K191926 · Vita Zanhnfabrik H. Rauter GmbH Co. · EBI · Jan 29, 2020 · Dental
Device Facts
Record ID
K191926
Device Name
VITA VIONIC Bond
Applicant
Vita Zanhnfabrik H. Rauter GmbH Co.
Product Code
EBI · Dental
Decision Date
Jan 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3760
Device Class
Class 2
Indications for Use
VITA VIONIC Bond is used to fix custom-fit prosthetic teeth in appropriately milled cavities of prosthetic bases.
Device Story
VITA VIONIC Bond is a two-component adhesive system composed of acrylate polymers based on methyl methacrylate. It is used by dental professionals to fix artificial teeth into milled denture bases. The device consists of two liquid components supplied in glass containers; these are mixed together until uniform. The resulting mixture is applied as a thin film to bonding surfaces, after which the artificial teeth are positioned in the denture base and allowed to cure via a self-curing chemical polymerization reaction. The device is provided non-sterile and is intended for use in a dental laboratory or clinical setting to facilitate the assembly of prosthetic dentures.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing only, including biocompatibility assessments (cytotoxicity, sensitization, irritation, intracutaneous reactivity, and genotoxicity) and performance testing per ISO 20795-1 (bonding to synthetic polymer teeth, water solubility, and water sorption). Chemical analysis confirmed residual monomer content is within acceptable limits.
Technological Characteristics
Two-component adhesive system; acrylate polymers based on methyl methacrylate (MMA). Self-curing via chemical polymerization. Provided non-sterile. Tested per ISO 20795-1:2013 (denture base polymers), ISO 10993-1:2009 (biocompatibility), and ISO 7405:2004 (biocompatibility).
Indications for Use
Indicated for fixing custom-fit prosthetic teeth into milled cavities of prosthetic bases for patients requiring dental prosthetics.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 29, 2020
VITA Zanhnfabrik H. Rauter GmbH Co. % Lindsay Tilton Regulatory Affairs & Compliance (Consultant) VITA North America 22705 Savi Ranch Parkway, Sutie 100 Yorba Linda. California 92887
Re: K191926
Trade/Device Name: VITA VIONIC Bond Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: December 20, 2019 Received: December 30, 2019
Dear Lindsay Tilton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K191926
Device Name VITA VIONIC Bond
Indications for Use (Describe)
VITA VIONIC Bond is used to fix custom-fit prosthetic teeth in appropriately milled cavities of prosthetic bases.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K191926 510(k) Summary
| Submitter | VITA Zahnfabrik H.Rauter GmbH Co.<br>Spitelgasse 3<br>Bad Sackingen, D-79713<br>Germany<br>Establishment Reg. No. 1000625496 | |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact | Bernd Walker<br>Head of Regulatory Affairs and Quality Systems<br>Phone (+49) 7761 562-361<br>Fax (+49) 7761 562-384<br>B.Walker@vita-zahnfabrik.com | |
| Official Correspondent | Lindsay Tilton<br>Regulatory Affairs & Compliance Manager<br>VITA North America, Inc.<br>22705 Savi Ranch Parkway, Suite 100<br>Yorba Linda, CA 92887<br>Establishment Reg. No. 2082832<br>Phone (925) 699-9091<br>E-mail: ltilton@vitanorthamerica.com | |
| Date Prepared | January 24, 2020 | |
- Trade/Device Name VITA VIONIC® Bond
- י Classification Name Denture relining, repairing, or rebasing resin
- . Regulation Number 21 CFR 872.3760
- Product Code EBI
{4}------------------------------------------------
### Predicate and Reference Predicate Devices
lvoBase CAD Bond, and Modelling Liquid – K151142 – Primary Predicate Device
Tokuyama Curegrace – K170549 – Reference Predicate Device
#### Device Description
VITA VIONIC® Bond is a two-component glue, mainly composed of acrylate polymers based on methyl methacrylate and used to fix artificial teeth in the milled denture base. VITA VIONIC® component I comes in a glass jar with a screw top lid and VITA VIONIC® Bond component II comes in a glass bottle with a screw top cap. Component II is mixed into component I by stirring until no streaks are visible. A thin film is then applied to all Bonding surfaces and then the artificial teeth are placed into the denture base and VITA VIONIC® Bond is allowed to cure. Comes in one, colorless shade.
