K191896 · Belimed AG · FLE · Oct 31, 2019 · General Hospital
Device Facts
Record ID
K191896
Device Name
Belimed Steam Sterilizer MST-H TOP 5000
Applicant
Belimed AG
Product Code
FLE · General Hospital
Decision Date
Oct 31, 2019
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6880
Device Class
Class 2
Indications for Use
The Belimed Steam Sterilizers MST-H TOP 5000, models 9-6-9 HS1, 9-6-9 HS2, 9-6-12 HS1, 9-6-12 HS2, 9-6-15 HS1, 9-6-15 HS2, 9-6-18 HS1, 9-6-18 HS2, GR 9-6-9 HS1, GR 9-6-9 HS2, GR 9-6-12 HS1, GR 9-6-12 HS2, GR 9-6-15 HS1, GR 9-6-15 HS2, GR 9-6-18 HS1 and GR 9-6-18 HS2 are designed for the sterilization of non-porous and porous heat and moisture stable materials used in healthcare facilities. The Belimed Steam Sterilizer MST-H TOP 5000 is equipped with the following factory programmed sterilization cycles and cycle values. This is shown in Table 1.
Device Story
The Belimed Steam Sterilizer MST-H TOP 5000 is a steam-based sterilization system for healthcare facilities. It processes non-porous and porous heat/moisture-stable materials, including instrument trays and fabric packs. The device operates via factory-programmed cycles (PreVac, Gravity, Liquid, Air Leak, Bowie-Dick) controlled by a PLC. Users load materials into the chamber (single or double door, carriage or ground loader versions). The system uses steam as the sterilizing agent. The HMI features a 5.7" color LCD. The device provides automated sterilization, which helps ensure the sterility of medical instruments, thereby supporting infection control and patient safety.
Clinical Evidence
No clinical data. Bench testing only. Sterilization efficacy validated per ANSI/AAMI ST8:2013 using biological indicators (SAL 10^-6) and physical temperature profiling. All tests passed.
Technological Characteristics
Chamber/piping: 316L stainless steel. Pressure: 39 psi. HMI: 5.7" color LCD. Control: PLC. Vacuum pump: 95-130 m3/h. Air filter: 0.3 microns. Standards: ANSI/AAMI ST8:2013, ASME Section VIII Div 1, IEC 61010-2-040, UL 61010-1.
Indications for Use
Indicated for sterilization of non-porous and porous heat and moisture stable materials in healthcare facilities.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 31, 2019
Belimed AG Marc Lüthi Head of R&D Sustaining Grienbachstrasse 11 CH-6300 Zug / Switzerland
Re: K191896
Trade/Device Name: Belimed Steam Sterilizer MST-H TOP 5000 Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE Dated: September 20, 2019 Received: September 25, 2019
Dear Marc Lüthi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K191896
Device Name
Belimed Steam Sterilizer MST-H TOP 5000
#### Indications for Use (Describe)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
The Belimed Steam Sterilizer MST-H TOP 5000 models 9-6-9 HS1, 9-6-12 HS1, 9-6-12 HS1, 9-6-15 HS1, 9-6-15 HS2, 9-6-18 HS1, 9-6-18 HS2, GR 9-6-9 HS1, GR 9-6-12 HS1, GR 9-6-12 HS1, GR 9-6-15 HS1, GR 9-6-15 HS2, GR 9-6-18 HS1, and GR 9-6-18 HS2 are designed for the sterilization of non-porous and porous heat and moisture stable materials used in healthcare facilities.
The Belimed Steam Sterilizer MST-H TOP 5000 is equipped with the following factory programmed sterilization cycles and cycle values. This is shown in Table 1.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Z Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Table 1: Factory programmed sterilization cycles and cycle values.
.
:
-
.
