feature to protect the placer from sharps injury. After the needle is inserted, the stylet is separated from the needle and needle hub. Upon separation of the stylet from the needle hub, the passive safety feature is released onto the stylet tip and can be safely discarded into a sharps container. Following needle insertion, the securement dressing can be applied to secure the needle hub to the skin. An extension set is available for access to the needle hub to support fluid exchange.
The subject device Piper GO-IO® Intraosseous Infusion System will be offered in needle set (for use with the powered driver) and manual driver needle kit configurations. Each kit configuration will include a securement dressing and an extension set.
### Intended Use
The Piper GO-IO® is intended to provide clinicians and emergency personnel with access to the intraosseous space.
### Indications for Use
The Piper GO-IO® Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
### Technological Characteristics
The technological characteristics of the subject Piper GO-IO® Intraosseous Infusion System are substantially equivalent with respect to the basic design and function as compared to the predicate EZ-IO® Intraosseous Infusion System. The differences between the subject and predicate devices are not critical to the intended use of the device and do not raise different questions of safety and effectiveness.
{5}------------------------------------------------
The following table provides a comparison of the technological characteristics between the subject and predicate device in sufficient detail to provide an understanding of the basis for a determination of substantial equivalence.
| Subject and Predicate Device Comparison Table | | |
|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | Piper GO-IO® Intraosseous<br>Infusion System<br>(Subject Device) | EZ-IO® Intraosseous<br>Infusion System<br>(Predicate) |
| 510(k) | Subject of this Premarket<br>Notification | K141117 |
| Product Code | Same as predicate | FMI |
| Intended Use | Same as predicate | Intended to provide clinicians and<br>emergency personnel with access to<br>the intraosseous space. |
| Indications for Use | Same as predicate with the<br>exception of the subject device<br>brand name | The EZ-IO® Intraosseous Infusion<br>System provides intraosseous access<br>in the proximal tibia, distal tibia and<br>humeral head (proximal humerus) of<br>adult and pediatric patients, and the<br>distal femur in pediatric patients<br>when intravenous access is difficult<br>or impossible to obtain in emergent,<br>urgent, or medically necessary cases<br>for up to 24 hours. |
| Target Patient<br>Population | Same as predicate | Adults and Pediatrics |
| Anatomical<br>Insertion Site | Same as predicate | Adults: Proximal tibia, distal tibia,<br>proximal humerus<br>Pediatrics: Proximal tibia, distal<br>tibia, proximal humerus, distal<br>femur |
| Primary IO System<br>Components | Same as predicate with the<br>exception of the subject stylet,<br>which includes a passive safety<br>sharps prevention feature | • Hypodermic Needle w/Stylet<br>· Needle Safety Cap<br>· Securement Dressing<br>· Extension Set<br>• Powered Driver<br>· Manual Driver |
| Needle: Dwell Time | Same as predicate | 24 hours or less |
| Needle: Use | Same as predicate | Single Use |
| Needle Lengths | 15mm (3-39kg)<br>25mm (>3kg)<br>45mm (>40kg) | 15 mm (3-39kg)<br>25 mm (3kg and over)<br>45 mm (40kg and over) |
| Subject and Predicate Device Comparison Table | | |
| Attribute | Piper GO-IO® Intraosseous<br>Infusion System<br>(Subject Device) | EZ-IO® Intraosseous<br>Infusion System<br>(Predicate) |
| Needle: Outer<br>Diameter | Same as predicate | 15 gauge |
| Needle: Materials | Same as predicate | 304 Stainless Steel |
| Needle: Tip Design | Touhy/Huber Style Needle Tip | Faceted Tip |
| Needle: Depth<br>Markers | Same as predicate | Depth markers every 1 cm |
| Needle: Hub<br>Material | Same as predicate | Medical grade polycarbonate |
| Needle: Hub<br>Connection | Same as predicate | Standard Luer Lock |
