The Dual Safety Pen Needle is intended for use with pen injector devices for subcutaneous injection of insulin. The product has two safety shields which lock in place after use (patient-end) and upon removal of the needle from the pen (pen connector-end). The lock shields help reduce the occurrence of needle sticks from both ends of the needle.
Device Story
Dual Safety Pen Needle; single-use hypodermic needle for subcutaneous insulin delivery via pen injectors. Features dual safety shields: patient-end shield retracts during insertion, automatically extends/locks post-injection; pen-connector-end shield activates upon removal from pen. Audible click confirms patient-end shield activation. Used by patients or clinicians in home/clinical settings. Prevents accidental needlesticks at both ends of the cannula. Device is sterile, single-use, and disposable.
Clinical Evidence
Clinical simulated use study with 8 participants (endocrine doctors/nurses) testing 500 samples. Evaluated effectiveness of the posterior safety protective mechanism. Results showed zero failures. No clinical patient trials performed; device relies on bench testing and simulated use validation.
Technological Characteristics
Materials: POM (fixer/hub), X5CrNi18-10 (needle tube), 0Cr18Mn8Ni5N (springs), ABS (shield), PP (container). Standards: ISO 7864, ISO 9626, ISO 11608-2, ISO 23908, ISO 10993 series. Sterilization: Ethylene Oxide (SAL 10^-6). Dimensions: 29G-31G, 4mm-8mm lengths. Mechanical activation: 1N-5N force; override force >20N.
Indications for Use
Indicated for patients requiring subcutaneous insulin injection using pen injector devices. No specific age or gender restrictions stated.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
Verifine ® Safety Type Insulin Pen Needle (K161950)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 5, 2019
Promisemed Hangzhou Meditech Co., Ltd. % John Beasley Senior Consultant MedTech Review, LLC 257 Garnet Garden Street Henderson, Nevada 98015
Re: K191853
Trade/Device Name: Dual Safety Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: July 10, 2019 Received: July 10, 2019
Dear John Beasley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191853
Device Name Dual Safety Pen Needle
### Indications for Use (Describe)
The Dual Safety Pen Needle is intended for use with pen injector devices for subcutaneous injection of insulin. The product has two safety shields which lock in place after use (patient-end) and upon removal of the needle from the pen (pen connector-end). The lock shields help reduce the occurrence of needle sticks from both ends of the needle.
| Type of Use (Select one or both, as applicable) | <div> <span>☐</span> <span>☐</span> </div> |
|-------------------------------------------------|--------------------------------------------|
|-------------------------------------------------|--------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Preparation Date: 09/05/2019
| Contact Details | | 21 CFR 807.92(a)(1) | |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Name | Promisemed Hangzhou Meditech Co., Ltd. | | |
| Applicant Address | Bldg. 1, No. 12, Longtan Road, Cangqian Street, Yuhang District, Hangzhou City, Zhejiang, 311121, CH | | |
| Applicant Telephone Number | 865-718-8772985 | | |
| Applicant Contact | Mr. Zearou Yang | | |
| Applicant Contact Email | zearou.yang@promisemed.ca | | |
| Correspondent Name | MedTech Review, LLC | | |
| Correspondent Address | 257 Garnet Garden Street, Henderson, NV, 89015, US | | |
| Correspondent Telephone Number | 1-612-889-5168 | | |
| Correspondent Contact | Mr. John Beasley, RAC (US) | | |
| Correspondent Contact Email | john@medtechreview.com | | |
| Device Name | | 21 CFR 807.92(a)(2) | |
| Device Trade Name | Dual Safety Pen Needle | | |
| Common Name | HYPODERMIC SINGLE LUMEN NEEDLE. | | |
| Classification Name | NEEDLE, HYPODERMIC, SINGLE LUMEN | | |
| Regulation Number | 880.5570 | | |