:
:
:
| CYCLES | STERILIZE<br>TEMP | STERI<br>LIZE<br>TIME | DRY<br>TIME | RECOMMENDED LOAD |
|-------------------------------------|-------------------|-----------------------|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PreVac<br>270F 4S /<br>30Dry | 270°F | 4<br>minutes | 30<br>minutes | Double-wrapped instrument trays.<br>Maximum weight of 25 Ib each for the following<br>models: 9-6-12 HS2, 9-6-15 HS1, GR 9-6-15 HS2 and<br>GR 9-6-18 HS2.<br>Maximum weight of 17 Ib each for the following<br>models: 9-6-9 HS1, 9-6-9 HS2, 9-6-12 HS1, 9-6-15<br>HS2, 9-6-18 HS1, 9-6-18 HS2, GR 9-6-9 HS1, GR 9-<br>6-9 HS2, GR 9-6-12 HS1, GR 9-6-15 HS1 and GR 9-<br>6-18 HS1.<br>Fabric Packs. |
| PreVac<br>270F 4S /<br>5Dry | 270°F | 4<br>minutes | 5<br>minutes | Fabric Packs. |
| Liquid 250F | 250°F | 45<br>minutes | | Max 1000 ml of Liquid per bottle. Use only vented<br>bottles, type I, borosilicate glass<br>Cycle disabled |
| Express<br>270F | 270°F | 4<br>minutes | 3<br>minutes | Single wrapped instrument tray with non porous single<br>instrument |
| Immediate<br>Use 270F<br>Single Ins | 270°F | 4<br>minutes | 3<br>minutes | Unwrapped instrument tray with non porous single<br>instrument |
| Immediate<br>Use 270F<br>Multi Ins | 270°F | 10<br>minutes | 1<br>minute | Unwrapped instrument tray with non porous multiple<br>instruments.<br>Maximum weight of 25 lb each for the following<br>models: 9-6-12 HS2, 9-6-15 HS1, and GR 9-6-18<br>HS2.<br>Maximum weight of 17 lbs. each for the following<br>models: 9-6-9 HS1, 9-6-9 HS2, 9-6-12HS1, 9-6-15<br>HS2, 9-6-18 HS1, 9-6-18 HS2, GR 9-6-9 HS1, GR 9-<br>6-9 HS2, GR 9-6-12 HS1, GR 9-6-12 HS2, GR 9-6-15<br>HS1, GR 9-6-15 HS2, GR 9-6-18 HS1. Maximum of<br>4 instrument trays. |
.
:
.
:
.
{4}------------------------------------------------
# 510(k) Summary
| K191896 | |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | October 25, 2019 |
| Owner's Name / Address: | Belimed AG<br>Grienbachstrasse 11<br>CH-6300 Zug / Switzerland |
| Preparer's Name / Address: | Belimed AG<br>Grienbachstrasse 11<br>CH-6300 Zug / Switzerland |
| Contact Person: | Marc Lüthi<br>Email: marc.luethi@belimed.com<br>Phone: +41 71 644 87 80<br>Fax: +41 41 449 78 89 |
| Trade Name: | Belimed Steam Sterilizer MST-H TOP 5000 |
| Models: | 9-6-9 HS1, 9-6-9 HS2, 9-6-12 HS1,<br>9-6-12 HS2, 9-6-15 HS1, 9-6-15 HS2,<br>9-6-18 HS1, 9-6-18 HS2, GR 9-6-9 HS1,<br>GR 9-6-9 HS2, GR 9-6-12 HS1,<br>GR 9-6-12 HS2, GR 9-6-15 HS1,<br>GR 9-6-15 HS2, GR 9-6-18 HS1,<br>GR 9-6-18 HS2 |
| Classification: | Steam Sterilizer - Class II, as listed per<br>21 CFR 880.6880<br>Product Code: FLE |
| Predicate Device: | Belimed Steam Sterilizer MST-H TOP 5000<br>(K170228) |
{5}------------------------------------------------
### Intended Use:
The Belimed Steam Sterilizers MST-H TOP 5000, models 9-6-9 HS1, 9-6-9 HS2, 9-6-12 HS1, 9-6-12 HS2, 9-6-15 HS1, 9-6-15 HS2, 9-6-18 HS1, 9-6-18 HS2, GR 9-6-9 HS1, GR 9-6-9 HS2, GR 9-6-12 HS1, GR 9-6-12 HS2, GR 9-6-15 HS1, GR 9-6-15 HS2, GR 9-6-18 HS1 and GR 9-6-18 HS2 are designed for the sterilization of non-porous and porous heat and moisture stable materials used in healthcare facilities.
