The equipment is intended to be used for the fluoroscopy/radiography diagnosis in hospital. The equipment must only be operated by qualified personnel, such as radiography technicians or those with equivalent qualifications. The equipment is used for total patient population. The equipment is NOT intended to be used for Mammography screening. The equipment is NOT intended to be used for interventional procedure. The equipment is used for radiographic, fluoroscopic, angiographic and pediatric examinations. Stored images in the equipment can be used for re-monitoring, image processing, storing to optical media (CD/DVD), and sending to DICOM server.
Device Story
Universal X-ray RF system for radiographic, fluoroscopic, and angiographic imaging; utilizes flat panel detector (FPD) to capture X-ray signals; high-frequency inverter (50 kHz) generates high-voltage; system includes X-ray diagnostic table with adjustable height (65-110 cm) and wide examination area; operator controls via front panel or 10.4-inch LCD touch panel; features include automatic BH-filter selection, pulsed fluoroscopy wave-tail cut-off, and anti-scatter grid; used in hospitals for diagnostic imaging; operated by radiography technicians; output includes diagnostic images stored in DICOM format; supports DICOM network connectivity; benefits include high-quality imaging, low-dose exposure, and efficient workflow for general R/F and pediatric exams.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and compliance with FDA-recognized consensus standards (IEC 60601-1, 60601-1-3, 60601-2-54, etc.).
Indicated for fluoroscopy/radiography diagnosis in hospital settings for total patient population, including pediatric patients. Not for mammography screening or interventional procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
SHIMADZU Corporation % Daniel Kamm, P.E. Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
August 5, 2019
## Re: K191877
Trade/Device Name: FLUOROspeed Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: JAA Dated: July 1, 2019 Received: July 15, 2019
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K191877
Device Name FLUOROspeed
Indications for Use (Describe)
The equipment is intended to be used for the fluoroscopy/radiography diagnosis in hospital.
The equipment must only be operated by qualified personnel, such as radiography technicians or those with equivalent qualifications.
The equipment is used for total patient population.
The equipment is NOT intended to be used for Mammography screening.
The equipment is NOT intended to be used for interventional procedure.
The equipment is used for radiographic, fluoroscopic, angiographic and pediatric examinations.
Stored images in the equipment can be used for re-monitoring, image processing, storing to optical media (CD/DVD), and sending to DICOM server.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary: 510(k) Number K191877
Image /page/3/Picture/1 description: The image shows the Shimadzu logo. The logo consists of a square with a white cross inside, followed by the word "SHIMADZU" in bold, black letters. The cross is centered within the square, and the wordmark is aligned to the right of the square.
Shimadzu Corporation 1, Nishinokyo-Kuwabaracho, Nakagyo-ku, Kyoto, 604-8511, Japan Registration Number: 8030233 Phone : +81-75-823-1305 Fax : +81-75-823-1377 Date Prepared: July 26, 2019 Contact: Toshio Kadowaki; Phone: +81-75-823-1920 E-mail: kadowaki@shimadzu.co.jp
## 1. Identification of the Device:
Trade/Device Names: FLUOROspeed Regulation Number: 21CFR892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Common/Usual Name: Fluoroscopic x-ray system
## 2. Equivalent legally marketed device: K190373
Trade/Device Name: SONIALVISION G4 Manufacturer: Shimadzu. Regulation Number: 21CFR892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Common/Usual Name: Fluoroscopic x-ray system REFERENCE/SIMILARLY CONFIGURED DEVICES: K173639 (Siemens) and K081091 (GE) Trade/Device Names: Luminos (Siemens), Precision (GE) Manufacturers: Siemens and GE. Regulation Numbers: 21CFR892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB (Siemens) JAA (GE) Common/Usual Name: Fluoroscopic x-ray system
- 3. Indications for Use (intended use): The equipment is intended to be used for the fluoroscopy/radiography diagnosis in hospital. The equipment must only be operated by qualified personnel, such as radiography technicians or those with equivalent qualifications. The equipment is used for total patient population. The equipment is NOT intended to be used for Mammography screening. The equipment is NOT intended to be used for interventional procedure. The equipment is used for radiographic, fluoroscopic, angiographic and pediatric examinations. Stored images in the equipment can be used for re-monitoring, image processing, storing to optical media (CD/DVD), and sending to DICOM server.
