The QuantX Breast MRI Biopsy Guidance Plugin is a software application that assists users of the QuantX software device in planning MRI guided interventional procedures.
Device Story
Software plugin for QuantX platform; assists radiologists in planning MRI-guided breast biopsies. Input: user-specified region of interest (lesion) coordinates and fiducial marker coordinates from MR images. Operation: calculates needle insertion location and depth based on selected grid, needle size, and approach (medial/lateral). Output: visual overlays of biopsy grid, target lesion, and needle trajectory; numerical coordinates and depth measurements. Used in clinical settings by radiologists. Benefits: provides automated guidance for needle placement to improve accuracy of interventional procedures.
Clinical Evidence
No clinical data. Bench testing only. Verification included software requirement testing, interface display, grid/needle variable selection, fiducial marker coordinate accuracy, and comparison of lesion depth/grid cell/block hole calculations against the predicate device.
Technological Characteristics
Software-based image processing system. Functions as a plugin for QuantX SE or Advanced software. Features include user-defined image overlays, coordinate geometry calculations for needle trajectory, and biopsy grid visualization. Operates on standard medical imaging workstations. No specific hardware materials or sterilization required as it is a software-only device.
Indications for Use
Indicated for use by trained medical professionals to assist in planning MRI-guided breast interventional procedures by calculating biopsy needle insertion location and depth for a user-selected region of interest.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
November 8, 2019
Qlarity Imaging, LLC % Mr. Robert Tomek Chief Technology Officer 222 W Merchandise Mart Plaza Suite 1230 CHICAGO IL 60654
Re: K191959
Trade/Device Name: QuantX Breast MRI Biopsy Guidance Plugin Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: July 19, 2019 Received: October 16, 2019
Dear Mr. Tomek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K191959
Device Name
QuantX Breast MRI Biopsy Guidance Module
Indications for Use (Describe)
The QuantX Breast MRI Biopsy Guidance Plugin is a software application that assists users of the QuantX software device in planning MRI guided interventional procedures.
Type of Use (Select one or both, as applicable)
| <span></span> | |
|------------------------------------------------------------|---------------------------------------------|
| <span></span> <div></div> | |
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Premarket Notification 510(k) Summary
## 1.1 Submitter
| Manufacturer: | Qlarity Imaging, LLC |
|-----------------------|------------------------------------|
| Address: | 222 W Merchandise Mart Plaza #1230 |
| | Chicago, IL 60654 |
| Contact Person: | Robert Tomek |
| Contact Person Title: | Chief Technology Officer |
| Telephone: | 872.529.2239 |
| Email: | rtomek@qlarityimaging.com |
| Date Prepared: | August 28, 2019 |
## 1.2 Device Information
510(k) Submission type: Traditional
Device Trade Name: QuantX Breast MRI Biopsy Guidance Plugin
Device Common Name: Image Processing System
510(k) Submitter: Qlarity Imaging, LLC
Classification Regulation: 21 CFR 892.2050
Classification Regulation Name: Radiology
Regulatory Class: Class II
Classification Product Code: LLZ
### 1.3 Leqally Marketed Predicate Device
The legally marketed predicate device is:
| Device Name | Manufacturer<br>Name | 510(k)<br>References | Product<br>Code | Classification |
|-------------|----------------------------|----------------------|-----------------|----------------|
| DynaCAD | MRI Devices<br>Corporation | K041286 | LLZ | Class II |
### 1.4 Device Description
The QuantX Breast MRI Biopsy Guidance Plugin assists users in planning MRI quided interventional procedures. Using information from MR images regarding the coordinates of a user-specified region of interest and fiducial coordinates, the software provides an automatic calculation of the location and depth of the targeted region of interest, such as a lesion or suspected lesion.
{4}------------------------------------------------
lts primary goal is to identify where and how deep a biopsy needle should be inserted into an imaged breast in order to strike a targeted lesion or region of interest, as chosen by a trained medical professional.
The QuantX Breast MRI Biopsy Guidance Plugin may be used in either the SE or Advanced version of QuantX, a software program used for the display and analysis of medical images.
# 1.5 Indications for Use
The QuantX Breast MRI Biopsy Guidance Plugin is a software application that assists users of the QuantX software device in planning MRI quided interventional procedures.
# 1.6 Technological Characteristics
The QuantX Breast MRI Biopsy Guidance Plugin has the following same technological characteristics as the predicate device.
- User-initiated selection of target breast (left or right) and direction of ● approach (medial or lateral).
- Customized image display depending on side and direction of approach. ●
- User-initiated placement of fiducial marker and biopsy grid. ●
- User-initiated selection of needle size and block combination within the biopsy panel.
- Display of image overlay for fiducial marker and biopsy grid. ●
- Display of image overlay for target region of interest location and sample ● needle trajectory.
- Calculation and display of coordinates and depth to reqion of interest. .
The following technological differences exist between the subject and predicate device; these differences do not affect the safety and effectiveness of the device when used as labeled:
- Subject device is a plugin that can be used in either the SE or Advanced versions of QuantX, a software program that analyzes patient breast images, and is designed to aid radiologists in the characterization of lesions as part of high-risk screening or diagnostic work-up. The predicate device is an image processing system designed to assist in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) studies. Additionally, the predicate device assists users in planning MRI guided interventional procedures.
These differences do not affect the safety and effectiveness of the device when used as labeled.
{5}------------------------------------------------
# 1.7 NON-CLINICAL PERFORMANCE DATA TESTING AND REVIEWS - as required by 807.92(b)(1)
The QuantX Breast MRI Biopsv Guidance Plugin has been successfully tested for its software requirements and features. Nonclinical tests included:
- Verification of proper activation of biopsy guidance mode and interface . display;
- Verification of proper creation of needle block images; ●
- Verification of proper loading of image series to be used in biopsy guidance; ●
- Verification of complete and correct selection of grid type variables; ●
- Verification of complete and correct needle type variables; ●
- Verification of correct fiducial marker image-space coordinates;
- Verification of correct size of grid image overlay and location;
- Verification of correct lesion marker overlay display and image-space ● coordinates;
- Verification of proper display of selected breast, grid cell, and block hole;
- Verification of proper display of needle block image and needle depth: ●
- . Verification of correct patient orientation indicators:
- Verification of lesion depth calculation by comparison to predicate; ●
- . Verification of correct needle block hole by comparison to predicate;
- Verification of correct grid cell by comparison to predicate; ●
- Verification of quidance worksheet output; ●
- Validation testing demonstrates that the device conforms to user needs and ● intended use.
Based on these tests, we conclude that the subject device is safe and effective and substantially equivalent to the predicate device.
# 1.8 Conclusion
Based on the information supplied in this 510(k), we conclude that the subject device is safe, effective, and substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.