The Belimed Steam Sterilizer MST-H TOP 5000 is equipped with the following factory programmed sterilization cycles and cycle values. This is shown in Table 1.
| CYCLES | STERILIZE<br>TEMP | STERI<br>LIZE<br>TIME | DRY<br>TIME | RECOMMENDED LOAD |
|-------------------------------------|-------------------|-----------------------|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PreVac<br>270F 4S /<br>30Dry | 270°F | 4<br>minutes | 30<br>minutes | Double-wrapped instrument trays.<br>Maximum weight of 25 lb each for the following<br>models: 9-6-12 HS2, 9-6-15 HS1, GR 9-6-15 HS2 and<br>GR 9-6-18 HS2.<br>Maximum weight of 17 lb each for the following<br>models: 9-6-9 HS1, 9-6-9 HS2, 9-6-12 HS1, 9-6-15<br>HS2, 9-6-18 HS1, 9-6-18 HS2, GR 9-6-9 HS1, GR 9-<br>6-9 HS2, GR 9-6-12 HS1, GR 9-6-15 HS1 and GR 9-<br>6-18 HS1.<br>Fabric Packs. |
| PreVac<br>270F 4S /<br>5Dry | 270°F | 4<br>minutes | 5<br>minutes | Fabric Packs. |
| Liquid 250F | 250°F | 45<br>minutes | - | Max 1000 ml of Liquid per bottle. Use only vented<br>bottles, type I, borosilicate glass<br>Cycle disabled |
| Express<br>270F | 270°F | 4<br>minutes | 3<br>minutes | Single wrapped instrument tray with non porous single<br>instrument |
| Immediate<br>Use 270F<br>Single Ins | 270°F | 4<br>minutes | 3<br>minutes | Unwrapped instrument tray with non porous single<br>instrument |
| Immediate<br>Use 270F<br>Multi Ins | 270°F | 10<br>minutes | 1<br>minute | Unwrapped instrument tray with non porous multiple<br>instruments.<br>Maximum weight of 25 lb each for the following<br>models: 9-6-12 HS2, 9-6-15 HS1, and GR 9-6-18<br>HS2.<br>Maximum weight of 17 lb each for the following<br>models: 9-6-9 HS1, 9-6-9 HS2, 9-6-12HS1, 9-6-15<br>HS2, 9-6-18 HS1, 9-6-18 HS2, GR 9-6-9 HS1, GR 9-<br>6-9 HS2, GR 9-6-12 HS1, GR 9-6-12 HS2, GR 9-6-15<br>HS1, GR 9-6-15 HS2, GR 9-6-18 HS1. Maximum of<br>4 instrument trays. |
Table 1: Factory programmed sterilization cycles and cycle values.
{6}------------------------------------------------
### Description of Device:
The Belimed Steam Sterilizers MST-H TOP 5000, models 9-6-9, 9-6-12, 9-6-15, and 9-6-18 are available in a single door (HS1) or double door (HS2) version.
The sterilizers are available in a carriage loader and a ground loader (GR) version.
The sterilizers are equipped with Prevacuum, Gravity, Liquid, Air Leak Test and BowieDick Test cycles.