| Stylet: Materials | Same as predicate | Stainless Steel |
| Stylet: Sharps<br>Injury Prevention<br>Feature | Includes a stylet tip safety feature | Does not include a sharps<br>prevention feature on the stylet |
| Drive Adapter Hub:<br>Materials | Same as predicate | Polycarbonate and stainless steel |
| Inclusion of a<br>Needle Protective<br>Cover | Yes, includes a cover made of<br>polypropylene | Yes, includes a cover made of<br>unknown material |
| Needle Set<br>Sterilization Method<br>& SAL | Same as predicate | EO, 10-6 |
| Manual Driver<br>Attachment | Manual driver handle with<br>integrated stylet mates with internal<br>lumen of needle and needle hub<br>attaches to manual driver | Manual handle attaches to<br>needle set (i.e. assembly of needle<br>and stylet) with magnet |
| Manual Driver<br>Component<br>Materials | Handle: ABS<br>Stylet: Stainless Steel (304) | Handle: Polycarbonate<br>Stylet: The manual handle does not<br>include an integrated stylet |
| Manual Driver<br>Sterilization Method | Same as predicate | EO |
| Manual Driver SAL | Same as predicate | 10-6 |
| Powered Driver<br>Features | Same as predicate | Cordless, Battery-powered |
| Powered Driver Use | Same as predicate | Reusable |
| Powered Driver<br>Materials | Copolyester | Polycarbonate |
| Powered Driver<br>Energy Source | Rechargeable Lithium Batteries | Lithium Batteries |
| Subject and Predicate Device Comparison Table | | |
| Attribute | Piper GO-IO® Intraosseous<br>Infusion System<br>(Subject Device) | EZ-IO® Intraosseous<br>Infusion System<br>(Predicate) |
| Powered Driver<br>Battery Light<br>Indicator | 4 battery light indicators to<br>represent battery charge level | One battery light indicator |
| Powered Driver<br>Needle Attachment | Same as predicate | Magnetic |
| Powered Driver<br>Cleaning Method | Same as predicate | High level disinfectant |
| Powered Driver:<br>Type of protection<br>against electric<br>shock | Internally Powered and Class II | Internally Powered |
| Powered Driver:<br>Degree of protection<br>against electric<br>shock | Same as predicate | Type BF applied part |
| Powered Driver:<br>Degree of protection<br>against ingress of<br>water | IP33, Spraying water and solid<br>foreign body protection (objects<br>>2.5mm diameter) | IPX0 Ordinary Protection |
| Means to Insert<br>Needle | Same as predicate | Manual or Powered Driver |
| General Method of<br>Insertion | Same as predicate | Push needle through soft tissue until<br>it contacts bone. Confirm depth<br>markings. Insert needle set through<br>bone until change in pressure is felt<br>or to desired depth. Remove stylet.<br>Connect IV extension set. |
| Degree of safety<br>or application in<br>the presence of a<br>flammable<br>anesthetic<br>mixture: | Not intended for use in an oxygen<br>rich environment | Equipment not suitable for use in<br>the presence of a flammable<br>anesthetic mixture with air or with<br>oxygen or nitrous oxide |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| Subject and Predicate Device Comparison Table | | | | |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Attribute | Piper GO-IO® Intraosseous<br>Infusion System<br>(Subject Device) | EZ-IO® Intraosseous<br>Infusion System<br>(Predicate) | | |
| Expected Service<br>Life (Powered<br>Driver) | Expected service life is<br>approximately 400 insertions.<br>Service life expectancy is<br>dependent on actual usage (bone<br>density and average insertion time),<br>storage, and frequency of testing. | The predicate device is warranted to<br>a service life of 500 insertions.<br>Service Life expectancy is<br>dependent on actual usage (bone<br>density and average insertion<br>time), storage, and frequency of<br>testing. | | |
| Electromagnetic<br>Immunity and<br>Emissions (IEC<br>60601-1-2) | Same as predicate | Complies with standard | | |
| Maximum RPM | Same as predicate | 1587 or less | | |
{9}------------------------------------------------
# Performance Testing
The following tables identify the performance tests completed on the subject device system, including the standard followed for each test.