| Product Code | FMI | | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | |
| Predicate [510(k)] # | Predicate Trade Name | Product Code | |
| K161950 | Verifine ® Safety Type Insulin Pen Needle | FMI | |
| Device Description Summary | | 21 CFR 807.92(a)(4) | |
| Item | Subject Device<br>K191853 | Predicate Device<br>K161950 | Similarities/Differences |
| Product Code | FMI | FMI | No difference in FDA product code. |
| Intended Use | The Dual Safety Pen<br>Needle is intended for<br>use with pen injector<br>devices for<br>subcutaneous injection<br>of insulin. The product<br>has two safety shields<br>which lock in place after<br>use (patient-end) and<br>upon removal of the<br>needle from the pen<br>(pen connector-end).<br>The lock shields help<br>reduce the occurrence<br>of needle sticks from<br>both ends of the<br>needle. | The Safety Type Insulin<br>Pen Needle is intended<br>for use with pen injector<br>devices for<br>subcutaneous injection<br>of insulin. Additionally,<br>after withdrawal of the<br>Safety Type Insulin Pen<br>Needle from the body,<br>the attached needle<br>safety shield<br>automatically covers<br>the needle to minimize<br>the risk of accidental<br>needlestick. | The safety pen needle products (subject<br>device and predicate device) help protect<br>against accidental needle sticks and are<br>substantially equivalent in their intended use. |
| Operating<br>Principle | As the user proceeds<br>with inserting the<br>needle into the skin the<br>shield will retract.<br>After the injection is<br>completed and needle<br>is removed from the<br>skin, the shield will<br>automatically extend to<br>cover the needle and<br>lock in place.<br>A click of the patient<br>end shield indicates the<br>needle has fully<br>penetrated the skin. A<br>second shield covers<br>the pen connector<br>needle when the needle<br>is removed from the<br>pen.<br>Once the Dual Safety<br>Pen Needle is in the<br>locked mode, it can no<br>longer be used. | As the user proceeds<br>with inserting the<br>needle into the skin the<br>shield will retract.<br>After the injection is<br>completed and needle<br>is removed from the<br>skin, the shield will<br>automatically extend to<br>cover the needle and<br>lock in place.<br>This product features<br>an audible cue (a click)<br>when the injection is<br>complete.<br>Once the Safety Type<br>Insulin Pen Needle is in<br>the locked mode, it can<br>no longer be used. | The safety pen needle products (subject<br>device and predicate device) use the same<br>operating principles: the use of lock shields<br>that help reduce the occurrence of needle<br>sticks. |
| Length | 4 ± 0.5 mm<br>5 ± 0.5 mm<br>6 ± 0.5 mm<br>8 ± 1.2 mm | 4 ± 0.5 mm<br>5 ± 0.5 mm<br>6 ± 0.5 mm<br>8 ± 1.2 mm | The needle tube lengths of the subject device<br>are the same as the predicate device. |
| Item | Subject Device<br>K191853 | Predicate Device<br>K161950 | Similarities/Differences |
| Gauge | 29G*6mm<br>29G*8mm | 29G*6mm<br>29G*8mm | The available needle lengths per gauge of the<br>subject device are the same as in the<br>predicate device. |
| | 30G*4mm<br>30G*5mm<br>30G*6mm<br>30G*8mm<br>31G*4mm<br>31G*5mm<br>31G*6mm | 30G*4mm<br>30G*5mm<br>30G*6mm<br>30G*8mm<br>31G*4mm<br>31G*5mm<br>31G*6mm | |
| | 31G*8mm | 31G*8mm | |
| Sharps Injury<br>Prevention<br>Features | The patient end of the<br>device has a<br>mechanism that allows<br>the needle to be<br>shielded and locked<br>after use.<br>The non-patient (pen<br>connection) end of the<br>cannula is visible prior<br>to attachment to the<br>pen injector. Following<br>removal of the device<br>from the pen injector,<br>the needle is shielded<br>with a mechanism that<br>is designed to reduce<br>the occurrence of<br>accidental needle stick<br>injuries. | The patient end of the<br>device has a<br>mechanism that allows<br>the needle to be<br>shielded and locked<br>after use. | The sharps injury prevention features of the<br>subject device includes both ends of the<br>needle as compared to only the patient end of<br>the needle in the predicate device. |