The Belimed Steam Sterilizers TOP 5000 are equipped with the following predefined cycles
(Table 2):
{7}------------------------------------------------
| No | CYCLE | PRE-<br>TREATM<br>ENT | STERILIZE<br>TEMP | STERI<br>LIZE<br>TIME | DRY<br>TIME | RECOMMENDED LOAD |
|----|-------------------------------|--------------------------------------------|-------------------|-----------------------|---------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | PreVac<br>270F 4S /<br>30Dry | 4 pulses | 270°F | 4<br>minutes | 30<br>minutes | Double-wrapped instrument trays.<br>Maximum weight of 25 lb each for the<br>following models: 9-6-12 HS2, 9-6-15<br>HS1, GR 9-6-15 HS2 and GR 9-6-18 HS2.<br>Maximum weight of 17 lb each for the<br>following models: 9-6-9 HS1, 9-6-9 HS2,<br>9-6-12 HS1, 9-6-15 HS2, 9-6-18 HS1, 9-6-<br>18 HS2, GR 9-6-9 HS1, GR 9-6-9 HS2,<br>GR 9-6-12 HS1, GR 9-6-15 HS1 and GR<br>9-6-18 HS1.<br>Fabric Packs. |
| 2 | PreVac<br>270F 4S /<br>5Dry | 4 pulses | 270°F | 4<br>minutes | 5<br>minutes | Fabric Packs. |
| 3 | Flash<br>270F -<br>Single Ins | Purge | 270°F | 3<br>minutes | 1 minute | Unwrapped instrument tray with a single<br>instrument |
| 4 | Flash<br>270F -<br>Multi Ins | Purge | 270°F | 10<br>minutes | 1 minute | Unwrapped instrument tray with non<br>porous multiple instruments.<br>Maximum weight of 25 Ib each for the<br>following models: 9-6-12 HS2, 9-6-15<br>HS1, and GR 9-6-18 HS2.<br>Maximum weight of 17 Ib each for the<br>following models: 9-6-9 HS1, 9-6-9 HS2,<br>9-6-12HS1, 9-6-15 HS2, 9-6-18 HS1, 9-6-<br>18 HS2, GR 9-6-9 HS1, GR 9-6-9 HS2,<br>GR 9-6-12 HS1, GR 9-6-12 HS2, GR 9-6-<br>15 HS1, GR 9-6-15 HS2, GR 9-6-18 HS1.<br>Maximum of 4 instrument trays. |
| 5 | Express<br>270F | 2 pulses | 270°F | 4<br>minutes | 3<br>minutes | Single wrapped instrument tray with non<br>porous single instrument |
| 6 | Bowie-<br>Dick Test | 4 pulses | 273°F | 3.5<br>minutes | 1 minute | One DART or Bowie-Dick-Test-Pack |
| 7 | Leak Test | Vacuum:<br>65 mbar<br>Test time:<br>15 min | - | - | - | Empty chamber |
| 8 | Warm up<br>& Leak<br>Test | Vacuum:<br>65 mbar<br>Test time:<br>15 min | 270°F | 4<br>minutes | 3<br>minutes | Empty chamber |
| 9 | - | - | - | - | - | - |
| 10 | Liquid<br>250F | Purge | 250°F | 45<br>minutes | - | Max 1000 ml of Liquid per bottle. Use only<br>vented bottles, type I, borosilicate glass<br>Cycle disabled |
### Table 2: Factory programmed Sterilization cycles
Notes on Table 2:
- 1. Factory set sterilize time is recommended.
- 2. Factory set dry time.
- 3. Recommended load: Refer to table 3 and 4.
- 4. Fabric load should be preconditioned between 68°F and 75°F and at a relative humidity of at least 35% to 60% for at least 2 hours.
- 5. The liquid cycle is for non-patient contact use only.