| Needle Set Kit and Manual Driver Kit<br>Performance Tests | Standard Followed |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Needle Outer Diameter (OD) | ISO 9626: 2016 |
| Effective Needle Length | ISO 7864: 2016 |
| Needle Lubricity | ISO 7864: 2016 |
| Needle Cleanliness | ISO 9626: 2016 |
| Needle to Hub Assembly Tensile | Internal Protocol/Standard |
| Stylet to Drive Adapter Hub Tensile | Internal Protocol/Standard |
| Needle and Stylet Disassembly Force | ISO 23908: 2011 |
| Safety Activation | FDA Guidance for Sharps Injury Prevention<br>Features & ISO 23908: 2011 |
| Stylet Safety Override (force to failure) | ISO 23908: 2011 |
| Securement Dressing - Pinch Force | Internal Protocol/Standard |
| Securement Dressing – Peel Strength | Internal Protocol/Standard |
| Securement Dressing – Liner Removal | Internal Protocol/Standard |
| Manual Driver Hub to Stylet Tensile | Internal Protocol/Standard |
| Cannula Needle Resistance to Corrosion | ISO 9626: 2016 |
| Cannula Needle Surface Finish | ISO 7864: 2016 |
| Needle Surface Finish and Visual Appearance | ISO 9626: 2016 |
| Needle Hub Luer | ISO 594-1: 1986 and ISO 594-2: 1998 |
| Needle Hub Cleanliness | ISO 7864: 2016 |
| Needle Point | ISO 7864: 2016 |
| Needle Resistance to Breakage | ISO 9626: 2016 |
| Needle Stiffness | ISO 9626: 2016 |
| Gravity Flow Rate | Internal Protocol/Standard |
| Liquid Leak Needle Hub | Internal Protocol/Standard |
| Limits for Acidity or Alkalinity (Needle) | ISO 7864: 2016 |
| Limits for Extractable Metals (Needle) | ISO 7864: 2016 |
| Depth Markings | Internal Protocol/Standard |
| Chemical Resistance | Internal Protocol/Standard |
| Needle Durability | Internal Protocol/Standard |
| Manual Drilling | Internal Protocol/Standard |
| Packaging Integrity and Seal Strength | ISO 11607-1:2006<br>ASTM F88/F88M: 2015<br>ASTM F1886/F1886M: 2016<br>ASTM F1929: 2015 |
| Sharps Injury Prevention Feature (in Simulated<br>Clinical Use) | FDA Guidance for Sharps Injury Prevention<br>Features & ISO 23908: 2011 |
{10}------------------------------------------------
| Powered Drill Performance Tests | Standards Followed |
|--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Needle Set Coupling | Internal Protocol/Standard |
| Battery Capacity/Indicator | Internal Protocol/Standard |
| Battery Usable Life | Internal Protocol/Standard |
| Battery Indicator / State of Charge | Internal Protocol/Standard |
| Drill High Temperature Shut Down | Internal Protocol/Standard |
| Use Life | Internal Protocol/Standard |
| Duty Cycle | Internal Protocol/Standard |
| Motor Stall/Stuck Shutoff | Internal Protocol/Standard |
| Usability | IEC 60601-1-6: 2013 |
| Electrical Safety and Electromagnetic<br>Compatibility | • ANSI AAMI ES60601-1:2005/(R)2012,<br>A1:2012, C1:2009/(R)2012,<br>A2:2010/(R)2012<br>• IEC 60601-1-2: 2014<br>• IEC 60601-1-12: 2014<br>• IEC 62133: 2012 |
| Firmware Verification and Validation | Guidance for the Content of<br>Premarket Submissions for Software Contained<br>in Medical Devices |
| Sterilization, Packaging, and Shelf-Life | Standard Followed |
|------------------------------------------|-----------------------------------------------------------------------------------------------|
| Sterilization Validation/Adoption | ISO 11135:2014 |
| Packaging/Shelf-Life Validations | ISO 11607-1 AMD 1: 2014<br>ASTM F88/F88M: 2015<br>ASTM F1886/F1886M: 2016<br>ASTM F1929: 2015 |
| Residuals | ISO 10993-7: 2008 |
| Bacterial Endotoxin | USP <85><br>USP <161> |
A biocompatibility evaluation was conducted on the subject device per ISO 10993-1:2009, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process. According to the evaluation, the biological tests in the table below were conducted.
| Biological Endpoint | Standard Followed |
|--------------------------------------|--------------------|
| Cytotoxicity | ISO 10993-05: 2009 |
| Sensitization | ISO 10993-10: 2010 |
| Irritation/Intracutaneous Reactivity | |
| Acute Systemic Toxicity | ISO 10993-11: 2006 |
| Material Mediated Pyrogenicity | |
| Hemocompatibility | ISO 10993-4: 2017 |
{11}------------------------------------------------
#### Summary of Substantial Equivalence
The subject device, Piper GO-IO® Intraosseous Infusion System, has the same intended use and the same fundamental scientific technology as the predicate device, EZ-IO Intraosseous Infusion System. The results of performance and biological tests conducted on the Piper GO-IO® Intraosseous Infusion System met all predetermined acceptance criteria and demonstrated that the different technological characteristics of the subject device do not raise different questions of safety and effectiveness. Based on the intended use, technological characteristics, performance and biological test results, the Piper GO-IO® Intraosseous Infusion System can be considered substantially equivalent to the cited predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.