HepaShield Bacterial Viral Breathing System Filter
K191909 · Flexicare Medical Limited. · CAH · Mar 23, 2020 · Anesthesiology
Device Facts
Record ID
K191909
Device Name
HepaShield Bacterial Viral Breathing System Filter
Applicant
Flexicare Medical Limited.
Product Code
CAH · Anesthesiology
Decision Date
Mar 23, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5260
Device Class
Class 2
Indications for Use
Flexicare's HepaShield Bacterial Viral Breathing System Filters are intended to reduce the transmission of bacteria and viruses to/from a patient during anesthesia. For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and/or expired gases is desired. Flexicare's HepaShield Bacterial Viral Breathing System Filters are single patient for use on a single patient for up to 24hrs and are available in Adult size. FlepaShield Bacterial Viral Breathing System Filters are designed to be used in hospital environments by trained personnel.
Device Story
HepaShield is a mechanical bacterial/viral breathing system filter; used in hospital settings by trained personnel during anesthesia or assisted ventilation. Device consists of a housing (top/bottom), pleated paper filter pack, and a tethered luer port cap for CO2 monitoring. Input is inspired/expired breathing gases; filter media mechanically traps pathogens; output is filtered gas. Device connects to standard 22mm/15mm conical connectors (ISO 5356-1). Single-use, disposable (max 24 hours). Benefits include reduced cross-contamination risk for patients and equipment. Healthcare providers integrate it into the breathing circuit to maintain sterility of the gas pathway.
Clinical Evidence
Bench testing only. No clinical data. Performance evaluated against ISO 9360-1, ISO 23328-2, ISO 5356-1, ISO 10993, and EN ISO 18562. Metrics included BFE (99.99999%), VFE (99.9999%), pressure drop, gas leakage, and biocompatibility. All tests passed.
Technological Characteristics
Mechanical filtration via pleated paper filter pack. Housing: ultrasonic welded. Connectors: 22mm/15mm conical (ISO 5356-1). Luer port: ISO 80369-7. Biocompatibility: ISO 10993, EN ISO 18562. Shelf life: 5 years. Sterile and non-sterile variants available.
Indications for Use
Indicated for adult patients requiring anesthesia, artificial respiration, or assisted respiration in hospital environments to reduce bacterial/viral transmission during gas filtration. Contraindicated for use between a humidification/nebulizing device and the patient, or at the patient end when used with a humidified breathing system.
Regulatory Classification
Identification
A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.
Predicate Devices
Pall's Ultipor 25 Bacterial Viral Breathing System Filter (K013093)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 23, 2020
Flexicare Medical Limited Joel Biddle Regulatory Manager Cynon Vallev Business Park Mountain Ash, cf45 4er UK
Re: K191909
Trade/Device Name: HepaShield Bacterial Viral Breathing System Filter Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: Class II Product Code: CAH Dated: February 19, 2020 Received: February 24, 2020
Dear Joel Biddle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
#### K191909
Device Name HepaShield Bacterial Viral Breathing System Filter
Indications for Use (Describe)
Flexicare's HepaShield Bacterial Viral Breathing System Filters are intended to reduce the transmission of bacteria and viruses to/from a patient during anesthesia. For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and/or expired gases is desired.
Flexicare's HepaShield Bacterial Viral Breathing System Filters are single patient for use on a single patient for up to 24hrs and are available in Adult size. FlepaShield Bacterial Viral Breathing System Filters are designed to be used in hospital environments by trained personnel.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for "flexicare". The logo is in blue and consists of a stylized letter "f" with three curved lines above it. Below the word "flexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller font size. The logo is clean and modern, and the blue color gives it a sense of trustworthiness and reliability.
# 510(k) Summary
# K191909
### 510(k) Sponsor, Contact Person and Date Summary Prepared:
Flexicare Medical Limited Cynon Valley Business Park Mountain Ash. CF45 4ER. United Kingdom
Joel Biddle Senior Compliance Engineer Telephone: 00 44 1443 474 647 Fax: 00 44 1443 474 222
Summary prepared on: March 16th, 2020. Device Name:
Trade Name: HepaShield Bacterial Viral Breathing System Filter
Common/Usual Name: Breathing circuit bacterial filter.