- 4. Description of the Device: The Shimadzu FLUOROspeed is a universal X-ray RF system offering radiographic, fluoroscopic, and angiographic modes. As compared to our recently cleared system the SONIALVISION G4, (K190373) the positions of the tube head and the digital x-ray receptor panel have
{4}------------------------------------------------
been swapped. This was done to respond to customer demand. This configuration is identical to products offered by Siemens (K173639) and GE Healthcare (K081091). The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a flat panel detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. FLUOROspeed may be used for emergency treatment on an outpatient basis, as well as for bedside examinations. Features:
• A large-field 17-inch FPD and digital image processing enable you to acquire high quality diagnostic images. A second FPD can be added as an option. This is the wireless AeroDR SYSTEM 2 1417HQ cleared by FDA in K151465.
• The equipment employs a high-frequency inverter system with maximum frequency of 50 kHz for high-voltage generation and ensures high efficiency by low-ripple output.
• The X-ray diagnostic table holds a wide examination area and allows radiography from head to foot without the need to move the patient.
• The tabletop can be raised or lowered in the height range from 65 cm to 110 cm to help the patient get on or off and allow the operator to take an unforced pose according to required examinations. • The operator can assist the patient while operating the X-ray diagnostic table on its front control
panel.
• The control panel is equipped with a 10.4-inch LCD touch panel, where the operator can view the patient information or change the fluoroscopy or radiography condition. The screen layout or display options can be switched as appropriate according to the purpose of examination. Customizable hard switches help you access to the frequently used functions with a single touch of a button.
• Low-dose exposure and high image quality are achieved at the same time by incorporating automatic selection of BH-filter (X-ray filter), wave-tail cut-off mechanism for pulsed fluoroscopy, detachable anti-scatter grid, virtual collimation, and single-mask collimator (option).
• A series of radiography conditions for examination can be registered in a program to repeat the same series of radiography.
• Image data can be stored for a prolonged period by saving it on DVD-R or CD-R disks in DICOM format.
• The equipment supports DICOM3.0 and allows smooth connection to a hospital network. DICOM Print, Media Storage, DICOM Storage, MWM/MPPS (option) and RDSR (option) are also supported
- 5. Safety and Effectiveness, comparison to predicate device. The results of bench and standards testing indicates that the new device is as safe and effective as the predicate device. Proper system operation is fully verified upon installation. Most of the components employed are identical to the predicate device.
## 6. Substantial Equivalence Chart: Below.
{5}------------------------------------------------
| | SONIALVISION G4 K190373 | FLUOROspeed |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use: | The equipment is intended to be used for the<br>fluoroscopy/radiography diagnosis in hospital.<br>The equipment must only be operated by<br>qualified personnel, such as radiography<br>technicians or those with equivalent<br>qualifications. The equipment is used for total<br>patient population. The equipment is NOT<br>intended to be used for Mammography<br>screening. The equipment is NOT intended to be<br>used for interventional procedure. The<br>equipment is used for radiographic, fluoroscopic,<br>angiographic and pediatric examinations. Stored<br>images in the equipment can be used for re-<br>monitoring, image processing, storing to optical<br>media (CD/DVD), and sending to DICOM server.<br>The Tomosynthesis option for the SONIALVISION<br>G4 is intended to generate tomographic images of<br>human anatomy including chest or extremities.<br>Tomosynthesis technique is used to produce a<br>specific cross-sectional plane of the body by<br>reconstruction of tomographic acquisition. | The equipment is intended to be used for the<br>fluoroscopy/radiography diagnosis in hospital.<br>The equipment must only be operated by<br>qualified personnel, such as radiography<br>technicians or those with equivalent<br>qualifications. The equipment is used for total<br>patient population. The equipment is NOT<br>intended to be used for Mammography<br>screening. The equipment is NOT intended to be<br>used for interventional procedure. The<br>equipment is used for radiographic,<br>fluoroscopic, angiographic and pediatric<br>examinations. Stored images in the equipment<br>can be used for re-monitoring, image<br>processing, storing to optical media (CD/DVD),<br>and sending to DICOM server.. THE SAME<br>EXCEPT: No Tomosynthesis. Tomosynthesis is<br>not possible because of the fixed position of<br>the tubehead. |
