K191972 · Jeil Medical Corporation · HRS · Nov 12, 2019 · Orthopedic
Device Facts
Record ID
K191972
Device Name
ARIX Wrist System
Applicant
Jeil Medical Corporation
Product Code
HRS · Orthopedic
Decision Date
Nov 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ARIX Wrist System(Radius) is intended for use in forearm fractures, osteotomies and arthrodesis. The ARIX Wrist System(Ulna) is intended for fractures and osteotomies, in particular for the ulna.
Device Story
ARIX Wrist System consists of rigid metallic bone plates and screws for internal fixation of forearm fractures, osteotomies, and arthrodesis. Plates vary in length and hole configuration; screws include self-tapping locking, non-locking, and smooth peg variants. Device implanted by orthopedic surgeons in clinical settings to stabilize bone segments. System provides mechanical support to facilitate bone healing. No software or electronic components involved.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including 4-point bending and fatigue tests for plates, and driving torque, torsion, and axial pull-out tests for screws, demonstrating compliance with ASTM F382 and ASTM F543.
Technological Characteristics
Materials: Unalloyed Titanium (ASTM F67) and Ti-6Al-4V ELI (ASTM F136). Components: Bone plates and screws (locking, non-locking, smooth peg). Mechanical fixation via variable locking design. Standards: ASTM F382 (plates), ASTM F543 (screws). Non-electronic, mechanical device.
Indications for Use
Indicated for patients requiring internal fixation for forearm fractures, osteotomies, and arthrodesis (Radius) or fractures and osteotomies (Ulna).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 12, 2019
Jeil Medical Corporation Ahhyeon Woo RA Specialist 702,703,704,705,706,804,805,807,812,815-ho,55 Digital-ro 34-gil. Guro-gu Seoul, 08378 KR
Re: K191972
Trade/Device Name: ARIX Wrist System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC, LXT Dated: July 22, 2019 Received: July 24, 2019
Dear Ahhyeon Woo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K191972
Device Name The ARIX Wrist System
Indications for Use (Describe)
The ARIX Wrist System(Radius) is intended for use in forearm fractures, osteotomies and arthrodesis. The ARIX Wrist System(Ulna) is intended for fractures and osteotomies, in particular for the ulna.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
702·703·704·705·706· Diaital-ro 34-ail Guro-au Tel : +82 2 850 3591 / Fax
K191972 Page 1 of 4
# 510(k) Summary
[As required by 21 CRF 807.92]
### 1. Date Prepared [21 CRF 807.92(a)(1)]
6th November 2019
-
-
### 2. Submitter's Information [21 CFR 807.92(a)(1)]
- Name of Sponsor: . Jeil Medical Corporation 702-703-704-705-706-804-805-807-812-815-ho,55 -Address: Digital-ro34-gil, Guro-gu, Seoul, 08378, Korea
- . Contact Name: Ahhyeon Woo / RA Specialist
- Telephone No. : +82 2 850 3591
- Fax No. : +82 2 850 3536
- -Email Address : uah0606@jeilmed.co.kr
- . Registration Number: 3004049923
- Name of Manufacturer: Same as Sponsor ● Same as Sponsor -Address:
### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
- . Trade Name: ARIX Wrist System ● Common Name: Bone Plate and Bone Screw Classification Name: . Plate, Fixation, Bone // Screw, Fixation, Bone // Appliance, Fixation, Nail/Blade/Plate combination, Multiple component, Metal composite . Classification Description: Single/multiple component metallic bone fixation appliances and accessories . Classification Panel: Orthopedic Classification Regulation: . 21 CFR 888.3030 . Product Code: HRS, HWC, LXT ll ● Device Class:
{4}------------------------------------------------
702-703-704-705-706 Digital-ro 34-ail Guro-au Tel · +82 2 850 3591 / Fax
### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
K191972 Page 2 of 4
The identified predicate devices within this submission are shown as follow;
| Primary Predicate | K170705 - ARIX Wrist System<br>Jeil Medical Corporation |
|-----------------------|---------------------------------------------------------------------------------|
| Additional Predicates | K151468 - ARIX Wrist System<br>Jeil Medical Corporation |
| | K161746 - ARIX Hand System<br>Jeil Medical Corporation |
| | K050932 - DISTAL VOLAR RADIUS ANATOMICAL PLATE SYSTEM<br>HAND INNOVATIONS, INC. |
| | K092247 - Synthes Locking Hand Plates<br>Synthes (USA) |
| | K062498 - Profyle System<br>Howmedica Osteonics Corp |
| Reference Predicates | K180972 - ARIX Clavicle System<br>Jeil Medical Corporation |
There are no significant differences between the subject device and the predicate devices (K170705, K151468, K050932, K092247, K062498, K161746 and K180972) that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, and operational principles as internal fixation components.