| Device Materials | Fixer, inside fixer:<br>POM<br>Hub: POM<br>Needle hub:<br>X5CrNi18-10<br>Springs:<br>0Cr18Mn8Ni5N<br>Shield: AB<br>Container: PP | Fixer: POM<br>Hub: POM<br>Needle tube:<br>X5CrNi18-10<br>Spring:<br>0Cr18Mn8Ni5N<br>Shield: ABS<br>Container: PP | The subject device contains an additional<br>posterior trigger spring in comparison to the<br>predicate device. Device materials and both<br>the nature of body contact and the duration of<br>body with the patient/user are the same. |
| Performance | Complies with<br>ISO 7864,<br>ISO 9626,<br>ISO 11608-2, and<br>ISO 23908 | Complies with<br>ISO 7864,<br>ISO 9626,<br>ISO 11608-2, and<br>ISO 23908 | The performance characteristics of the subject<br>device are the same as the predicate device. |
| Force to Activate<br>Sharps Injury<br>Prevention<br>Feature | Complies with ISO<br>23908. The force to<br>activate the sharps<br>injury prevention<br>feature is 1N to 5N for<br>both ends of the Dual<br>Safety Pen Needle. | Complies with ISO<br>23908. The force to<br>activate the sharps<br>injury prevention<br>feature is 1N to 5N for<br>the Safety Type Insulin<br>Pen Needle. | The forces for activate the patient-end sharps<br>injury prevention feature of the subject device<br>are the same as the predicate device. The<br>predicate device does not have a pen-end<br>sharps prevention feature. |
| Item | Subject Device<br>K191853 | Predicate Device<br>K161950 | Similarities/Differences |
| Force to<br>Override Sharps<br>Injury Prevention<br>Feature | Complies with ISO<br>23908. The force<br>required to override the<br>sharps injury prevention<br>feature greater than<br>20N for both ends of<br>the Dual Safety Pen<br>Needle. | Complies with ISO<br>23908. The force<br>required to override the<br>sharps injury prevention<br>feature is greater than<br>20N for the Safety Type<br>Insulin Pen Needle. | The forces for overriding the patient-end<br>sharps injury prevention feature of the subject<br>device are the same as the predicate device.<br>The predicate device does not have a pen-<br>end sharps prevention feature. |
| Sterilization | SAL: 10-6<br>Ethylene Oxide<br>Sterilization | SAL: 10-6<br>Gamma Radiation<br>Sterilization | The sterilization methods of the subject and<br>predicate devices are different; however, both<br>are traditional methods of processing medical<br>devices so as to provide sterility assurance<br>levels of 10-6. |
| Shelf Life | 5 years | 5 years | Device stability and expiry date of the subject<br>device are the same as the predicate device. |
| Single Use | Yes | Yes | The safety pen needle products (subject<br>device and predicate device) are single use<br>devices. |
| Biocompatibility | Complies with ISO<br>10993 Series<br>Standards for<br>• Cytotoxicity<br>• Skin Irritation<br>• Skin Sensitization<br>• Acute Systemic<br>• Toxicity<br>• Hemolysis<br>• Non pyrogenic<br>(material mediated<br>pyrogenicity) | Complies with ISO<br>10993 Series<br>Standards for<br>• Cytotoxicity<br>• Skin Irritation<br>• Skin Sensitization<br>• Acute Systemic<br>• Toxicity<br>• Hemolysis<br>• Non pyrogenic<br>(material mediated<br>pyrogenicity) | The biological safety of the subject device is<br>identical to the predicate device. |
| | Non Clinical and/or Clinical Tests Summary & Conclusions | | 21 CFR 807.92(b) |
The device is a pen needle with shields at both the patient end and the pen connector end which are intended to reduce risk of needle stick before and after injections. The patient end mechanism shields the needle before and after an injection, and a second mechanism passively covers the pen connector needle following removal from the patient end shield indicates the needle has fully penetrated the skin. A second shield covers the pen connector needle when it is removed from the pen.