| Position of the X-<br>ray tube head | On top, aimed down at the patient resting on<br>the table | Underneath the table aimed up through the<br>table |
| Position of the<br>digital x-ray<br>receptor panel | In a slot in the table | Overhead where the tubehead used to be. |
| Digital X-Ray Receptor Panel (Previously Cleared in K171194) | | |
| Manufacturer | Canon Medical Equipment Group | SAME |
| Model | Canon AS-10, CXDI-401RF (K171194)<br>We call it the SFD-1717BF | SAME |
| Size | 43 x 43 CM | SAME |
| Pixel Size | 160 μm | SAME |
| Number of Pixels | 2,688 x 2,688 | SAME |
| Acquisition<br>Modes (Binning) | Up to 3 fps (1x1)<br>Up to 15 fps (1x1)<br>Up to 30 fps (2x2) | SAME |
| DQE @ 1 µGy in 0<br>lp/mm, RQA5 | 0.75 | SAME |
| Spatial Resolution<br>[MTF@2cycle/mm<br>, RQA5] | 0.28 | SAME |
| A/D Conversion | 16 Bit | SAME |
| Digital Radiography<br>Unit | DR-300 | SAME |
| | SONIALVISION G4 K190373 | FLUOROspeed |
| Rad/Fluoro Characteristics | | |
| General<br>Appearance | Image: SONIALVISION G4 K190373 | Image: FLUOROspeed |
| Generator | 80 kW | 80 kW SAME |
| Collimator | Type R-300 | SAME |
| Ceiling Tube<br>Support | CH-200 | SAME |
| Radiography (Unchanged from Predicate) | | |
| Tube Voltage | 40 to 150 kV | 40 to 150 kV SAME |
| Tube Current | 10 to 800 mA | 10 to 800 mA SAME |
| mAs | 0.5 to 800 mAs | 0.5 to 800 mAs SAME |
| Time | 0.001 to 10 sec | 0.001 to 10 sec SAME |
| Fluoroscopy (Unchanged from Predicate) | | |
| Tube Voltage | 50 to 125 kV | 50 to 125 kV SAME |
| Tube Current | Max. 20 mA | Max. 20 mA SAME |
| Time | Total time display: 99 min 59 sec.,<br>continuous fluoroscopy time: 10 min | Total time display: 99 min 59 sec.,<br>continuous fluoroscopy time: 10 min<br>SAME |
| Pulsed<br>Fluoroscopy | Max. 30 fps, grid-controlled. | Max. 30 fps, grid-controlled. SAME |
{6}------------------------------------------------
- 7. Summary of non-clinical testing: This device has been tested and is certified to comply with the US Radiation Safety Performance Standard. Performance and safety testing was conducted by third party NRTL certified testing laboratories and the device was found to comply with the following FDA recognized standards:
- IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (or IEC 60601-1: 2012 . reprint); IEC 60601-1 :2005+Corrigendum 1+Corrigendum 2+A1 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance, FDA 19-4
- . IEC 60601-1-3:2008+A1 Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment FDA 12-269
{7}------------------------------------------------
- IEC 60601-1-6:2010+A1 General requirements for basic safety and essential performance ● Collateral standard: Usability FDA 5-89
- IEC/EN 60601-2-28:2010 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis FDA 12-204
- IEC 60601-2-54:2009+A1 Medical electrical equipment – Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy, FDA 12-296
- IEC 62304:2006 Medical device software – Software life cycle processes, FDA 13-32
- IEC 62366:2007+A1 Application of usability engineering to medical devices, FDA 5-114
- EN 60601-1-2:2015, IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests FDA 19-8
We consulted the following FDA guidance documents in the development of the FLUOROspeed: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, Guidance for Industry and Food and Drug Administration Staff; and Pediatric Information for X-ray Imaging Device Premarket Notifications.
- 8. Summary of clinical testing: Not applicable. Clinical testing was not deemed to show substantial equivalence. We relied on non-clinical testing and compliance with standards.
- 9. Conclusion: After analyzing standards compliance results and bench tests, it is the conclusion of Shimadzu Corporation that the MODIFIED FLUOROspeed is as safe and effective as the predicate device, has few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.