### 5. Description of the Device [21 CFR 807.92(a)(4)]
The ARIX Wrist System is rigid fixation consisting of plates and screws in various configurations, shapes and sizes.
The ARIX Wrist System is made of Unalloyed Titanium Alloy (Ti-6AI-4V), which meet ASTM F67, Standard Specification for Unalloyed Titanium for Surgical Implant Applications, and ASTM F136, Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications, which are widely used for surgical implants with well-known biocompatibility.
The plates vary essentially through different lengths and number of plate holes. The screws are selftapping, which are applied with the reconstruction locking screws together. The Locking Screws are provided with diameter 1.5mm and 2.5mm. The Non-locking Screws are provided with diameter 2.5mm. The Smooth Peg Screws are provided with diameter 2.0mm. And Locking Screws are provided with lenaths from 4mm to 56mm. Non-locking Screws are provided with lengths from 6mm to 56mm and Smooth Peg Screws are provided with lengths from 10mm to 30mm.
{5}------------------------------------------------
702-703-704-705-706 Diaital-ro 34-ail Guro-au +82 2 850 3591 /
### 6. Indication for use [21 CFR 807.92(a)(5)]
K191972 Page 3 of 4
The ARIX Wrist System(Radius) is intended for use in forearm fractures, osteotomies and arthrodesis. The ARIX Wrist System(Ulna) is intended for fractures and osteotomies, in particular for the ulna.
### 7. Technological Characteristics [21 CFR 807.92(a)(6)]
### ARIX Wrist System, Bone Plate:
Based on a technical feature comparison, the subject device was found to be similar to all predicate devices with regard to design and materials. The subject plates also have a variable locking feature, similar to the design used in the predicate device (K151468).
### ARIX Wrist System, Bone Screw:
ARIX Wrist System Screws consist of the cleared "ARIX Hand System (K161746)" and "ARIX Wrist System (K170705)". Also, they share similar head, neck and thread designs that are currently cleared under the predicate device (K050932).
### Non-Clinical Test Summary:
Bench tests were conducted to verify that the subject device met all design specifications. The test result demonstrated that the subject device complies with the following standards:
- . ASTM F382, Standard Specification and Test Method for Metallic Bone Plates
- ASTM F543, Standard Specification and Test Method for Metallic Medical Bone Screws .
The following tests were performed with the predicate device:
- Plate .
- 4-Point Bending Test
- 4-Point Fatigue Test
- Screw ●
- Driving Torque Test
- Torsion Test
- Axial Pull-out Test
- Smooth Peg Screw .
- 3-Point Bending Test
The results of this testing indicate that the ARIX Wrist System is equivalent to predicate device.
### Clinical Test Summary:
No clinical studies were considered necessary and performed.
### Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92] 8.
When compared to the predicate device (K151468), the ARIX Wrist System presented in this submission has the same:
- . Indication for Use
- . Technological characteristics
- Operating principle
{6}------------------------------------------------
702·703·704·705·706 Digital-ro 34-gil, Guro-gu, S Tel : +82 2 850 3591 / Fax : +82 2
- K191972 Page 4 of 4
- Design features ●
- Performance ●
- Biocompatibility
- Materials
- . Method of sterilization
## 9. Conclusion [21 CFR 807.92(b)(3)]
In all respects, the ARIX Wrist System is the equivalent of currently marketed devices. This device is made of same materials and has similar dimensions and characteristics. The ARIX Wrist System is manufactured from the unalloyed titanium and titanium alloy that is used generally in this kind of bone plate and bone screw system. Based on the information submitted, ARIX Wrist System is substantially equivalent to the currently marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.