Classification Name: Breathing circuit bacterial filter: 21 CFR 868.5260
Product Codes: CAH (Breathing circuit bacterial filter)
#### Predicate Device:
Pall's Ultipor 25 Bacterial Viral Breathing System Filter (K013093)
#### Device Description:
Flexicare's HepaShield Bacterial Viral Breathing System Filters are designed to reduce the transmission of bacteria and viruses to and from a patient who requires aesthesia, artificial respiration or other types of assisted respiration
Flexicare's HepaShield Bacterial Viral Breathing System Filters consist of a top/bottom housing, pleated paper filter pack and a luer port cap. The Female luer-lock port allows connection of a luer-terminating monitoring line for monitoring of patient exhaled CO2 & is supplied with a tethered cap to seal port when not in use.
The patient side of the Flexicare's HepaShield Bacterial Viral Breathing System Filters feature a 22mm Male and a 15mm Female conical. The machine side features a 22mm Female and a 15mm Male conical.
All conical connectors comply with dimensions stated within BS EN ISO 5356-1 standard.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Flexicare. The logo is blue and features a stylized "F" followed by the word "lexicare." Below the word "lexicare" is the phrase "TOTAL QUALITY - TOTAL CARE." The logo is clean and professional, and the blue color gives it a sense of trustworthiness.
Flexicare's HepaShield Bacterial Viral Breathing System Filters are available in Adult sizes.
Flexicare's HepaShield Bacterial Viral Breathing System Filters are available in a both Sterile and non-sterile state, packaged in individually sealed polybags, or blister trays (if sterile).
#### Indications For Use:
Flexicare's HepaShield Bacterial Viral Breathing System Filters are intended to reduce the transmission of bacteria and viruses to/from a patient during anesthesia. For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and/or expired gases is desired.
Flexicare's HepaShield Bacterial Viral Breathing System Filters are single use devices for use on a single patient for up to 24hrs and are available in Adult size. Flexicare's HepaShield Bacterial Viral Breathing System Filters are designed to be used in hospital environments by trained personnel.
### Technoloqical Characteristic Comparison Table:
The table below shows the similarities and differences between the Flexicare's HepaShield Bacterial Viral Breathing System Filter and the predicate device manufactured by Pall.
| | Flexicare's HepaShield Breathing System<br>Filter | Pall's Ultipor 25 Breathing System Filter |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | K191909 | K013093 |
| Components | Filter housing top<br>Filter housing bottom<br>Filter Media pack<br>Tethered luer port cap<br>Outer shrink sleeve | Filter housing top<br>Filter housing bottom<br>Filter Media pack<br>Tethered luer port cap |
| Assembly Method | Ultrasonic welded housing | Ultrasonic welded housing |
| Target population | Adult | Adult |
| Emergency use | Yes | Yes |
| Environment of use | Hospital | Hospital |
| Patient use/Duration of use | Single use, disposable, <24hrs | Single use, disposable, <24hrs |
| Contraindications | - DO NOT place the HepaShield<br>Bacterial Viral Breathing System<br>Filter between a humidification<br>device or nebulizing device and the<br>patient.<br><br>- Single Use. Do not reuse.<br><br>- DO NOT attempt to decontaminate<br>this product in any way. This<br>includes rinsing, washing or<br>decontamination using gas, heat,<br>steam, or boiling water.<br><br>- When used in conjunction with a | - Do not install the filter in either<br>positions A or B in conjunction with<br>heater water-bath humidifiers or<br>nebulizers.<br><br>- Do not Reuse.<br><br>- Do not Soak, rinse, wash, sterilize or<br>treat with liquid disinfectants.<br><br>- Note: This Product is free of natural<br>rubber latex. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Flexicare. The logo is blue and features a stylized "f" with three curved lines above it. The word "flexicare" is written in a sans-serif font to the right of the "f". Below the word "flexicare" is the tagline "TOTAL QUALITY - TOTAL CARE".
