K191948 · Verathon Medical (Canada) Ulc · EOQ · Aug 21, 2019 · Ear, Nose, Throat
Device Facts
Record ID
K191948
Device Name
GlideScope BFlex 5.8 Single-Use Bronchoscope
Applicant
Verathon Medical (Canada) Ulc
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Aug 21, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
GlideScope® BFlex™ Single-Use bronchoscopes are intended to work with a video monitor, in conjunction with non-powered endoscopic accessories and other ancillary equipment, for endoscopy within the airways and tracheobronchial tree.
Device Story
The GlideScope BFlex 5.8 is a single-use, flexible bronchoscope designed for airway procedures. It functions as part of a system including a reusable video monitor and cable. The device features a distal tip with an integrated camera and LED light source, allowing real-time visualization of the tracheobronchial tree. A suction button and tip maneuverability controls are provided for the operator. The device is used in hospital settings by clinicians to navigate airways and facilitate the use of non-powered endoscopic accessories. The video monitor displays the camera feed, assisting the clinician in diagnostic or therapeutic airway interventions. The single-use nature of the bronchoscope eliminates the need for reprocessing, reducing cross-contamination risks. The 5.8mm diameter model is specifically sized for use with endotracheal tubes having a minimum internal diameter of 7.0mm and endoscopic accessories with a minimum working channel width of 3.0mm.
Clinical Evidence
No clinical data was required or provided. Substantial equivalence was supported by bench testing, including system requirements, electrical safety (AAMI/ANSI ES60601-1, IEC 60601-2-18), electromagnetic compatibility (ANSI AAMI IEC 60601-1-2), optical testing (ISO 8600-1, -3, -4), biocompatibility (ISO 10993-1), aging performance, sterile packaging integrity, and cleaning testing.
Technological Characteristics
Flexible endoscope; 5.8mm insertion tube diameter; 3.0mm working channel; 0° direction of view; 120° diagonal field of view; 5-50mm depth of field; 640x480 resolution; LED light source; sterile (EO); rechargeable lithium-ion battery (in monitor). Complies with ISO 8600-1, ISO 8600-3, ISO 8600-4, and ISO 10993-1.
Indications for Use
Indicated for use in adult patients in a hospital environment for endoscopy within the airways and tracheobronchial tree.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
GlideScope® BFlex™ 5.0 Single-Use Bronchoscope System (K183256)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 21, 2019
Verathon Medical (Canada) ULC Haleh Ghassemi Technical Manager, Regulatory Affairs 2227 Douglas Road Burnaby, V5C 5A9 Ca
Re: K191948
Trade/Device Name: GlideScope BFlex 5.8 Single-Use Bronchoscope Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: July 19, 2019 Received: July 22, 2019
Dear Haleh Ghassemi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
Device Name
GlideScope® BFlex™ 5.8 Single-Use Bronchoscope
Indications for Use (Describe)
GlideScope® BFlex™ Single-Use Bronchoscopes are intended to work with the video monitor, in conjunction with nonpowered endoscopic accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Verathon. The logo consists of a stylized "V" shape in blue and green, followed by the word "VERATHON" in blue, sans-serif font. The "V" shape is formed by two intersecting lines, with the left line in blue and the right line in green.
# 510(K) SUMMARY
This summary of Safety and Effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR 807, Subpart E, section 807.92.
## Submitter:
Verathon Medical (Canada) ULC 2227 Douglas Road Burnaby, BC V3W 1P2 Canada
#### Contact Person:
Haleh Ghassemi Technical Manager, Regulatory Affairs Phone: (425) 629-5517 Email: haleh.ghassemi@verathon.com
#### Date Summary Prepared:
July 19, 2019
### Establishment Registration Number:
Verathon Medical (Canada) ULC Registration Number: 9615393 Owner/Operator Number: 9095489
#### Device Trade or Proprietary Name:
GlideScope® BFlex™ 5.8 Single-Use Bronchoscope
#### Device Common or Usual Name:
Flexible Bronchoscope
| Device Trade or Proprietary Name | Device Common or Usual Name |
|---------------------------------------------------|-----------------------------|
| GlideScope® BFlex™ 5.8 Single-Use<br>Bronchoscope | Flexible Bronchoscope |
#### Device Classification:
| Classification Name | Class | Product Code | Classification Regulation |
|-----------------------------------------------------|-------|--------------|---------------------------|
| Bronchoscope (Flexible or<br>Rigid) and Accessories | | EOQ | 21 CFR 874.4680 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Verathon logo. The logo consists of a stylized "V" shape in blue and green, followed by the word "VERATHON" in blue, with a registered trademark symbol next to it. The logo is simple and modern, with a clean design.
# Review Panel:
Ear. Nose, and Throat
# Predicate Device:
The features and functions of the proposed GlideScope® BFlex™ 5.8 Single-Use Bronchoscope is substantially equivalent to the previously cleared GlideScope® BFlex™ 5.0 Single-Use Bronchoscope. The 510(k) clearance number and respective clearance date for the predicate device is included in the table below:
| Predicate Device | 510(k) Number | Clearance Date |
|----------------------------------------------------------|---------------|------------------|
| GlideScope® BFlex™ 5.0 Single-Use<br>Bronchoscope System | K183256 | January 04, 2019 |
# Device Description:
The GlideScope® BFlex™ 5.8 Single-Use Bronchoscope is one component of the GlideScope® BFlex™ Single-Use Bronchoscope System. The system consists of a single-use flexible bronchoscope, a reusable monitor, and a reusable cable. The GlideScope® BFlex™ Single-Use Bronchoscope System is intended to provide real time viewing and recording for a wide range of airway procedures.