{8}------------------------------------------------
The Belimed Steam Sterilizers MST-H TOP 5000 are offered in the following versions:
| Model | Configuration | Chamber Size (H x W x D) (mm) | Overall Size (H x W x D) (mm) |
|---------------|---------------|-------------------------------|-------------------------------|
| Single Door | | | |
| Double Door | | | |
| 9-6-9 HS1 | 1 door | 1080 x 660 x 1040 | 2000 x 1700 x 1280 |
| GR 9-6-9 HS1 | | 1230 x 660 x 1100 | 2000 x 1900 x 1600 |
| 9-6-9 HS2 | 2 door | 1080 x 660 x 1040 | 2000 x 1700 x 1300 |
| GR 9-6-9 HS2 | | 1230 x 660 x 1100 | 2000 x 1900 x 1600 |
| 9-6-12 HS1 | 1 door | 1080 x 660 x 1398 | 1970 x 1700 x 1714 |
| GR 9-6-12 HS1 | | 1230 x 660 x 1400 | 2000 x 1900 x 1900 |
| 9-6-12 HS2 | 2 door | 1080 x 660 x 1398 | 1970 x 1700 x 1714 |
| GR 9-6-12 HS2 | | 1230 x 660 x 1400 | 2000 x 1900 x 1900 |
| 9-6-15 HS1 | 1 door | 1080 x 660 x 1700 | 1970 x 1700 x 2022 |
| GR 9-6-15 HS1 | | 1230 x 660 x 1706 | 2000 x 1900 x 2200 |
| 9-6-15 HS2 | 2 door | 1080 x 660 x 1700 | 1970 x 1700 x 2022 |
| GR 9-6-15 HS2 | | 1230 x 660 x 1706 | 2000 x 1900 x 2200 |
| 9-6-18 HS1 | 1 door | 1080 x 660 x 2014 | 1970 x 1700 x 2330 |
| GR 9-6-18 HS1 | | 1230 x 660 x 2000 | 2000 x 1900 x 2500 |
| 9-6-18 HS2 | 2 door | 1080 x 660 x 2014 | 1970 x 1700 x 2330 |
| GR 9-6-18 HS2 | | 1230 x 660 x 2000 | 2000 x 1900 x 2500 |
Table 3. Versions and dimensions.
{9}------------------------------------------------
The following tables show the recommended loads for the Belimed Steam Sterilizers MST-H TOP 5000:
| Model<br>single door<br>double door | Wrapped<br>instrument<br>trays, max. 17<br>lb each | Fabric Packs<br>11"*x11"x9"<br>max.6.6 lb each | Fabric Packs<br>23"x11"x11" max.<br>17 lb each |
|-------------------------------------|----------------------------------------------------|------------------------------------------------|------------------------------------------------|
| 9-6-9 HS1 | 9 | 18 | 9 |
| 9-6-9 HS2 | | | |
| GR 9-6-9 HS1 | | | |
| GR 9-6-9 HS2 | | | |
| 9-6-12 HS1 | 12 | 30 | 12 |
| GR 9-6-12HS1 | | | |
| GR 9-6-12HS2 | | | |
| 9-6-15 HS2 | 15 | 36 | 15 |
| GR 9-6-15 HS1 | | | |
| 9-6-18 HS1 | 18 | 42 | 18 |
| 9-6-18 HS2 | | | |
| GR 9-6-18 HS1 | | | |
Table 4: Recommended loads by sterilizer for 17 lb. instrument trays and fabric packs.
Table 5: Recommended loads by sterilizer for 25 lb. instrument trays and fabric packs.
| Model<br>single door<br>double door | Wrapped<br>instrument<br>trays, max. 25<br>lb each | Fabric Packs<br>11*x11"x9"<br>max.6.6 lb each | Fabric Packs<br>23"x11"x11" max.<br>17 lb each |
|-------------------------------------|----------------------------------------------------|-----------------------------------------------|------------------------------------------------|
| 9-6-12 HS2 | 12 | 30 | 12 |
| 9-6-15 HS1 | 15 | 36 | 15 |
| GR 9-6-15 HS2 | | | |
| GR 9-6-18 HS2 | 18 | 42 | 18 |
{10}------------------------------------------------
| Model | Volume of Liquid in One<br>Container | Number of Containers |
|---------------------|--------------------------------------|----------------------|
| 9-6-9<br>GR 9-6-9 | 1000 mL | 126 |
| 9-6-12<br>GR 9-6-12 | 1000 mL | 168 |
| 9-6-15<br>GR 9-6-15 | 1000 mL | 210 |
| 9-6-18<br>GR 9-6-18 | 1000 mL | 252 |
Table 6. Recommended loads for liquid cycle 250F.