| | humidified breathing system, the<br>HepaShield Bacterial Viral Breathing<br>System Filter is contraindicated for<br>use at the patient end. | |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Supplied sterile | Both Non-sterile & sterile variants | Non-sterile |
| Product labelling | HepaShield Bacterial Viral Breathing System<br>Filter | PALL Ultipor BREATHING SYSTEM FILTER |
| Indications for use | Flexicare's HepaShield Bacterial Viral<br>Breathing System Filters are intended to<br>reduce the transmission of bacteria and<br>viruses to/from a patient during anesthesia.<br>For use with ventilators, anesthesia<br>machines and open flow systems where<br>filtration of inspired and/or expired gases is<br>desired.<br>Flexicare's HepaShield Bacterial Viral<br>Breathing System Filters are single use<br>devices for use on a single patient for up to<br>24hrs and are available in Adult size.<br>Flexicare's HepaShield Bacterial Viral | The Pall Ultipor Anesthesia Breathing Circuit<br>System with a Breathing System Filter<br>("BSF") is intended for use in the<br>administration of medical gases during<br>anesthesia. The circuit connects the<br>anesthesia gas machine to the patient, by<br>means of an oronasal facemask or by a<br>connection to an artificial airway, such as an<br>endotracheal tube or laryngeal mask. The<br>Pall Ultipor 25 BSF minimizes viral and<br>bacterial contamination of the inspiratory<br>and expiratory limbs of the circuit with a<br>minimum efficiency of 99.999%. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for "flexicare". The logo is in blue and consists of a stylized "f" followed by the word "lexicare" in a sans-serif font. Below the word "lexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller sans-serif font. The logo is clean and modern, and the blue color gives it a sense of trustworthiness and reliability.
| | Breathing System Filters are designed to be<br>used in hospital environments by trained<br>personnel. | Source: K013093 510(k) summary |
|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Intended use | Instruction leaflet | Instruction leaflet |
| Patient connection | ET tube, Laryngeal Mask Airway, Catheter<br>mount, Breathing circuit | ET tube, Laryngeal Mask Airway, Catheter<br>mount, Breathing circuit |
| Standard 22/15mm<br>connections in compliance<br>with ISO 5356-1 | Yes | Yes |
| Luer port for gas sampling<br>in compliance with ISO<br>80369-7 | Yes | Yes |
| Configurations | Straight with luer port | Straight with luer port |
| Filtration Method | Mechanical | Mechanical |
| Placement within circuit | Patient side<br>Machine side | Patient side<br>Machine side |
| Weight (g) | 42g | 26g |
| Internal Volume/dead<br>space as per ISO 9360-1 | 47ml | 35ml |
| Bacterial/ Viral Filtration<br>efficiency/ filter integrity<br>(Fresh) | BFE - 99.99999%<br>VFE - 99.9999% | BFE - 99.999% |
| Salt Method filtration<br>efficiency | 99.89% | Not stated |
| Tidal Volume range (ml) | 141ml-800ml | Not stated |
| Pressure Drop/ Flow<br>resistance per BS EN ISO<br>23328-2/ BS EN ISO 9360-1 | 1.70 cmH2O @ 30LPM<br>3.92 cmH2O @ 60LPM<br>6.56 cmH2O @ 90LPM | 1.43 cmH2O @ 30LPM<br>3.10 cmH2O @ 60LPM<br>4.96 cmH2O @ 90LPM |
| Leakage per BS EN ISO<br>23328-2/ BS EN ISO 9360-1 | <2 ml/min | <2 ml/min |
| Compliance per BS EN ISO<br>9360-1 | 0.057 ml/cmH2O | 0.029 ml/cmH2O |
| Reuse, Cleaning &<br>Disinfection | N/A - Single use (max 24hrs) | N/A - Single use (max 24hrs) |
| Shelf Life | 5 years | 5 years |
| Packaging | Polybag | Polybag |
| Temp/humidity req's | Not stated | Not stated |
| Standards met | ISO 5356-1<br>ISO 10993-1<br>BS EN ISO 23328-1<br>BS EN ISO 80369-7 | 510(K) Summary does not specify |
| Biocompatibility | ISO 10993 compliant<br>EN ISO 18562 compliant | Not stated |
Summary of Nonclinical Testing: Flexicare's HepaShield Bacterial Viral Breathing System Filter has been evaluated in accordance with standards listed in table:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Flexicare. The logo is blue and features a stylized "f" with three curved lines above it. To the right of the "f" is the word "lexicare" in a sans-serif font. Below the word "lexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in a smaller font.