This product is provided non-sterile.
#### Indications for use
VITA VIONIC® Bond is used to fix custom-fit prosthetic teeth in appropriately milled cavities of prosthetic bases.
#### Material Composition
VITA VIONIC® Bond is mainly composed of acrylate polymers based on methyl methacylate (MMA)
#### Non-Clinical Performance Testing
VITA VIONIC® Bond was evaluated according to the following FDA recognized standards:
- . DIN EN ISO 20795-1:2013 - Dentistry - Base polymers - Part 1: Denture base polymers (ISO 20795-1:2013)
- o This standard mainly refers to the requirements for a denture base. VITA VIONIC® Bond, bonds the artificial teeth to the denture base, which is tested under section 8.7 "Bonding to synthetic polymer teeth"
- Section 8.9: Water solubility and water sorption
- ISO 10993-1:2009 - Biological evaluation of medical devices
- I ISO 7405:2004 – Dentistry – Evaluation of biocompatibility of medical devices
- . A Chemical Analysis was conducted on VITA VIONIC® Bond and the Residual monomer content were found to be below the limit values.
{5}------------------------------------------------
Bench test results allowed us to conclude that VITA VIONIC® Bond is substantially equivalent to the predicate device for its intended use.
#### Biocompatibility
A biocompatibility assessment was performed on VITA VIONIC® Bond in accordance with ISO 10993-1:2009 – Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process, and ISO 7405:2008 - Dentistry -Evaluation of Biocompatibility of Medical Devices Used in Dentistry. This assessment supports that VITA VIONIC® Bond is biocompatible and concludes that the device is substantially equivalent to the predicate device for its intended use based on the following:
Biocompatibility tests were conducted on VITA VIONIC Bond for cytotoxicity, sensitization, irritation and intracutaneous reactivity, and genotoxicity.
- I Cytotoxicity and genotoxicity and the results showed no cytotoxic or genotoxic results
- . Irritation and sensitization risks were evaluated and found to be low
#### Clinical Performance Data
Not applicable. No human clinical testing was performed to support the substantial equivalence of VITA VIONIC® Bond.
| | Subject Device | Primary Predicate<br>Device | Reference Predicate<br>Device |
|-----------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Device Name | VITA VIONIC® Bond | IvoBase Bond and<br>Modelling Liquid | Tokuyama Cure<br>Grace |
| Device Classification | Denture relining,<br>repairing, or<br>rebasing resin | Denture relining,<br>repairing, or<br>rebasing resin | Denture relining,<br>repairing, or<br>rebasing resin |
| Product Code | EBI | EBI | EBI,EBG |
| Indication for use | Used to fix custom-<br>fit prosthetic teeth<br>in appropriately<br>milled cavities of<br>prosthetic bases | For the fabrication<br>of removable<br>dentures, e.g.:<br>-partial and<br>complete denture | An acrylic resin used<br>in a variety of<br>dental application<br>such as:<br>-Temporary inlay, |
#### Comparison to Predicate Devices
{6}------------------------------------------------
| | | prosthetics | crowns and bridges |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| | Hybrid denture<br>prosthetics | | -Repair of broken or<br>cracked dentures |
| | -Combined denture<br>prosthetics | | -Denture border<br>extension |
| | -Mouth guards | | -Replacement of<br>lost denture teeth |
| | -Implant-supported<br>denture prosthetics | | -Adjustment of<br>occlusal height of<br>resin teeth |
| Components | Liquid and Liquid | Powder and Liquid | Powder and Liquid |
| Principle of<br>operation | Cured by chemical<br>polymerization<br>reaction starting<br>with mixing the<br>liquid and liquid<br>component. (Self-<br>curing) | Cured by chemical<br>polymerization<br>reaction starting<br>with mixing the<br>powder and liquid<br>component. (Self-<br>curing) | Cured by chemical<br>polymerization<br>reaction starting<br>with mixing the<br>powder and liquid<br>component. (Self-<br>curing) |
## Conclusion
Information provided in this application demonstrates that VITA VIONIC® Bond is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.