The Belimed Steam Sterilizer MST-H TOP 5000 is designed for sterilization of non-porous and porous heat and moisture stable materials used in healthcare facilities.
The Belimed Steam Sterilizer MST-H TOP 5000 is offered in the following medium-size configurations: 26"x 42.5"x 41" (660 mm x 1080 mm x 1040 mm) Single Door, Prevacuum 26"x 42.5"x 41" (660 mm x 1080 mm x 1040 mm) Double Door, Prevacuum 26"x 42.5"x 55" (660 mm x 1080 mm x 1400 mm) Single Door, Prevacuum 26"x 42.5"x 55" (660 mm x 1080 mm x 1400 mm) Double Door, Prevacuum 26"x 42.5"x 67" (660 mm x 1080 mm x 1700 mm) Single Door, Prevacuum 26"x 42.5"x 67" (660 mm x 1080 mm x 1700 mm) Double Door, Prevacuum 26"x 42.5"x 79" (660 mm x 1080 mm x 2000 mm) Single Door, Prevacuum 26"x 42.5"x 79" (660 mm x 1080 mm x 2000 mm) Double Door, Prevacuum 26"x 48.5"x 43" (660 mm x 1230 mm x 1100 mm) Single Door. Floor Flush Design. Prevacuum 26"x 48.5"x 43" (660 mm x 1230 mm x 1100 mm) Double Door, Floor Flush Design, Prevacuum 26"x 48.5"x 55" (660 mm x 1230 mm x 1400 mm) Single Door, Floor Flush Design, Prevacuum 26"x 48.5"x 55" (660 mm x 1230 mm x 1400 mm) Double Door, Floor Flush Design, Prevacuum 26"x 48.5"x 67" (660 mm x 1230 mm x 1700 mm) Single Door, Floor Flush Design, Prevacuum 26"x 48.5"x 67" (660 mm x 1230 mm x 1700 mm) Double Door, Floor Flush Design, Prevacuum 26"x 48.5"x 79" (660 mm x 1230 mm x 2000 mm) Single Door, Floor Flush Design, Prevacuum 26"x 48.5"x 79" (660 mm x 1230 mm x 2000 mm) Double Door, Floor Flush Design, Prevacuum
{11}------------------------------------------------
# Technological Characteristics Table:
| Shown below is a comparison of the subject device versus the predicate device: | | | |
|--------------------------------------------------------------------------------|--|--|--|
| | | | |
| General Sterilizer<br>Features | Belimed Steam Sterilizer<br>MST-H TOP 5000 (K170228) | Belimed Steam Sterilizer<br>MST-H TOP 5000 (K191896) | Comparison |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Intended Use | Terminal Sterilization of non-<br>porous and porous heat and<br>moisture-stable materials<br>used in healthcare facilities | Terminal Sterilization of non-<br>porous and porous heat and<br>moisture-stable materials<br>used in healthcare facilities | Same |
| Operating<br>principle | The sterilizing agent is steam. | The sterilizing agent is steam. | Same |
| Built according to<br>standard | ANSI/AAMI ST8:2013<br>Hospital Steam Sterilizers | ANSI/AAMI ST8:2013<br>Hospital Steam Sterilizers | Same |
| Chamber sizes<br>Model: | (H x W x D) in mm: | (H x W x D) in mm: | Same |
| 9-6-9<br>9-6-12<br>GR9-6-12 | 1080x660x1040<br>1080x660x1400<br>1230x660x1400 | 1080x660x1040<br>1080x660x1400<br>1230x660x1400 | |