Similar to the predicate GlideScope® BFlex™ 5.0 Single-Use Bronchoscope, the GlideScope® BFlex™ 5.8 Single-Use Bronchoscope is distributed sterile and is for single use only. The GlideScope® BFlex™ bronchoscopes operate with a portable reusable GlideScope video monitor (GVM or Core monitor) for purposes of image display.
# Intended Use:
GlideScope® BFlex™ Single-Use bronchoscopes are intended to work with a video monitor, in coniunction with non-powered endoscopic accessories and other ancillary equipment, for endoscopy within the airways and tracheobronchial tree.
# Intended Patient Population:
The GlideScope® BFIex™ Single-Use system is for use in a hospital environment. The GlideScope® BFlex™ bronchoscope is a single-use device designed for use in adults. It has been verified and validated for the following endotracheal tube (ETT) and endoscope accessory (EA) sizes:
| Model | Minimum ETT Internal Diameter | EA Minimum Working Channel Width |
|-----------|-------------------------------|----------------------------------|
| BFlex 5.0 | 6.0 mm | 2.1 mm |
| BFlex 5.8 | 7.0 mm | 3.0 mm |
{5}------------------------------------------------
# Technological Characteristics:
The proposed subject GlideScope® BFlex™ 5.8 Single-Use Bronchoscope when compared to the predicate bronchoscope has similar technological characteristics. See the comparison table below for similarities and differences between the proposed and predicate devices:
| Technological Characteristic | Predicate GlideScope®<br>BFlex™ 5.0 Single-Use<br>Bronchoscope | Proposed GlideScope®<br>BFlex™ 5.8 Single-Use<br>Bronchoscope |
|------------------------------------------------------------------------|----------------------------------------------------------------|---------------------------------------------------------------|
| Flexible Endoscope | Yes | Yes |
| Outside Diameter of Flexible<br>Insertion Tube/Shaft and<br>Distal Tip | 5.0mm | 5.8mm |
| Minimum Internal Diameter<br>of Working Channel | 2.1mm | 3.0mm |
| Suction Button | Yes | Yes |
| Single Use Bronchoscope | Yes | Yes |
| Sterility | Sterile by Ethylene Oxide<br>(EO) | Same |
| Control Button for Tip<br>maneuverability | Yes | Yes |
| Power Source | Rechargeable Lithium-ion<br>Battery | Same |
| Camera | Yes | Yes |
| Direction of View, Relative to<br>Center Line of Distal tip | 0° | Same |
| Field of view,<br>horizontal/vertical | 85° | Same |
| Field of View, diagonal | 120° | Same |
| Depth of Field | 5-50mm | Same |
| Image Resolution | 640x480 | Same |
| LED Light Source | Yes | Yes |
| Image Display | Displays image on a<br>Reusable Video Monitor | Same |
| Extended Viewing | Yes | Yes |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Verathon. The logo consists of a stylized "V" shape in green, with the word "VERATHON" in blue, sans-serif font to the right of the "V". The "V" shape is formed by two intersecting lines, with the left line extending slightly higher than the right line. The logo is simple and modern, with a clean and professional look.
# Performance Testing:
Performance testing has been completed to demonstrate that the proposed GlideScope® BFlex™ 5.8 Single-Use Bronchoscope meets the safety and performance requirements established in the design specifications. Comprehensive verification and validation testing included the following:
- Full System Requirements Testing
- Electrical Safety according to ●
- AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance)
- IEC 60601-2-18: Edition 3.0 2009-08 o Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
- Electromagnetic Compatibility according to ●
- ANSI AAMI IEC 60601-1-2:2014 Edition 4.0 o Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 4)
- Optical testing according to ●
- ISO 8600-1 Fourth Edition 2015-10-15 Endoscopes - Medical endoscopes and endotherapy devices -- Part 1: General requirements
- o ISO 8600-3 First edition 1997-07-01 (Amendment 1 2003-12-01 Optics and Optical instruments - Medical endoscopes and endoscopic accessories - Part 3: Determination of field of view and direction of view of endoscopes with optics [Including: Amendment 1 (2003)]
- ISO 8600-4 Second Edition 2014-03-15 Endoscopes - Medical endoscopes and certain accessories - Part 4: Determination of maximum width of insertion portion
- Biocompatibility according to ●
- ANSI AAMI ISO 10993-1:2009/(R)2013 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- Aging Performance Testing ●
- Sterile Packaging Integrity Testing
- Cleaning Testing ●
- Design Validation ●
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Verathon. The logo consists of a stylized green plant-like symbol on the left, followed by the word "VERATHON" in blue, with a registered trademark symbol next to it. The logo is simple and modern, with a focus on the company name.
Results: All testing resulted in acceptance criteria passed.
# Summary of Clinical Tests:
The GlideScope® BFlex™ 5.8 Single-Use Bronchoscope, subject of this submission, did not require clinical studies to support the determination of substantial equivalence.
#### Conclusion:
The information in this 510(k) Premarket Notification demonstrates that the proposed GlideScope® BFlex™ 5.8 Single-Use Bronchoscope is substantially equivalent to the previously cleared predicate GlideScope® BFlex™ 5.0 Single-Use Bronchoscope with respect to safety, effectiveness, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.