| Test | Standard / Pre Determined Acceptance<br>Criteria | Results |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Visual Inspection | Pre-determined Acceptance Criteria* | Pass |
| Compliance testing | BS EN ISO 9360: 2009 | Methodology only, no pass<br>criteria - Comparable<br>performance outcome between<br>Flexicare's new devices and<br>the predicate devices |
| Pressure drop | BS EN ISO 23328-2 (Refers to: BS EN ISO<br>9360: 2009) - Methodology only, no pass criteria | Methodology only, no pass<br>criteria - Comparable |
| Gas Leakage | - for comparison only.<br>HME aspects not applicable. | performance outcome between<br>Flexicare's new devices and<br>the predicate devices |
| Conical Connector compliance | BS EN ISO 5356-1 2004 | Pass |
| Leak testing | | Pass |
| Drop testing | | Pass |
| Cytotoxicity, Irritation, Sensitization,<br>Systemic toxicity, Material Mediated | BS EN ISO 10993-10:2010<br>BS EN ISO 10993-5:2009 | Pass |
| Pyrogenicity, | BS EN ISO 10993-11:2009 | Pass |
| Bacterial Endotoxin | (USP) guidelines <161> and <85>.<br>ANSI/AAMI ST72:2011, | Pass |
| Particulate emissions (gas<br>pathway) | EN ISO 18562-2:12017 | Pass |
| VOC emissions (gas pathway) | EN ISO 18562-3:2017 | Pass |
| Gauging tests on luer | | Pass |
| Liquid leakage from luer | | Pass |
| Air leakage from luer | | Pass |
| Luer separation force | BS EN ISO 80369-7:2016 | Pass |
| Luer unscrewing torque | | Pass |
| Luer ease of assembly | | Pass |
| Luer resistance to overriding | | Pass |
| Luer testing for stress cracking | | Pass |
| Shelf life testing | Standards included within this table | Pass |
| Filter integrity | Based on ASTM F2101 (Nelson Labs Protocol) | Methodology only, no pass<br>criteria within standard. |
| | BS EN ISO 23328-1 (2008) | Methodology only, no pass<br>criteria within standard. |
| Pressure Drop | BS EN ISO 23328-2 (2009) | Methodology only, no pass<br>criteria within standard |
| Housing Leakage | BS EN ISO 23328-2(2009) | Methodology only, no pass<br>criteria within standard |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for "flexicare" in blue. The logo features a stylized "f" with three curved lines above the word "flexicare". Below the word "flexicare" is the text "TOTAL QUALITY - TOTAL CARE".
#### * Pre-determined Acceptance Criteria - Visual inspection
Criteria: Product packaging must remain sealed & undamaged. All components of devices must be free of flash, cracking, warping, discoloration or contamination.
Methodology: Visual inspection under normal or corrected vision of device packaging (inspection of panels, seams and welds for unsealed areas/openings, pint quality, tares) and assembled device (inspection of device for cracking, warping, discolouration, cosmetic defects, flash and contamination within material).
Flexicare's HepaShield Bacterial Viral Breathing System Filter passed the performance testing when tested against methods and criteria from both pre-determined acceptance criteria methods and relevant FDA Recognized standards.
The results of this testing show that Flexicare's HepaShield Bacterial Viral Breathing System Filter passes all performance tests and perform at least as well as the marketed predicate device.
#### Consensus Standards
The recognized consensus standard for devices classified through FDA product code CAH are:
- ISO 5356-1: 2004 Anaesthetic And Respiratory Equipment Conical Connectors: ● Part 1: Cones And Sockets.
- EN ISO 18652: 2017 Biocompatibility Evaluation Of Breathing Gas Pathways In ● Healthcare Applications
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Flexicare. The logo is blue and features a stylized "F" with three curved lines. The word "Flexicare" is written in a sans-serif font to the right of the "F". Below the word "Flexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in a smaller font.
# Conclusion:
The conclusions drawn from the nonclinical tests demonstrates that the Flexicare HepaShield Bacterial Viral Breathing System Filter device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, Pall's Ultipor 25 Bacterial Viral Breathing System Filter (K013093). .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.