| 9-6-15<br>GR9-6-15 | 1080x660x1700<br>1230x660x1700 | 1080x660x1700<br>1230x660x1700 | |
| 9-6-18<br>GR9-6-18 | 1080x660x2000<br>1230x660x2000 | 1080x660x2000<br>1230x660x2000 | |
| Sterilization<br>cycles | PreVac 270 °F | PreVac 270 °F | Same |
| | Gravity 270 °F | Gravity 270 °F | |
| | Liquid 250 °F | Liquid 250 °F | |
| | Bowie Dick Test | Bowie Dick Test | |
| | Air Leak Test | Air Leak Test | |
| Max. instrument<br>tray weight | Double-wrapped instrument<br>trays.<br>Maximum weight of 25 lb each<br>for the following models: 9-6-12<br>HS2, 9-6-15 HS1 and<br>GR 9-6-18 HS2.<br>Maximum weight of 17 lb each<br>for the following models: 9-6-9<br>HS1, 9-6-9 HS2, 9-6-12 HS1, 9-<br>6-15 HS2, 9-6-18 HS1, 9-6-18<br>HS2, GR 9-6-9 HS1, GR 9-6-9<br>HS2, GR 9-6-12 HS1, GR 9-6-<br>15 HS1, GR 9-6-15 HS2 and<br>GR 9-6-18 HS1.<br>Fabric Packs | Double-wrapped instrument<br>trays.<br>Maximum weight of 25 lb each<br>for the following models: 9-6-12<br>HS2, 9-6-15 HS1, GR 9-6-15<br>HS2 and GR 9-6-18 HS2.<br>Maximum weight of 17 lb each<br>for the following models: 9-6-9<br>HS1, 9-6-9 HS2, 9-6-12 HS1, 9-<br>6-15 HS2, 9-6-18 HS1, 9-6-18<br>HS2, GR 9-6-9 HS1, GR 9-6-9<br>HS2, GR 9-6-12 HS1, GR 9-6-<br>15 HS1 and GR 9-6-18 HS1.<br>Fabric Packs | Similar |
| General Sterilizer<br>Features | Belimed Steam Sterilizer<br>MST-H TOP 5000 (K170228) | Belimed Steam Sterilizer<br>MST-H TOP 5000 (K191896) | Comparison |
| Load capacity<br>instrument trays | | | |
| Model: | | | |
| 9-6-9 / 9-6-12 /<br>GR9-6-12 | 9 | 9 | same |
| 9-6-15 / GR9-6-15 | 12 | 12 | |
| 9-6-18 / GR9-6-18 | 15 | 15 | |
| | 18 | 18 | |
| Chamber<br>material | Stainless steel 316L | Stainless steel 316L | same |
| Chamber design<br>pressure | 39 psi | 39 psi | same |
| Door construction | Automated horizontal sliding<br>door(s) | Automated horizontal sliding<br>door(s) | same |
| Safety (pressure<br>relief) valves | ASME approved | ASME approved | same |
| Piping | Stainless steel 316 L | Stainless steel 316 L | same |
| Electrical | 3 phase 208V 60Hz | 3 phase 208V 60Hz | same |
| HMI Technology | LCD color / 5.7" | LCD color / 5.7" | same |
| Control<br>Technology | PLC | PLC | Same |
| Printer<br>Technology | Matrix Dot Technology | Matrix Dot Technology | Same |
| Vacuum pump | 9-6-9 / GR 9-6-9: 95m3/h<br>9-6-12 / GR 9-6-12: 95m3/h<br>9-6-15 / GR 9-6-15: 130m3/h<br>9-6-18 / GR 9-6-18: 130m3/h | 9-6-9 / GR 9-6-9: 95m3/h<br>9-6-12 / GR 9-6-12: 95m3/h<br>9-6-15 / GR 9-6-15: 130m3/h<br>9-6-18 / GR 9-6-18: 130m3/h | Same |
| Air filter | 0.3 microns | 0.3 microns | Same |
{12}------------------------------------------------
Based on the comparison above the subject device and the predicate device demonstrate many of the same technological characteristics.