## Intended Use/Indications for Use
The Dual Safety Pen Needle is intended for use with pen injector devices for subcutaneous injection of insulin. The product has two safety shields which lock in place after use (patient-end) and upon removal of the needle from the pen (pen connector-end). The lock shields help reduce the occurrence of needle sticks from both ends of the needle.
## Indications for Use Comparison
Promisemed adds an additional safety shield to the predicate device such that the subject device's two safety shields lock in place (i) after use (patient-end) and (ii) upon removal of the needle from the pen (pen connector-end). The lock shields help reduce the occurrence of needle sticks from both ends of the needle. This additional protection does not constitute a new intended use. However, the Indications For Use' has been revised to reflect the additional safety shield. Both safety pen needle products (the predicate device and the subject device) help protect against accidental needle sticks.
# 21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
{4}------------------------------------------------
## Technological Comparison
21 CFR 807.92(a)(6)
For both the subject and predicate devices, the non-patient (pen connector) ends of the cannula are visible prior to attachment to the pen injector. However, the predicate device is modified such that following removal of the subject device from the pen injector, its needle is shielded with a mechanism designed to reduce the occurrence of accidental needle stick injuries.
{5}------------------------------------------------
{6}------------------------------------------------
### Non-Clinical Summary:
Product functionality: After sterilization, the device meets criteria specified in ISO 11608-2.
Shelf life: Sterile devices meet criteria specified in ISO 11608-2 after accelerated aging. The product is labeled with an expiry date of 5 years.
Biocompatibility: Device meets cytotoxicity requirements of ISO 10093-5 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity
Biocompatibility: Device meets sensitization requirements of ISO 10993-10 Biological evaluation of medical devices -Part 10:Tests for irritation and skin sensitization.
Biocompatibility: Device meets intracutaneous reactivity requirements of ISO 10993-10 Biological evaluation of medical devices - Part 10:Tests for irritation and skin sensitization.
Biocompatibility: Device meets hemocompatibility requirements of both ISO 10993-4 Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood and ASTM F756 Standard Practice for Assessment of Hemolytic Properties of Materials
Biocompatibility: Device meets system toxicity requirements of ISO 10993-11:2006(R)2010, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
Sharps injury protection: Device meets safety mechanism activation requirements as per ISO 23908.internal protocol and test results
Sharps injury protection: Device meets safety overriding/unlocking force after activation requirements as per ISO 23908, internal protocol and test results
{7}------------------------------------------------
Clinical Simulated Use testing evaluated whether the posterior safety protective mechanism of the Dual Safety Pen Needle is effective. A total of 8 participants were selected to evaluate the subject device, and included endocrine doctors and nurses who provide training to patients in the use of pen needles for insulin injection. A total of 500 samples were tested and the evaluations resulted in zero failures of each of the Dual Safety Pen Needle.
Sterilization: A Sterility Assurance Level (SAL) of 10° has been validated in accordance with the requirements of ISO 11135:2014 for Ethylene Oxide.
LAL Pyrogen Testing: Device meets endotoxin requirements of USP39_NF34<85> Bacterial Endotoxins Test.
Packaging: Shipping package meets requirements of ISTA 3A. Unit package functionality meets acceptance criteria specified in ASTM F1929-98 (2004) for dye penetration and peel strength after EO sterilization. The unit package also provides a seal having a microbial barrier that meets log reduction value (LRV) requirements of ASTM F1608-16.
### Conclusion:
Based on device comparison information and functional performance bench testing, the proposed device is substantially equivalent to the legally marketed predicate device (K161950).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.