{13}------------------------------------------------
## Clinical Data:
No clinical data is required for this device classification submission.
## Summary of Performance Testing:
Show below is a summary of the non-clinical testing that was performed with this device.
| Test Performed | Device<br>Describtion/<br>Sample Size | Test Method/<br>Applicable<br>Standards | Acceptance<br>Criteria | Results |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Biological Performance Tests: | | | | |
| Sterilization Efficacy<br>Validation<br><br>PreVac 132°C (270°F)<br>for 4 minutes Exposure<br>and 30 min. Dry 25 lbs<br>wrapped instrument<br>PCD<br><br>1: Full cycle / 3 runs | MST-H<br>TOP 5000<br>GR 9-6-15<br>HS2 | ANSI AAMI<br>ST8:2013<br><br>5.5.4. Biological<br>performance with<br>wrapped instrument<br>PCD<br><br>BI's used in testing<br>comply with<br>ANSI/AAMI/ISO<br>11138-3 | Sterility assurance level<br>(SAL) 10-6<br><br>time-at-temperature<br>sufficient to produce an<br>F0 of at least 12 minutes<br>by half cycle, moisture<br>retention of less than 20%<br>increase<br><br>ANSI/AAMI ST8:2013<br>5.5.4.5<br><br>ANSI/AAMI/ISO 14161 | Pass |
| Sterilization Efficacy<br>Validation<br><br>PreVac 132°C (270°F)<br>for 4 minutes Exposure<br>25 lbs wrapped<br>instrument PCD<br><br>1:<br>Full cycle / 3 runs<br>2:<br>Half cycle 2 minutes /<br>3 runs | MST-H<br>TOP 5000<br>GR 9-6-15<br>HS2 | ANSI AAMI<br>ST8:2013<br><br>5.5.4. Biological<br>performance with<br>wrapped instrument<br>PCD<br><br>BI's used in testing<br>comply with<br>ANSI/AAMI/ISO<br>11138-3 | Sterility assurance level<br>(SAL) 10-6<br><br>time-at-temperature<br>sufficient to produce an<br>F0 of at least 12 minutes<br>by half cycle<br><br>ANSI/AAMI ST8:2013<br>5.5.4.5<br><br>ANSI/AAMI/ISO 14161 | Pass |
| Physical Performance Test: | | | | |
| Chamber temperature<br>profile | MST-H<br>TOP 5000<br>GR 9-6-15<br>HS2 | ANSI AAMI<br>ST8:2013<br><br>5.4.2.5 with empty<br>chamber | +3°C (or +6°F) and<br>-0°C (or -0°F) of the<br>selected sterilization<br>exposure temperature | Pass |
{14}------------------------------------------------
### Standards Used:
Belimed's sterilizers, including the Belimed Steam Sterilizer MST-H TOP 5000 series have been designed and constructed to meet the safety and performance requirements of various national safety codes and standards. The Belimed Steam Sterilizer MST-H TOP 5000 complies with the following standards:
- 1. ANSI/AAMI ST8:2013
- 2. ANSI/AAMI ST79:2017
- 3. IEC EN 62304:2006
- 4. EN ISO 14971:2012
- 5. ISO 13485:2016 Medical Devices - Quality management systems. Requirements for regulatory purposes
- 6. American Society of Mechanical Engineers (ASME), Section VIII, Division 1 for unfired pressure vessels Ed. 2017.
- UL 61010-1:2004 Safety requirements for electrical equipment for measurement, control 7. and laboratory use -Part 1: General requirements.
- 8. IEC 61010-2-040:2005 Safety requirements for electrical equipment for measurement, control and laboratory use -Part 2: Particular requirements for sterilizers and washerdisinfectors used to treat medical materials.
- 9. IEC 61326-1 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
### Conclusion:
The nonclinical tests in this premarket notification demonstrate that this device is as safe, as effective, and performs as well or better than the predicate device series (K170228).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.